You are on page 1of 3

Temperature Measurement

Master Reference
Range Run 1 Run 2 Run 3 Limits
As Found

THEODOR FRIEDRICHS
Page 1 of 3
CALIBRATION CERTIFICATE
Certificate Number: VCS2023-6602
VC-RF-T-05,REV.NO:03, REV.DATE : JUNE 2023

Customer Name : AAMAL CEMENT INDUSTRIES Delivery Date : 22-Jul-2023


Address : Doha Qatar Calibration Date : 20-Jul-2023
CSR Number : CSR/1137/01 Suggested Due Date : 19-Jul-2024
Date Received : 19-Jul-2023 Validity : 1 Year
Description : Digital Thermometer Received Condition : In Tolerance
Brand / Make : Testo Returned Condition : In Tolerance
Model / Type : 104 Calibration Result : Pass
Serial Number : 35745219 Location : Vertex Lab
Equipment ID Number : N/A Temperature : 23.6 °C
Size / Range : -50 to +250°C Relative Humidity : 40.5 %RH
Resolution/Least Count : 0.1°C Page Number : 1 of 2
Calibration Procedure : VC/CM/T/01

DETAILS OF MASTER INSTRUMENTS USED FOR CALIBRATION


DESCRIPTION ASSET NUMBER BRAND/MAKE MODEL/TYPE DUE DATE TRACEABILITY NUMBER
Dry Block Calibrator VER-CAL-DBC-001 GE Druck DRYTC650 4-Aug-2023 22070465
Dry Block Calibrator VER-CAL-DBC-002 GE Druck DRYTC165 12-Sep-2023 22091128
Barometric Pressure VER-CAL-BAR-001 Extech SD 700 12-Jun-2024 2022061129

Remarks:
1. Averages of minimum three readings are reported.
2. Suggested due date and test point were given based on customer requirements.
3. These results are obtained at the time of calibration.
4. The user should determine the suitability of the instrument for its intended use.
5. Uncertainty of measurement at 95% confidence level by multiplying coverage factor k = 2.
6. Condition of instrument found satisfactory during receipt.
7. Our masters are directly calibrated through ISO/IEC 17025 accredited calibration lab having direct traceability with national /
International standard.
8. Vertex Calibration Laboratory is accredited in accordance with the recognized International Standard as per ISO/IEC 17025:2017.
9. UUC- Unit Under Calibration
10.Pass / Fail criteria is decided by comparing the error of the unit under calibration and its associated by measurement uncertainty with
the tolerance specified by the manufacturer / customer.
11.Accuracy, tolerance and acceptance criteria stated as per customer requirements
12.Any hand written corrections (except @) or photocopies of the report invalidate this certificate.
13.Results reported outside the scope of accreditation are mark with asterisk (*).

Calibrated By: Checked By: Approved By:

Muhammed shihab Mohideen Maajith Ronaldo Acosta


Calibration Technician QA/QC Engineer Technical In-Charge

Calibration cycle are resulting due dates were submitted/approved by the customer.Any number of factors may cause an instrument to drift out of tolerance before the next scheduled of
calibration.Recalibration cycles should be based of frequency of used, environmental condition and customer's established systmatic accuracy.The information on this report, pertains only to the
instrument identified.
CALIBRATION RESULT
Certificate Number: VCS2023-6602
VC-RF-T-05,REV.NO:03, REV.DATE : JUNE 2023

Description : Digital Thermometer CSR Number : CSR/1137/01


Calibration Procedure : VC/CM/T/01 Calibration Date : 20-Jul-2023
Serail Number : 35745219 Suggested Due Date : 19-Jul-2024
Equipment ID Number : N/A Model / Type : 104
Brand / Make : Testo Size / Range : -50 to +250°C
Temperature : 23.6°C Location : Vertex Lab
Humidity : 40.5%RH Page Number : 2 of 2

Temperature Measurement
Master
Range Reference As UUC As Found UUC As Left Deviation Tolerance Result Uncertainty
Found
-50 to +250 °C -10.00 °C -10.1 °C -10.1 °C -0.10 °C ± 0.50 °C PASS ± 0.2500 °C
-50 to +250 °C 0.00 °C 0.2 °C 0.2 °C 0.20 °C ± 0.50 °C PASS ± 0.2500 °C
-50 to +250 °C 25.00 °C 24.9 °C 24.9 °C -0.10 °C ± 0.50 °C PASS ± 0.2500 °C
-50 to +250 °C 50.00 °C 49.9 °C 49.9 °C -0.10 °C ± 0.50 °C PASS ± 0.2500 °C
-50 to +250 °C 100.00 °C 99.9 °C 99.9 °C -0.10 °C ± 1.00 °C PASS ± 0.2500 °C
-50 to +250 °C 150.00 °C 149.7 °C 149.7 °C -0.30 °C ± 1.00 °C PASS ± 0.2500 °C
-50 to +250 °C 200.00 °C 200.1 °C 200.1 °C 0.10 °C ± 1.00 °C PASS ± 0.2500 °C
-50 to +250 °C 240.00 °C 240.1 °C 240.1 °C 0.10 °C ± 1.00 °C PASS ± 0.2500 °C
***End of Calibration***

Where statements of conformity are made in this report, the following decision rules are applied:
PASS - Results ± expanded uncertainty within limits / specification
FAIL - Results ± expanded uncertainty exceeds limit / specification
INDETERMINATE - Results ± expanded uncertainty overlaps limit / specification

Calibrated By: Checked By: Approved By:

Muhammed shihab Mohideen Maajith Ronaldo Acosta


Calibration Technician QA/QC Engineer Technical In-Charge