You are on page 1of 4

Coriolis Smart Meter Verification

Sunday, February 25, 2024


4:45:01 PM
Diagnostic Results
Smart Meter Verification Result Pass
Current Alerts Status
Sensor Circuits Pass
Transmitter/Electronics Pass
User-Configured Critical Alarms Pass
Maintenance Alarms Pass
Other Alarms Pass
Changes from the last SMV run Status
Calibration Coefficients No Change
Zero Coefficient No Change
Smart Meter Verification may Pass with active device alerts. If any of the alerts are active (Status is yellow and shows
“Warning” or “Changed”), refer to the active alerts in the interface used to generate this report for more information.

Customer Identification
Instrument Owner: Carlos Contact Name/Tested By:
Company: SOAP-Conoco Telephone:
Location: Kermit Application/Service:
Note: Run Counter: 5

Meter Information
Transmitter Tag: FIT-2200 Sensor Model: RFH/100
Transmitter Model: 1700A Sensor Serial Number: 14980304
Transmitter Serial Number: 0 Core Processor Unique ID: 26063777
Core Processor Serial Number: 0

Customer Signature: ______________________________________________________________________

Witness Signature: ______________________________________________________________________

Tested By Signature: ______________________________________________________________________

All Micro Motion Coriolis meters produced by Emerson are factory calibrated in ISO/IEC 17025 accredited and ISO 9001
certified laboratories. With a factory baseline, Smart Meter Verification is traceable to one or more of the following
national metrological institutes: CENAM-Mexico, INM-Romania, NIM-China, NIST-USA, VSL-The Netherlands.
Test Conditions and Meter Configuration
Flow Configuration Sensor Operating Conditions
Flow Damping: 3.20000 Sec Mass Flow Rate: 0.00000 lbs/day
Flow Calibration Factor: 272.674.67 Gas Standard Volume Flow Rate: 0.00000 MSCFD
Mass Factor: 1.00000 Density: 5.54257 lbs/ft3
Density Meter Factor: 1.00000 Temperature: 108.61600 °F
Volume Factor: 1.00000 External Pressure: 1250.00000 PSIG
Factory Zero: -0.00193 µsec Current Zero: -0.00193 µsec
DT or TC: 4.40000 Drive Gain: 2.44642 %
FT: 4.67000

Density Configuration Field Reference Point Information


Density Damping: 1.60000 Sec As-Found Inlet Stiffness Factor: 0.99901
K1: 3615.55200 µsec As-Found Outlet Stiffness Factor: 0.99921
K2: 4274.77393 µsec As-Left Inlet Stiffness Factor: 0.99901
D1: 0.00000 g/cm3 As-Left Outlet Stiffness Factor: 0.99921
D2: 1.00000 g/cm3
FD: 2400.00000
Version Information
Transmitter Revision: 8.02
Core Processor Revision: 4.7
ProLink III Revision: 4.9.553.5
Previous Test Results

Run Number Date/Time Result Reason


5 2/25/2024 4:45:01 PM Meter is unchanged
4 2/19/2024 10:37:16 AM Meter is unchanged
3 8/22/2023 1:25:09 PM Meter is unchanged
2 2/17/2023 12:35:57 PM Meter is unchanged
1 8/19/2022 12:52:04 AM Meter is unchanged

BACKGROUND

A Coriolis meter has no moving parts so there is nothing to wear or break. Therefore, in clean fluid services, the
expectation is that the meter will not exhibit a calibration change over time.

Fluids that corrode, erode, or coat the flow tube can alter the mechanical characteristics of the meter and may
change the meter’s calibration. Other mechanical damage (such as over pressurization) can also affect the meter’s
calibration. Smart Meter Verification measures the mechanical characteristics of the meter and detects subtle
changes that affect the calibration constants.

METHOD

Smart Meter Verification compares the mechanical properties of the meter at operating conditions to the factory
measured values. When the first Smart Meter Verification test is performed in the field a Field Reference Point is
applied. The Field Reference Point reduces external biases on the stiffness measurement, such as temperature and
pressure. It is highly advisable that the first Smart Meter Verification test be performed at normal operating
conditions. In general, Micro Motion recommends running Smart Meter Verification periodically (at least once per
month) using the scheduling feature in the transmitter.

DESCRIPTION OF RESULTS
When the Smart Meter Verification test is completed, the result is reported as Pass, Fail, or Abort. In all cases where a
Smart Meter Verification Professional test aborts, no report will be generated.

Pass Indicates that the meter is functioning properly and performing within specifications.
Fail The current SMV value is statistically different from the factory baseline value. Verify the sensor condition
by performing a visual inspection, density verification, or field proving. If possible, run Smart Meter
Verification with ProLink III Professional and save the report and Smart Meter Verification results. Contact
the factory for further evaluation and instructions.
RECOMMENDED ACTIONS

Calibration Coefficients Diagnostic


This indicates if any calibration coefficients have changed from the last time Smart Meter Verification was run.
Recommendations Verify the transmitter configuration to ensure all information is correct. If using a 5700, it is
recommended to review the configuration audit log in its historian.
Zero Coefficient Diagnostic
This indicates the meter zero calibration has changed from the last time Smart Meter Verification was run.
Recommendations If the new zero calibration was expected, do nothing. Otherwise, stop flow block in the meter
at operating conditions and run the Zero Verification Tool.

Using ProLink III Basic or Professional: Device Tools → Calibration → Smart Zero Verification
and Calibration → Verify Zero