Certificate No.: CCTPL/0350/01 Field: Mechanical SRF No.: 335 Date of Receipt: 16/07/2022 SRF Date: 16/07/2022 Date of Calibration: 16/07/2022 Company Name & Address M/s. Achintya Healthcare Pvt. Ltd Recommended Due Date.: 15/07/2023 (As per agreed by customer) Khasra No. 376, Gangol Road Behind Shopia School Partapur, Meerut - 250103 Certificate Issue Date: 18/07/2022 DUC* Details Instrument Name: Standard Weight Box Instrument Sr. No.: 942 Make:/Class -------/E2 Instrument Id No.: SWB-01 Material: S.S Location: Lab Range: 1 mg to 200 g DUC Condition: Ok Readability: NA Calibration Performed at: At Lab Standard Equipments Used(Traceable to National Standard) Sr. No Instrument Name Cert. No Cal. By Calibration Due Date BLUEBOX 1 Standard Weight (E2) BBT/172/JUL/21 26/08/2022 TECHNO BLUEBOX 2 Analytical Balance BBT/283/APR/22 05/05/2023 TECHNO Temperature Humidity Reference Standard Calibration Procedure No. Environment Condition 24.8 ºC 54 %RH OIML R 111-1 CCTPL/M/MV/01 CALIBRATION RESULT Nominal Value Denomination Value Deviation Uncertainity Sr.No. g g g (± mg) 1 0.001 0.00100 0.00000 0.01 2 0.002 0.00200 0.00000 0.01 3* 0.002 0.00200 0.00000 0.01 4 0.005 0.00500 0.00000 0.01 5 0.01 0.01000 0.00000 0.01 6 0.02 0.02001 0.00001 0.01 7* 0.02 0.02001 0.00001 0.01 8 0.05 0.04999 -0.00001 0.01 9 0.1 0.09999 -0.00001 0.01 10 0.2 0.19998 -0.00002 0.01 11* 0.2 0.19998 -0.00002 0.01 12 0.5 0.50002 0.00002 0.01 13 1 1.00003 0.00003 0.02 14 2 2.00004 0.00004 0.02 15* 2 2.00003 0.00003 0.02 16 5 5.00005 0.00005 0.02 17 10 9.99996 -0.00004 0.02 18 20 19.99995 -0.00005 0.02 19* 20 19.99994 -0.00006 0.02 20 50 49.99992 -0.00008 0.025 21 100 99.9999 -0.0001 0.025 22 200 199.9998 -0.0002 0.069 23* 200 199.9997 -0.0003 0.069 Note : DUC* : Device Under Calibration Uncertainty of measurment at approx 95% Confidence Level with coverage factor k = 2,
Calibrated By : Priyank Sharma Approved By : Sandeep Sharma
Calibration Engineer Director Note. 1.This Certificate only to the particular instrument submitted for calibration. 2.This Caliibration Result reported in this certificate valid at the time of and under the stated condition of measurement. 3.This particular certificate can not be reproduce except in full written, without prior permission of CCTPL. `---End of Certificate--- Page 1 of 1