Professional Documents
Culture Documents
,
Calibration Services, No.1 SIDCO
Industrial Estate, Ambattur, Chennai.
CERTIFICATE OF CALIBRATION Tamil Nadu - 600098. India.
Dir.: +91-44-66544651/52/53
FR-QT-21
Certificate No : C-200605-7-20
Date of Calibration : 05-Jun-20 Recom. Due Date : 05-Jun-21
Customer Details : SRF No. : 200605-7
M/s. UL India Pvt. Ltd., Laboratory Building, Calibrated at : Onsite
Sy.No.129/4, Kalyani Platina, Phase II, Date of Receipt : 05-Jun-20
EPIP Zone, White Field, Bangalore-560066. Cond. On Receipt : Satisfactory
Page 1 of 16
Godrej & Boyce Mfg. Co. Ltd.
Godrej & Boyce Mfg. Co. Ltd.,
Calibration Services, No.1 SIDCO
Industrial Estate, Ambattur, Chennai.
CERTIFICATE OF CALIBRATION Tamil Nadu - 600098. India.
Dir.: +91-44-66544651/52/53
Certificate No : C-200605-7-20
ULR NO : CC255920000010864 F
Calibration Results
1. Repeatability Test :
Sl. No 50% of Range 100% of Range
g g
1 49.95 99.93
2 49.95 99.93
3 49.95 99.93
4 49.95 99.93
5 49.95 99.93
6 49.95 99.93
7 49.95 99.93
8 49.95 99.93
9 49.95 99.93
10 49.95 99.93
Mean 49.95 99.93
Standard Deviation 0.00 0.00
Max. Difference 0.00 0.00
2. Linearity Test :
S.No. Applied Mass ( g ) Test Reading ( g ) Error ( g ) Expanded Uncertainty ± ( g )
1 0.099999 0.10 0.000001 0.0012
2 0.200007 0.20 -0.000007 0.0012
3 0.500000 0.50 0.000000 0.0012
4 1.000000 1.00 0.000000 0.0012
5 2.000060 2.00 -0.000060 0.0012
6 4.999990 5.00 0.000010 0.0012
7 9.999870 9.99 -0.009870 0.0012
8 20.000100 19.98 -0.020100 0.0012
9 50.000140 49.95 -0.050140 0.0012
10 100.000130 99.93 -0.070130 0.0012
11 149.999970 149.85 -0.149970 0.0012
Page 2 of 16
Godrej & Boyce Mfg. Co. Ltd.
Godrej & Boyce Mfg. Co. Ltd.,
Calibration Services, No.1 SIDCO
Industrial Estate, Ambattur, Chennai.
CERTIFICATE OF CALIBRATION Tamil Nadu - 600098. India.
Dir.: +91-44-66544651/52/53
Certificate No : C-200605-7-20
ULR NO : CC255920000010864 F
Calibration Results
3. Eccentricity Test :
Page 3 of 16
Godrej & Boyce Mfg. Co. Ltd.
FR-QT-21 Page 1 of 3
Certificate No : C-200605-7-21 Date of Issue : 08-Jun-20
Date of Calibration : 05-Jun-20 Recom. Due Date : 05-Jun-21
Customer Details : SRF No. : 200605-7
M/s. UL India Pvt. Ltd., Laboratory Building, Calibrated at : Onsite
Sy.No.129/4, Kalyani Platina, Phase II, Date of Receipt : 05-Jun-20
EPIP Zone, White Field, Bangalore-560066. Cond. On Receipt : Satisfactory
Remarks
1. Overall uncertainy ± 0.0024 g
2. This calibration certificate will not be legal for the purpose of the Standard of
“Weights & Measure (enforcement) act 1985”.
3. This calibration certificate is issued as per customer request.
4. The reported Expanded Uncertainty is calculated at 95.45 % C.L with coverage factor k=2
5. This calibration certificate is valid for scientific & industrial purpose only.
2. Linearity Test :
S.No. Applied Mass ( g ) Test Reading ( g ) Error ( g ) Expanded Uncertainty ± ( g )
1 0.005000 0.0050 0.000000 0.0012
2 0.010009 0.0099 -0.000109 0.0012
3 0.019985 0.0200 0.000015 0.0012
4 0.049998 0.0501 0.000102 0.0012
5 0.099999 0.1002 0.000201 0.0012
6 0.200007 0.2000 -0.000007 0.0012
7 0.500000 0.5001 0.000100 0.0012
8 1.000000 0.9998 -0.000200 0.0012
9 2.000060 1.9995 -0.000560 0.0012
10 4.999990 4.9999 -0.000090 0.0012
11 9.999870 9.9995 -0.000370 0.0012
12 20.000100 19.9994 -0.000700 0.0012
13 50.000140 49.9998 -0.000340 0.0012
14 100.000130 100.0009 0.000770 0.0012
15 200.000140 200.0006 0.000460 0.0012
16 219.999930 220.0010 0.001070 0.0012
Remarks
1. Overall uncertainy ± 0.0018 g
2. This calibration certificate will not be legal for the purpose of the Standard of
“Weights & Measure (enforcement) act 1985”.
3. This calibration certificate is issued as per customer request.
4. The reported Expanded Uncertainty is calculated at 95.45 % C.L with coverage factor k=2
5. This calibration certificate is valid for scientific & industrial purpose only.
2. Linearity Test :
S.No. Applied Mass ( g ) Test Reading ( g ) Error ( g ) Expanded Uncertainty ± ( g )
1 0.005000 0.0049 -0.000100 0.0012
2 0.010009 0.0098 -0.000209 0.0012
3 0.019985 0.0202 0.000215 0.0012
4 0.049998 0.0500 0.000002 0.0012
5 0.099999 0.1001 0.000101 0.0012
6 0.200007 0.2000 -0.000007 0.0012
7 0.500000 0.4997 -0.000300 0.0012
8 1.000000 0.9998 -0.000200 0.0012
9 2.000060 2.0000 -0.000060 0.0012
10 4.999990 4.9999 -0.000090 0.0012
11 9.999870 9.9999 0.000030 0.0012
12 20.000100 19.9998 -0.000300 0.0012
13 50.000140 49.9995 -0.000640 0.0012
14 100.000130 99.9990 -0.001130 0.0012
15 200.000140 199.9998 -0.000340 0.0012
16 220.000240 219.9998 -0.000440 0.0126
Remarks
1. Overall uncertainy ± 0.0146 g
2. This calibration certificate will not be legal for the purpose of the Standard of
“Weights & Measure (enforcement) act 1985”.
3. This calibration certificate is issued as per customer request.
4. The reported Expanded Uncertainty is calculated at 95.45 % C.L with coverage factor k=2
5. This calibration certificate is valid for scientific & industrial purpose only.
2. Linearity Test :
S.No. Applied Mass ( g ) Test Reading ( g ) Error ( g ) Expanded Uncertainty ± ( g )
1 0.005000 0.0051 0.000100 0.0012
2 0.010009 0.0101 0.000091 0.0012
3 0.019985 0.0199 -0.000085 0.0012
4 0.049998 0.0501 0.000102 0.0012
5 0.099999 0.1000 0.000001 0.0012
6 0.200007 0.2000 -0.000007 0.0012
7 0.500000 0.5002 0.000200 0.0012
8 1.000000 1.0001 0.000100 0.0012
9 2.000060 2.0003 0.000240 0.0012
10 4.999990 5.0002 0.000210 0.0012
11 9.999870 9.9998 -0.000070 0.0012
12 20.000100 20.0004 0.000300 0.0012
13 50.000140 50.0008 0.000660 0.0012
14 100.000130 99.9999 -0.000230 0.0012
15 200.000140 200.0017 0.001560 0.0012
16 220.000240 220.0021 0.001860 0.0126