Prima Calibration Services


F.199, MANGAL BAZAR LAXMI NAGAR DELHI.I1OO92
Ph. : 0 1 1 -42657 237 Mob. : 981 0804449, 8368671 530
ffi
\&\4/,y
}K
E- mail : primac2l @gmail.com cc-2339

Format No.
F01(7.8/01)

t
:
ULR No cc233920000001 303F Calibration Done on 08.08.2020 SRF.No- &Date : 0001/08 & 08.08.2020
lssue Date 10.08.2020 Calibration Valid Upto 07.08.2021 Page : 1of1
UUC Receipt Date:
Calibrated for M/S.
Punjab Genral Industries Pvt. Ltd.
Plot No-149-150, Sec -24,
Faridabad-1 21 005 (Haryana)

lnstrument Name Dig.Weighing Balance Size/Range Upto 100 Kg


Least Count 0.00s Kg Calibration ltem I Good
IVlanufactured By lVarun lCalibration
performed at
lVlanufacturer Sr. No Location Railway Broduction Office
Party lD tMark WB-o1 Ref. Standard/Procedure orML-R76-1

Calibration Procedure / Method : As Calibration Procedure Document No. PCS/CP/01 Section No.CP-S -03

Erivi ronmental condition Tem 30t C. Relative Humid 55+/-15% RH


I
t Standard & Used
Standard Calibrating Traceability Cedificate Valid
Make / Range
Equipments Agency Cert. No. Date Up to
Standard Weight 'l mg to 200 g wMcLtEt2020-
Weightronics 03.0612020 02.06.2023
06/3369

500 9,1 kg, 2kg, 5 kg, 10 wMcL/M/2020-


Standard Weight Weightronics 03.06.2020 02.06 2022
kg, 20 kg (7 Nos) 06i3368
I
Calibration Results:
Related Uncer.
Standard Value Observed Value on C.F.(K=2) at 95%
(Ks) Error (Kg)
on UUC(Kg) Confidence Level
(ts)
0.5 0.500 0.000
1.0 1.000 0.000
2.0 2 000 0.000
5.0 s.000 0.000
'10.0 10.000 0.000 18
20.0 20.000 0.000
40.0 39.995 -0.005
50.0 49.995 -0.005
80.0 79.990 -0 010
100.0 99.98s -0.015
ll Load at 100 Kg = 0.003 Kg
Repeatibility Test
Half Load at 50 Kg = 0.002 Kg
Off - Centre Load at 30 Kg = 0.005 Kg

NOTE:
************
1. The result are valid only for this Calibrateilfttifii:** END THE REPORT
2, The calibration report shall nol be reproduced without the wrilten approval of calibration laboratory.
3. Result reported are valid at the time of & under the stated conditions of measurement
4. Laboratory standards are traceable to National/lnternational Standards.
5. Calibration Certificates issued for Weight & Measure parameters i.e;,Mass, Balance, Volumelric
equipment, Measuring Scales / Tapes etc. are for scientific pqrpose only and should not be used for Trade / Gommercial use.

CALIBRATED BY _4iL
APPROVED BY

Jw
(CALIBRATION ENGINEER)
ANWARA. SIDDIQUI