(Rev) UCDS(D)_QC-002511_08-07-2021_Start_v01_00
ULTRASONIC Form. Rev. 6
CALIBRATION DATA
Date 08/07/2021
SHEET
Page 1 of 6
1. EQUIPMENT
1.1 Instrument
Model No. Manufacturer Serial number Calibration Date
Mainframe OMNISCAN-MX2 Olympus OMNI2-101098 1/26/2021
UT module OMNI-M2-PA32128PR Olympus QC-073166 1/26/2021
1.2 Probes and Wedges
TECHNIQUE T1-T3 Model No. Manufacturer Serial number
Probe 2.25DM7X4PM-19X12-A17 Olympus U4107
Wedge SA17-DN55L0-IHC Olympus ABWX2786
TECHNIQUE T5 Model No. Manufacturer Serial number
Probe 2.25DM7X4PM-19X12-A17 NOT REQUIRED Olympus
Wedge SA17-DN55L0-IHC Olympus
2. PHYSICAL STATE
Visual Inspection for Physycal Damage Instrument Probes Cables Calibration Blocks
Results of calibration OK OK OK OK
3. ENCODER
For linear scanners
1
Doc. No. UCDS(D)_QC-002511_08-07-2021_Start_v01_00
ULTRASONIC Form. Rev. 6
CALIBRATION DATA
Date 08/07/2021
SHEET
Page 2 of 6
4. ELEMENT ACTIVITY
Active Elements 2x28
Dead Elements 0
Max. Consecutive Dead Elements 0
Results of Calibration OK
Data file DMA_ACTIVITY.EL._08-07-2021_01
5. THICKNESS CHECK
Calibration Blocks IP-01
Delay law 0
Reflectors BACKWALL (AREA WITHOUT REFLECTORS)
Sound Path (mm) 59:61
Before OK
After OK
Tolerance: The measured thickness shall be in tolerance with ± 1 mm of block thickness (nominal value 60 mm)
Doc. No. UCDS(D)_QC-002511_08-07-2021_Start_v01_00
ULTRASONIC Form. Rev. 6
CALIBRATION DATA
Date 08/07/2021
SHEET
Page 3 of 6
6. ELEMENT ACTIVITY
Active Elements 2x28
Dead Elements 0
Max. Consecutive Dead Elements 0
Results of Calibration OK
Data file DMA_ACTIVITY.EL._08-07-2021_01
Tolerance: The probe index position shall be within ± 1 mm of the existing index.
Tolerance: The measured beam angle shall be within ± 2º of the required refraction angle (± 3º for 62º).
8. TCG CALIBRATION
Calibration Blocks SCB-04
Delay law 30 48 66
(Required refraction angle) (30º) (48º) (62º)
Reflectors H20 H40 H60 H20 H40 H60 H20 H40 H60
Sound Path (mm) 20:24 43:48 66:71 26:32 56:62 85:92 39:48 81:94 123:141
Before OK OK OK OK OK OK OK OK OK
After OK OK OK OK OK OK OK OK OK
Amplitude (%) 63.6 - 100.6
Before OK OK OK OK OK OK OK OK OK
After OK OK OK OK OK OK OK OK OK
Tolerance: The measured sound path shall be in tolerance with ± 2º of the required refraction angle (± 3º for 62º) or
within ± 2 mm of the nominal value.
Doc. No. UCDS(D)_QC-002511_08-07-2021_Start_v01_00
ULTRASONIC Form. Rev. 6
CALIBRATION DATA
Date 08/07/2021
SHEET
Page 4 of 6
9. ELEMENT ACTIVITY
Active Elements 2x28
Dead Elements 0
Max. Consecutive Dead Elements 0
Results of Calibration OK
Data file DMA_ACTIVITY.EL._08-07-2021_01
Tolerance: The probe index position shall be within ± 1 mm of the existing index.
Tolerance: The measured beam angle shall be within ± 2º of the required refraction angle (± 3º for 64º).
Reflectors H05 H10 H20 H05 H10 H20 H05 H10 H20
Sound Path (mm) 4 : 10 11 : 18 27 : 33 5:13 15:22 32:40 7 : 18 20 : 31 42 : 51
Before OK OK OK OK OK OK OK OK OK
After OK OK OK OK OK OK OK OK OK
Amplitude (%) 63.6 - 100.6
Before OK OK OK OK OK OK OK OK OK
After OK OK OK OK OK OK OK OK OK
Tolerance: The measured sound path shall be in tolerance with ± 2º of the required refraction angle (± 3º for 64º) or
within ± 2 mm of the nominal value.
Doc. No. UCDS(D)_QC-002511_08-07-2021_Start_v01_00
ULTRASONIC Form. Rev. 6
CALIBRATION DATA
Date 08/07/2021
SHEET
Page 5 of 6
R (mm) NA NA NA
Results NA NA NA
Tolerance: The probe index position shall be within ± 1 mm of the existing index.
Tolerance: The measured beam angle shall be within ± 2º of the required refraction angle.
Tolerance: The measured sound path shall be in tolerance with ± 2º of the required refraction angle or within ± 2 mm of the nominal value.
Tolerance: The measured sound path shall be in tolerance with ± 2º of the required refraction angle
(± 5º for 66º) or within ± 2 mm of the nominal value.