You are on page 1of 6

Emerson Automation Solutions

Gmbh & Co. OHG Weβling


Argelsrieder Feld 3
D-82234 Weβling, Germany
Tel: +49 (8153) 939-0

Calibration Certificate - consistent with ISO10474-3.1:2013 / EN10204-3.1:2004

Customer Information Manufacturer Information


Name: MONSUD SPA Sales Order: 39361571
PO: 361/2021 Line/Unit: 7

Device Information Calibration Information


Device Type: In-line Pressure Transmitter Factory: Emerson Process Management Weßling
Station Name: IDM16
Serial #: 21WEPJ0665782 Operator ID: 021
Model #: 2088G3S33A1M5D4B4Q4Q8S5HR5 Calibration Date: Sep 1 2021 8:20AM
Module Serial #: 67892681
Output: Linear
Device ID: 10671798

Permanent Tag Information Wire-on Tag Information


MNNDA05CP001

Attached Models
0306RT22BA11

Equipment Used
Equipment # Equipment Name Cal Due Date
IDMB02 Multimeter 28.04.2022 12:16:14
IDMG88 Pressure Controller 24.12.2021 19:32:14
LB-IDM103 Load Box 28.04.2022 05:27:19

Calibration Data Reference of Calibration: Gauge


Range: 0,000 TO 25,000 bar Damping: 0,40 seconds
% of Range Applied Pressure Requested Applied Pressure Analog Output (mA) % Span Error Pass/Fail

99,996 24,999 bar 24,9990 bar 20,00010 0,00438 PASS


79,996 19,999 bar 19,9990 bar 16,80002 0,00388 PASS
60,000 15,000 bar 15,0000 bar 13,60110 0,00688 PASS
40,000 10,000 bar 10,0000 bar 10,40098 0,00613 PASS
20,000 5,000 bar 5,0000 bar 7,20078 0,00488 PASS
0,000 0,000 bar 0,0000 bar 4,00022 0,00138 PASS

Issued Date: September 1, 2021 21WEPJ0665782_Q4-Pressure_WSS_RevA


Form #: Q4-Pressure_Ver 10 Page 1 of 2
This document certifies that the listed serialized instrument meets applicable Emerson specifications, has been
calibrated per Emerson’s calibration procedures and that the instrument conforms fully to all data and
engineering requirements provided to Emerson and acknowledged by Emerson. The calibration of measuring and
test equipment used in the manufacture and inspection of the listed product is traceable, through an unbroken
chain of calibrations or comparisons, to standards maintained by internationally recognized National Metrology
Institutes (NMI) such as NIM, NIST, PTB, etc.

Mihai Blaga
Quality Manager

Issued Date: September 1, 2021 21WEPJ0665782_Q4-Pressure_WSS_RevA


Form #: Q4-Pressure_Ver 10 Page 2 of 2
Emerson Automation Solutions
Gmbh & Co. OHG Weβling
Argelsrieder Feld 3
D-82234 Weβling, Germany
Tel: +49 (8153) 939-0

Calibration Certificate - consistent with ISO10474-3.1:2013 / EN10204-3.1:2004

Customer Information Manufacturer Information


Name: MONSUD SPA Sales Order: 39361571
PO: 361/2021 Line/Unit: 8

Device Information Calibration Information


Device Type: In-line Pressure Transmitter Factory: Emerson Process Management Weßling
Station Name: CALIBRATION_IDM_14
Serial #: 21WEPJ0665723 Operator ID: 2804
Model #: 2088G3S33A1M5D4B4Q4Q8S5HR5 Calibration Date: Aug 24 2021 1:02PM
Module Serial #: 67899054
Output: Linear
Device ID: 10666612

Permanent Tag Information Wire-on Tag Information


MNNDA05CP002

Attached Models
0306RT22BA11

Equipment Used
Equipment # Equipment Name Cal Due Date
IDMG79 Pressure Controller 23.12.2021 19:28:32
LB-IDM10 Load Box 28.04.2022 04:52:39
MM-IDM07 Multimeter 28.04.2022 03:46:46

Calibration Data Reference of Calibration: Gauge


Range: 0,000 TO 25,000 bar Damping: 0,40 seconds
% of Range Applied Pressure Requested Applied Pressure Analog Output (mA) % Span Error Pass/Fail

99,996 24,999 bar 24,9990 bar 20,00036 0,00600 PASS


79,996 19,999 bar 19,9990 bar 16,79912 -0,00175 PASS
60,000 15,000 bar 15,0000 bar 13,59948 -0,00325 PASS
40,000 10,000 bar 10,0000 bar 10,39944 -0,00350 PASS
20,000 5,000 bar 5,0000 bar 7,19930 -0,00437 PASS
0,000 0,000 bar 0,0000 bar 3,99952 -0,00300 PASS

Issued Date: August 24, 2021 21WEPJ0665723_Q4-Pressure_WSS_RevA


Form #: Q4-Pressure_Ver 10 Page 1 of 2
This document certifies that the listed serialized instrument meets applicable Emerson specifications, has been
calibrated per Emerson’s calibration procedures and that the instrument conforms fully to all data and
engineering requirements provided to Emerson and acknowledged by Emerson. The calibration of measuring and
test equipment used in the manufacture and inspection of the listed product is traceable, through an unbroken
chain of calibrations or comparisons, to standards maintained by internationally recognized National Metrology
Institutes (NMI) such as NIM, NIST, PTB, etc.

Mihai Blaga
Quality Manager

Issued Date: August 24, 2021 21WEPJ0665723_Q4-Pressure_WSS_RevA


Form #: Q4-Pressure_Ver 10 Page 2 of 2
Emerson Automation Solutions
Gmbh & Co. OHG Weβling
Argelsrieder Feld 3
D-82234 Weβling, Germany
Tel: +49 (8153) 939-0

Calibration Certificate - consistent with ISO10474-3.1:2013 / EN10204-3.1:2004

Customer Information Manufacturer Information


Name: MONSUD SPA Sales Order: 39361571
PO: 361/2021 Line/Unit: 8

Device Information Calibration Information


Device Type: In-line Pressure Transmitter Factory: Emerson Process Management Weßling
Station Name: CALIBRATION_IDM_14
Serial #: 21WEPJ0665724 Operator ID: 2804
Model #: 2088G3S33A1M5D4B4Q4Q8S5HR5 Calibration Date: Aug 24 2021 11:07AM
Module Serial #: 67903097
Output: Linear
Device ID: 10651185

Permanent Tag Information Wire-on Tag Information


MNNDA05CP003

Attached Models
0306RT22BA11

Equipment Used
Equipment # Equipment Name Cal Due Date
IDMG79 Pressure Controller 23.12.2021 19:28:32
LB-IDM10 Load Box 28.04.2022 04:52:39
MM-IDM07 Multimeter 28.04.2022 03:46:46

Calibration Data Reference of Calibration: Gauge


Range: 0,000 TO 25,000 bar Damping: 0,40 seconds
% of Range Applied Pressure Requested Applied Pressure Analog Output (mA) % Span Error Pass/Fail

99,996 24,999 bar 24,9990 bar 19,99868 -0,00450 PASS


79,996 19,999 bar 19,9990 bar 16,79782 -0,00987 PASS
60,000 15,000 bar 15,0000 bar 13,59856 -0,00900 PASS
40,000 10,000 bar 10,0000 bar 10,39886 -0,00712 PASS
20,000 5,000 bar 5,0000 bar 7,19950 -0,00312 PASS
0,000 0,000 bar 0,0000 bar 3,99956 -0,00275 PASS

Issued Date: August 24, 2021 21WEPJ0665724_Q4-Pressure_WSS_RevA


Form #: Q4-Pressure_Ver 10 Page 1 of 2
This document certifies that the listed serialized instrument meets applicable Emerson specifications, has been
calibrated per Emerson’s calibration procedures and that the instrument conforms fully to all data and
engineering requirements provided to Emerson and acknowledged by Emerson. The calibration of measuring and
test equipment used in the manufacture and inspection of the listed product is traceable, through an unbroken
chain of calibrations or comparisons, to standards maintained by internationally recognized National Metrology
Institutes (NMI) such as NIM, NIST, PTB, etc.

Mihai Blaga
Quality Manager

Issued Date: August 24, 2021 21WEPJ0665724_Q4-Pressure_WSS_RevA


Form #: Q4-Pressure_Ver 10 Page 2 of 2

You might also like