Godrej & Boyce Mfg. Co. Ltd.

,
Calibration Services, No.1 SIDCO
Industrial Estate, Ambattur, Chennai.
CERTIFICATE OF CALIBRATION Tamil Nadu - 600098. India.
Dir.: +91-44-66544651/52/53

FR-QT-21
Certificate No : C-200605-7-20
Date of Calibration : 05-Jun-20 Recom. Due Date : 05-Jun-21
Customer Details : SRF No. : 200605-7
M/s. UL India Pvt. Ltd., Laboratory Building, Calibrated at : Onsite
Sy.No.129/4, Kalyani Platina, Phase II, Date of Receipt : 05-Jun-20
EPIP Zone, White Field, Bangalore-560066. Cond. On Receipt : Satisfactory

Details of Test Instrument:


Description : Fabric Weight Digital Balance Model No. : G204
Range : 0 g to 150 g Serial No : 150L0019
Least Count : 0.01 g Identification No. : WEI 01
Make : SDL Atlas Accuracy : As Per Manual
Location : Physical Lab
Details of Standard Used :
Name Id No. Valid upto Traceability
Weights QSC/M/01 26/Oct/20 LCGC, Hyderabad

Work Instruction : WI-MS-01


Environmental Details : Temperature : 25±5°C Relative Humidity : 50±15 % RH
Remarks:
1. Equipment used for calibration were calibrated & traceable to National & International Standards.
2. This report refers only to particular item (s) submitted for calibration.
3. Godrej Calibration Services is not liable for any change in calibration data & performance specifications on account
of malfunctioning of Standards / Instruments / Equipment covered by this report due to damage caused to it after
issuance of this report.
4. The calibration results reported are valid at the time of and under the stated conditions of the measurements.
5. This report shall not be reproduced in full / part without prior permission of Godrej Calibration Services.
6. Satisfactory calibration report in no ways implies that the equipment calibrated is approved by NABL.
7. All precautions have been taken for any error or omission while calibrating the instruments and issuing this
certificate. However Godrej Calibration Services shall not be liable for any loss or liability that may be arise to any
party in this regards.
8. Readings given are average of 3 Measurements.
9. The reported Expanded Uncertainty is calculated at 95.45 % Confidence level with coverage factor k=2.
10.Overall Uncertainty = ± 0.0018g

AUTHORIZED SIGNATORY : P.Muthukumar


DESIGNATION : SM
CALIBRATED BY : Lab Engineer : Chethan K N
ULR NO : CC255920000010864 F

Page 1 of 16
Godrej & Boyce Mfg. Co. Ltd.
Godrej & Boyce Mfg. Co. Ltd.,
Calibration Services, No.1 SIDCO
Industrial Estate, Ambattur, Chennai.
CERTIFICATE OF CALIBRATION Tamil Nadu - 600098. India.
Dir.: +91-44-66544651/52/53

Certificate No : C-200605-7-20
ULR NO : CC255920000010864 F
Calibration Results
1. Repeatability Test :
Sl. No 50% of Range 100% of Range
g g
1 49.95 99.93
2 49.95 99.93
3 49.95 99.93
4 49.95 99.93
5 49.95 99.93
6 49.95 99.93
7 49.95 99.93
8 49.95 99.93
9 49.95 99.93
10 49.95 99.93
Mean 49.95 99.93
Standard Deviation 0.00 0.00
Max. Difference 0.00 0.00

2. Linearity Test :
S.No. Applied Mass ( g ) Test Reading ( g ) Error ( g ) Expanded Uncertainty ± ( g )
1 0.099999 0.10 0.000001 0.0012
2 0.200007 0.20 -0.000007 0.0012
3 0.500000 0.50 0.000000 0.0012
4 1.000000 1.00 0.000000 0.0012
5 2.000060 2.00 -0.000060 0.0012
6 4.999990 5.00 0.000010 0.0012
7 9.999870 9.99 -0.009870 0.0012
8 20.000100 19.98 -0.020100 0.0012
9 50.000140 49.95 -0.050140 0.0012
10 100.000130 99.93 -0.070130 0.0012
11 149.999970 149.85 -0.149970 0.0012

Page 2 of 16
Godrej & Boyce Mfg. Co. Ltd.
Godrej & Boyce Mfg. Co. Ltd.,
Calibration Services, No.1 SIDCO
Industrial Estate, Ambattur, Chennai.
CERTIFICATE OF CALIBRATION Tamil Nadu - 600098. India.
Dir.: +91-44-66544651/52/53

Certificate No : C-200605-7-20
ULR NO : CC255920000010864 F
Calibration Results
3. Eccentricity Test :

S.No. Position of Test Reading Error between centre(3)


Weights used (g) and Other points ( g )
1 Top Left 49.95 0.00
2 Top Right 49.94 0.01
3 Centre 49.95 0.00
4 Bottom Left 49.95 0.00
5 Bottom Right 49.95 0.00
Maximum Error between centre (1) and other points = 0.01 g

Page 3 of 16
Godrej & Boyce Mfg. Co. Ltd.
FR-QT-21 Page 1 of 3
Certificate No : C-200605-7-21 Date of Issue : 08-Jun-20
Date of Calibration : 05-Jun-20 Recom. Due Date : 05-Jun-21
Customer Details : SRF No. : 200605-7
M/s. UL India Pvt. Ltd., Laboratory Building, Calibrated at : Onsite
Sy.No.129/4, Kalyani Platina, Phase II, Date of Receipt : 05-Jun-20
EPIP Zone, White Field, Bangalore-560066. Cond. On Receipt : Satisfactory

Details of Test Instrument:


Description : Digital Weighing Balance Model No. : BSA 224S
Range : 0 g to 220 g Serial No : 32890312
Least Count : 0.0001 g Identification No. : WEI 02
Make : Sartorius Accuracy : As Per Manual
Location : Physical Lab
Details of Standard Used :
Name Id No. Valid upto Traceability
Weights QSC/M/01 26/Oct/20 LCGC, Hyderabad

Work Instruction : WI-MS-01


Environmental Details : Temperature : 25±5°C Relative Humidity : 50±15 % RH

Remarks
1. Overall uncertainy ± 0.0024 g
2. This calibration certificate will not be legal for the purpose of the Standard of
“Weights & Measure (enforcement) act 1985”.
3. This calibration certificate is issued as per customer request.
4. The reported Expanded Uncertainty is calculated at 95.45 % C.L with coverage factor k=2
5. This calibration certificate is valid for scientific & industrial purpose only.

AUTHORIZED SIGNATORY : P.Muthukumar


DESIGNATION : SM
CALIBRATED BY : Lab Engineer : Chethan K N
ULR NO : CC255920000010865 F
Certificate No : C-200605-7-21 Page 2 of 3
Calibration Results
1. Repeatability Test :
Sl. No 50% of Range 100% of Range
g g
1 100.0009 200.0005
2 100.0010 200.0006
3 100.0009 200.0006
4 100.0009 200.0005
5 100.0010 200.0004
6 100.0009 200.0004
7 100.0010 200.0006
8 100.0009 200.0004
9 100.0009 200.0005
10 100.0009 200.0005
Mean 100.0009 200.0005
Standard Deviation 0.0000 0.0001
Max. Difference 0.0001 0.0002

2. Linearity Test :
S.No. Applied Mass ( g ) Test Reading ( g ) Error ( g ) Expanded Uncertainty ± ( g )
1 0.005000 0.0050 0.000000 0.0012
2 0.010009 0.0099 -0.000109 0.0012
3 0.019985 0.0200 0.000015 0.0012
4 0.049998 0.0501 0.000102 0.0012
5 0.099999 0.1002 0.000201 0.0012
6 0.200007 0.2000 -0.000007 0.0012
7 0.500000 0.5001 0.000100 0.0012
8 1.000000 0.9998 -0.000200 0.0012
9 2.000060 1.9995 -0.000560 0.0012
10 4.999990 4.9999 -0.000090 0.0012
11 9.999870 9.9995 -0.000370 0.0012
12 20.000100 19.9994 -0.000700 0.0012
13 50.000140 49.9998 -0.000340 0.0012
14 100.000130 100.0009 0.000770 0.0012
15 200.000140 200.0006 0.000460 0.0012
16 219.999930 220.0010 0.001070 0.0012

AUTHORIZED SIGNATORY : P.Muthukumar


DESIGNATION : SM
CALIBRATED BY : Lab Engineer : Chethan K N
ULR NO : CC255920000010865 F
Certificate No : C-200605-7-21 Page 3 of 3
Calibration Results
3. Eccentricity Test :

S.No. Position of Test Reading Error between centre(3)


Weights used (g) and Other points ( g )
1 Top Left 100.0009 0.0000
2 Top Right 100.0009 0.0001
3 Centre 100.0009 0.0000
4 Bottom Left 100.0008 0.0001
5 Bottom Right 100.0008 0.0001
Maximum Error between centre (1) and other points = 0.0001 g

AUTHORIZED SIGNATORY : P.Muthukumar


DESIGNATION : SM
CALIBRATED BY : Lab Engineer : Chethan K N
ULR NO : CC255920000010865 F
FR-QT-21 Page 1 of 3
Certificate No : C-200605-7-22 Date of Issue : 08-Jun-20
Date of Calibration : 05-Jun-20 Recom. Due Date : 05-Jun-21
Customer Details : SRF No. : 200605-7
M/s. UL India Pvt. Ltd., Laboratory Building, Calibrated at : Onsite
Sy.No.129/4, Kalyani Platina, Phase II, Date of Receipt : 05-Jun-20
EPIP Zone, White Field, Bangalore-560066. Cond. On Receipt : Satisfactory

Details of Test Instrument:


Description : Digital Weighing Balance Model No. : BSA 224S
Range : 0 g to 220 g Serial No : 32890313
Least Count : 0.0001 g Identification No. : WEI 03
Make : Sartorius Accuracy : ± 0.01 g
Location : Colour Fastness Lab
Details of Standard Used :
Name Id No. Valid upto Traceability
Weights QSC/M/01 26/Oct/20 LCGC, Hyderabad

Work Instruction : WI-MS-01


Environmental Details : Temperature : 25±5°C Relative Humidity : 50±15 % RH

Remarks
1. Overall uncertainy ± 0.0018 g
2. This calibration certificate will not be legal for the purpose of the Standard of
“Weights & Measure (enforcement) act 1985”.
3. This calibration certificate is issued as per customer request.
4. The reported Expanded Uncertainty is calculated at 95.45 % C.L with coverage factor k=2
5. This calibration certificate is valid for scientific & industrial purpose only.

AUTHORIZED SIGNATORY : P.Muthukumar


DESIGNATION : SM
CALIBRATED BY : Lab Engineer : Chethan K N
ULR NO : CC255920000010866 F
Certificate No : C-200605-7-22 Page 2 of 3
Calibration Results
1. Repeatability Test :
Sl. No 50% of Range 100% of Range
g g
1 99.9986 200.0001
2 99.9985 200.0002
3 99.9986 200.0001
4 99.9986 200.0002
5 99.9985 200.0002
6 99.9985 200.0002
7 99.9985 200.0001
8 99.9986 200.0002
9 99.9985 200.0002
10 99.9986 200.0002
Mean 99.9986 200.0002
Standard Deviation 0.0001 0.0000
Max. Difference 0.0001 0.0001

2. Linearity Test :
S.No. Applied Mass ( g ) Test Reading ( g ) Error ( g ) Expanded Uncertainty ± ( g )
1 0.005000 0.0049 -0.000100 0.0012
2 0.010009 0.0098 -0.000209 0.0012
3 0.019985 0.0202 0.000215 0.0012
4 0.049998 0.0500 0.000002 0.0012
5 0.099999 0.1001 0.000101 0.0012
6 0.200007 0.2000 -0.000007 0.0012
7 0.500000 0.4997 -0.000300 0.0012
8 1.000000 0.9998 -0.000200 0.0012
9 2.000060 2.0000 -0.000060 0.0012
10 4.999990 4.9999 -0.000090 0.0012
11 9.999870 9.9999 0.000030 0.0012
12 20.000100 19.9998 -0.000300 0.0012
13 50.000140 49.9995 -0.000640 0.0012
14 100.000130 99.9990 -0.001130 0.0012
15 200.000140 199.9998 -0.000340 0.0012
16 220.000240 219.9998 -0.000440 0.0126

AUTHORIZED SIGNATORY : P.Muthukumar


DESIGNATION : SM
CALIBRATED BY : Lab Engineer : Chethan K N
ULR NO : CC255920000010866 F
Certificate No : C-200605-7-22 Page 3 of 3
Calibration Results
3. Eccentricity Test :

S.No. Position of Test Reading Error between centre(3)


Weights used (g) and Other points ( g )
1 Top Left 99.9986 0.0000
2 Top Right 99.9985 0.0001
3 Centre 99.9986 0.0000
4 Bottom Left 99.9986 0.0000
5 Bottom Right 99.9986 0.0000
Maximum Error between centre (1) and other points = 0.0001 g

AUTHORIZED SIGNATORY : P.Muthukumar


DESIGNATION : SM
CALIBRATED BY : Lab Engineer : Chethan K N
ULR NO : CC255920000010866 F
FR-QT-21 Page 1 of 3
Certificate No : C-200605-7-23 Date of Issue : 09-Jun-20
Date of Calibration : 06-Jun-20 Recom. Due Date : 06-Jun-21
Customer Details : SRF No. : 200605-7
M/s. UL India Pvt. Ltd., Laboratory Building, Calibrated at : Onsite
Sy.No.129/4, Kalyani Platina, Phase II, Date of Receipt : 06-Jun-20
EPIP Zone, White Field, Bangalore-560066. Cond. On Receipt : Satisfactory

Details of Test Instrument:


Description : Digital Weighing Balance Model No. : BSA 224S-CW
Range : 0 g to 220 g Serial No : 29392276
Least Count : 0.0001 g Identification No. : WEI 04
Make : Sartorius Accuracy : As Per Manual
Location : Analytical Wet Lab
Details of Standard Used :
Name Id No. Valid upto Traceability
Weights QSC/M/01 26/Oct/20 LCGC, Hyderabad

Work Instruction : WI-MS-01


Environmental Details : Temperature : 25±5°C Relative Humidity : 50±15 % RH

Remarks
1. Overall uncertainy ± 0.0146 g
2. This calibration certificate will not be legal for the purpose of the Standard of
“Weights & Measure (enforcement) act 1985”.
3. This calibration certificate is issued as per customer request.
4. The reported Expanded Uncertainty is calculated at 95.45 % C.L with coverage factor k=2
5. This calibration certificate is valid for scientific & industrial purpose only.

AUTHORIZED SIGNATORY : P.Muthukumar


DESIGNATION : SM
CALIBRATED BY : Lab Engineer : Chethan K N
ULR NO : CC255920000010867 F
Certificate No : C-200605-7-23 Page 2 of 3
Calibration Results
1. Repeatability Test :
Sl. No 50% of Range 100% of Range
g g
1 100.0003 200.0018
2 100.0002 200.0017
3 100.0003 200.0017
4 100.0003 200.0017
5 100.0002 200.0018
6 100.0003 200.0018
7 100.0003 200.0017
8 100.0002 200.0017
9 100.0002 200.0018
10 100.0002 200.0017
Mean 100.0003 200.0017
Standard Deviation 0.0001 0.0001
Max. Difference 0.0001 0.0001

2. Linearity Test :
S.No. Applied Mass ( g ) Test Reading ( g ) Error ( g ) Expanded Uncertainty ± ( g )
1 0.005000 0.0051 0.000100 0.0012
2 0.010009 0.0101 0.000091 0.0012
3 0.019985 0.0199 -0.000085 0.0012
4 0.049998 0.0501 0.000102 0.0012
5 0.099999 0.1000 0.000001 0.0012
6 0.200007 0.2000 -0.000007 0.0012
7 0.500000 0.5002 0.000200 0.0012
8 1.000000 1.0001 0.000100 0.0012
9 2.000060 2.0003 0.000240 0.0012
10 4.999990 5.0002 0.000210 0.0012
11 9.999870 9.9998 -0.000070 0.0012
12 20.000100 20.0004 0.000300 0.0012
13 50.000140 50.0008 0.000660 0.0012
14 100.000130 99.9999 -0.000230 0.0012
15 200.000140 200.0017 0.001560 0.0012
16 220.000240 220.0021 0.001860 0.0126

AUTHORIZED SIGNATORY : P.Muthukumar


DESIGNATION : SM
CALIBRATED BY : Lab Engineer : Chethan K N
ULR NO : CC255920000010867 F
Certificate No : C-200605-7-23 Page 3 of 3
Calibration Results
3. Eccentricity Test :

S.No. Position of Test Reading Error between centre(3)


Weights used (g) and Other points ( g )
1 Top Left 100.0003 0.0000
2 Top Right 100.0003 0.0000
3 Centre 100.0003 0.0000
4 Bottom Left 100.0003 0.0000
5 Bottom Right 100.0002 0.0001
Maximum Error between centre (1) and other points = 0.0001 g

AUTHORIZED SIGNATORY : P.Muthukumar


DESIGNATION : SM
CALIBRATED BY : Lab Engineer : Chethan K N
ULR NO : CC255920000010867 F