~ Calibration Certificate ~

Model Number: 442A102 Customer: Research Design & Standard

Serial Number: 604 Organisation (RDSO)

Description: Signal Conditioner

Manufacturer: PCB ID #:

Calibration Data

Temperature: 72 °F ( 22 °C) Humidity: 55.30%

Volts Current
(mA) Gain X1
Specification: 24 ± 0.5 4 ± 15% 1 ± 1%
Measured: 24.2 4.18 1.000

Condition of Unit

As Found: In Tolerance
As Left: In Tolerance

n/a
New unit, in tolerance Notes

1. Calibration is traceable through NIST.


2. This certificate shall not be reproduced, except in full, without written approval from The Modal Shop, Inc.
3. Calibration is performed in compliance with ANSI/NCSL Z540-1-1994 and ISO 17025 using PRD-P292.
4. Measurement uncertainty (95% confidence level with a coverage factor of 2) for the sensitivity ratios are +/- 0.2 %
5. See Manufacturer’s Specification Sheet for a detailed listing of performance specifications.
6. The results documented in this certificate relate only to the items tested or calibrated.

Technician: Ed Devlin Cal Date: 25-Aug-20

Approval: Due Date: