Repair and Metrology Services (Pty) Ltd

29 Galaxy Avenue
Linbro Business Park
Linbro Park
Tel: 011 608 8550
cathye@repmet.co.za
Directors: C Elias; L Juju
Co. Reg: 2004/021865/07

Certificate No : 124096-1 The South African National Accreditation Systems


(SANAS) is a member of the International
Customer : JB Madagascar Laboratory Accreditation Co-Operation (ILAC) for
The Mutual Recognition Agreement (MRA). The
MRA allows for mutual recognition of technical
test and calibration data by the member
Address : 24, rue RADAMA accreditation bodies worldwide. For more
BP207- information on the MRA please consult
ANTANANARIVO 101 www.ilac.org
Madagascar
This certificate is issued without alteration and
accordance with the conditions of accreditation
granted by SANAS. Copyright of this certificate is
Calibration date : 03 April 2023 owned jointly by SANAS and Repair and Metrology
Services and may not be reproduced other than in
Expiry date : April 2025 full except with the prior written approval of the
aforementioned.
The results given in this certificate were correct at
Issue date : 03 April 2023 the time of calibration and relate to UUT identified on
this certificate. Subsequently the accuracy
will depend on factors such as care exercised in
handling the instrument and frequency of use.
Manufacturer : Fluke
Recalibration should be performed after a period,
which has been chosen to ensure that, under
Model : TiS45 normal circumstances, the instrument's accuracy
remains within the desired limits. The accuracies of
Description : Thermal Imager all measurements are traceable to national
or international measuring standards. The reported
Camera expanded uncertainty of measurement is stated as
Serial No : TiS45-16070241 the standard uncertainty of measurement
multiplied by the coverage factor k=2, which for a
Asset No : None normal distribution corresponds to a coverage
probability of approximately 95 %

Calibrated by Checked by SANAS Technical Signatory


MJ. Frade SP. Mare MJ. Frade

Page 1 of 6
Certificate of Calibration
Certificate No: 124096-1

1. Standards and equipment used

Fixed
Make Model Description Serial Number Certificate No Cal Due Date
Asset ID
Hart 9132 IR Calibrator A59418 Std 12 120685-3 10/2023
Infrared
Fluke 9133 B2C421 Std 50 123037-3 01/2024
Calibrator

2. Procedure

2.1 The Fluke TiS20 was calibrated according to Section 1, 2 & 3.1 of the Fluke
TiS10, TiS20, TiS40, TiS45, TiS50, TiS55, TiS60, TiS65 & TiS75 Thermal Imager
Service Manual, Revision 1.2, dated 09 December 2015.

2.2 The Emissivity of the source and Thermal Imager were set 0.95.

3. Results

3.1 Preliminary inspection

Visual Inspection Report


Housing No obvious damage
Infrared Coated Lens No obvious damage
LCD Lens No scratches visible
Battery Was fully charged
Charger Present
SD Card None
Carry Case/ Soft carry case Present
Keypad Test Pass
Primary Trigger Test Pass
Backlight Test Pass
HDMI Output Test N/A

Page 2 of 6
Certificate of Calibration
Certificate No: 124096-1

3.2 Temperature Measurements

Set TiS20 TiS20 Tolerance TiS20 Thermal


Temperature Display Display (°C) Display Images
(°C) Minimum Maximum Average
(°C) (°C) (°C)
-5.0 -6.0 -4.5 -7.0 to -3.0 -5.3 Fig 1
30.0 30.1 30.7 28.0 to 32.0 30.4 Fig 2
70.0 70.1 70.9 68.0 to 72.0 70.6 Fig 3
100.0 98.9 99.9 98.0 to 102.0 99.5 Fig 4
165.0 167.4 168.7 161.7 to 168.3 168.1 Fig 5
240.0 234.8 237.5 235.2 to 244.8 236.3 Fig 6

Uncertainty of Measurement: -5 to 100 C  2 °C


100 to 240 C  5 °C

Fig1

Page 3 of 6
Certificate of Calibration
Certificate No: 124096-1

Fig 2

Fig 3

Page 4 of 6
Certificate of Calibration
Certificate No: 124096-1

Fig 4.

Fig 5

Page 5 of 6
Certificate of Calibration
Certificate No: 124096-1

Fig 6

4. Comments

4.1 The calibration was performed at the RMS laboratory in Linbro Park in a
controlled environment. The ambient temperature was at 23 °C ± 5 °C and
relative humidity 30 % rh to 60 % rh.
4.2 The UUT was in good condition.
4.3 The above results were entered directly on to a computer.
4.4 The diameter for STD 12 and STD 50 is 57 mm.
4.5 Results in BOLD fall outside of the manufacturer’s accuracy specifications.

------End of Certificate------