CERTIFICATE OF CALIBRATION

2102-13241
Customer information
Client : Goodyear Dunlop Dudelange Tires Operations S.A.
Contact : Ahmet KOSE
Address : Zone Industrielle Riedgen
3451, Dudelange
Luxembourg

Reference Trescal : TP-37823

Instrument information
Make : JUMO
Type : 902050/10
Description : PT100
Serial number :
Identification number : 99T00073
Range : 20 .. 120 °C
Accuracy : Class B (EN 60751): ± (0,3 + 0,005.|t|) °C

Date of calibration : 15 February 2021

Used method
P2-02-S.002 Calibration of temperature sensors (with or without readout) in calibrators.

Environmental conditions
During this service the following conditions were not exceeded :
Ambient temperature : 23°C ± 5°C
Relative humidity : 45%rh ± 20%rh

Used reference
The equipment used is traceable to National and/or International standards.
R1965 Hygro- / Thermometer cert n° 200119422
R3615 Dry Block Calibrator & Ref. Probe cert n° 201012087
R3731 Dry Block / Liquid Bath Calibrator & Ref. Probe cert n° 200119421
R4400 Dry Block / Liquid Bath Calibrator & Ref. Probe cert n° 191105454
R4452 Multifunctional calibrator cert n° 200413969
Note
This service was performed at the clients location.
The instrument is measured but not adjusted, so the results are both 'as found' as 'as left'.

Technician Issue date 15 February 2021 Head of the laboratory


Gregory Gaspart Luc Van Pelt

This document may not be reproduced other than in full, except with the prior written approval of the head of the issuing laboratory.
Unless otherwise stated, the calibration was performed at the address mentioned in the footnote.

Trescal nv | Vosstraat 200 | Berchem (Antwerpen) | Belgium | Telephone + 32 3 542 62 90 | Email: info.benelux@trescal.com

Page 1 to 2
CERTIFICATE OF CALIBRATION
2102-13241
Initial results:
- Reference Units Instrument Units Calculated Difference Uncertainty Units
value value value ±
1 20.00 °C 107.76 Ohm 19.91 -0.10 0.12 °C
2 49.96 °C 119.35 Ohm 49.87 -0.08 0.12 °C
3 90.02 °C 134.69 Ohm 89.94 -0.07 0.16 °C
4 120.01 °C 146.02 Ohm 119.86 -0.15 0.16 °C
5 20.00 °C 107.76 Ohm 19.90 -0.09 0.12 °C

The stated uncertainty is that of the entire set-up including the object under test.
The reported uncertainty is based on a standard uncertainty multiplied by a coverage factor k=2, providing a
level of confidence of approximately 95% .
The uncertainty is calculated following EA-4/02 in accordance with the requirements of the ISO/IEC 17025.