Customer
Instrument
Output system :
Software Description : Compugraph Serial Number : N/A Version : 6.5.26
Device 1 Description : N/A Serial Number : N/A .
Device 2 Description : N/A Serial Number : N/A
Calibration Method
The above system (Instrument and Output Device(s), as shown above) has been calibrated in accordance with ASTM D5289-17 Standard Test
Method for Rubber Property - Vulcanization Using Rotorless Cure Meter, using apparatus calibrated and traceable to NIST ( National Institute
of Standards and Technology ).
Calibrated By
The work has been done on site. See customer address for details.
The results indicated on this certificate relate only to the items calibrated.
This Certificate and attachments shall not be reproduced, except in full, without the written approval of Alpha Technologies.
The calibration has been performed following the Company ISO procedure : SVC0105
Calibration Conditions
Calibration Temperature : 177.0 Temperature Units : °C Temperature Tolerance : 0.5
Calibration Results
Reference Torque Standard Serial Number : T10S0085 Warmup time (min) : 30
Pre-Calibration Recorded Value (as Found) : 121.85 Final Recorded Value (as Left) : 120.80
The expanded uncertainty expressed at approximately the 95% confidence level using a coverage factor of k=2 is 0 lbinch.
Verification Apparatus
Verification Apparatus : Reference Torque Standard
Comments / Notes