CAL-MATRIX Technical Services Inc.

Certificate Number 0111C27-6


1495 Bonhill Road Unit 1
Mississauga, Ontario, Canada Page Number 1 of 2
L5T 1M2
Tel.: (905) 564-6235 / Fax: (905) 564-6399 Calibration Procedure Mfr’s Specs Ver. 1.0

CALIBRATION CERTIFICATE
UNIT UNDER TEST (UUT) CUSTOMER
Manufacturer Avo Name Goodrich Landing Gear
Model 359985 Address 1400 South Service Rd.
Description Hand Tachometer Oakville, Ontario
Serial Number 2895-103048 (MA8576) L5L 5Y7
Calibration Date 15 November 2001 Contact Helen Varelas
Calibration Interval 1 Year Purchase Order 135248 OJ
Next Calibration Due 15 November 2002 Work Order 0111-27-6

ENVIRONMENTAL CONDITIONS:
Ambient Temperature 23.0 oC ± 2 oC Barometric Pressure N/A PSIA ± 0.1 PSIA
Relative Humidity 40.0 % ±5%

UNIT UNDER TEST CONDITION:


Received
Within Specs  Outside of Specs Good Condition  “As Received”
readings taken

Returned
Within Specs  Outside of Specs Required Repair “As Returned”
readings taken

This is to certify that the instrument listed above was calibrated against Cal-Matrix Technical Services Inc. Company Standards which are directly traceable to the “National Research Council”
(NRC), Canada, “National Institute of Standards” (NIST), U.S.A. and/or other accredited laboratories. This calibration complies with MIL-STD-45662A and ANSI/NCSL Z540-1-1994. The
Test Uncertainty Ratio (TUR) exceeds four to one unless otherwise indicated.

STANDARDS AND EQUIPMENT USED


Equipment Description Manufacturer Model / Type Serial Number Cal Due Date
Calibrator Fluke 5520A 7225203 11 December 2001

Remarks: This unit was calibrated at customer’s premises.


Replaced 3 “AA” batteries.

Steve Gerassimou 15 Nov 2001 15 Nov 2001


CALIBRATED BY DATE APPROVED BY DATE
Cal-Matrix Inc. Standards Calibration Laboratory Quality Manager

© Copyright of this Certificate is owned by the issuing laboratory and may not be reproduced other than in full except with the prior written approval of the issuing laboratory.
CAL-MATRIX Technical Services Inc. Certificate Number 0111C27-6
1495 Bonhill Road Unit 1
Mississauga, Ontario, Canada Page Number 2 of 2
L5T 1M2
Tel.: (905) 564-6235 / Fax: (905) 564-6399 Calibration Procedure Mfr’s Specs Ver. 1.0

CALIBRATION DATA
Manufacturer Avo Description Hand Tachometer
Model 359985 Serial Number 2895-103048 (MA8576)

Frequency Input Calibration:


Specifications: (0.02% of measured value + 1 lsd from 10rpm to 19999rpm)

Note: Apply 3Vp-p @ Freq. Square wave to designated points inside unit.

UUT Setting Applied to UUT UUT Indication Specification Tolerance Limits


Range (Internal optocouple) As Received As Returned +/- Lower Upper

10.00RPM to 99.99RPM 1.00 Hz light source / RPM [0.02] 59.98 to 60.02 RPM

100.0RPM to 999.9RPM 2.00 Hz light source / RPM [0.1] 119.9 to 120.1 RPM

3.00 Hz light source / RPM [0.1] 179.9 to 180.1 RPM

4.00 Hz light source / RPM [0.1] 239.8 to 240.1 RPM

5.00 Hz light source / RPM [0.2] 299.8 to 300.2 RPM

10.0 Hz light source / RPM [0.2] 599.8 to 600.2 RPM

15.0 Hz light source / RPM [0.2] 899.8 to 900.2 RPM

1000RPM to 19999RPM 20.0 Hz light source / RPM [1] 1199 to 1201 RPM

30.0 Hz light source / RPM [1] 1799 to 1801 RPM

40.0 Hz light source / RPM [1] 2399 to 2401 RPM

50.0 Hz light source / RPM [2] 2998 to 3002 RPM

100 Hz light source / RPM [2] 5998 to 6002 RPM

150 Hz light source / RPM [3] 8997 to 9003 RPM

200 Hz light source / RPM [3] 11997 to 12003 RPM

300 Hz light source / RPM [5] 17995 to 18005 RPM

Steve Gerassimou 15 Nov 2001 15 Nov 2001


CALIBRATED BY DATE APPROVED BY DATE
Cal-Matrix Inc. Standards Calibration Laboratory Cal-Matrix Inc. Quality Manager