You are on page 1of 1

Certificado de Calibração

Laboratório Medição Ribeirão Preto


Certificate: 219596/21 Calibration Date: 01/11/2022 Due Date: 11/2023

1 / 1
Customer: GK ESTRUTURAS METÁLICAS E CALDEIRARIA
Rua Andre Camacho, 552 - Jaboticabal - SP - Brasil

Contracting:
Instrument Identification
Description: TERMÔMETRO DIGITAL Identification: TD-1107
Brand: MINIPA Model: MT401A Department:---
Serial Nº: ---
Ambient Conditions
Service performed at the permanent facilities of the Laboratory.
Temperature: 27.2 ºC ± 1.0 ºC Humidity: 40.0 %ur ± 5.0 %ur
Procedure
Calibration performed as: ITTEC019 Revision: 2
Standards Used
Identification: Brand: Cert. Number: Calibrated by: Due Date:
PTO-1051 TERMOHIGRÔMETRO PADRÃO MINIPA PTO105106/02/18 MEDIÇÃO-CAL0183 12/2021
PTT-0881 CALIBRADOR PORTÁTIL ECIL 8312/17 ECIL-CAL0026 12/2021
PTT-0882 TERMORRESISTÊNCIA PADRÃO ECIL 7902/17 ECIL-CAL0026 12/2021

Test Results
TERMOMETRO
Range: -50 a 750 °C
Indication Range: -50 a 750 °C Resolution: 1 °C

Expanded (k) Veff


Measurement Expanded
V.I V.R Uncertainty +
Error Uncertainty
Error
°C °C °C °C °C
-10 -8,9 -1,1 0,6 1,7 2,00 Infinito
20 19,1 0,9 0,6 1,5 2,00 Infinito
50 50,0 0,0 0,6 0,6 2,00 Infinito
100 100,1 -0,1 0,6 0,7 2,00 Infinito
150 149,2 0,8 0,6 1,4 2,00 Infinito

Notes
NÃO HOUVE AJUSTE

- V.I: Value indicated on the instrument in the same unit as the instrument .
- V.R: Reference Value in the unit of measurement of the Standard Equipment .
- The reported expanded measurement uncertainty is declared as the standard measurement uncertainty multiplied by the
coverage factor k, which for a t-distribution with Veff effective degrees of freedom corresponds to a coverage probability of
approximately 95%.
- The standard measurement uncertainty was determined in accordance with publication EA-4/02.
- The Approved / Disapproved condition is restricted to the metrological quantities of the instrument, and the error limit specified for
this condition is the responsibility of the Client.
- The adjustment is not part of the scope of services.
- The instrument calibration validity, when presented in this certificate, is the responsibility of the Client.