1, SIDCO Industrial Estate,


CC-2559 Ambattur, Chennai - 600098.
CERTIFICATE OF CALIBRATION Telefax : +91-44-66544651/52/53

FR-QT-21
Certificate No : C-200218-10-4
Date of Calibration : 19-Feb-20 Recom. Due Date : 19-Feb-21
Customer Details : SRF No. : C-200218-10
M/s. UL India Private Limited, Calibrated at : Onsite
Sy.No.129/4, Kalyani Platina, Phase II, Date of Receipt : 18-Feb-20
EPIPZone, WhiteField,Bangalore-560 066 Cond. On Receipt : Satisfactory

Details of Test Instrument:


Description : Voltage Probe Model No. : 701938
Range : 600 V / DC to 200 MHz Serial No : --
Least Count : -- Identification No. : MVP01
Make : Yokogawa Location : Medical Lab

Details of Standard Used :


Name Sr.No. Valid upto Traceability
Multifunction calibrator F105318 27-Jul-20 G&B, Mumbai
Digital Multimeter (6½) 1613015 26-Feb-20 G&B, Mumbai
Digital Oscilloscope C022786 29-Aug-20 G&B, Chennai
Work Instruction /IS No. : WI-ET-11 &05
Environmental Details : Temperature: 25±5 °C Relative Humidity: 50±15 % RH
Remarks
1. Equipment used for calibration were calibrated & traceable to National & International Standards.
2. This report refers only to particular item (s) submitted for calibration.
3. Lawkim Motors group is not liable for any change in calibration data & performance specifications on account of malfunctioning of Standards /
Instruments / Equipment covered by this report due to damage caused to it after issuance of this report.
4. The calibration results reported are valid at the time of and under the stated conditions of the measurements.
5. This report shall not be reproduced in full / part without prior permission of Lawkim Motors Group.
6. Satisfactory calibtration report in no ways implies that the equipment calibrated is approved by NABL.
7. All precautions have been taken for any error or omission while calibrating the instruments and issuing this certificate. However Lawkim Motors
Group shall not be liable for any loss or liability that may be arise to any party in this regards.
8. Readings given are average of 3 Measurements.
9. The reported Expanded Uncertainty is calculated at 95.45 % C.L with coverage factor k=2.

AUTHORIZED SIGNATORY :C. Praveen


DESIGNATION : D M
CALIBRATED BY : Lab Engineer : S.Senthamizh Selvan
ULR NO : CC255920000006522 F

Page 1 of 11
Godrej & Boyce Mfg. Co. Ltd.,
No. 1, SIDCO Industrial Estate,
CC-2559 Ambattur, Chennai - 600098.
CERTIFICATE OF CALIBRATION Telefax : +91-44-66544651/52/53

Certificate No : C-200218-10-4
ULR NO : CC255920000006522 F
Calibration Results
SL NO PARAMETER / STANDARD EXPECTED MEASURED CALCULATED ERROR MEASUREMENT CTL limits Remarks
RANGE INPUT OUTPUT OUTPUT READING UNCERTAINTY ±
V V V V % % %
1 DC VOLTAGE 1.0000 0.10 0.100 1.000 0.000 0.006 1.5
2 600 V 50.0000 5.00 5.016 50.156 0.313 0.006 1.5
3 100.0000 10.00 10.02 100.23 0.230 0.006 1.5
4 200.0000 20.00 20.13 201.30 0.650 0.006 1.5
5 300.000 30.00 30.3 303.00 1.000 0.006 1.5
6 500.000 50.00 50.5 505.00 1.000 0.006 1.5
7 600.000 60.00 60.7 607.00 1.167 0.006 1.5
V V V V % % %
8 AC VOLTAGE 1.000000 0.1000 0.100489 1.005 0.4890 0.156 1.5
9 600 V @ 50Hz 50.00000 5.0000 5.03098 50.310 0.6196 0.156 1.5
10 100.0000 10.0000 10.0610 100.610 0.6100 0.156 1.5
11 200.0000 20.0000 20.1206 201.206 0.6030 0.156 1.5
12 300.000 30.0000 30.1839 301.839 0.6130 0.156 1.5
13 500.000 50.0000 50.3135 503.135 0.6270 0.156 1.5
14 600.000 60.0000 60.3755 603.755 0.6258 0.156 1.5
V V V V % % %
15 AC VOLTAGE 1.000000 0.1000 0.100521 1.00521 0.5210 0.241 1.5
16 600 V @ 60Hz 50.00000 5.0000 5.02910 50.2910 0.5820 0.241 1.5
17 100.0000 10.0000 10.0563 100.563 0.5630 0.241 1.5
18 200.0000 20.0000 20.1114 201.114 0.5570 0.241 1.5
19 300.000 30.0000 30.1672 301.672 0.5573 0.241 1.5
20 500.000 50.0000 50.2859 502.859 0.5718 0.241 1.5
21 600.000 60.0000 60.3476 603.476 0.5793 0.241 1.5

BANDWIDTH SL NO STANDARD TEST MEASUREMENT


READING READING UNCERTAINTY
MHz dB ± dB
22 100 -2.64 0.39
23 200 -2.78 0.39

Page 2 of 11
Godrej & Boyce Mfg. Co. Ltd.,
Godrej & Boyce Mfg. Co. Ltd., Lawkim Motors Group (Calibration and Inspection)
RAW DATA SHEET
Certificate No.: C-200218-10-4 SRF No : C-200218-10
Date of Calibration : 19/Feb/20 Calibration Frequency: 12 Months
Calibrated at: Onsite
Details of Test Instrument :
Description: Voltage Probe Model: 701938
Range: 600 V / DC to 200 MHz Serial No: --
Least count: -- ID No: MVP01
Make: Yokogawa Location: Medical Lab
Details of Standard Used:
Sr.No. Name Sr. No/ I.D. No. Valid Up to
1 Multifunction calibrator F105318 27/Jul/20
2 Digital Ocilloscope C011303 26/Dec/17
3 Digital Multimeter (6½) 1613015 26/Feb/20

Environmental Temperature: Start: 25 °C End: 25 °C


Condition: Humidity: Start: 52% End: 52%
Work Instruction /IS No. WI-ET-11 &05
OBSERVATIONS
SL NO PARAMETER / STANDARD EXPECTED MEASURED
RANGE INPUT OUTPUT OUTPUT
V V V V V
1 DC VOLTAGE 1.000000 0.1000 0.100000 0.100000 0.100000
2 1000 V 50.000000 5.0000 5.015640 5.015640 5.015640
3 100.000000 10.0000 10.023000 10.023000 10.023000
4 200.000000 20.0000 20.130000 20.130000 20.130000
5 300.000000 30.0000 30.300000 30.300000 30.300000
6 500.000000 50.0000 50.500000 50.500000 50.500000
7 600.000000 60.0000 60.700000 60.700000 60.700000
V V V V V
8 AC VOLTAGE 1.000000 0.1000 0.100489 0.100489 0.100489
9 1000 V @ 50Hz 50.000000 5.0000 5.030980 5.030980 5.030980
10 100.000000 10.0000 10.061000 10.061000 10.061000
11 200.000000 20.0000 20.120600 20.120600 20.120600
12 300.000000 30.0000 30.183900 30.183900 30.183900
13 500.000000 50.0000 50.313500 50.313500 50.313500
14 600.000000 60.0000 60.375500 60.375500 60.375500
V V V V V
15 AC VOLTAGE 1.000000 0.1000 0.100521 0.100521 0.100521
16 1000 V @ 60Hz 50.000000 5.0000 5.029100 5.029100 5.029100
17 100.000000 10.0000 10.056300 10.056300 10.056300
18 200.000000 20.0000 20.111400 20.111400 20.111400
19 300.000000 30.0000 30.167200 30.167200 30.167200
20 500.000000 50.0000 50.285900 50.285900 50.285900
21 600.000000 60.0000 60.347600 60.347600 60.347600
SL NO STANDARD TES T
READING READING
MHz dB dB dB
22 100 -2.64 -2.64 -2.64
23 200 -2.78 -2.78 -2.78

Calibrated by: Approved By:


Name & Signature: Name & Signature:
FR-GR-01
Issue No.01,Date: 16.05.2015
Rev. No. 00 , Date:--
No. 1, SIDCO Industrial Estate,
CC-2559 Ambattur, Chennai - 600098.
CERTIFICATE OF CALIBRATION Telefax : +91-44-66544651/52/53

FR-QT-21
Certificate No : C-200218-10-5
Date of Calibration : 19-Feb-20 Recom. Due Date : 19-Feb-21
Customer Details : SRF No. : C-200218-10
M/s. UL India Private Limited, Calibrated at : Onsite
Sy.No.129/4, Kalyani Platina, Phase II, Date of Receipt : 18-Feb-20
EPIPZone, WhiteField,Bangalore-560 066 Cond. On Receipt : Satisfactory

Details of Test Instrument:


Description : Voltage Probe Model No. : 701938
Range : 600 V / DC to 200 MHz Serial No : --
Least Count : -- Identification No. : MVP02
Make : Yokogawa Location : Medical Lab

Details of Standard Used :


Name Sr.No. Valid upto Traceability
Multifunction calibrator F105318 27-Jul-20 G&B, Mumbai
Digital Multimeter (6½) 1613015 26-Feb-20 G&B, Mumbai
Digital Oscilloscope C022786 29-Aug-20 G&B, Chennai
Work Instruction /IS No. : WI-ET-11 &05
Environmental Details : Temperature: 25±5 °C Relative Humidity: 50±15 % RH
Remarks
1. Equipment used for calibration were calibrated & traceable to National & International Standards.
2. This report refers only to particular item (s) submitted for calibration.
3. Lawkim Motors group is not liable for any change in calibration data & performance specifications on account of malfunctioning of Standards /
Instruments / Equipment covered by this report due to damage caused to it after issuance of this report.
4. The calibration results reported are valid at the time of and under the stated conditions of the measurements.
5. This report shall not be reproduced in full / part without prior permission of Lawkim Motors Group.
6. Satisfactory calibtration report in no ways implies that the equipment calibrated is approved by NABL.
7. All precautions have been taken for any error or omission while calibrating the instruments and issuing this certificate. However Lawkim Motors
Group shall not be liable for any loss or liability that may be arise to any party in this regards.
8. Readings given are average of 3 Measurements.
9. The reported Expanded Uncertainty is calculated at 95.45 % C.L with coverage factor k=2.

AUTHORIZED SIGNATORY :C. Praveen


DESIGNATION : D M
CALIBRATED BY : Lab Engineer : S.Senthamizh Selvan
Page 5 of 11
Godrej & Boyce Mfg. Co. Ltd.,
Regd. Office: Pirojshanagar, L.B.S. Marg, Vikhroli, Mumbai - 400 079.T.:+91-22-67961700/1800 W:www.godrej.com
(This document is digitally signed on first page and does not require signature)
No. 1, SIDCO Industrial Estate,
CC-2559 Ambattur, Chennai - 600098.
CERTIFICATE OF CALIBRATION Telefax : +91-44-66544651/52/53

ULR NO : CC255920000006523 F

Page 6 of 11
Godrej & Boyce Mfg. Co. Ltd.,
Regd. Office: Pirojshanagar, L.B.S. Marg, Vikhroli, Mumbai - 400 079.T.:+91-22-67961700/1800 W:www.godrej.com
(This document is digitally signed on first page and does not require signature)
No. 1, SIDCO Industrial Estate,
CC-2559 Ambattur, Chennai - 600098.
CERTIFICATE OF CALIBRATION Telefax : +91-44-66544651/52/53

Certificate No : C-200218-10-5
ULR NO : CC255920000006523 F
Calibration Results
SL NO PARAMETER / STANDARD EXPECTED MEASURED CALCULATED ERROR MEASUREMENT CTL limits Remarks
RANGE INPUT OUTPUT OUTPUT READING UNCERTAINTY ±
V V V V % % %
1 DC VOLTAGE 1.000000 0.10 0.100 1.00 0.00 0.006 1.5
2 600 V 50.00000 5.00 5.016 50.16 0.31 0.006 1.5
3 100.0000 10.00 10.02 100.23 0.23 0.006 1.5
4 200.0000 20.00 20.13 201.30 0.65 0.006 1.5
5 300.000 30.00 30.3 303.0 1.00 0.006 1.5
6 500.000 50.00 50.5 505.0 1.00 0.006 1.5
7 600.000 60.00 60.7 607.0 1.17 0.006 1.5
V V V V % % %
8 AC VOLTAGE 1.000000 0.1000 0.100718 1.007 0.7180 0.156 1.5
9 600 V @ 50Hz 50.00000 5.0000 5.03224 50.322 0.6448 0.156 1.5
10 100.0000 10.0000 10.0665 100.665 0.6650 0.156 1.5
11 200.0000 20.0000 20.1301 201.301 0.6505 0.156 1.5
12 300.000 30.0000 30.1951 301.951 0.6503 0.156 1.5
13 500.000 50.0000 50.3287 503.287 0.6574 0.156 1.5
14 600.000 60.0000 60.3942 603.942 0.6570 0.156 1.5
V V V V % % %
15 AC VOLTAGE 1.000000 0.1000 0.100591 1.00591 0.5910 0.241 1.5
16 600 V @ 60Hz 50.00000 5.0000 5.03052 50.3052 0.6104 0.241 1.5
17 100.0000 10.0000 10.0625 100.625 0.6250 0.241 1.5
18 200.0000 20.0000 20.1203 201.203 0.6015 0.241 1.5
19 300.000 30.0000 30.1807 301.807 0.6023 0.241 1.5
20 500.000 50.0000 50.3056 503.056 0.6112 0.241 1.5
21 600.000 60.0000 60.3684 603.684 0.6140 0.241 1.5

BANDWIDTH SL NO STANDARD TEST MEASUREMENT


READING READING UNCERTAINTY
MHz dB ± dB
22 100 -2.64 0.39
23 200 -2.78 0.39

Page 7 of 11
Godrej & Boyce Mfg. Co. Ltd.,
Regd. Office: Pirojshanagar, L.B.S. Marg, Vikhroli, Mumbai - 400 079.T.:+91-22-67961700/1800 W:www.godrej.com
(This document is digitally signed on first page and does not require signature)
Godrej & Boyce Mfg. Co. Ltd., Lawkim Motors Group (Calibration and Inspection)
RAW DATA SHEET
Certificate No.: C-200218-10-5 SRF No : C-200218-10
Date of Calibration : 19/Feb/20 Calibration Frequency: 12 Months
Calibrated at: Onsite
Details of Test Instrument :
Description: Voltage Probe Model: 701938
Range: 600 V / DC to 200 MHz Serial No: --
Least count: -- ID No: MVP02
Make: Yokogawa Location: Medical Lab
Details of Standard Used:
Sr.No. Name Sr. No/ I.D. No. Valid Up to
1 Multifunction calibrator F105318 27/Jul/20
2 Digital Ocilloscope C011303 26/Dec/17
3 Digital Multimeter (6½) 1613015 26/Feb/20

Environmental Temperature: Start: 25 °C End: 25 °C


Condition: Humidity: Start: 52% End: 52%
Work Instruction /IS No. WI-ET-11 &05
OBSERVATIONS
SL NO PARAMETER / STANDARD EXPECTED MEASURED
RANGE INPUT OUTPUT OUTPUT
V V V V V
1 DC VOLTAGE 1.000000 0.1000 0.100000 0.100000 0.100000
2 600 V 50.000000 5.0000 5.015640 5.015640 5.015640
3 100.000000 10.0000 10.023000 10.023000 10.023000
4 200.000000 20.0000 20.130000 20.130000 20.130000
5 300.000000 30.0000 30.300000 30.300000 30.300000
6 500.000000 50.0000 50.500000 50.500000 50.500000
7 600.000000 60.0000 60.700000 60.700000 60.700000
V V V V V
8 AC VOLTAGE 1.000000 0.1000 0.100718 0.100718 0.100718
9 600 V @ 50Hz 50.000000 5.0000 5.032240 5.032240 5.032240
10 100.000000 10.0000 10.066500 10.066500 10.066500
11 200.000000 20.0000 20.130100 20.130100 20.130100
12 300.000000 30.0000 30.195100 30.195100 30.195100
13 500.000000 50.0000 50.328700 50.328700 50.328700
14 600.000000 60.0000 60.394200 60.394200 60.394200
V V V V V
15 AC VOLTAGE 1.000000 0.1000 0.100591 0.100591 0.100591
16 600 V @ 60Hz 50.000000 5.0000 5.030520 5.030520 5.030520
17 100.000000 10.0000 10.062500 10.062500 10.062500
18 200.000000 20.0000 20.120300 20.120300 20.120300
19 300.000000 30.0000 30.180700 30.180700 30.180700
20 500.000000 50.0000 50.305600 50.305600 50.305600
21 600.000000 60.0000 60.368400 60.368400 60.368400
SL NO STANDARD TES T
READING READING
MHz dB dB dB
22 100.00 -2.64 -2.64 -2.64
23 200 -2.78 -2.78 -2.78

Calibrated by: Approved By:


Name & Signature: Name & Signature:
FR-GR-01
Issue No.01,Date: 16.05.2015
Rev. No. 00 , Date:--