MTMS-FM-03/B-IV-IS REV 10

SITE CALIBRATION CERTIFICATE FOR BRINELL HARDNESS TESTING MACHINE


(INDIRECT VERIFICATION)
ULR - CC229320000003182F Date of Issue: 28/12/2020 O. A. No.: 10604
Certificate No. MTMS/NABL/12/2020/23491 Page No.: 1 / 2
Discipline : MECHANICAL CALIBRATION

NAME OF CUSTOMER DETAILS OF EQUIPMENT


MAKE: SAROJ IS : 1500-2 :2013
M/s. Hi - Mac Casting Pvt. Ltd., REFERED STD
MODEL: B-3000 / O ISO: 6506-2 :2017
Plot No. 272 / A, Gate No.1,
SR. NO.: 08/03/160 WORK INSTR. NO. MTMS-WI-02
Lodhika GIDC, Kalawad Road,
ID. NO.: BHT – 01 CALIBRATED ON 25/12/2020
Lodhika. ROOM TEMP.:28ºC AGREED DUE DATE 24/12/2021
LOCATION: HARDNESS ROOM CALIBRATED BY S S Patil
CONDITION OF THE EQUIPMENT AT THE TIME OF
SATISFACTORY
CALIBRATION
DETAILS OF STANDARD TEST BLOCKS USED FOR CALIBRATION:
BLOCK VALUE WITH SR. NO. TRACEABILITY UNCERTAINTY IN MEASUREMENT
194 HBW 5/750 - 0775708 NIST ± 1.6 %
309 HBW 5/750 - 0775402 NIST ± 1.7 %
188.1 HBW 10/3000 -1220616 Through FRI to NPL India, ± 1.50 %
337 HBW 10/3000 - 6002146 NIST ± 1.0 %

Calibration Results are enclosed on next page & permissible Values are as below:
Permissible Error in reference Indentation is 0.5 % as per Cl. No. 5.2 of IS 1500-2
Permissible Error in Hardness value is as per Cl. No. 5.8 of IS 1500-2
Permissible Repeatability in Diagonal is as per Cl. No. 5.7 of IS 1500-2

Statement of Conformity:
Overall deviation is within specified Limit respective to reference standard without considering Uncertainty.

Measuring Device: Display on Computer

Disclaimer: - - -

Contd…2

MTMS-FM-03/B-IV-IS REV 10
ULR - CC229320000003182F Date of Issue: 28/12/2020 O. A. No.: 10604
Certificate No. MTMS/NABL/12/2020/23491 Page No. 2 / 2
Discipline : MECHANICAL CALIBRATION
INDIRECT VERIFICATION
Mean
Mean value Diameter of Impressions in (mm) Observed Permissible
Dia. (mm)
Dia. of reference Repeatability Repeatability
Ball Size HBW / Indentations in Diagonal in Diagonal
d1 d2 d3 d4 d5 d
(mm)/ (mm) (rrel) % (rrel) % UIM
Test (%)
Observed Average
Force Std. Corresponding Hardness Value Observed Permissible (±)
Error in HBW
(kgf) Test Error of Error of
reference
Block Hardness Hardness
Indentation H1 H2 H3 H4 H5 H
Value (Erel) % (Erel) %
(d) %
5/7.355 2.164 2.171 2.171 2.160 2.165 2.167 2.173 2.165 0.60 2.50
2.1
(750) 194 0.32 193 195 195 195 192 194.00 0.00 2.50
5/7.355 1.713 1.710 1.740 1.733 1.736 1.740 1.738 1.735 0.58 2.00
2.1
(750) 309 -0.18 306 308 309 306 309 307.60 -0.45 2.00
10/29.42 4.3897 4.397 4.400 4.397 4.410 4.407 4.408 4.400 0.45 2.50
2.0
(3000) 188.1 0.17 187 188 186 187 187 187.00 -0.58 2.50
10/29.42 3.316 3.320 3.320 3.330 3.327 3.332 3.330 3.326 0.36 2.00
2.0
(3000) 337 0.12 337 335 337 334 335 335.60 -0.42 2.00
U. I. M. is calculated with the coverage factor 2 corresponding to 95.45 % confidence level.
* The new accreditation number allocated by NABL is: CC-2293 in lieu of NABL C-0262

Calibrated by: Checked by: Authorised by: