You are on page 1of 2

Form No.

: FM/22
ELECTRO METER CORPORATION
P- 5, C.I.T ROAD, CALIBRATION CERTIFICATE OF:
KOLKATA - 700014
Weight Scale
Certificate No.:
CC - 2147
CALIBRATION CERTIFICATE NO.: EMC2022-KOL-00000285 ULR NO.: CC214722000000228F Page: 1/2
1.0 ISSUED TO: ARMA ELECTRIC COMPANY
ALIPUR, VOBERCHAR ( DHAKA CHITTAGONG HIGHWAY )
GAZARIA, MUNSHIGANJ, BANGLADESH

1.1 Service Request / Order No.: -


1.2 Description identification of item to be calibrated:
i Name: Weight Scale ii Code No.: WS-01
iii Sl. No.: 808996 iv Make: RFL
v Model / Type No.: - vi Range: Max 100 kg
vii Sensor: - viii Resolution: 0.1 kg
ix Location: Office Room
1.3 Applicable specification of item to be calibrated: Accuracy / permissible limit: Not Specified
1.4 Date of receipt of item: 16-Jan-2022 1.5 Date of calibration: 16-Jan-2022
Recommended date for next
1.6 16-Jan-2023 1.7 Frequency of calibration: 12 months
calibration
Temperature: 25.2°C
1.8a) Environmental condition
during calibration Humidity: 55% RH
1.8b) Recommended Temperature: 25°C ± 10°C
Environmental condition
Humidity: 50% RH ± 20% RH
S.O.P Number S.O.P/Mass/02
Standard Followed OIML R76-1: 2006
1.9 Basis of calibration:
Using E2 Class Standard Weights > 200 g to 3 kg for Calibration of Class II Weighing
Description of Method
Balances and Coarser as Per OIML R-76
2.0 Thermal Stabilization Time : 3 hrs 2.1 Exact Location of Balance (from Ground Floor) : 1
2.2 Traceability : Standards Used for Calibration are Traceable to National/International Standards through ISO/IEC: 17025 Accredited Laboratory.

Name of the standard Recommended


Traceable to /
Sl. No. instrument Range Resolution date for
Certificate Nos.
used (Sl. No./Code Nos.) next calibration

EMC (Kolkata) /
A Set of Weight, Grade F1, Code No:
1 1 mg to 200 g - 01-07-2022 EMC2021-KOL-
EMC/AWB/152
00032771

WMCL /
Standard Weight, Grade E2, Code No.
2 500 g - 13-12-2022 WMCL/E/2020-
EMC/SWt-500/248
12/1067

WMCL /
Standard Weight, Grade E2, Code No.
3 1000 g - 13-12-2022 WMCL/E/2020-
EMC/SWt-1/247
12/1067

EMC (Kolkata) /
Standard Weight, F2 Grade, 2 kg ,
4 2000 g - 27-02-2022 EMC2021-KOL-
Code No. EMC/SW/539
00015859

WMCL /
Standard Weight, Grade F2, Code
5 5 kg - 11-12-2022 WMCL/F/2020-
No:, EMC/SW-5kg/582
12/1066

WMCL /
Standard Weight, Grade F2, Code No:
6 10 kg - 11-12-2022 WMCL/F/2020-
EMC/SW-10kg/841
12/1066

WMCL /
Standard Weight, Grade F2, Code No:
7 20 kg - 11-12-2022 WMCL/F/2020-
EMC/SW-20kg/576
12/1066

WMCL /
Standard Weight, Grade F2, Code No.
8 50 kg - 11-12-2022 WMCL/F/2020-
EMC/SW-50kg/867
12/1066

WMCL /
Standard Weight, Grade F2, Code No.
9 50 kg - 11-12-2022 WMCL/F/2020-
EMC/SW-50kg/868
12/1066

"Digital Temperature & Humidity


EMC (Kolkata) /
Meter, Make: Metravi, Model No. HT- "19°C to 22°C (for Temperature) & "0.1°C (for Temperature) &
10 20-03-2022 EMC2021-KOL-
3005, Sl. No. 141017137, Code No. 40% RH to 80% RH (for Humidity)" 0.1% RH (for Humidity)"
00017277
EMC/DTHM/840
Form No.: FM/22
ELECTRO METER CORPORATION
P- 5, C.I.T ROAD, CALIBRATION CERTIFICATE OF:
KOLKATA - 700014. Weight Scale
Certificate No.:
CC - 2147
CALIBRATION CERTIFICATE NO.: EMC2022-KOL-00000285 ULR NO.: CC214722000000228F Page: 2/2
CALIBRATION RESULT
Linearity Test :
Parameter / Nominal Value on DUC Std. Value Error
Range (kg) (kg) (kg)
100.2 100.000032 0.199968
80.2 80.000169 0.199831
50.1 50.000031 0.099969
Mass 30.1 30.000058 0.099942
Max 100 kg 20.1 20.000024 0.099976
10.2 10.000034 0.199966
5.1 5.000001 0.099999
2.0 1.999996 0.000004
1.0 0.999999 0.000001
0.5 0.500000 0.000000
0.2 0.200000 0.000000
0.0 0.000000 0.000000
Repeatability at Max Load : 100 kg Repeatability at Half Load : 50 kg
1.0 100.2 kg 1 50.1 kg
2.0 100.0 kg 2 50.0 kg
3.0 100.0 kg 3 50.1 kg
4.0 100.1 kg 4 50.1 kg
5.0 100.0 kg 5 50.1 kg
6.0 100.1 kg 6 50.2 kg
7.0 100.2 kg 7 50.1 kg
8.0 100.2 kg 8 50.2 kg
9.0 100.1 kg 9 50.0 kg
10.0 100.2 kg 10 50.0 kg
Standard deviation: 0.000 kg Standard deviation: 0.000 kg
Eccentricity: at 50 kg is 0 kg
Note: i) The Calibration result is only to be used for scientific purpose.
ii) Average of 10 readings had been taken in DUC/STD.
iii) The calibration certificate is issued by Electro Meter Corporation. This certificate refers only for the particular item submitted to calibrate.
iv) The certificate if produced for any purpose commercial or otherwise, should be reproduced in full. Partial production of the certificate or an abstract
thereof has to be got specifically approved from the technical and the quality manager of this lab.
v) The results in the calibration certificate are valid at the time of and under the stated conditions on measurement.
vi) The calibration result is only to be used for scientific & industrial purpose. This result can not be used for commercial purpose.
Measurement uncertainty at approx 95 % of confidence level & at a coverage factor of 2.00 is ± 0.24 kg
Remarks: Statement of conformity not mentioned.
DUC- Device Under Calibration F.S.-Full Scale Physical status of the DUC: OK
Calibrated by: Saykat Mallik Checked & Approved by: Saptarshi Das
Date : 16-1-2022 Date 17-01-2022
Report Issue Date : 17-01-2022

END OF CERTIFICATE

You might also like