e vYOeUNAG NGG. & CALIBRATION SERVICES PVT LTD.

ff: 033 26771412,26671909, 82329 85881


Tele Fax: 033 2677 1411

Admn. Office & Lab: 44, Shahid Khudiram Bose Sarani (formerly Kamardanga Road), lchapur, Howrah-711 104
· - - -

E-mail: calibration@yeagroup.org *Visit us at: www.yeagroup.org ·


CIN: U74140WB2005PTC1 06389 *Toll Free.Help Desk: 1800 345 9001

Certificate No.: YEA(C)I l90515SAID5691001 Dated : 18.05.2019


ULR No.: CC222119000018684F

Calibrated for : Dee Vee Projects Limited


Medical College Site, Ganga pur,
Ambikapur, Chhattisgarh.

Calibrated on. Next Calibration Due On. Page No. No. of Pages
15.05.2019 15.05.2020 1 2

Description of Instrument
Nam e : Compression Testing Machine
Jdeutijicatiou : OVPL I CTM - 0 I
Capacity : 2000 kN
Make : NIF

Service Request No. I Order Reference 190515 SA/0569100 I -00 I

Receipt of Instrument 15.05.20 19

Location where calibration performed : At Client Premises

Applicable Specification of Instrument to be Calibrated : Not Specified

Condition oflnstrument to be Ca librated : Satisfactory

Basis o f Ca libration : Comparison method with reference to YEAIWI/M-59

Environmental Condition during Calibration : Temperature : 29 °C


Equipment used for Calibration : Standard Force Proving Instrument

Traceability : To National Standard vide Certificate No:


18050 15 6101.051C-65 Amend valid till 24.07.2020

Measurement Uncertainty : The figures are stated in the result at coverage factor
k = 2 corresponding to confidence probability of
95% fo r normal distribution.

Laboratory Identification : Sticker is affixed for authentication of cal ibration.

co,ro_
YEA(C)/F-08

~
~ ~eA V: 033 2677 1412, 2667 1909, 82329 85881
~ ( · Telefax:03326771411
, ...L•IIIH$1=1 ·~ [C;TCJI;Sffl'! I :J-;1.;)1 l•JIEi =t j\!4 [!J :If-' I :t\!j IIi j •J
Admn. Office & Lab : 44, Shahid Khudirani Bose Sarani (formerly Kamardanga Road), lchapur, Howrah-711 104
.E-mail: calibration@yeagroup.org *Visit us at: www.yeagroup.org
CIN : U74140WB2005PTC1 06389 *Toll Free Help Desk: 1800 345 9001
Certificate No.: YEA(C)/ 190SISSA/0569/00I Date: IS.05.2019
Calibrated By A. Pradhan Page 2 of 2 Pages

Calibration Results

Mechanical Calibration
A) Calibration Results of Compression Testing Machine :
Temperature Actual Observed Reading
Standa rd Complementary
Force Co rrected Average Reading
Reading Positio n o• Position 120• Position 240• Reading
Reading
(l{N) (d iv) (div) (d iv) (div) (div) (div) (div)
Scale : 0 - 2000 kN (Using 2000 kN Force Proving Instrument)
200 101.5 101.7 101.0 10 1.0 100.0 ---------- 100.7
400 202.6 202.9 201.0 202.0 202.0 ..................... 201.7
600 303.1 303.6 301.0 302.0 302.0 ---------- 301.7
800 403.3 404.0 403.0 404.0 403.0 ---------- 403.3
1000 503.1 503 .9 505.0 504.0 503.0 ..................... 504.0
1200 60 1.8 602.8 603.0 603.0 604.0 ---------- 603.3
1400 700.1 701.2 700.0 702.0 70 1.0 ---------- 701.0
1600 798.5 799.8 799.0 798.0 799.0 ---------- 798.7
1800 896.3 897.8 898.0 897.0 897.0 ..................... 897.3
---------- ...................... ---------- ---------- ---------- ---------- ---------- ----------

ouc Relative Error


Reading Overall
Force Afte•· Measurement
Set Zero Resolution Acct11·acy Repeatability
Removal of Unce rtainty
Force
(I{N) (I{N) (%) ('Yo) (%) (%) (± %)
Scale : 0 - 2000 kN (Using 2000 kN Force Proving Instrument)
200 I st. 0.0000 0.000 1.000 0.991 0.993
400 2nd. 0.0000 0.000 0.500 0.626 0.496
600 3rd. 0.0000 0.000 0.333 0.638 0.33 1
800 ..................... ---------- .. .................. 0.250 0.1 54 0.248
1000 ---------- ..................... ---------- 0.200 -0.0 17 0.397 1.82
1200 ---------- ---------- ---------- 0.167 -0.093 0.166
1400 ---------- ---------- ---------- 0.143 0.033 0.285
1600 ---------- ---------- ---------- 0.125 0. 14 1 0.125
1800 ---------- ---------- ---------- 0.111 0.047 0.111
.................... ---------- ---------- ---------- ---------- ---------- ---------- ----------
OUC: Device Under Calibration
REMARKS: The above Compression Testi ng Mach ine has been calibrated and the readings are
tabulated above.
NOTES This certificate refers to the value observed at the time of calibration and this should not be
reproduced other than in full except with prior written approval from the head of
calibration laboratory.