Godrej & Boyce Mfg. Co. Ltd.

,
CALIBRATION CERTIFICATE Calibration Services, No.1 SIDCO
Industrial Estate, Ambattur, Chennai.
Tamil Nadu - 600098. India.
Dir.: +91-44-66544651/52/53

FR-QT-21
Certificate No : C-220519-5-2 Date of Issue : 21-May-22
Date of Calibration : 19-May-22 Recom. Due Date : 19-May-23
Customer Details : Receipt No. : C-220519-5
IAI MALAT DIVISION., Challan No : Lab
C/O Elcom Systems Pvt Ltd,SP 86-88, RICO Industrial Area, Date of Receipt : 19-May-22
Bhiwadi,Alwar, Cond. On Receipt : Satisfactory
Rajasthan-301019.
Details of Machine Under Calibration:
Description : Wireless Weighing machine With Indicator Sr. No. : 0311PK2001
Max. Capacity : 700 kg Model : 100406-RFX
Make : Intercomp Mode : Compression
Load cell Sl.No : 0210LB202272 Channel : 2

Details of Force Proving Instrument (Master Load cell )Used For Calibration:
Name Load Cell with Indicator
ID No QSC/F/06
Make/Model/capacity Sensortron / SEN-10KN-C / 10KN
Certificate Number TP/FCM/2021/004
Due Date 3-Nov-22
Uncertainty in % at 95% C.L @k=2 0.11
Traceability Techno Products
Calibrated at Temp 26 °C

Work Instruction No: WI-FC-01,02 IS Std: 1828(Part1):2015 /ISO 7500-1


Environmental Details: Temperature : 26.9 °C Relative Humidity: 57% RH

AUTHORIZED SIGNATORY : P. Muthukumar


DESIGNATION : S M
CALIBRATED BY : Lab Engineer : T. Krishnan
ULR No.: CC255922000012167 N

Godrej & Boyce Mfg. Co. Ltd.


Regd. Office: Pirojshanagar, L.B.S. Marg, Vikhroli, Mumbai - 400 079.T.:+91-22-67961700/1800 W:www.godrej.com
(This document is digitally signed on first page and does not require signature) Page 1 of 2
Godrej & Boyce Mfg. Co. Ltd.,
CALIBRATION CERTIFICATE Calibration Services, No.1 SIDCO
Industrial Estate, Ambattur, Chennai.
Tamil Nadu - 600098. India.
Dir.: +91-44-66544651/52/53

FR-QT-21 ULR No.: CC255922000012167 N


Certificate No : C-220519-5-2 Date of Issue : 21-May-22
Date of Calibration : 19-May-22 Recom. Due Date : 19-May-23
Calibration Results (Mechanical - Force)
Load cell (0210LB202274)
Mode Load Cell Range : 700 kg Least Count Temperature in °C: 26.9
Compession Certified range : 700 kg 0.5 kg Humidity in % RH: 57
Force Expanded
Observed Observed Observed Avg. Test
Proving Force Applied Std In Equvalent Force Uncert-
Test Reading Test Reading in Test Reading Reading Error in kg
Instrum- N in Kg ainty %
in kg(1) kg (2) in kg(3) in -kg
ent Used (±)
686.5 70.0 70.0 70.0 70.0 70.0 0.00 0.30
10 kN Compression Load cell

1372.9 140.0 140.0 140.0 140.0 140.0 0.00 0.30


2059.4 210.0 210.0 210.0 210.0 210.0 0.00 0.30
2745.9 280.0 280.0 280.0 280.0 280.0 0.00 0.30
3432.3 350.0 350.0 350.0 350.0 350.0 0.00 0.30
4118.8 420.0 420.0 420.0 420.0 420.0 0.00 0.30
4805.3 490.0 490.0 490.0 490.0 490.0 0.00 0.30
5491.7 560.0 560.0 560.0 560.5 560.2 0.17 0.31
6178.2 630.0 630.5 630.5 630.5 630.5 0.50 0.41
6864.7 700.0 700.5 700.5 700.5 700.5 0.50 0.41

Load Test Nos 1 2 3 Relative zero error in %


Residual Indication on Force Indicator in kg 0.0 0.0 0.0 0.00
Relative Resolution error in %
Resolution of the force Indicator in kg 0.5
0.01

Remarks:
1. Equipment used for calibration were calibrated & traceable to National & International Standards.
2. This report refers only to particular item (s) submitted for calibration.
3. Godrej Calibration Services is not liable for any change in calibration data & performance specifications on account of malfunctioning of Standards /
Instruments / Equipment covered by this report due to damage caused to it after issuance of this report.
4. The calibration results reported are valid at the time of and under the stated conditions of the measurements.
5. This report shall not be reproduced in full / part without prior permission of Godrej Calibration Services.
6. Satisfactory calibration report in no ways implies that the equipment calibrated is approved by NABL.
7. All precautions have been taken for any error or omission while calibrating the instruments and issuing this certificate. However Godrej Calibration
Services shall not be liable for any loss or liability that may be arise to any party in this regards.
8. Readings given are average of 3 Measurements.
9. The reported Expanded Uncertainty is calculated at 95.45 % Confidence level with coverage factor k=2.
10.This report does not cover under the scope of Accreditation.

Godrej & Boyce Mfg. Co. Ltd.


Regd. Office: Pirojshanagar, L.B.S. Marg, Vikhroli, Mumbai - 400 079.T.:+91-22-67961700/1800 W:www.godrej.com
(This document is digitally signed on first page and does not require signature) Page 2 of 2