Professional Documents
Culture Documents
29 Galaxy Avenue
Linbro Business Park
Linbro Park
Tel: 011 608 8550
cathye@repmet.co.za
Directors: C Elias; L Juju
Co. Reg: 2004/021865/07
Page 1 of 6
Certificate of Calibration
Certificate No: 124096-1
Fixed
Make Model Description Serial Number Certificate No Cal Due Date
Asset ID
Hart 9132 IR Calibrator A59418 Std 12 120685-3 10/2023
Infrared
Fluke 9133 B2C421 Std 50 123037-3 01/2024
Calibrator
2. Procedure
2.1 The Fluke TiS20 was calibrated according to Section 1, 2 & 3.1 of the Fluke
TiS10, TiS20, TiS40, TiS45, TiS50, TiS55, TiS60, TiS65 & TiS75 Thermal Imager
Service Manual, Revision 1.2, dated 09 December 2015.
2.2 The Emissivity of the source and Thermal Imager were set 0.95.
3. Results
Page 2 of 6
Certificate of Calibration
Certificate No: 124096-1
Fig1
Page 3 of 6
Certificate of Calibration
Certificate No: 124096-1
Fig 2
Fig 3
Page 4 of 6
Certificate of Calibration
Certificate No: 124096-1
Fig 4.
Fig 5
Page 5 of 6
Certificate of Calibration
Certificate No: 124096-1
Fig 6
4. Comments
4.1 The calibration was performed at the RMS laboratory in Linbro Park in a
controlled environment. The ambient temperature was at 23 °C ± 5 °C and
relative humidity 30 % rh to 60 % rh.
4.2 The UUT was in good condition.
4.3 The above results were entered directly on to a computer.
4.4 The diameter for STD 12 and STD 50 is 57 mm.
4.5 Results in BOLD fall outside of the manufacturer’s accuracy specifications.
------End of Certificate------