You are on page 1of 2

Form No. : SVCF2085 Revision Level : 0.

00

Certificate of Calibration Page 1 of 2

Alpha Technologies Services, USLP


6279 Hudson Crossing Parkway, #200
Hudson, Ohio 44236, USA
Phone : +1 330 745 1641
Fax : +1 330 848 7326
www.alpha-technologies.com

Certificate Type : Torque calibration Ticket No. : 283-18-001156


Certificate Number : DRA-283-18-001156
Applicable Standards : ASTM D2084-19 / ISO 3417 : 2008
Calibration Date : 19-Aug-2020 Due Date : 19-Feb-2021
(dd-mmm-yyyy) (dd-mmm-yyyy)

Customer

Company : Maquilados de Elastomeros


Address 1 : Calz. San Juan de Aragon
Address 2 : Col. Granjas Modernas
Zip : 7460 City : Gustavo A. Madero State : Ciudad de Mexico
Country : Mexico
Contact : Guillermo Andrade E-mail : j.calidad@mae-sa.com.mx

Customer specific name for location : Laboratorio

Instrument

Main Unit : ODR2000 Serial Number : 10SHH2027


Manufacturer : Alpha Technologies Building Year : 1988
Firmware Version : 6.00
Customer Asset No. : N/A

Output system :
Software Description : Compugraph Serial Number : N/A Version : 6.5.26
Device 1 Description : N/A Serial Number : N/A .
Device 2 Description : N/A Serial Number : N/A

Calibration Method
The above system (Instrument and Output Device(s), as shown above) has been calibrated in accordance with ASTM D5289-17 Standard Test
Method for Rubber Property - Vulcanization Using Rotorless Cure Meter, using apparatus calibrated and traceable to NIST ( National Institute
of Standards and Technology ).

National / International Standards


ASTM D2084-19 Standard Test Method for Rubber Property - Vulcanization Using Oscilating Disc Cure Meter
ISO 3417 : 2008 Rubber - Measurement of Vulcanization Characteristics with Oscilating Disc Rheometer

Calibrated By
The work has been done on site. See customer address for details.
The results indicated on this certificate relate only to the items calibrated.
This Certificate and attachments shall not be reproduced, except in full, without the written approval of Alpha Technologies.

The calibration has been performed following the Company ISO procedure : SVC0105

Authorized Field Service Engineer : David Ramirez

Date of Issue : 31-Aug-2020 Signature :


(dd-mmm-yyyy) E-mail : david.ramirez@alpha-technologies.com
Form No. : SVCF2085 Revision Level : 0.00

Certificate of Calibration Page 2 of 2

Alpha Technologies Services, USLP


6279 Hudson Crossing Parkway, #200
Hudson, Ohio 44236, USA
Phone : +1 330 745 1641
Fax : +1 330 848 7326
www.alpha-technologies.com

Certificate Type : Torque calibration Ticket No. : 283-18-001156


Certificate Number : DRA-283-18-001156

Calibration Conditions
Calibration Temperature : 177.0 Temperature Units : °C Temperature Tolerance : 0.5

Degree of Arc : 3.0 Torque Units : lbinch Ambient Temperature : 25 °C

Calibration Results
Reference Torque Standard Serial Number : T10S0085 Warmup time (min) : 30

Reference Torque Standard Value : 120.80 Units : lbinch

Pre-Calibration Recorded Value (as Found) : 121.85 Final Recorded Value (as Left) : 120.80

Torque Value Control Limits : 1% Calibration Status :


PASSED
Resolution of main indicator : 0.01

The expanded uncertainty expressed at approximately the 95% confidence level using a coverage factor of k=2 is 0 lbinch.

Verification Apparatus
Verification Apparatus : Reference Torque Standard

Manufacturer : Alpha Technologies

Model / Type : ODR Serial Number : T10S0085

Reference Value @ 177 °C : 120.80 lbinch Degree of Arc : 3

Calibration Date : 09-Mar-2020 Due Date : 09-Mar-2021


(dd-mmm-yyyy) (dd-mmm-yyyy)

Certificate Number : 2003ODR1319

Comments / Notes

End of the ISO17025 Calibration certificate