Oil Dispenser Accuracy

UUC Applied Value Run 1 Run 2 Run 3 Tolerance


1000 L 996.000 996.000 996.000 ±10.000

Refer to Cal data sheets front pages


Page 1 of 3
CALIBRATION CERTIFICATE
Certificate Number: VOC2023-4614
VC-RF-M-106,REV,NO: 03, REV.DATE: JUNE 2023

Customer Name : AAL IMRAN HOUSING WLL


Address : Industrial area, Street 41، Gate 77, Doha Delivery Date : 14-Aug-2023
CSR Number : CSR/OS/383/01 Calibration Date : 12-Aug-2023
Date Received : 12-Aug-2023 Suggested Due Date : 11-Aug-2024
Description : Flow Meter Validity : 1 Year
Brand / Make : B METERS Received Condition : In Tolerance
Model / Type : SK 14-MI001-SMU036 Returned Condition : In Tolerance
Serial Number : 210400097 Calibration Result : Pass
Equipment ID Number : N/A Location : On site
Size / Range : 999999 m3 Temperature : 24.5 °C
Resolution/Least Count : 0.0001 m3 Relative Humidity : 50.8 %RH
Calibration Procedure : As per manufacturer specification Page Number : 1 of 2

DETAILS OF MASTER INSTRUMENTS USED FOR CALIBRATION


DESCRIPTION ASSET NUMBER BRAND/MAKE MODEL/TYPE CAL DUE TRACEABILITY NUMBER
Ultrasonic Flow Meter VER-CAL-UFM-002 Taosonics TUF-2000H 6-Jul-2024 551220084451089
Digital Protractor VER-CAL-DP-001 Absolute IP65 20-Jun-2024 551220085153109
Barometric Pressure VER-CAL-BAR-001 Extech SD700 12-Jun-2024 2023061129

Remarks:
1. Averages of minimum three readings are reported.
2. Suggested due date and test point were given based on customer requirements.
3. These results are obtained at the time of calibration.
4. The user should determine the suitability of the instrument for its intended use.
5. Uncertainty of measurement at 95% confidence level by multiplying coverage factor k = 2.
6. Condition of instrument found satisfactory during receipt.
7. Our masters are directly calibrated through ISO/IEC 17025 accredited calibration lab having direct traceability with national /
International standard.
8. Vertex Calibration Laboratory is accredited in accordance with the recognized International Standard as per ISO/IEC 17025:2017.
9. UUC- Unit Under Calibration
10.Pass / Fail criteria is decided by comparing the error of the unit under calibration and its associated by measurement uncertainty
with the tolerance specified by the manufacturer / customer.
11.Accuracy, tolerance and acceptance criteria stated as per customer requirements.
12.Any hand written corrections (except @) or photocopies of the report invalidate this certificate.
13.Results reported outside the scope of accreditation are mark with asterisk (*).

Calibrated By: Approved By: Approved By:

Bekhdadi Youcef Mohideen Maajith Ronaldo Acosta


Calibration Engineer QA/QC Engineer Technical In-Charge

Calibration cycle are resulting due dates were submitted/approved by the customer.Any number of factors may cause an instrument to drift out of tolerance before the next
scheduled of calibration.Recalibration cycles should be based of frequency of used, environmental condition and customer's established systmatic accuracy.The information on
this report, pertains only to the instrument identified.

This certificate may not be reproduced other than full,except with prior written approval of issuing laboratory. This report shall not be valid if not accomplished with Vertex calibration official seal.
CALIBRATION RESULT
Certificate Number: VOC2023-4614
VC-RF-M-106,REV,NO: 03, REV.DATE: JUNE 2023

Description : Flow Meter CSR Number : CSR/OS/383/01


Calibration Procedure : As per manufacturer specification Calibration Date : 12-Aug-2023
Serial Number : 210400097 Suggested Due Date : 11-Aug-2024
Equipment ID Number : N/A Model / Type : SK 14-MI001-SMU036
Brand / Make : B METERS Size / Range : 999999.0°C
Temperature : 24.5 °C Location : On site
Humidity : 50.8 %RH Page Number : 2 of 2

Oil Dispenser Accuracy

Observed Reading of UUC


UUC Applied Value Deviation Tolerance Result Uncertainty
As Found As Left

1000 L 996.000 L 996.000 L -4.000 L ±20.000 L PASS ±4.3100 L

*** End of Calibration ***

Where statements of conformity are made in this report, the following decision rules are applied:
PASS - Results ± expanded uncertainty within limits / specification
FAIL - Results ± expanded uncertainty exceeds limit / specification
INDETERMINATE - Results ± expanded uncertainty overlaps limit / specification

Calibrated By: Approved By: Approved By:

Bekhdadi Youcef Mohideen Maajith Ronaldo Acosta


Calibration Engineer QA/QC Engineer Technical In-Charge