UUC Details:

Instrument Name: Multifunction Calibrator Range: 0 -20 V Date: 4/24/2020


Make/Model: Fluke/754 ID No:
Least Count: 0.1 Type: Digital 2
STANDARDS DETAILS
Instrument Name: Multifunction Calibrator Traceability:
Range: 0 - 1000 V Due on:
Resolution: 0.10 V Sr.No
Unc in Measurement: 0.01 V RDG %+D:
Drift / Accuracy : 0.02 V ( RDG %+D)

Environmental Conditions: Avg. Temp: 24.3 °C Avg. Humidity: 54 % RH

TYPE:A Maximum Range: 1000 Reading Taken on: 5 Type A Results are in
Step:1
Test Readings X1 X2 X3 X4 X5 Mean 5.1
5.1 5.1 5.1 5.1 5

Mean Ẍ V
Step:2 Xi Mean Xi -Mean (Xi -Mean)2
Standard Deviation 5.1 5.1 0.02 0.000400
5.1 5.1 0.02 0.000400
S.D = 0.044721 V 5.1 5.1 0.02 0.000400
5.1 5.1 0.02 0.000400
5 5.1 -0.08 0.006400
Total 0.008000
Step:3

Type A= 0.022361 V
Step:4
Degree of Freedom for Type A
DOF= 4
TYPE:B
Step:5
Uncertainity from Certificate (Master) b1= 0.005 V
Uncertainity due to Accuracy of master(Master) b2= 0.011547 V
Uncertainity due to Test Resolution (Test) b3= 0.029 V
Step:6
Combined Uncertainty Uc = 0.038622
Step:7
Effective Degree of freedom Veff= 35.60111
Step:8
Expanded Uncertainity
U= 0V
1+2D V

Air Pressure: --

e A Results are in V

n 5
n-1 4 2

35.60111

0.05 Digital 50% 2


Analog 20% 5