CERTIFICATE OF

CALIBRATION

NO. 28177-501

Page I of 1

CalibrationDate: OglOAl2O14.
Customeri

RecalibrationDue: O2lOBl2OlS

Description: Gau6sMeter

CalibrationLocation: CCt

Model:

ObservedCondition: good

Manufacturer: R.B.Annis

Size:

MfrSerialNo:

Calibration

25120

+/-20 x I gauss

03217
lD
n/a
As Reoeived: Wilhin Tolerance

Number

Uncertainty: +/-,25%FS

Previous

Ambient 68 F., R.H.46%

Cerl n/a

Procedure: 118-A - DC Gauss Meter, Analog


Techniciani Roger K. Mefford
As Returned: Within Tolerance

CALIBRATION RESULTS:

Test Poi4t: -20.0 -10.0 -5.0 0.0 +5.C +1C.0 +20.0 ga-usE PointE
Check: -20.0 -L0.2 -5.0 0.0 +5.0 +10.1 +20.1 gauss checka accurate
to +/-[5* of ful1 sca].el at tst points.
liieets Asmrr EL444t4-L2 publisbd specificati.ons
No adjustruent fequi.Eed.
Accpta.b1e.

Descriotion
Gauss lvleter
Gap [.4agnet F]xtlre

APPLICABLE MEASUREIV]ENT STANDARDS

L4fr.
Model
l.D Nlrnber Due Date
N I S T Ref. No
Alphalab Gl\,11 HS
151
OTnAnA 2/165-9A1
MTl
MXF|25 153
07119/15 27165 3O1

AS FOUND AND FTNAL RESULTS ARE THE SAII/IE, UNLESS OTHERWISE NOTED. The catibEtion described abov was pedomed
in compllance wilh ANSUNCSL 2540.3-2006: ISO 10012-1 & ISO/IEC Guide 25. Accuraciss of standads used a tE@able to the Nalional lnstitute

of Standads & Technology (N.l.S.T.) or have been derived trom accepted vatues of natuEtty occuiring physictconsrants.

Cerlined Calibraiions'rosporsibilitiss shallin no event, norfor any cause whatsoever exceed ths cosl ot the setuice epEsented.

This,epod applies only to the item(s) identifed abovo, atlhelime of testing. This pod shallnot be rcprcduced, excepl in full,
wilhout witlen pemission from Cedified Calibrations, ln6.

Certified Calibrations,

lnc.

FPd,l727)734-9137

Reviewed & Approved: MkhapL L. R.epdt