You are on page 1of 3

Device Information

Parameter Settings
Order information
Customer ECONOMY PROCESS SOLUTIONS PVT LTD
Customer purchase order EPSPL/1213
Sales order number / Item 62/3016367493/1023385510
Internal order number / Item 3002266356/0020

Device information
Description iTEMP TMT162, Field Transmitter
TAG TT-102
Serial number TB044597223
Order code TMT162-65A8/101
Extended order code TMT162-E21331BBA

Additional information
Box / Boxes: 2/6
TAG, metal TT-102

Electronics

Electronic type 4...20 mA, HART®


Device revision 4
Firmware version 04.01.11
Hardware version 04.01
Model name TMT162
Device type 0x11CE
SIL NO

Input 1

Sensor type Pt100, IEC 751


Sensor connection 3- wire

Output

Output signal direction 4...20 mA


PV-Mode PV=Ch1, Ch2 not
Measuring range 0...100 °C
Output failure mode <= 3.6 mA
Damping 0s
Ofset 0K

Statement

The device operates within the manufacturers stated tolerances

Aurangabad, 2022-11-09
Endress+Hauser Wetzer (India) Pvt Ltd
Plot No. M 171/173; MIDC, Waluj,
431 136 Aurangabad
www.endress.com Document ID: TB044597223_2022 Page 1 of 1
Endress+Hauser Wetzer (India) Pvt Ltd
Plot No. M 171/173; MIDC, Waluj,
431 136 Aurangabad
Phone: +91 240 256 3600
Fax: +91 240 255 5179
www.endress.com

Factory calibration certifcate

Certifcate-No. TB044597223
Date of issue 09.11.2022
Order code TMT162-65A8/101
Extended order code TMT162-E21331BBA
Serial number TB044597223

Tagging (TAG), metal TT-102


Box / Boxes: 2/6

Standards used

Description of the standard ID-number of the standard Standard calibrated to


Keithley 2010 20130018 02/2023
Digistant 4462 20130016 02/2023
RTD M-612 20130017 02/2023

Procedure of calibration/measurement and ambient conditions

All measurements have been made in ambient controlled at 25°C +/- 3K and R.H. 0%...80%

Measurement results

Digital output: Pt100 IEC

Nominal Value Measured temperature Deviation Max allowed deviation

(°C) (°C) (K) (±K)

Channel 1
-200,0 - -200,00 0,00 0,06
10,0 - 10,00 0,00 0,07
220,0 - 220,01 0,01 0,08
430,0 - 430,01 0,01 0,09
640,0 - 640,01 0,01 0,10
850,0 - 850,02 0,02 0,11

Analog output: Range 0...100 °C

Nominal Value Measured output Measured temperature Deviation Max allowed deviation

(°C) (mA) (°C) (K) (±K)


0,0 4,001 0,01 0,01 0,03
20,0 7,199 19,99 -0,01 0,03
40,0 10,399 39,99 -0,01 0,03
60,0 13,599 59,99 -0,01 0,03
80,0 16,801 80,01 0,01 0,03
100,0 20,001 100,01 0,01 0,03

TB044597223 / 09.11.2022 Page 1 of 2


Measurement results

The test unit has been calibrated and complies with the tolerances stated by manufacturer. This report has been created
automatically and is valid without a signature.

TB044597223 / 09.11.2022 Page 2 of 2

You might also like