Calibration Report

REPORT FOR IM-MT-013


Details of the IMTE
Instrument's name: Analog Vernier caliper
Serial Number: 11335638
Manufacturer: Mitutoyo

1.0 Calibration History


Actual Reading = Instrument Reading + Correction
Correction = Actual Reading - Instrument Reading
Correction
Uncertainty = x100
Instruments Re ading

1st calibration: 20.November.2014 2nd calibration: 22.January.2015 3rd calibration: 12.May.2015


Uncertainty: ±0.09mm Uncertainty: ±0.09mm Uncertainty: ±0.09mm
Description: Outside&Inside Description: Outside&Inside Description: Outside&Inside
Calibration Method : JIS B7507 Calibration Method : JIS B7507 Calibration Method : JIS B7507
Uncertainty Correction Uncertainty Correction Uncertainty Correction
No (mm) No (mm) No (mm)
+ - + - + -
1 0.09 -0.09 0 1 0.09 -0.09 0 1 0.09 -0.09 0
2 0.09 -0.09 0 2 0.09 -0.09 0 2 0.09 -0.09 0.02
3 0.09 -0.09 0 3 0.09 -0.09 0 3 0.09 -0.09 0.02
4 0.09 -0.09 0 4 0.09 -0.09 0 4 0.09 -0.09 0.02
5 0.09 -0.09 0 5 0.09 -0.09 0 5 0.09 -0.09 0.02
6 0.09 -0.09 0 6 0.09 -0.09 0 6 0.09 -0.09 0.02
7 0.09 -0.09 0 1 0.09 -0.09 0 7 0.09 -0.09 0.01
8 0.09 -0.09 0 2 0.09 -0.09 -0.05 1 0.09 -0.09 0
1 0.09 -0.09 0 3 0.09 -0.09 -0.05 2 0.09 -0.09 0
2 0.09 -0.09 0 4 0.09 -0.09 -0.05 3 0.09 -0.09 0
3 0.09 -0.09 0 5 0.09 -0.09 -0.05 4 0.09 -0.09 0
4 0.09 -0.09 0 6 0.09 -0.09 -0.05 5 0.09 -0.09 0
5 0.09 -0.09 0 6 0.09 -0.09 0
6 0.09 -0.09 0
7 0.09 -0.09 0
8 0.09 -0.09 0
4th calibration: 4.September.2015
Uncertainty: ±0.09mm
Description: Outside&Inside
Calibration Method : JIS B7507
Uncertainty Correction
No (mm)
+ -
1 0.09 -0.09 0
2 0.09 -0.09 0
3 0.09 -0.09 0
4 0.09 -0.09 0
5 0.09 -0.09 0
6 0.09 -0.09 0
7 0.09 -0.09 0
1 0.09 -0.09 0
2 0.09 -0.09 0
3 0.09 -0.09 0
4 0.09 -0.09 0
5 0.09 -0.09 0

QA/QC Department Page 1 PT Borneo Mandiri Investment


Calibration Report

6 0.09 -0.09 0

QA/QC Department Page 2 PT Borneo Mandiri Investment


Calibration Report

REPORT FOR IM-MT-013

2.0 Calibration Graph

1st calibration (Outside) 1st calibration (Inside)


0.1 0.1
0.08 0.08
0.06 0.06
0.04 0.04
0.02 0.02
0 0
-0.02 1 2 3 4 5 6 7 8 -0.02 1 2 3 4 5 6 7 8
-0.04 -0.04
-0.06 -0.06
-0.08 -0.08
-0.1 -0.1

2nd calibration 2nd calibration (Inside)


(Outside) 0.1
0.08
0.1
0.06
0.08
0.06 0.04
0.04 0.02
0.02 0
0 -0.02 1 2 3 4 5 6
-0.02 1 2 3 4 5 6 -0.04
-0.04 -0.06
-0.06 -0.08
-0.08
-0.1
-0.1

3rd calibration 3rd calibration (Inside)


(Outside) 0.1
0.08
0.1
0.08 0.06
0.06 0.04
0.04 0.02
0.02 0
0 -0.02 1 2 3 4 5 6
-0.02 1 2 3 4 5 6 7
-0.04 -0.04
-0.06 -0.06
-0.08 -0.08
-0.1 -0.1

QA/QC Department Page 3 PT Borneo Mandiri Investment


Calibration Report

REPORT FOR IM-MT-013

2.0 Calibration Graph

4th calibration 4th calibration (Inside)


(Outside) 0.1
0.08
0.1
0.08 0.06
0.06 0.04
0.04 0.02
0.02 0
0 -0.02 1 2 3 4 5 6
-0.02 1 2 3 4 5 6 7 -0.04
-0.04
-0.06 -0.06
-0.08 -0.08
-0.1 -0.1

QA/QC Department Page 4 PT Borneo Mandiri Investment


Calibration Report

REPORT FOR IM-MT-013

3.0 Calibration Frequency

According to the report from QC-580-02 Masterlist of IMTE (By usage Category) this instrument is used.

QA/QC Department Page 5 PT Borneo Mandiri Investment


Calibration Report

REPORT FOR IM-MT-013

4.0 Calibration Report

Report on: January 2016


In section 2.0 the graph being construct is from the calibration certificate on the 20.11.2014, 22.1.2015,
12.5.2015, and 4.9.2015. All the graph shows the point is within the upper limit and lower limit of uncertainty.
The instrument is being use in the production/QC section which needed high reliability of the instrument. The
calibration certificate available has reach 3 times and the interval also within 3 month period. So, the analysis by
using these data is valid. Due to the result obtained from the graph the interval of the calibration is increase for
another 3 month. So, the total calibration interval for this instrument after last calibration is 6 month until futher
calibration data is obtained.

Suggestion:
Calibration is being extend to 6 month interval until futher data obtained.

QA/QC Department Page 6 PT Borneo Mandiri Investment


Calibration Report

REPORT FOR IM-MT-013

5.0 Reference

QA/QC Department Page 7 PT Borneo Mandiri Investment