Testing Machines Department
Certificate No: HT / 2021-2022 - SEC - 005 ULR No.: CC242220000003057F Page 1/1
CALIBRATION CERTIFICATE
DATE OF VERIFICATION : 01.07.2021 NEXT VERIFICATION RECOMMENDED ON : 30.06.2022

DATE OF ISSUE : 02.07.2021 DATE OF RECEIVED : 01.07.2021

EQUIPMENT : ROCKWELL CUM BRINELL HARDNESS TESTER


MODEL : RASNE-B
SERIAL NUMBER : 05/2017 -3496 M/s Sanray Laboratories Pvt Ltd
MAKE : FIE, Maharastra. No 302/3,Ec Extension
METHOD OF VERIFICATON : INDIRECT ECIL Hyderabad 500062
MEANS OF VERIFICATION : TEST BLOCK
ROOM TEMPRATURE Deg.C : 24°C

HARDNESS TEST STD.TEST UTB AVG. ERROR RPTTY Uncertanity*


OBSERVED READINGS
BLOCK BLOCK (%) H(max)- (+/-)
(H1,H2,H3,H4,H5) (+/- %)
SCALES SR.NO VALUE(H) HA HA-H H(min) IN Units
HRA 08198545 55.54 0.51 55.20 54.80 55.00 55.80 55.60 55.28 -0.26 1.00 1.50 N/A
HRA 08198543 82.22 0.50 82.80 82.60 81.80 82.60 82.60 82.48 0.26 1.00 1.50 N/A
HRBW 08198807 65.09 0.62 66.00 65.80 65.60 66.20 66.40 66.00 0.91 0.80 1.50 N/A
HRBW 08198551 89.96 0.52 90.40 90.20 90.80 89.80 90.60 90.36 0.40 1.00 1.50 N/A
HRC 08198561 20.73 0.51 21.20 20.80 21.80 21.60 21.60 21.40 0.67 1.00 1.68 N/A
HRC 08198536 42.76 0.51 43.80 43.60 43.60 43.60 43.40 43.60 0.84 0.40 1.68 N/A
HRC 08198562 61.88 0.51 62.80 63.00 62.40 62.20 62.40 62.56 0.68 0.80 1.68 N/A
HBW 2.5/187.5 08198682 205.50 0.23 206.00 207.00 207.00 207.00 207.00 206.80 1.30 1.00 N/A 2.70

# This Machine has been verified and the results have been found within permissible Limits.
As per IS-1586 (II) 2012 ISO 6508 - 1 : 2005 (Rockwell) & IS-1500 = 2013 ISO 6506 -1 : 2005.(Brinell Hardness)
* The Reference Test Block used for calibration is having unbroken chain of Traceability to NPL.
* The machine complied with the requirements of the standards with regards to Tests for Repeatability & Accuracy
* This Certificate may not be reproduced except in full unless written permission has been obtained
from M/s Blue Star Engineering & Electronics Limited. New Delhi.
* The overall estimated uncertainty has been computed at 95% confidence level
* k=2 for 95% confidence level for Veff >30 or as per NABL-141
# Readings are valid at the time of Calibration Only
# Results and Calibration Certificate related to the Item received and Calibrated.
Blue Star Engineering & Electronics Limited.

S.Sivaramakrishnan
Manager
Testing Machines Department
Accredited by NABL ( Ministry of Science & Technology ) for calibration of Testing Machines
End of Report