Certificate of Calibration
Manufacturer: Kingston Report #: TJZ-22-0006314-R1
Description: Valvula de Seguridad P.O. #: N/A
This certifies that the above instrument was calibrated in compliance with the Calibration System Requirements of ISO /IEC
17025:2017, ANSI/NCSL Z540-1-1994 in accordance with referenced procedures. Standards used to perform this calibration
are traceable to SI units; their source of traceability derives from a National Metrology Institute such as NIST, CENAM, NPL,
DIN, from natural physical constants, consensus standards or derived by the ratio type of calibrations. Collective
uncertainties are determined as required with a distribution that corresponds to a probability of approximately 95% (k=2).
Unless otherwise noted calibrations are performed to manufacturer's specifications. Compliance statements are in
conformance with ILAC-G8:2019 simple acceptance decision rule. This form shall not be reproduced, except in full, without
the expressed written consent of Techmaster. Contact our customer service representative for clarification of this document.
F. Viaña
Ernesto Matamoros Ing. Mauricio Garayzar Fernando Viaña
Quality Assurance Technical Manager Certified By
Tijuana, B.C. Calle Seminario No. 8610-11, Col. Niños Héroes Este, CP 22120, Tijuana, BC, México Tels.: (664) 624-4444 / 624-3128 / 624-4445 / 624-0440
Mexicali, B.C. Calle Gales Nº 1201, Col. Villa Fontana, C.P. 21180, Mexicali, BC, México Tels.:(686) 555-1660 / 557-6117 / 555-1766
Cd. Juárez,Chih. Ave. Antonio J. Bermudez No 2151-5 Fracc. Parque Ind J.Bermudez, C.P32470 Cd. Juarez Chih. Tels.:(656) 687-2471 / 648-1181
Monterrey, N.L. Ave. Morones Prieto No. 914, Oriente Int. 112, Col. La Huerta, C.P. 67144. Tels.:(81)1334-0701 / 1334-0702
Queretaro, QRO. Ctra. Federal 57,Mex-Qro Lateral Norte Km 210+100,Advance Business Park Conin, int.26 CP.76240,Col. Cumbres de Conin, Tels.:(442)241 51 04
Issued on: 2021-09-15 14:42:44.7600000 -07:00 540.1 2101
Page 1 of 2
TJZ-22-0006314 33
Calibration Results
Manufacturer: Kingston Asset No. V1RSP5
Step # Function Tested / Nominal Value As Found ± MU* In Tol As Left ± MU* In Tol Calibration Tolerance
1 Pressure 150.00 psi 151.200 ± 0.091 YES 151.200 ± 0.091 YES 145.50 to 154.50 psi
2 150.00 psi 150.600 ± 0.091 YES 150.600 ± 0.091 YES 145.50 to 154.50 psi
3 150.00 psi 149.960 ± 0.091 YES 149.960 ± 0.091 YES 145.50 to 154.50 psi
*********************************End of Report***********************************
Manufacturer
540.6 1503
Page 2 of 2