Certificate of Calibration

Calibration Certificate No. 0104/T/2022


LB-CAL-062 TAMM Request No. 04-EMI-CAL-22-000809

Abhath Weights & Measurement Laboratory Digital thermometer


Customer Name of equipment

P.O. Box 48402,


Dubai, United Arab Emirates Fluke 1523-566-12
Customer Address Make/Model

The thermometer was calibrated by


comparison to a reference Platinum Resistance
Thermometer in accordance with EMI
procedure No: CP-T-01. 2001153 / 41748
Calibration method Serial number

Calibration results are given on Page 2 20 December 2022


Result of calibration Date received

The calibration was performed using


measurement standards which are traceable to
the International System of Units. The
measurement standards used are listed on
Page 2 30 December 2022
Traceability Calibration date

30 December 2022
Date issued

Note 1: This calibration certificate shall not be reproduced except in full, without the written consent of the
Emirates Metrology Institute.
Note 2: The reported expanded uncertainty of measurement is stated as the standard uncertainty of measurement
multiplied by the coverage factor k=2 such that the coverage probability corresponds to approximately
95%. The uncertainty evaluation has been carried out in accordance with JCGM 100:2008 Evaluation of
measurement data — Guide to the expression of uncertainty in measurement. Long term variations of the
item under calibration are not included in the uncertainty estimate.
Note 3: The result of calibration is valid only for the item submitted to the Emirates Metrology Institute.
Note 4: The certificate is consistent with the capabilities that are included in Appendix C of the MRA drawn up by
the International Committee for Weights and Measures (CIPM). Under the MRA, all participating institutes
measurement uncertainties specified in Appendix C (for detail see www.bipm.org).

Kryptolabs Building, Masdar City ∣ PO Box 853, Abu Dhabi, UAE


Tel: +971 (0)2 406 6666 ∣ Fax: +971 (0)2 406 6677
emi.customerservice@qcc.gov.ae ∣ https://emi.qcc.gov.ae
Page 1 of 2
Form : EMI-QF-18(A) Revision no. 11 Revision date: 18 October 2022
Calibration Certificate No. 0104/T/2022
TAMM Request No. 04-EMI-CAL-22-000809

Result of calibration
The temperature deviation of the units under test (UUT) from the reference thermometer are given in the
following table:

Reference Temperature UUT Display Temperature UUT Deviation Temperature Uncertainty


(˚C) (˚C) (˚C) (˚C)

-29.753 -29.736 0.017 0.02

-9.735 -9.717 0.018 0.02

0.000 0.018 0.018 0.02

49.860 49.880 0.020 0.02

99.981 99.983 0.002 0.03

200.043 200.037 -0.006 0.03

399.961 399.973 0.012 0.22

The measurement uncertainty referred in the table, is the combined expanded uncertainty (k=2) of the
measurement. It includes the uncertainty of the reference temperature and the uncertainty of the UUT.

The immersion of the thermometer in the calibration bath was approximately 28 cm.

Environmental conditions

From To
Air Temperature (°C) 22.0 24.0

Relative Humidity (%rh) 40.0 60.0

Measurement standards used

Name of Standard Make/Model Serial number Traceability

Platinum Resistance
Fluke 5628 4311 0020/T/2022, EMI, UAE
Thermometer
Super Thermometer Fluke 1595A B2B108 0121/E/2022, EMI, UAE

Ahmed Al Breiki Maryam Al Senaidi


Calibrated by Checked by Acting Head of Laboratory

End of Calibration Certificate

This calibration certificate shall not be reproduced except in full, without the written consent of the Emirates Metrology Institute.

Form : EMI-QF-18(A) Revision no. 11 Revision date: 18 October 2022 Page 2 of 2