You are on page 1of 1

Issued For : M/S KDM ENGINEERS INDIA PVT. LTD. ULR No.

: CC319321000000513F
VIZAG
Cal. Certificate No. : DRS/08/21/513
SRF No. & Date : DRS/SRF/21/318
Date of Calibration : 05.08.2021

Calibration Due on : 04.08.2022

Details of Unit Under Calibration Issue Date of Certificate : 06.08.2021

Type : PROVING RING


Make : - Date of Reciept : 02.08.2021
Mode of Calibration : Compression Customer Ref. No. & Date : : BY COURIER
Serial Number : PR.10KN.377 Condition on receipt : OK
ID Number : -
Capacity : 10 KN
Cable Length : N/A Environmental Condition
Details of Dial gauge Room Temperature : 24±1°C
Indicator Type : Dial Gauge Humidity
: 43.1% RH
Make : MERCER
Correction of Temperature Variation:-
Model : -
Serial Number : - Apply ±0.027% Correction to each Reading for each
Resolution : 0.1 Div. (1 Div. = 0.002mm) degree celcius rise or fall of temperature.

Accessories : Self-aligning Compression Pads.


Standard used : 10 kN DEAD Weight Force Calibration Machine (DRS/DWFCM-02)
Associated Uncertainty : 0.041% at (k=2)
Traceability : FRI/09/20/DWCM/021 Dated 03/09/2020 Valid upto 03/09/2022.
Principle/Methodology of Calibration and Calibration Procedure:
1. Work Instruction Number: DRS/SOP/01F Based on IS 4169-2014 & ISO 376 : 2011
2. Preloading : Before the application of the calibration forces, the instrument was preloaded thrice to its maximum capacity and kept at
full load for about 90 seconds.

3. Calibration : The sequence of applied calibration force in Compression is given below:

At 0° : Two series of calibration forces in increasing values. At 120° and 240° positions: One series of calibration forces each increasing
values. Creep test is performed by calculating the difference in output i30 obtained at 30s and i300 obtained at 300s after the removal of
the maximum calibration force and express this difference as percentage of the maximum deflection. The calibration was made by using
self aligning Compression Pads provided along with the instrument to ensure axial application of the force. Between each series, the
instrument was rotated along its axis so as to occupy during calibration three positions (0°,120°,240°) and the instrument was subjected to
the full load once for about 90 seconds each time before starting in a new position. Between the loadings, readings corresponding to no
load after waiting at least 30 seconds for the return to zero were noted.

Calibrated by Approved by
Gurdeep Singh Kaushal Kishore
(Technical Manager)

FORMAT NO. : DRS/CF/18.01 Page 1 of 2

You might also like