Calibration Report

REPORT FOR IM-MT-018 (UNUSED)

Details of the IMTE


Instrument's name: Digital caliper
gauge
Serial Number: HE07L005
Manufacturer: Mitutoyo

1.0 Calibration History


Actual Reading = Instrument Reading + Correction
Correction = Actual Reading - Instrument Reading
Correction
Uncertainty = x100
Instruments Re ading
3rd calibration:
1st calibration: 3.December.2013 2nd calibration: 22.April.201424.September.2014
Uncertainty: ±0.02mm Uncertainty: ±0.012mm Uncertainty: ±0.016mm
Description: - Description: Right Scale Description:-
Calibration Method : JIS B7507 Calibration Method : JIS B7502 Calibration Method : JIS B7502
Uncertainty Correction Uncertainty Correction Uncertainty Correction
No (mm) No (mm) No (mm)
+ - + - + -
1 0.02 -0.02 0 1 0.012 -0.012 -0.0014 1 0.016 -0.016 -0.014
2 0.02 -0.02 0 2 0.012 -0.012 0.009 2 0.016 -0.016 -0.002
3 0.02 -0.02 0 3 0.012 -0.012 0.01 3 0.016 -0.016 0
4 0.012 -0.012 0.004 4 0.016 -0.016 0.004

4th calibration: 26.January.2015


Uncertainty: ±0.007mm
Description: -
Calibration Method : JIS B7507
Uncertainty Correction
No (mm)
+ -
1 0.007 -0.007 -0.011
2 0.007 -0.007 -0.005
3 0.007 -0.007 0.001

QA/QC Department Page 1 PT Borneo Mandiri Investment


Calibration Report

REPORT FOR IM-MT-018

2.0 Calibration Graph

1st calibration 2nd calibration


0.025 0.015
0.02
0.015 0.01
0.01 0.005
0.005
0 0
-0.005 1 2 3 1 2 3 4
-0.01 -0.005
-0.015 -0.01
-0.02
-0.025 -0.015

3rd calibration 4th calibration


0.02 0.008
0.015 0.006
0.004
0.01
0.002
0.005 0
0 -0.002 1 2 3
1 2 3 4 -0.004
-0.005
-0.006
-0.01
-0.008
-0.015 -0.01
-0.02 -0.012

QA/QC Department Page 2 PT Borneo Mandiri Investment


Calibration Report

REPORT FOR IM-MT-018

3.0 Calibration Frequency

According to the report from QC-580-02 Masterlist of IMTE (By usage Category) this
instrument is unused.

QA/QC Department Page 3 PT Borneo Mandiri Investment


Calibration Report

REPORT FOR IM-MT-018

4.0 Calibration Report

Report on: January 2016


As shown in section 2.0 above the graphs being constructed by using the data obtained
form the calibration certificate from 3.12.2013, 22.4.2014, 24.9.2014, and 26.1.2015.
From the graph, it shows that most of the point is within the upper limit and lower limit
of the uncertainty. This indicate that the instrument is still in calibrate. Apart from that,
the instrument also use in production/QC section but from section 3.0 stated that the
instrument is unused which indicate that it not necessary to be calibrate.

Suggestion:
No calibration is necessary.

QA/QC Department Page 4 PT Borneo Mandiri Investment


Calibration Report

REPORT FOR IM-MT-018

5.0 Reference

QA/QC Department Page 5 PT Borneo Mandiri Investment