Calibration Method : This Proving ring is Calibrated in Compression Mode as ULR No.

CC319321000000513F
per DRS Engineering & Calibration Services Procedure No. : DRS/SOP/01F
Based on IS 4169-2014/ISO 376-2011 Cal. Certificate No. DRS/08/21/513

Result : Compression Mode


The Calibration Data is only for following Dial Gauge Setting:-
Large Pointer at 12'O Clock position & Small Pointer at 1.
(Dial Gauge reading in division)
Applied Force in
SL No Position 0° Position 0° Position 120° Position 240° Average
kN
series 1 series 2 series 3 series 4 (series 1,3,4)
1 0 0.0 0.0 0.0 0.0 ----
2 1.0 70.2 70.1 70.1 70.2 70.2
3 2.0 150.5 150.6 150.5 150.5 150.5
4 3.0 220.7 220.6 220.6 220.7 220.7
5 4.0 291.1 291.0 291.1 291.2 291.1
6 5.0 361.2 361.3 361.1 361.2 361.2
7 6.0 431.4 431.5 431.6 431.5 431.5
8 7.0 501.7 501.8 501.6 501.7 501.7
9 8.0 572.4 572.5 572.3 572.4 572.4
10 9.0 642.2 642.3 642.2 642.1 642.2
11 10.0 712.5 712.6 712.6 712.5 712.5
12 0 0.0 0.0 0.0 0.0 ----

Creep at No Load
Reading After 30 Sec. = 0.1 Div.
Reading After 300 Sec. = 0.0 Div.

Classification :

The Force proving instrument is found to comply with the requirements of IS:4169-2014 classified as follows:

Class Mode From To Uncertainty of Measurement

Class 1 Compression 10 KN 2 KN ±0.139%

Class 2 Compression 10 KN 1 KN ±0.240%

NOTE :-
1 The Reported Uncertainty is at coverage factor k=2 which corresponds probability of approximately 95% for a normal distribution. Considering
the relative deviation of difference components such as creep, zero, resolution, repeatibilty, reproducibility, & combining with the uncertainty of
applied force.
2 Calibration has been performed using self-alinging compression pad provided by the customer.
3 Result relate only the item calibrated
4 Creep is measured at zero load.
5 This certificate shall not be reproduced, expect in full.