Calibration Certificate

ULR No.: CC30962200005754F


Certificate No.: CCTPL/MP/0676/03 Field: Mechanical
SRF No.: 600/22 Date of Receipt: 12/12/2022
SRF Date: 12/12/2022 Date of Calibration: 13/12/2022
Company Name & M/S : Suguna Foods Pvt.
Recommended Due Date.:
Address Ltd.(Central Lab QA) 12/12/2023
(As per agreed by customer)
Plot No.-1144, Near Moon Hospital,
Mahanadi Vihar, Cuttack-753004 Certificate Issue Date: 14/12/2022

DUC* Details

Instrument Name: Micro Pipette Instrument Sr. No.: RC634892


Make: Riviera Instrument Id No.: SF/MP-01
Model No.: --- Location: ---
Range: 100 to 1000 µl DUC Condition: Ok
Readability: 5 µl Calibration Performed at: At Lab
Standard Equipments Used(Traceable to National Standard)
Sr. No Instrument Name Cert. No Cal. By Calibration Due Date
1 Analytical Balance BBT/283/APR/22 Blue Box 05/05/2023

Temperature Humidity Reference Standard Calibration Procedure


Environment Condition
23.8ºC 56%RH ISO: 8655-6 CCTPLP/M/MV-02

CALIBRATION RESULT
Observed Value on
DUC* Value Error Mean Value at 27⁰C
Sr. No. Standard
(in µl) (in mg) (in µl) (in µl)
1 100 100.32 -0.32 100.45

2 500 500.42 -0.42 501.04

3 1000 1000.62 -0.62 1001.87

Note : DUC* : Device Under Calibration


Uncertainty of measurment at approx 95% Confidence Level with coverage factor k = 2, is ± 0.9 µl.

Calibrated By : Priyank Sharma Approved By : Sandeep Sharma


Calibration Engineer Director

Note. 1.This Certificate only to the particular instrument submitted for calibration.
2.This calibration Result reported in this certificate valid at the time of and under the stated condition of measurement.
3.This particular certificate can not be reproduce except in full, without prior permission of CCTPL

`---End of Certificate--- Page 1 of 1