Date: 11 – 08 – 2021 Sheet: 1 OF 1

Format No. : APT/FF/25-11 Report No.APT/CC/D/2021/820 ULR No.:CC242621000001119


M/s: Mahindra University
Customer Name and Address: Survey No. 62/1A, Qutubullapur, Bahadurpally Village, RR District, Hyderabad - 500043.
Work Order Through: M/s. RK Labs
SRF. No. and Date: APT/2021/085 & 05/08/2021 Received Condition: Satisfactory
CALIBRATED INSTRUMENT / EQUIPMENT DETAILS
Nomenclature: Radius Gauge Make : Insize
Range / Resolution: 1 to 7 mm (17 Nos) N/A Sr. No.: 1512168203
Instrument Location: Metallurgy Lab Temperature / Humidity : 20oC±2oC 50±10% (RH)
Calibrated on: 10/08/2021 Calibration due on: 09/08/2022
MASTER EQUIPMENT TRACEABILITY DETAILS
Sl. No. Nomenclature Make & Model Sl. No. Traceable Cert. No. Validity
1 Profile Projector OptoMech / PP 300TLE 04 / 2016 MME/CAL/022/21-22 08-01-2022
The master equipment’s used are traceable to national standards Ref. Doc. SOP-APT/WI/11
CALIBRATION OF RADIUS GAUGES: (All values in mm)
Nominal Calibrated values Calibrated values
Sl. No. Sl. No. Nominal size
size EXTERNAL INTERNAL EXTERNAL INTERNAL
1 1.00 1.02 1.02 10 3.50 3.52 3.51
2 1.25 1.22 1.25 11 4.00 4.01 4.03
3 1.50 1.51 1.49 12 4.50 4.52 4.52
4 1.75 1.76 1.74 13 5.00 5.01 5.00
5 2.00 2.03 1.97 14 5.50 5.50 5.49
6 2.25 2.28 2.22 15 6.00 6.03 6.02
7 2.50 2.54 2.52 16 6.50 6.51 6.49
8 2.75 2.77 2.78 17 7.00 7.02 7.00
9 3.00 3.03 3.02 --- --- --- ---

Allowable Values for Radius Gauges As Per IS: 5273-1969 (All Values in mm)
NOMINAL SIZE PERMISSIBLE ERROR
0.6 upto 0.8 mm ±0.050
1 upto 6 mm ±0.063
8 upto 25 mm ±0.100
Note:
● Determination of Checking Arc of Circle of Radius Gauge by direct method using Profile Projector.
Conclusion:
●Uncertainty of calibration at 95 % Confidence level and Coverage Factor K = 2 : ± 7.7µm
●The Reported Results are valid only for the conditions of the received Industries /gauges at the time of and
under the stated conditions of the calibration
********---********

Authorized Signatory

(Name: Chandradeep B.)


Date: 11 – 08 – 2021 Sheet: 1 OF 1

Format No. : APT/FF/25-11 Report No.APT/CC/D/2021/821 ULR No.:CC242621000001120


M/s: Mahindra University
Customer Name and Address: Survey No. 62/1A, Qutubullapur, Bahadurpally Village, RR District, Hyderabad - 500043.
Work Order Through: M/s. RK Labs
SRF. No. and Date: APT/2021/085 & 05/08/2021 Received Condition: Satisfactory
CALIBRATED INSTRUMENT / EQUIPMENT DETAILS
Nomenclature: Radius Gauge Make : Insize
Range / Resolution: 7.5 to 15 mm (16 Nos) N/A Sr. No.: 2805168283
Instrument Location: Metallurgy Lab Temperature / Humidity : 20oC±2oC 50±10% (RH)
Calibrated on: 10/08/2021 Calibration due on: 09/08/2022
MASTER EQUIPMENT TRACEABILITY DETAILS
Sl. No. Nomenclature Make & Model Sl. No. Traceable Cert. No. Validity
1 Profile Projector OptoMech / PP 300TLE 04 / 2016 MME/CAL/022/21-22 08-01-2022
The master equipment’s used are traceable to national standards Ref. Doc. SOP-APT/WI/11
CALIBRATION OF RADIUS GAUGES: (All values in mm)
Nominal Calibrated values Calibrated values
Sl. No. Sl. No. Nominal size
size EXTERNAL INTERNAL EXTERNAL INTERNAL
1 7.5 7.49 7.51 10 12.0 11.99 12.01
2 8.0 8.01 7.99 11 12.5 12.51 12.51
3 8.5 8.51 8.49 12 13.0 13.01 13.01
4 9.0 9.01 9.01 13 13.5 13.52 13.49
5 9.5 9.52 9.51 14 14.0 14.02 13.99
6 10.0 10.01 10.02 15 14.5 14.51 14.48
7 10.5 10.51 10.52 16 15.0 15.01 14.98
8 11.0 10.99 11.01 --- --- --- ---
9 11.5 11.49 11.51 --- --- --- ---

Allowable Values for Radius Gauges As Per IS: 5273-1969 (All Values in mm)
NOMINAL SIZE PERMISSIBLE ERROR
0.6 upto 0.8 mm ±0.050
1 upto 6 mm ±0.063
8 upto 25 mm ±0.100
Note:
● Determination of Checking Arc of Circle of Radius Gauge by direct method using Profile Projector.
Conclusion:
●Uncertainty of calibration at 95 % Confidence level and Coverage Factor K = 2 : ± 7.7µm
●The Reported Results are valid only for the conditions of the received Industries /gauges at the time of and
under the stated conditions of the calibration
********---********

Authorized Signatory

(Name: Chandradeep B.)


Date: 11 – 08 – 2021 Sheet: 1 OF 1

Format No. : APT/FF/25-13 Report No.APT/CC/D/2021/822 ULR No.:CC242621000001121


M/s: Mahindra University
Customer Name and
Survey No. 62/1A, Qutubullapur, Bahadurpally Village, RR District, Hyderabad - 500043.
Address:
Work Order Through: M/s. RK Labs
SRF. No. and Date: APT/2021/085 & 05/08/2021 Received Condition: Satisfactory
CALIBRATED INSTRUMENT / EQUIPMENT DETAILS
Nomenclature: Bevel Protector Make : Mitutoyo
Range / Resolution: 0 –360 ˚ 5’ Sr. No.: 173338
Instrument Location: Metallurgy Lab Temperature / Humidity : 20oC±2oC 50±10% (RH)
Calibrated on: 10/08/2021 Calibration due on: 09/08/2022
MASTER EQUIPMENT TRACEABILITY DETAILS
Sl. No. Nomenclature Make & Model Sl. No. Traceable Cert. No. Validity
1 Profile Projector Opto Mech / PP 300TLE 04 / 2016 MME/CAL/022/21-22 08-01-2022
The master equipment’s used are traceable to national standards Ref. Doc. IS : 4239 & SOP- APT/WI/13
CALIBRATION OF BEVEL PROTRACTOR

CALIBRATION RESULTS
SL No. NOMINAL ERRORS
VALUES TOP RIGHT TOP LEFT BOTTOM RIGHT BOTTOM LEFT
1 0° 0 0 +1’ +1’
2 1° -1’ -1’ +1’ +1’
3 5° -1’ -1’ +1’ +1’
4 20° -2’ -1’ +2’ +3’
5 30° -1’ -2’ +2’ +2’
6 50° -1’ -1’ +2’ +3’
7 70° -1’ -2’ +1’ +2’
8 90° -2’ -2’ +2’ +1’
Note:
 Determination of various Angle measured by direct measurement Using Profile Projector.
Conclusion:
 Uncertainty of calibration at 95 % Confidence level and Coverage Factor K = 2 : ± 8.3 min of Arc
 The Reported Results are valid only for the conditions of the received instruments/gauges at the time of and
under the stated conditions of the calibration

Authorized Signatory

(Name: Chandradeep B.)


Date: 11-08-2021 Sheet: 1 OF 1

Format No. : APT/FF/25-12 Report No.APT/CC/D/2021/823 ULR No.:CC242621000001122


M/s. Mahindra University
Customer Name and
Survey No. 62/1A, Qutubullapur, Bahadurpally Village, RR District, Hyderabad - 500043.
Address:
Work Order Through: M/s. RK Labs
SRF. No. and Date: APT/2021/085 & 05/08/2021 Received Condition: Satisfactory
CALIBRATED INSTRUMENT / EQUIPMENT DETAILS
Nomenclature: Height Gauge Make : Mitutoyo
Range / Resolution: 0 to 300 mm 0.02 mm Sr. No.: 17702395
Instrument Location: Metallurgy Lab Temperature / Humidity : 20oC±2oC 50±10% (RH)
Calibrated on: 10/08/2021 Calibration due on: 09/08/2022
MASTER EQUIPMENT TRACEABILITY DETAILS
Sl. No. Nomenclature Make & Model Sl. No. Traceable Cert. No. Validity
1 Caliper Checker KC PRECISION / CC600 2584 VI/20-21/0415-01 28-10-2021
The master equipment’s used are traceable to national
Ref. Doc. IS 2921-1988 & SOP- APT/WI/12
standards
CALIBRATION OF RADIUS GAUGES: (All values in mm)
CALIBRATION RESULTS All values are in mm
NOMINAL VALUES HEIGHT GAUGE READINGS Error Permissible Error ( +/-)
0 0(Set) 0.00 0.02
20.00 20.00 0.00 0.02
50.00 50.00 0.00 0.03
75.00 74.99 -0.01 0.03
100.00 99.99 -0.01 0.03
150.00 149.99 -0.01 0.03
200.00 199.99 -0.01 0.03
250.00 249.99 -0.01 0.03
300.00 299.99 -0.01 0.04
Note:
 Determination of various sizes a from respective reference faces were measured by direct measurement against
Caliper Checker.
Conclusion:
 Uncertainty of calibration at 95 % Confidence level and Coverage Factor K = 2 : ± 10.20µm
 The Reported Results are valid only for the conditions of the received instruments/gauges at the time of and
under the stated conditions of the calibration

Authorized Signatory

(Name: Chandradeep B.)


Date: 11-08-2021 Sheet: 1 OF 1

Format No. : APT/FF/25-12 Report No.APT/CC/D/2021/824 ULR No.:CC242621000001123


M/s. Mahindra University
Customer Name and Address: Survey No. 62/1A, Qutubullapur, Bahadurpally Village, RR District, Hyderabad - 500043.
Work Order Through: M/s. RK Labs
SRF. No. and Date: APT/2021/085 & 05/08/2021 Received Condition: Satisfactory
CALIBRATED INSTRUMENT / EQUIPMENT DETAILS
Nomenclature: Digital Height Gauge Make : Mitutoyo
Range / Resolution: 0 to 300 mm 0.01 mm Id. No.: HG-01
Instrument Location: Metallurgy Lab Temperature / Humidity : 20oC±2oC 50±10% (RH)
Calibrated on: 10/08/2021 Calibration due on: 09/08/2022
MASTER EQUIPMENT TRACEABILITY DETAILS
Sl. No. Nomenclature Make & Model Sl. No. Traceable Cert. No. Validity
1 Caliper Checker KC PRECISION / CC600 2584 VI/20-21/0415-01 28-10-2021
The master equipment’s used are traceable to national standards Ref. Doc. IS 2921-1988 & SOP- APT/WI/12
CALIBRATION OF RADIUS GAUGES: (All values in mm)
CALIBRATION RESULTS All values are in mm
NOMINAL VALUES HEIGHT GAUGE READINGS Error Permissible Error ( +/-)
0 0(Set) 0.00 0.02
20.00 20.00 0.00 0.02
50.00 50.00 0.00 0.03
75.00 75.00 0.00 0.03
100.00 100.00 0.00 0.03
150.00 150.00 0.00 0.03
200.00 200.00 0.00 0.03
250.00 250.00 0.00 0.03
300.00 299.99 -0.01 0.04
Note:
 Determination of various sizes a from respective reference faces were measured by direct measurement against
Caliper Checker.
Conclusion:
 Uncertainty of calibration at 95 % Confidence level and Coverage Factor K = 2 : ± 10.20µm
 The Reported Results are valid only for the conditions of the received instruments/gauges at the time of and
under the stated conditions of the calibration

Authorized Signatory

(Name: Chandradeep B.)