Calibration Report

REPORT FOR IM-MT-021


Details of the IMTE
Instrument's name: Dial indicator & Holder
Serial Number: 7010S-10
Manufacturer: Mitutoyo

1.0 Calibration History


Actual Reading = Instrument Reading + Correction
Correction = Actual Reading - Instrument Reading
Correction
Uncertainty = x100
Instruments Re ading
1st calibration: 22.January.2015
Uncertainty: ±0.002mm Uncertainty: ±0.002mm
Description: Reverse Description: Forward
Calibration Method : JIS B7533
Uncertainty Correction Uncertainty Correction
No (mm) No
+ - + - (mm)
1 0.002 -0.002 0.003 1 0.002 -0.002 0
2 0.002 -0.002 0.002 2 0.002 -0.002 0
3 0.002 -0.002 0.002 3 0.002 -0.002 0
4 0.002 -0.002 0.004 4 0.002 -0.002 0
5 0.002 -0.002 0.002 5 0.002 -0.002 -0.001
6 0.002 -0.002 0.003 6 0.002 -0.002 -0.001
7 0.002 -0.002 0.003 7 0.002 -0.002 -0.001
8 0.002 -0.002 0.001 8 0.002 -0.002 0
9 0.002 -0.002 0.002 9 0.002 -0.002 -0.001
10 0.002 -0.002 0.002 10 0.002 -0.002 0
11 0.002 -0.002 0.002 11 0.002 -0.002 0
12 0.002 -0.002 0.002 12 0.002 -0.002 0.002
13 0.002 -0.002 0.002 13 0.002 -0.002 0.002
14 0.002 -0.002 0.001 14 0.002 -0.002 0.002
15 0.002 -0.002 0.002 15 0.002 -0.002 0.002
16 0.002 -0.002 0.001 16 0.002 -0.002 0.002
17 0.002 -0.002 0.001 17 0.002 -0.002 0.003
18 0.002 -0.002 0.001 18 0.002 -0.002 0.003
19 0.002 -0.002 0.002 19 0.002 -0.002 0.003
20 0.002 -0.002 0.001 20 0.002 -0.002 0.004
21 0.002 -0.002 0.001 21 0.002 -0.002 0.002
22 0.002 -0.002 -0.001 22 0.002 -0.002 0.003

QA/QC Department Page 1 PT Borneo Mandiri Investment


Calibration Report

REPORT FOR IM-MT-021

1.0 Calibration History


2nd calibration: 5.June.2015
Uncertainty: ±0.002mm Uncertainty: ±0.002mm
Description: Reverse Description: Forward
Calibration Method : JIS B7533
Uncertainty Correction Uncertainty Correction
No (mm) No
+ - + - (mm)
1 0.002 -0.002 0.002 1 0.002 -0.002 0
2 0.002 -0.002 0.004 2 0.002 -0.002 0.001
3 0.002 -0.002 0.004 3 0.002 -0.002 0
4 0.002 -0.002 0.004 4 0.002 -0.002 0.001
5 0.002 -0.002 0.004 5 0.002 -0.002 0
6 0.002 -0.002 0.004 6 0.002 -0.002 0
7 0.002 -0.002 0.004 7 0.002 -0.002 0
8 0.002 -0.002 0.003 8 0.002 -0.002 -0.001
9 0.002 -0.002 0.004 9 0.002 -0.002 0
10 0.002 -0.002 0.004 10 0.002 -0.002 0
11 0.002 -0.002 0.003 11 0.002 -0.002 0.001
12 0.002 -0.002 0.003 12 0.002 -0.002 0.002
13 0.002 -0.002 0.003 13 0.002 -0.002 0.002
14 0.002 -0.002 0.002 14 0.002 -0.002 0.003
15 0.002 -0.002 0.002 15 0.002 -0.002 0.004
16 0.002 -0.002 0.002 16 0.002 -0.002 0.003
17 0.002 -0.002 0.003 17 0.002 -0.002 0.003
18 0.002 -0.002 0.002 18 0.002 -0.002 0.003
19 0.002 -0.002 0.002 19 0.002 -0.002 0.004
20 0.002 -0.002 0.002 20 0.002 -0.002 0.002
21 0.002 -0.002 0.002 21 0.002 -0.002 0.002
22 0.002 -0.002 0.002 22 0.002 -0.002 0.002

QA/QC Department Page 2 PT Borneo Mandiri Investment


Calibration Report

REPORT FOR IM-MT-021

2.0 Calibration Graph

1st calibration (Reverse) 1st calibration (Forward)


0.005 0.005

0.004 0.004

0.003 0.003

0.002 0.002

0.001 0.001

0 0.000

-0.001
1 3 5 7 9 11 13 15 17 19 21 -0.001
1 3 5 7 9 11 13 15 17 19 21

-0.002 -0.002

-0.003 -0.003

2nd calibration 2nd calibration (Forward)


(Reverse) 0.005
0.005 0.004
0.004 0.003
0.003
0.002
0.002
0.001
0.001
0.000
0
-0.001
1 3 5 7 9 11 13 15 17 19 21
-0.001 1 3 5 7 9 11 13 15 17 19 21
-0.002 -0.002

-0.003 -0.003

QA/QC Department Page 3 PT Borneo Mandiri Investment


Calibration Report

REPORT FOR IM-MT-021

3.0 Calibration Frequency

According to the report from QC-580-02 Masterlist of IMTE (By usage Category) this
instrument is used.

QA/QC Department Page 4 PT Borneo Mandiri Investment


Calibration Report

REPORT FOR IM-MT-021

4.0 Calibration Report

Report on: January 2016


Section 2.0 shows that the graph being construct is obtained from the calibration
certificate on the 22.1.2015 and 5.6.2015. The graph shows that most of the point is
outside the upper limit of the uncertainty. This instrument also use in the production/QC
section where it accuracy mus be high. This instrument need to be calibrate again for
another 3 times with consistent 3 month interval each and the 1st and 2nd calibration is
being ignore due to the interval from the last calibration until now (22.1.2016) is
exceeded 3 month.
Suggestion:
3 times calibration for 3 month interval each need to be done.

QA/QC Department Page 5 PT Borneo Mandiri Investment


Calibration Report

REPORT FOR IM-MT-021

5.0 Reference