CERTIFICATE OF CALIBRATION

Date of Issue : 07 January 2023 Certificate Number : QR20-2284


Order Item No.: 2012068 Page : 1 of 6

www.qreborn.com

Quality Reborn Co.,Ltd.


42/267 Leab Klong pasricharoen fangnue 8/1
Nongkham, Bangkok 10160 NSC-TISI-TIS17025
Tel : +662-4447-382, Fax: +662-4447-383 CALIBRATION 0292

Customer: DKSH (Myanmar) Ltd


No.13, Thitsar Road, Yankin Township, Yangon, Myanmar.
Date Received: 27 December 2022
Date of Calibration: 28 December 2022
Due of Calibration: 28 December 2024
Instrument: Description: Data Acquisition/Switch Unit with Sensor
Model: 34972A
Serial Number: MY59002090
ID Number: -
Manufacturer: Keysight
Site: Quality Reborn Office
Location: Calibration Room 1
Environmental Conditions
Temperature: 25 ̊C ± 3 ̊C
Relative Humidity: 55% ± 25 %
Calibration Method Used
This instrument was calibrated by comparison of indication with the Standard Platinum
Resistance Thermometer in liquid calibration bath according to calibration procedure no. CP-T01-01.
Traceability of Measurement
This certificate of calibration documents the traceability to national standards, which
realize the units of measurement according to the International System of Units (SI).
Calibrated By: Mr.Wutthichai Rujiwut Approved By: .

[ ]Mr. Thanat Sutthinate

[ ]Mr. Jatuporn Juijai-ngam

This certificate may not be reproduced other than in full except with the prior written approval of Quality Reborn Co.,Ltd
Certificate No. : QR20-2284
www.qreborn.com
Order Item No.: 2012058
Page : 2 of 6

Details of Calibration
1. Reference Standard Equipment Used
Description Certificate No. Due Date
Platinum Resistance Thermometer “Accu Mac” S/N 1621951 91037 03-April 2023
Reference Thermometer “Isotech” milliK S/N39622-1 C20/00002 08-January 2023
2. The results reported in this certificate refer to the condition of the instrument on the date of calibration
and carry no implication regarding the long-term stability of the instrument.
3. Condition of calibration item: normal condition, no indication find for any damage or malfunction
4. Immersion Depth of Calibration = 150 mm.
Result of Calibration: (Without Adjustment)
Function: Temperature Measurement of Slot 100 S/N MY60000817
Description of Sensor: RTD Model:Pt100 Diameter 1.6mm Length: 200mm
Ch: 101 SensorS/N. 5002915-001 Parameter Setting: Gain=1.0 Offset=0.0000
Standard Temperature UUC* Temperature Uncertainty of
(C̊ ) Reading ( C
̊ ) Correction ( C
̊ ) Measurement (± ̊C)
-39.9953 -40.10 0.1047 0.040
-20.0098 -20.05 0.0402 0.040
-0.0088 0.02 -0.0288 0.040
500084 50.20 -0.0181 0.040
99.9917 100.33 -0.3383 0.040

Ch: 102 SensorS/N.: 5002915-002 Parameter Setting: Gain=1.0 Offset=0.0000


Standard Temperature UUC* Temperature Uncertainty of
(C̊ ) Reading ( C
̊ ) Correction ( C
̊ ) Measurement (± ̊C)
-39.9953 -40.10 0.1147 0.040
-20.0098 -20.05 0.0402 0.040
-0.0088 0.03 -0.0388 0.040
50.0084 50.21 -0.2016 0.040
99.9917 100.34 -0.3483 0.040

Ch: 103 SensorS/N.: 5002915-003 Parameter Setting: Gain=1.0 Offset=0.0000


Standard Temperature UUC* Temperature Uncertainty of
(C̊ ) Reading ( C
̊ ) Correction ( C
̊ ) Measurement (± ̊C)
-39.9953 -40.09 0.0947 0.040
-20.0098 -20.03 0.0202 0.040
-0.0088 0.04 -0.0488 0.040
50.0084 50.18 -0.1716 0.040
99.9917 100.26 -0.2683 0.040

UUC*: Unit Under Calibration


Certificate No. : QR20-2284
www.qreborn.com
Order Item No.: 2012058
Page : 3 of 6

Ch: 104 SensorS/N. 5002915-004 Parameter Setting: Gain=1.0 Offset=0.0000

Standard Temperature UUC* Temperature Uncertainty of


(C̊ ) Reading ( C
̊ ) Correction ( C
̊ ) Measurement (± ̊C)
-39.9953 -40.10 0.1047 0.040
-20.0098 -20.03 0.0202 0.040
-0.0088 0.04 -0.0488 0.040
500084 50.18 -0.1716 0.040
99.9917 100.28 -0.2883 0.040

Ch: 105 SensorS/N.: 5002915-005 Parameter Setting: Gain=1.0 Offset=0.0000


Standard Temperature UUC* Temperature Uncertainty of
(C̊ ) Reading ( C
̊ ) Correction ( C
̊ ) Measurement (± ̊C)
-39.9953 -40.10 0.1147 0.040
-20.0098 -20.05 0.0402 0.040
-0.0088 0.03 -0.0388 0.040
50.0084 50.21 -0.2016 0.040
99.9917 100.34 -0.3483 0.040
Ch: 103 SensorS/N.: 5002915-002 Parameter Setting: Gain=1.0 Offset=0.0000
Standard Temperature UUC* Temperature Uncertainty of
(C̊ ) Reading ( C
̊ ) Correction ( C
̊ ) Measurement (± ̊C)
-39.9953 -40.10 0.1047 0.040
-20.0098 -20.04 0.0302 0.040
-0.0088 0.03 -0.0388 0.040
50.0084 50.19 -0.1816 0.040
99.9917 100.30 -0.3083 0.040

UUC*: Unit Under Calibration


Certificate No. : QR20-2284
www.qreborn.com
Order Item No.: 2012058
Page : 4 of 6

Result of Calibration: ( Without Adjustment)


Function: Temperature Measurement of Slot 200 S/N MY60001201
Description of Sensor: RTD Model: Pt 100 Diameter 3.2 mm Length: 50 mm
Ch: 201 SensorS/N. 1 Parameter Setting: Gain=1.0 Offset=0.0000

Standard Temperature UUC* Temperature Uncertainty of


(C̊ ) Reading ( C
̊ ) Correction ( C
̊ ) Measurement (± ̊C)
-39.9954 -40.25 0.2553 0.040
-20.0017 -20.22 0.2183 0.040
-0.0105 -0.19 0.1795 0.040
99.9949 99.89 0.1049 0.040
200.0061 199.65 0.3561 0.040

Ch:202 SensorS/N.: 2 Parameter Setting: Gain=1.0 Offset=0.0000

Standard Temperature UUC* Temperature Uncertainty of


(C̊ ) Reading ( C
̊ ) Correction ( C
̊ ) Measurement (± ̊C)
-39.9947 -40.23 0.2353 0.040
-20.0017 -20.17 0.1683 0.040
-0.0105 -0.12 0.1095 0.040
99.9949 99.98 0.0149 0.040
200.0061 199.78 0.2261 0.040

Ch: 203 SensorS/N.:3 Parameter Setting: Gain=1.0 Offset=0.0000


Standard Temperature UUC* Temperature Uncertainty of
(C̊ ) Reading ( C
̊ ) Correction ( C
̊ ) Measurement (± ̊C)
-39.9947 -40.24 0.2453 0.040
-20.0017 -20.20 0.1983 0.040
-0.0105 -0.17 0.1595 0.040
99.9949 99.87 0.1249 0.040
200.0061 199.62 0.3861 0.040

Ch: 204 SensorS/N.:4 Parameter Setting: Gain=1.0 Offset=0.0000

Standard Temperature UUC* Temperature Uncertainty of


(C̊ ) Reading ( C
̊ ) Correction ( C
̊ ) Measurement (± ̊C)
-39.9947 -40.22 0.2253 0.040
-20.0017 -20.20 0.1983 0.040
-0.0105 -0.18 0.1695 0.040
99.9949 99.88 0.1149 0.040
200.0061 199.63 0.3761 0.040

UUC*: Unit Under Calibration


Certificate No. : QR20-2284
www.qreborn.com
Order Item No.: 2012058
Page : 5 of 6

Ch: 205 SensorS/N. 5 Parameter Setting: Gain=1.0 Offset=0.0000

Standard Temperature UUC* Temperature Uncertainty of


(C̊ ) Reading ( C
̊ ) Correction ( C
̊ ) Measurement (± ̊C)
-39.9954 -40.26 0.2653 0.040
-20.0017 -20.23 0.2283 0.040
-0.0105 -0.20 0.1895 0.040
99.9949 99.86 0.1349 0.040
200.0061 199.64 0.3661 0.040

Ch:206 SensorS/N.: 6 Parameter Setting: Gain=1.0 Offset=0.0000


Standard Temperature UUC* Temperature Uncertainty of
(C̊ ) Reading ( C
̊ ) Correction ( C
̊ ) Measurement (± ̊C)
-39.9947 -40.20 0.2053 0.040
-20.0017 -20.18 0.1783 0.040
-0.0105 -0.15 0.1395 0.040
99.9949 99.94 0.0549 0.040
200.0061 199.73 0.2761 0.040

Ch: 207 SensorS/N.:7 Parameter Setting: Gain=1.0 Offset=0.0000


Standard Temperature UUC* Temperature Uncertainty of
(C̊ ) Reading ( C
̊ ) Correction ( C
̊ ) Measurement (± ̊C)
-39.9947 -40.24 0.2453 0.040
-20.0017 -20.18 0.1783 0.040
-0.0105 -0.12 0.1095 0.040
99.9949 99.99 0.0049 0.040
200.0061 199.81 0.1961 0.040

Ch: 208 SensorS/N.:8 Parameter Setting: Gain=1.0 Offset=0.0000


Standard Temperature UUC* Temperature Uncertainty of
(C̊ ) Reading ( C
̊ ) Correction ( C
̊ ) Measurement (± ̊C)
-39.9947 -40.28 0.2853 0.040
-20.0017 -20.24 0.2383 0.040
-0.0105 -0.18 0.1695 0.040
99.9949 99.89 0.1049 0.040
200.0061 199.67 0.3361 0.040

UUC*: Unit Under Calibration


Certificate No. : QR20-2284
www.qreborn.com
Order Item No.: 2012058
Page : 6 of 6

Ch: 209 SensorS/No.9 Parameter Setting: Gain=1.0 Offset=0.0000

Standard Temperature UUC* Temperature Uncertainty of


(C̊ ) Reading ( C
̊ ) Correction ( C
̊ ) Measurement (± ̊C)
-39.9947 -40.24 0.2453 0.040
-20.0017 -20.21 0.2083 0.040
-0.0105 -0.17 0.1595 0.040
99.9949 99.90 0.0949 0.040
200.0061 199.66 0.3461 0.040

Ch:210 SensorS/N.: 10 Parameter Setting: Gain=1.0 Offset=0.0000


Standard Temperature UUC* Temperature Uncertainty of
(C̊ ) Reading ( C
̊ ) Correction ( C
̊ ) Measurement (± ̊C)
-39.9947 -40.21 0.2153 0.040
-20.0017 -20.16 0.1583 0.040
-0.0105 -0.12 0.1095 0.040
99.9949 99.99 0.0049 0.040
200.0061 199.80 0.2061 0.040

UUC*: Unit Under Calibration


This reported uncertainty of measurement was based on a standard uncertainty multiplied by a
coverage factor k=2, providing a level of confidence of approximately 95%.