0.01
Upper limit
0.005 Lower limit
Correction
0
1 2 3 4 5 6
-0.005
-0.01
-0.015
0.01
Upper limit
0.005 Lower limit
Correction
0
1 2 3 4 5 6
-0.005
-0.01
-0.015
According to the report from QC-580-02 Masterlist of IMTE (By usage Category) this
instrument is seldom use.
Suggestion:
To analyze an accurate data, 3 calibration history must be obtained. But due to the time
interval between the 1st calibration certificate until now (18.1.2016), the data obtained
from the 1st calibration cannot be used and the data collection need to be start again
and the suggested time interval is 3 month for 3 times calibration.
5.0 Reference
Reference:
[1] Bare A. (2006), Simplified Calibration Interval Analysis. U.S
1st calibration: 15.April.2014 2nd Calibration: 1.October.2014 3rd calibration: 6 April 2015
Uncertainty: ±0.004mm Uncertainty: ±0.012mm Uncertainty: ±0.003mm
Description: - Description: - Description: -
Calibration Method : JIS B7502 Calibration Method : JIS B7502 Calibration Method : JIS B7502
Uncertainty Correction Uncertainty Correction Uncertainty Correction
No (mm) No (mm) No (mm)
+ - + - + -
1 0.004 -0.004 0 1 0.012 -0.012 0 1 0.003 -0.003 0
2 0.004 -0.004 0 2 0.012 -0.012 0 2 0.003 -0.003 0
3 0.004 -0.004 0 3 0.012 -0.012 0 3 0.003 -0.003 0
4 0.004 -0.004 0 4 0.012 -0.012 0.004 4 0.003 -0.003 0
5 0.004 -0.004 0 5 0.012 -0.012 0.004 5 0.003 -0.003 0.004
6 0.004 -0.004 0.002 6 0.012 -0.012 0.008 6 0.003 -0.003 0.008
7 0.004 -0.004 0.002 7 0.012 -0.012 0.010 7 0.003 -0.003 0.020
8 0.004 -0.004 0.010 8 0.012 -0.012 0.010 8 0.003 -0.003 0.020
9 0.004 -0.004 0.010 9 0.012 -0.012 0.010 9 0.003 -0.003 0.020
10 0.004 -0.004 0.010 10 0.012 -0.012 0.010 10 0.003 -0.003 0.020
11 0.004 -0.004 0.010 11 0.012 -0.012 0.010 11 0.003 -0.003 0.010
1st Calibration
0.012
0.01
0.008 Upper limit
0.006 Lower limit
0.004 Correction
0.002
0
-0.002 1 2 3 4 5 6 7 8 9 10 11
-0.004
-0.006
2nd calibration
0.015
0.010
Upper limit
0.005 Lower limit
Correction
0.000
1 2 3 4 5 6 7 8 9 10 11
-0.005
-0.010
-0.015
3rd calibration
0.025
0.020
Upper limit
0.015 Lower limit
Correction
0.010
0.005
0.000
1 2 3 4 5 6 7 8 9 10 11
-0.005
According to the report from QC-580-02 Masterlist of IMTE (By usage Category) this
instrument is unused.
5.0 Reference
1st calibration
0.015
0.01
Upper limit
0.005
Lower limit
Correction
0
1 2 3 4 5 6 7 8 9 10 11
-0.005
-0.01
-0.015
2nd calibration
0.015
0.01
Upper limit
0.005
Lower limit
Correction
0
1 2 3 4 5 6 7 8 9 10 11
-0.005
-0.01
-0.015
3rd calibration
0.025
0.02
0.015
Upper limit
0.01 Lower limit
Correction
0.005
0
1 2 3 4 5 6 7 8 9 10 11
-0.005
-0.01
-0.015
According to the report from QC-580-02 Masterlist of IMTE (By usage Category) this instrument is unused.
Suggestion:
Calibration does not need to be done. (Cost effective)
5.0 Reference
1st calibration
0.0030
0.0020
Upper limit
0.0010
Lower limit
Correction
0.0000
1 2 3 4 5 6 7 8 9 10 11
-0.0010
-0.0020
-0.0030
2nd calibration
0.004
0.003
0.002
Upper limit
0.001 Lower limit
Correction
0.000
1 2 3 4 5 6 7 8 9 10 11
-0.001
-0.002
-0.003
-0.004
3rd calibration
0.015
0.010
Upper limit
0.005
Lower limit
Correction
0.000
1 2 3 4 5 6 7 8 9 10 11
-0.005
-0.010
-0.015
According to the report from QC-580-02 Masterlist of IMTE (By usage Category) this instrument is unused.
5.0 Reference
1st calibration
0.015
0.01
Upper limit
0.005
Lower limit
Correction
0
1 2 3 4 5 6 7 8 9 10 11
-0.005
-0.01
-0.015
According to the report from QC-580-02 Masterlist of IMTE (By usage Category) this
instrument is unused.
5.0 Reference
5.0 Reference
Calibration Report
According to the report from QC-580-02 Masterlist of IMTE (By usage Category) this instrument
is being use.
5.0 Reference
According to the report from QC-580-02 Masterlist of IMTE (By usage Category) this
instrument is used.
Suggestion:
The instrument need to be calibrate at least 3 times for 3 month each time. The time
interval between the period must be constant to ensure the relevant and accurate data
can be obtained.
IM-MT-
5.0 Reference
According to the report from QC-580-02 Masterlist of IMTE (By usage Category) this
instrument is used.
5.0 Reference
According to the report from QC-580-02 Masterlist of IMTE (By usage Category) this instrument is used.
Suggestion:
Calibration interval is increase to 6 month. (until futher analysis)
5.0 Reference
-0.1
-0.1
According to the report from QC-580-02 Masterlist of IMTE (By usage Category) this
instrument is used.
5.0 Reference
According to the report from QC-580-02 Masterlist of IMTE (By usage Category) this instrument is
used.
Suggestion:
Calibration is needed for 3 times with 3 month interval period before the interval can be extend.
5.0 Reference
6 0.09 -0.09 0
According to the report from QC-580-02 Masterlist of IMTE (By usage Category) this instrument is used.
Suggestion:
Calibration is being extend to 6 month interval until futher data obtained.
5.0 Reference
According to the report from QC-580-02 Masterlist of IMTE (By usage Category) this
instrument is used.
5.0 Reference
4th calibration:
24.September.2014 5th calibration: 26.January.2015
Uncertainty: ±0.008mm Uncertainty: ±0.004mm
Description: - Description: -
Calibration Method : JIS B7502 Calibration Method : JIS B7507
Uncertainty Correction Uncertainty Correction
No (mm) No (mm)
+ - + -
1 0.008 -0.008 -0.002 1 0.004 -0.004 -0.002
2 0.008 -0.008 -0.004 2 0.004 -0.004 0.003
3 0.008 -0.008 -0.002 3 0.004 -0.004 0.005
4 0.008 -0.008 -0.004
5 0.008 -0.008 -0.012
6 0.008 -0.008 -0.012
7 0.008 -0.008 -0.012
5th calibration
0.006
0.004
0.002
0
1 2 3
-0.002
-0.004
-0.006
According to the report from QC-580-02 Masterlist of IMTE (By usage Category) this
instrument is unused.
5.0 Reference
3rd calibration:
1st calibration: 3.December.2013 2nd calibration: 22.April.2014 24.September.2014
Uncertainty: ±0.02mm Uncertainty: ±0.012mm Uncertainty: ±0.016mm
Description: - Description: Right Scale Description: -
Calibration Method : JIS B7507 Calibration Method : JIS B7502 Calibration Method : JIS B7502
Uncertainty Correction Uncertainty Correction Uncertainty Correction
No (mm) No (mm) No (mm)
+ - + - + -
1 0.02 -0.02 0 1 0.012 -0.012 -0.062 1 0.016 -0.016 -0.002
2 0.02 -0.02 -0.01 2 0.012 -0.012 -0.082 2 0.016 -0.016 -0.022
3 0.02 -0.02 0 3 0.012 -0.012 -0.1 3 0.016 -0.016 -0.022
4 0.012 -0.012 -0.04 4 0.016 -0.016 -0.01
5 0.012 -0.012 -0.041 5 0.016 -0.016 -0.011
6 0.012 -0.012 -0.04 6 0.016 -0.016 -0.01
7 0.012 -0.012 -0.037 7 0.016 -0.016 -0.017
8 0.012 -0.012 -0.029 8 0.016 -0.016 -0.009
According to the report from QC-580-02 Masterlist of IMTE (By usage Category) this
instrument is unused.
Suggestion:
Calibration is not necessary. (Cost effective)
5.0 Reference
Uncertainty Correction
No (mm)
+ -
1 0.007 -0.007 -0.002
2 0.007 -0.007 -0.004
3 0.007 -0.007 -0.011
According to the report from QC-580-02 Masterlist of IMTE (By usage Category) this instrument
is unused.
5.0 Reference
According to the report from QC-580-02 Masterlist of IMTE (By usage Category) this
instrument is unused.
Suggestion:
No calibration is necessary.
5.0 Reference
0.01 0.010
0.005 0.005
0 0.000
1 2 3 4 5 6 1 2 3 4 5 6
-0.005 -0.005
-0.01 -0.010
-0.015 -0.015
3rd calibration
0.025
0.020
0.015
0.010
0.005
0.000
-0.005 1 2 3 4 5 6
-0.010
-0.015
-0.020
-0.025
According to the report from QC-580-02 Masterlist of IMTE (By usage Category) this instrument is
used.
5.0 Reference
0.000 0.005
1 3 5 7 9 11 13 15 17 19 21 23 25 27 29 31
-0.005 0.000
1 4 7 10 13 16 19 22 25 28 31
-0.010 -0.005
-0.015 -0.010
-0.020 -0.015
0.01 0.015
0.010
0.005
0.005
0
0.000
1 4 7 10 13 16 19 22 25 28 31
-0.005
-0.005
1 4 7 10 13 16 19 22 25 28 31
-0.01 -0.010
-0.015 -0.015
-0.005
1 4 7 10 13 16 19 22 25 28 31
-0.010
-0.01
-0.015
-0.015
-0.020 -0.02
0.005 0.015
0.000 0.010
-0.005
1 4 7 10 13 16 19 22 25 28 31 0.005
-0.010 0.000
-0.015 -0.005
1 4 7 10 13 16 19 22 25 28 31
-0.020 -0.010
According to the report from QC-580-02 Masterlist of IMTE (By usage Category) this
instrument is used.
5.0 Reference
0.004 0.004
0.003 0.003
0.002 0.002
0.001 0.001
0 0.000
-0.001
1 3 5 7 9 11 13 15 17 19 21 -0.001
1 3 5 7 9 11 13 15 17 19 21
-0.002 -0.002
-0.003 -0.003
-0.003 -0.003
According to the report from QC-580-02 Masterlist of IMTE (By usage Category) this
instrument is used.
5.0 Reference
According to the report from QC-580-02 Masterlist of IMTE (By usage Category) this instrument is
used.
5.0 Reference
Page 121
Calibration Method
Page 122
Calibration Method
According to the report from QC-580-02 Masterlist of IMTE (By usage Category) this instrument
is used.
Page 123
Calibration Method
Suggestion:
Calibration is needed for 3 times and 3 month interval each.
Page 124
Calibration Method
5.0 Reference
Page 125
REPORT FOR IM-MT-024
Details of the IMTE
Instrument's name: Digimatic caliper
Serial Number: 108726
Manufacturer: Mitutoyo
According to the report from QC-580-02 Masterlist of IMTE (By usage Category) this
instrument is used.
5.0 Reference
According to the report from QC-580-02 Masterlist of IMTE (By usage Category) this
instrument is used.
Suggestion:
New calibration slot need to be done for 3 times and 3 month interval each.
5.0 Reference
1st calibration:
20.November.2014
Uncertainty: ±2.8Nm
Description: -
Calibration Method : JIS B4650
Uncertainty Correction
No
+ - (mm)
1 2.8 -2.8 1.4
2 2.8 -2.8 2.6
3 2.8 -2.8 3.2
4 2.8 -2.8 6.8
5 2.8 -2.8 6.9
6 2.8 -2.8 8.4
7 2.8 -2.8 8.8
1st calibration
10
8
6
4
2
0
1 2 3 4 5 6 7
-2
-4
According to the report from QC-580-02 Masterlist of IMTE (By usage Category) this
instrument is seldom unused.
5.0 Reference
According to the report from QC-580-02 Masterlist of IMTE (By usage Category) this instrument
is seldom unused.
REPORT FOR IM-MT-028
Suggestion:
Calibration is not necessary until futher report have been identified.
REPORT FOR IM-MT-028
5.0 Reference
REPORT FOR IM-MT-029
According to the report from QC-580-02 Masterlist of IMTE (By usage Category) this instrument
is used.
REPORT FOR IM-MT-029
Suggestion:
Calibration is necessary.
REPORT FOR IM-MT-029
5.0 Reference
REPORT FOR IM-MT-030
According to the report from QC-580-02 Masterlist of IMTE (By usage Category) this instrument
is seldom unused.
REPORT FOR IM-MT-030
Suggestion:
Calibration is not necessary until futher report have been identified.
REPORT FOR IM-MT-030
5.0 Reference
REPORT FOR IM-MT-031
According to the report from QC-580-02 Masterlist of IMTE (By usage Category) this instrument
is used.
REPORT FOR IM-MT-031
Suggestion:
Calibration is needed.
REPORT FOR IM-MT-031
5.0 Reference
REPORT FOR IM-NE-001
According to the report from QC-580-02 Masterlist of IMTE (By usage Category) this
instrument is being used.
REPORT FOR IM-NE-001
5.0 Reference
REPORT FOR IM-NE-002
5.0 Reference
Calibration Report
1st calibration
0.150
0.100
Upper limit
0.050 Lower limit
Correction
0.000
1 2 3
-0.050
-0.100
-0.150
According to the report from QC-580-02 Masterlist of IMTE (By usage Category) this
instrument is being unused.
Suggestion:
No calibration necessary.
5.0 Reference
1st calibration:
18.November.2015
Uncertainty: ±2.7-2.8µm
Description: -
Calibration Method :
Uncertainty Correction
No
+ - (µm)
1 2.7 -2.7 0
2 2.7 -2.7 0
3 2.7 -2.7 0
4 2.8 -2.8 1.2
5 2.8 -2.8 0.5
6 2.8 -2.8 0.3
7 2.8 -2.8 0.7
8 2.8 -2.8 0.6
1st calibration
4
3
2 Upper limit
1 Lower limit
Correction
0
-1 1 2 3 4 5 6 7 8
-2
-3
-4
According to the report from QC-580-02 Masterlist of IMTE (By usage Category) this
instrument is used.
Suggestion:
Calibration is needed.
5.0 Reference
Uncertainty: ±(146-179psi)
Description:
Calibration Method : Test Gauge
SM231
Uncertainty Correction
No
+ - (psi)
1 146 -146 0
2 152 -152 153
3 192 -192 -83
4 167 -167 -244
5 179 -179 -290
6 179 -179 -405
Page 181
Calibration Report
1st calibration
300
200
100 Upper limit
0 Lower limit
1 2 3 4 5 6 Correction
-100
-200
-300
-400
-500
Page 182
Calibration Report
According to the report from QC-580-02 Masterlist of IMTE (By usage Category) this
instrument is used.
Page 183
Calibration Report
Suggestion:
Another 2 times calibration with 3 month interval have to be done. Calibration is
neccesary.
Page 184
Calibration Report
5.0 Reference
Page 185
Calibration Report
According to the report from QC-580-02 Masterlist of IMTE (By usage Category) this
instrument is used.
Suggestion:
Calibration is necessary.
5.0 Reference
Uncertainty Correction
No (psi)
+ -
1 123 -123 -17
2 123 -123 33
3 123 -123 0
4 123 -123 67
5 123 -123 100
6 123 -123 67
7 123 -123 117
8 123 -123 83
9 123 -123 117
100
50
50
0
1 2 3 4 5 0
-50 1 2 3 4
-50
-100 -100
-150 -150
According to the report from QC-580-02 Masterlist of IMTE (By usage Category) this
instrument is Not In Use.
5.0 Reference
4 4
2 2
0 0
1 2 3 4 5 6 1 2 3 4 5 6
-2 -2
-4 -4
-6 -6
-8 -8
According to the report from QC-580-02 Masterlist of IMTE (By usage Category) this instrument
is not in use.
5.0 Reference
5.0 Reference
6 60
4 40
2 20
0 0
1 2 3 1 2 3 4 5 6 7
-2 -20
-4 -40
-6 -60
-8 -80
3rd calibration
60
40
20
0
1 2 3 4 5 6 7
-20
-40
-60
According to the report from QC-580-02 Masterlist of IMTE (By usage Category) this
instrument is seldom used.
Suggestion:
Calibration interval is increase to 6 month.
5.0 Reference
1st calibration
80
60
40
20
0
1 2 3 4 5 6 7
-20
-40
-60
-80
According to the report from QC-580-02 Masterlist of IMTE (By usage Category) this
instrument is seldom used.
5.0 Reference
1st calibration
60
40
20
0
1 2 3 4 5 6 7
-20
-40
-60
-80
-100
-120
According to the report from QC-580-02 Masterlist of IMTE (By usage Category) this
instrument is seldom used.
5.0 Reference
1st calibration:
30.November.2015
Uncertainty: ± 35-58psi
Description: -
Calibration Method : Test Gauge
SM054
Uncertainty Correction
No (psi)
+ -
1 35 -35 0
2 58 -58 18
3 58 -58 18
4 56 -56 17
5 43 -43 2
6 50 -50 3
7 45 -45 -100
1st calibration
80
60
40
20 Upper limit
0 Lower limit
1 2 3 4 5 6 7 Correction
-20
-40
-60
-80
-100
-120
According to the report from QC-580-02 Masterlist of IMTE (By usage Category) this
instrument is used.
5.0 Reference
1st calibration
100
80
60
Upper limit
40
Lower limit
20 Correction
0
-20 1 2 3 4 5 6 7
-40
-60
-80
According to the report from QC-580-02 Masterlist of IMTE (By usage Category) this
instrument is used.
Suggestion:
Calibration is needed.
5.0 Reference
1st calibration
5
4
3
Upper limit
2
Lower limit
1
Correction
0
-1 1 2 3 4 5
-2
-3
-4
-5
REPORT FOR IM-PG-011
According to the report from QC-580-02 Masterlist of IMTE (By usage Category) this
instrument is seldom used.
REPORT FOR IM-PG-011
5.0 Reference
Calibration Report
Uncertainty: ± 2.9-3.8bar
Description: -
Calibration Method : Test Gauge
SM 054
Uncertainty Correction
No (bar)
+ -
1 2.9 -2.9 0
2 3.8 -3.8 -1.6
3 3.8 -3.8 0.1
4 3.8 -3.8 2.4
5 3.8 -3.8 6.2
1st calibration
8
4 Upper limit
Lower limit
2 Correction
0
1 2 3 4 5
-2
-4
-6
According to the report from QC-580-02 Masterlist of IMTE (By usage Category) this
instrument is sedom used.
Suggestion:
Calibration is not necessary.
5.0 Reference
1st calibration
60
40
20 Upper limit
0 Lower limit
1 2 3 4 5 6 7 Correction
-20
-40
-60
-80
-100
2nd calibration
60
40
Upper limit
20
Lower limit
Correction
0
1 2 3 4 5 6 7
-20
-40
-60
According to the report from QC-580-02 Masterlist of IMTE (By usage Category) this
instrument is used.
5.0 Reference
1st calibration
1.5
1
0.5
0 Upper limit
1 2 3 4 5 Lower limit
-0.5 Correction
-1
-1.5
-2
-2.5
-3
According to the report from QC-580-02 Masterlist of IMTE (By usage Category) this
instrument is seldom used.
Suggestion:
Calibration is not necessary.
5.0 Reference
1st calibration
1.5
-1
-1.5
-2
According to the report from QC-580-02 Masterlist of IMTE (By usage Category) this instrument
is used.
Report on:
Section 2.0 shows the graph which construct from the calibration certificate on 18.11.2015. The
graph show that most of the point is inside the upper limit and lower limit of the uncertainty
except for the 2nd point which is lower than the lower limit. The instrument is use in the
production section. To have a sufficient and more reliable information to do the comparison at
least 3 calibration certificate need to be obtained with consistent 3 month interval.
5.0 Reference
2nd calibration:
1st calibration: 14.August.2015 30.November.2015
Uncertainty: ±0.6-0.7bar Uncertainty: ±6.1-8.8psi
Description: - Description: -
Calibration Method : Test Gauge Calibration Method : Test
SM064 Gauge SM064
Uncertainty Correction Uncertainty Correctio
No (bar) No
+ - + - n (psi)
1 0.6 -0.6 0 1 6.1 -6.1 0
2 0.6 -0.6 0.2 2 7.9 -7.9 5.3
3 0.7 -0.7 -0.4 3 8.8 -8.8 2.1
4 0.6 -0.6 -1.6 4 7.7 -7.7 -6.9
5 0.6 -0.6 -2 5 6.9 -6.9 4.2
6 0.7 -0.7 -1.9 6 8.8 -8.8 12.7
1st calibration
1
0.5
Upper limit
0
Lower limit
-0.5 1 2 3 4 5 6
Correction
-1
-1.5
-2
-2.5
2nd calibration
15
10
Upper limit
Lower limit
5
Correction
0
1 2 3 4 5 6
-5
-10
According to the report from QC-580-02 Masterlist of IMTE (By usage Category) this
instrument is used.
Calibration Report
Report on:
As shown in section 2.0, the graph is being construct by using the data obtained in the
calibration certificate on 14.8.2015 and 30.11.2015. The graph shows that most of the
point for both calibration graph is within the upper limit and lower limit of the
uncertainty. The instrument is used in the production section and as mentioned in
section 3.0 above the instrument is used. At least 3 calibration certificate is needed to
analyze and to make comparison with consistent 3 month interval.
The status is Used
Suggestion:
Calibration is needed.
5.0 Reference
2nd calibration:
1st calibration: 3.August.2015 18.November.2015
Uncertainty: ±3.6-7.6psi Uncertainty: ±2.9-4.9psi
Description: - Description: -
Calibration Method : Test Gauge Calibration Method : Test
SM104 Gauge SM055, SM104
Uncertainty Correction Uncertainty Correctio
No (psi) No n (psi)
+ - + -
1 3.6 -3.6 0 1 2.9 -2.9 0
2 7.5 -7.5 0.7 2 2.9 -2.9 4.8
3 4.6 -4.6 -4.2 3 2.9 -2.9 -0.5
4 5.2 -5.2 -10.7 4 2.9 -2.9 -4.7
5 4.7 -4.7 -8.5 5 4.7 -4.7 -12.0
6 7.6 -7.6 -4.7 6 4.9 -4.9 -7.2
1st calibration
10
5
Upper limit
0 Lower limit
1 2 3 4 5 6 Correction
-5
-10
-15
2nd calibration
6
4
2
0 Upper limit
-2 1 2 3 4 5 6 Lower limit
Correction
-4
-6
-8
-10
-12
-14
According to the report from QC-580-02 Masterlist of IMTE (By usage Category) this
instrument is used.
Suggestion:
Calibration is needed.
5.0 Reference
1st calibration
1.5
1
Upper limit
0.5 Lower limit
Correction
0
1 2 3 4 5 6
-0.5
-1
-1.5
According to the report from QC-580-02 Masterlist of IMTE (By usage Category) this
instrument is seldom used.
5.0 Reference
Category) this
n certificate on
and lower limit
o have a better
se in
t is sldom use.
1st calibration
0.25
0.2
0.15
Upper limit
0.1
Lower limit
0.05
Deviation
0
-0.05 1 2 3 4 5 6 7 8 9
-0.1
-0.15
-0.2
-0.25
2nd calibration
60
40
Upper limit
20 Lower limit
Correction
0
1 2 3 4 5 6 7
-20
-40
-60
According to the report from QC-580-02 Masterlist of IMTE (By usage Category) this
instrument is used.
Suggestion:
Calibration is necessary.
5.0 Reference
1st calibration
0.25
0.2
0.15
Upper limit
0.1
Lower limit
0.05
Deviation
0
-0.05 1 2 3 4 5 6 7 8 9
-0.1
-0.15
-0.2
-0.25
According to the report from QC-580-02 Masterlist of IMTE (By usage Category) this
instrument is seldom used.
5.0 Reference
1st calibration
0.1
0.08
0.06
Upper limit
0.04
Lower limit
0.02
Deviation
0
-0.02 1 2 3 4 5 6 7 8 9
-0.04
-0.06
-0.08
-0.1
According to the report from QC-580-02 Masterlist of IMTE (By usage Category) this
instrument is seldom used.
5.0 Reference
Uncertainty: ±0.0058-0.0069MPa
Description: -
Calibration Method : -
Uncertainty Deviation
No
+ - (MPa)
1 0.0058 -0.0058 0.000
2 0.0060 -0.0060 0.009
3 0.0067 -0.0067 0.007
4 0.0066 -0.0066 0.006
5 0.0069 -0.0069 0.003
6 0.0061 -0.0061 -0.003
7 0.0061 -0.0061 -0.003
8 0.0069 -0.0069 -0.004
9 0.0059 -0.0059 -0.004
1st calibration
0.0100
0.0080
0.0060
Upper limit
0.0040
Lower limit
0.0020 Deviation
0.0000
-0.0020 1 2 3 4 5 6 7 8 9
-0.0040
-0.0060
-0.0080
According to the report from QC-580-02 Masterlist of IMTE (By usage Category) this
instrument is seldom used.
Report on:
As shown in section 2.0 above, the graph construct is based on the calibration certificate on
9.2.2015. The graph show that most of the ppoint is within upper limit and lower limit,
except for the 2nd point which is above the upper limit. The instrument is used in
production section and as mentioned in section 3.0 the instrument is seldom used.
The status is : Not In Use
Suggestion:
Calibration is not necessary.
5.0 Reference
Uncertainty: ±0.0058-0.0066MPa
Description: -
Calibration Method : -
Uncertainty Deviation
No (MPa)
+ -
1 0.0058 -0.006 0.000
2 0.0063 -0.006 0.007
3 0.0066 -0.007 0.008
4 0.0059 -0.006 0.008
5 0.0065 -0.007 0.002
6 0.0064 -0.006 -0.004
7 0.0059 -0.006 -0.001
8 0.0062 -0.006 -0.003
9 0.0058 -0.006 -0.004
1st calibration
0.01
0.008
0.006
Upper limit
0.004 Lower limit
0.002 Deviation
0
-0.002 1 2 3 4 5 6 7 8 9
-0.004
-0.006
-0.008
According to the report from QC-580-02 Masterlist of IMTE (By usage Category) this
instrument is seldom used.
5.0 Reference
1st calibration
0.0150
0.0100
Upper limit
0.0050 Lower limit
Deviation
0.0000
1 2 3 4 5 6 7 8 9
-0.0050
-0.0100
-0.0150
According to the report from QC-580-02 Masterlist of IMTE (By usage Category) this
instrument is seldom used.
Suggestion:
Calibration is not necessary.
5.0 Reference
1st calibration
0.0100
0.0080
0.0060
Upper limit
0.0040
Lower limit
0.0020 Deviation
0.0000
-0.0020 1 2 3 4 5 6 7 8 9
-0.0040
-0.0060
-0.0080
According to the report from QC-580-02 Masterlist of IMTE (By usage Category) this
instrument is seldom used.
Report on:
As shows in section 2.0 above the graph is being construct by using the data from the calibration
certificate on 9.2.2015. The graph shows that most of the point is inside the uppe limit and lower
limit of uncertainty except for point number 4. The instrument is use at production section and it
is seldom used.
The status is : Not In Use
Suggestion:
Calibration is not necessary.
5.0 Reference
1st calibration
0.08
0.06
0.04 upper limit
0.02 Lower limit
Deviation
0
-0.02 1 2 3
-0.04
-0.06
-0.08
According to the report from QC-580-02 Masterlist of IMTE (By usage Category) this
instrument is seldom used.
Suggestion:
Calibration is not needed.
5.0 Reference
5.0 Reference
1st calibration
0.015
0.01
Upper limit
0.005
Lower limit
Correction
0
1 2 3 4 5 6 7
-0.005
-0.01
-0.015
REPORT FOR IM-TS-002
According to the report from QC-580-02 Masterlist of IMTE (By usage Category) this
instrument is seldom used.
REPORT FOR IM-TS-002
Suggestion:
No calibration is required.
REPORT FOR IM-TS-002
5.0 Reference