You are on page 1of 2

CERTIFICATE OF CALIBRATION

CITADEL ISO/IEC17025 ACCREDITION LABORATORY & API Q2 SPECS REGISTERED


Issued By: Citadel Company Certificate Number: CEC-2022/CAD/OS-0086
Date of Issue: 18-Jun-2022

Citadel Calibration Lab, Gazna Road, Erbil Kurdistan Region, Iraq


Phone Number: +964 -7506489336/+964 –7717023040
Website: www.citadel-engineering.com Email: info@citadel-engineering.com

Customer Name: OilServ Oilfield Services. Date of Calibration: 18-Jun-2022


Address: Kirkuk Road, South Industrial Area, Erbil, Kurdistan Region, Iraq. Due Date of Calibration: 17-Jun-2023

Description: Pressure Chart Recorder Model / Type: N/A


Manufacture: BARTON Serial No: 199-2120132761041
Location: N/A Tag No: N/A
Temperature: 20 °C ± 1°C Relative Humidity: 45 % RH ± 10% RH

Calibration Procedure

The Abovementioned Device Calibration According to The Validated CEC Method CEC-CAD-PRD-EN-0073, Calibration Procedure.

Traceability

The calibration results published in this certificate were obtained using equipment capable of producing results that are traceable
through UKAS to the International System of Units (SI)

Uncertainty
The reported expanded uncertainty is based on a standard uncertainty multiplied by a coverage factor k=2, providing a level of
confidence of approximately 95%. The uncertainty evaluation has been carried out in accordance with ISO/IEC GUIDE98-3:2008
requirements.
Standards Used
No. Description Model Manufacture Serial No. Cal.due date Certificate No.
Hydraulic Ultra High
1 ADT 949 ADDITEL 181A15030151
Calibration Pressure

2 Pressure Calibrator 721-3630 Fluke 2729038 05-Feb-2024 CEC-0042

Calibrated by: Reviewed by : Approved by :


Digitally Signed by Digitally Signed by Digitally Signed by
karkar M. Farooq S. Humam Isam
Date:18-Jun-2022 Date:18-Jun-2022 Date:18-Jun-2022

CECDOC SECURED
Please scan this QR code to verify this certificate on citadel-engineering.com only.

This certificate is issued in accordance with the laboratory accreditation requirements of the Iraqi Accreditation System (IQAS).It provides traceability of measurement to the SI
system of units and/or to units of measurement realized at the National Physical Laboratory or other recognized national metrology institutes. This certificate may not be
reproduced other than in full, except with the prior written approval of the issuing laboratory. The results contained within this certificate relate solely to the unit under test
identified on page one.

Doc No: CEC-CAD-FRM-EN-0119 Page 1 of 2 Revision Date: 09/20/2021


CERTIFICATE OF CALIBRATION
CITADEL ISO/IEC17025 ACCREDITION LABORATORY
Issued By: Citadel Company Certificate Number : CEC-2022/CAD/OS-0086
Date of Issue: 18-Jun-2022

AS-FOUND CALIBRATION RESULT

U.U.C Ref. Reading Ave.


Uncertainty
No. Reading Rising Falling Deviation Status
psi ± psi
1 0 0.0 0.0 0.0 8.7 Pass
2 300 292.4 292.1 7.8 8.7 Pass
3 600 591.8 591.6 8.3 8.7 Pass
4 900 891.5 891.3 8.6 8.7 Pass
5 1200 1186.5 1186.3 13.6 8.7 Pass
6 1500 1486.3 1486.3 13.7 8.7 Pass

Unit Under Calibration (U.U.C):


Range : 0 ~ 1500 psi
Resolution : 30 psi
Accuracy : ± 1.0% (x F.S) psi

End of Document

This certificate is issued in accordance with the laboratory accreditation requirements of the Iraqi Accreditation System (IQAS).It provides traceability
of measurement to the SI system of units and/or to units of measurement realised at the National Physical Laboratory or other recognised national
metrology institutes. This certificate may not be reproduced other than in full, except with the prior written approval of the issuing laboratory. The results
contained within this certificate relate solely to the unit under test identified on page one.
Doc No.: CEC-CAD-FRM-EN-0119 Page 2 of 2 Revision Date: 09/20/2021