Calibration Certificate

ULR No.: CC309622000002529F


Certificate No.: CCTPL/0350/01 Field: Mechanical
SRF No.: 335 Date of Receipt: 16/07/2022
SRF Date: 16/07/2022 Date of Calibration: 16/07/2022
Company Name &
Address M/s. Achintya Healthcare Pvt. Ltd Recommended Due Date.:
15/07/2023
(As per agreed by customer)
Khasra No. 376, Gangol Road Behind
Shopia School Partapur, Meerut - 250103 Certificate Issue Date: 18/07/2022
DUC* Details
Instrument Name: Standard Weight Box Instrument Sr. No.: 942
Make:/Class -------/E2 Instrument Id No.: SWB-01
Material: S.S Location: Lab
Range: 1 mg to 200 g DUC Condition: Ok
Readability: NA Calibration Performed at: At Lab
Standard Equipments Used(Traceable to National Standard)
Sr. No Instrument Name Cert. No Cal. By Calibration Due Date
BLUEBOX
1 Standard Weight (E2) BBT/172/JUL/21 26/08/2022
TECHNO
BLUEBOX
2 Analytical Balance BBT/283/APR/22 05/05/2023
TECHNO
Temperature Humidity Reference Standard Calibration Procedure No.
Environment Condition
24.8 ºC 54 %RH OIML R 111-1 CCTPL/M/MV/01
CALIBRATION RESULT
Nominal Value Denomination Value Deviation Uncertainity
Sr.No.
g g g (± mg)
1 0.001 0.00100 0.00000 0.01
2 0.002 0.00200 0.00000 0.01
3* 0.002 0.00200 0.00000 0.01
4 0.005 0.00500 0.00000 0.01
5 0.01 0.01000 0.00000 0.01
6 0.02 0.02001 0.00001 0.01
7* 0.02 0.02001 0.00001 0.01
8 0.05 0.04999 -0.00001 0.01
9 0.1 0.09999 -0.00001 0.01
10 0.2 0.19998 -0.00002 0.01
11* 0.2 0.19998 -0.00002 0.01
12 0.5 0.50002 0.00002 0.01
13 1 1.00003 0.00003 0.02
14 2 2.00004 0.00004 0.02
15* 2 2.00003 0.00003 0.02
16 5 5.00005 0.00005 0.02
17 10 9.99996 -0.00004 0.02
18 20 19.99995 -0.00005 0.02
19* 20 19.99994 -0.00006 0.02
20 50 49.99992 -0.00008 0.025
21 100 99.9999 -0.0001 0.025
22 200 199.9998 -0.0002 0.069
23* 200 199.9997 -0.0003 0.069
Note : DUC* : Device Under Calibration
Uncertainty of measurment at approx 95% Confidence Level with coverage factor k = 2,

Calibrated By : Priyank Sharma Approved By : Sandeep Sharma


Calibration Engineer Director
Note. 1.This Certificate only to the particular instrument submitted for calibration.
2.This Caliibration Result reported in this certificate valid at the time of and under the stated condition of measurement.
3.This particular certificate can not be reproduce except in full written, without prior permission of CCTPL.
`---End of Certificate--- Page 1 of 1