You are on page 1of 1

Certificate of

Calibration
8635 Washington Avenue • Racine, Wis. 53406-1580 • Phone: 877-243-1010 • indorders@badgermeter.com

Customer Name: CIMA LTDA Report # 993948 - 2106000032


Customer Address: CARRERA 96J # 23 A 55 BOGOTÁ COLUMBIA 110911
Customer PO # CIMA-DYM-OC-001-2021
Model # CPT84-AN-C-C06 Cal Date: 6/23/2021
Serial # 2106000032 Customer Re-Cal Date:
Signal: Pulse Output Lab Temp: 72 Deg. F
Calibration Procedure: QAP-035 Lab Relative Humidity: 52%
Calibration Tech: Tori
Fluid Specifications: MIL-PRF-7024 II Specific Gravity: 0.76
Temperature (F): 84.0 Viscosity (CSTKS): 1.12
Notes, Adjustments & Repairs
Calibration Results (Initial Calibration)

Test Point Frequency Flow Rate Freq./ Visc K Factor Flow Rate Mass Flow
# Hz GPM Hz/cstk pul/gal LPM PPH
1 20.995 0.0472 18.746 26685.7 0.1787 17.946
2 35.247 0.0759 31.470 27857.8 0.2874 28.861
3 58.574 0.1235 52.298 28464.1 0.4674 46.940
4 94.422 0.1970 84.305 28758.2 0.7457 74.893
5 148.30 0.3100 132.41 28699.9 1.1736 117.87
6 232.16 0.4867 207.29 28622.1 1.8423 185.02
7 362.63 0.7610 323.78 28590.4 2.8808 289.32
8 587.22 1.2337 524.30 28558.8 4.6701 469.02
9 950.10 1.9979 848.31 28533.0 7.5629 759.55
10 1557.0 3.2783 1390.2 28496.5 12.410 1246.3

Standards Used in Calibration


Standard # Description Serial # ReCal Date
FDI-130 3 GPM Liquid Prover N/A 11/20/2023
FDI-130 Cart 3 GPM Liquid Prover Cart (mA, Vdc, Temp, Freq) N/A 10/15/2021
The instrument referenced above was calibrated using standards traceable to the National Institute of Standards and Technology or SI. Calibration reports for
references maintained by Badger Meter, Inc. are available upon request to the customer of this calibration report. The overall Expanded Uncertainty of the
meter factor is +/- 0.0964% of reading. (Represents an expanded uncertainty using a coverage factor, k = 2.20, at an approximate level of confidence of
95%). This uncertainty includes +/- 0.0407% of reading for the volumetric flow rates reported, +/- 0.0100% of reading uncertainty of the UUT (Unit Under
Test), and +/- 0.0130% of reading for fluid properties effects.
Badger Meter, Inc. calibration services are accredited by NVLAP (NVLAP Lab Code 200668) to ISO/IEC 17025:2017 (NIST Handbook 150) and are
compliant to ANSI/NCSLZ540-l-1994; Part I. This certificate is not used to claim product certification, approval, or endorsement by NVLAP, NIST. or any
agency of the U.S. Government.
Statements of conformity to specifications are made in this report without taking measurement uncertainty into account, except when requested by the
customer. Where statements of conformity are made in this report, the following decision rules are applied: Pass or In-tolerance - results within
limits/specifications. Fail or Out-of-tolerance - results exceed limits/specifications.
The results reported relate only to the item(s) calibrated as described above. This report may not be reproduced, except in full, without the written
approval of Badger Meter, Inc.

Daniel Digitally signed


by Daniel Norton
Date: 2021.06.23

I certify the accuracy of this Calibration Report: Daniel Norton Calibration Engineer
Norton 16:19:04 -05'00'

Name Title Signature


End of Report

Doc Nbr: CRF-001 Rev: L


Report #:993948 - 2106000032-I Page 1 of 1