CALIBRATION CERTIFICATE

Certificate No. HT/CC/201210-11/001 ULR No. ULR - CC215120000038201F


CUSTOMER DETAILS DEVICE UNDER CALIBRATION DETAILS
M/s. L&T Construction Research and Testing
Name Description Pressure Indicator
Centre,
L&T Construction, Make Geo Technik
PB.No.979, Model --
Address Mount Poonamallee Road,
Manapakkam, Range 0 to 3000 kPa
Chennai - 600 089. Least Count 0.1 kPa
Job No. 201210-11 Accuracy --
Job Receipt Date 10-Dec-2020 Identification S-CP-02
Issue Date 22-Dec-2020 Serial No. --
Received Condition of DUC Satisfactory Location Soil Lab
ENVIRONMENTAL CONDITION Calibration On 10-Dec-2020
Temperature 25 °C ± 5 °C Due On --(Customer Defined)
Humidity 30 % RH to 75 % RH Calibrated By V. Cibiraja
Discipline & Group Mechanical Calibration & Pressure and Vacuum
Calibration Procedure HT/WI/M/PV (Section 01) & DKD-R 6-1 (Sequence C ) by Comparison Method
STANDARD DETAILS
Description Process Calibrator
Identification No. M/PV/PC/160/01 & M/PV/PCI/01
Certificate No. HT/CC/200526-09/006
Valid Till 26-May-2021
CALIBRATION RESULT
Standard Reading in kPa Average of Expanded
Hysteresis
DUC Reading in kPa Standard Reading Error in kPa Uncertainty in ±
Increasing Decreasing in kPa
in kPa kPa
0.0 0.00000 0.00000 0.00000 0.00000 0.00000 --
25.0 25.0214 25.0196 25.0205 -0.0205 0.0018 0.11
50.0 50.0380 50.0370 50.0375 -0.0375 0.0010 0.11
100.0 100.093 100.081 100.087 -0.087 0.012 0.11
150.0 149.878 149.868 149.873 0.127 0.010 0.11
200.0 199.743 199.733 199.738 0.262 0.010 0.11
500.0 499.654 499.646 499.650 0.350 0.008 0.15
600.0 599.526 599.515 599.521 0.479 0.011 0.15
700.0 699.534 699.526 699.530 0.470 0.008 0.15
800.0 799.438 799.430 799.434 0.566 0.008 0.15
900.0 899.303 899.294 899.299 0.701 0.009 0.15
1000.0 999.291 999.283 999.287 0.713 0.008 0.15
2000.0 1999.85 1999.80 1999.83 0.17 0.05 0.36
2500.0 2500.36 2500.30 2500.33 -0.33 0.06 0.36
3000.0 3000.97 3000.94 3000.96 -0.96 0.03 0.39
Remarks :
• Pressure Conversion 1 kPa = 1000 Pa.
• The Calibration Certificate is valid for scientific & industrial purpose only.
• The Calibration is performed with a Temperature Change of ± 1 °C at Customer's Facility.
• The Error is provided as the Difference of DUC Reading and Average of Standard Reading.
• The Expanded uncertainty is expressed at 95.45 % confidence level with coverage factor k=2.
• Pressure Transmission Medium used is Water and the position of DUC during Calibration is Vertical.
• The Calibration Points are taken as per Customer's Request.

Checked By Approved By
K. MUHILAN - CALIBRATION TECHNICIAN R.KOWSELYA - CALIBRATION ENGINEER
HT / FT / 7.8 / 01 End of Certificate Page 1 of 1
CALIBRATION CERTIFICATE
Certificate No. HT/CC/201210-11/002 ULR No. ULR - CC215120000038202F
CUSTOMER DETAILS DEVICE UNDER CALIBRATION DETAILS
M/s. L&T Construction Research and Testing
Name Description Pressure Indicator
Centre,
L&T Construction, Make Geo Technik
PB.No.979, Model --
Address Mount Poonamallee Road,
Manapakkam, Range 0 to 3000 kPa
Chennai - 600 089. Least Count 0.1 kPa
Job No. 201210-11 Accuracy --
Job Receipt Date 10-Dec-2020 Identification S-BP-02 (TOP)
Issue Date 22-Dec-2020 Serial No. --
Received Condition of DUC Satisfactory Location Soil Lab
ENVIRONMENTAL CONDITION Calibration On 10-Dec-2020
Temperature 25 °C ± 5 °C Due On --(Customer Defined)
Humidity 30 % RH to 75 % RH Calibrated By V. Cibiraja
Discipline & Group Mechanical Calibration & Pressure and Vacuum
Calibration Procedure HT/WI/M/PV (Section 01) & DKD-R 6-1 (Sequence C ) by Comparison Method
STANDARD DETAILS
Description Process Calibrator
Identification No. M/PV/PC/160/01 & M/PV/PCI/01
Certificate No. HT/CC/200526-09/006
Valid Till 26-May-2021
CALIBRATION RESULT
Standard Reading in kPa Average of Expanded
Hysteresis
DUC Reading in kPa Standard Reading Error in kPa Uncertainty in ±
Increasing Decreasing in kPa
in kPa kPa
0.0 0.00000 0.00000 0.00000 0.00000 0.00000 --
25.0 24.9898 24.9860 24.9879 0.0121 0.0038 0.11
50.0 49.9726 49.9695 49.9711 0.0289 0.0031 0.11
100.0 99.9586 99.9560 99.9573 0.0427 0.0026 0.11
150.0 149.940 149.903 149.922 0.078 0.037 0.11
200.0 199.896 199.860 199.878 0.122 0.036 0.11
500.0 499.637 499.604 499.621 0.379 0.033 0.15
600.0 599.209 599.174 599.192 0.808 0.035 0.15
700.0 698.995 698.961 698.978 1.022 0.034 0.15
800.0 798.874 798.850 798.862 1.138 0.024 0.15
900.0 898.721 898.708 898.715 1.285 0.013 0.15
1000.0 998.593 998.580 998.587 1.413 0.013 0.15
2000.0 1998.18 1998.12 1998.15 1.85 0.06 0.36
2500.0 2497.96 2497.94 2497.95 2.05 0.02 0.35
Remarks :
• Pressure Conversion 1 kPa = 1000 Pa.
• The Calibration Certificate is valid for scientific & industrial purpose only.
• The Calibration is performed with a Temperature Change of ± 1 °C at Customer's Facility.
• The Error is provided as the Difference of DUC Reading and Average of Standard Reading.
• The Expanded uncertainty is expressed at 95.45 % confidence level with coverage factor k=2.
• Pressure Transmission Medium used is Water and the position of DUC during Calibration is Vertical.
• The Calibration Points are taken as per Customer's Request.

Checked By Approved By
K. MUHILAN - CALIBRATION TECHNICIAN R.KOWSELYA - CALIBRATION ENGINEER
HT / FT / 7.8 / 01 End of Certificate Page 1 of 1
CALIBRATION CERTIFICATE
Certificate No. HT/CC/201210-11/003 ULR No. ULR - CC215120000038203F
CUSTOMER DETAILS DEVICE UNDER CALIBRATION DETAILS
M/s. L&T Construction Research and Testing
Name Description Pressure Indicator
Centre,
L&T Construction, Make Geo Technik
PB.No.979, Model --
Address Mount Poonamallee Road,
Manapakkam, Range 0 to 3000 kPa
Chennai - 600 089. Least Count 0.1 kPa
Job No. 201210-11 Accuracy --
Job Receipt Date 10-Dec-2020 Identification S-BP-02 (Bottom)
Issue Date 22-Dec-2020 Serial No. --
Received Condition of DUC Satisfactory Location Soil Lab
ENVIRONMENTAL CONDITION Calibration On 10-Dec-2020
Temperature 25 °C ± 5 °C Due On --(Customer Defined)
Humidity 30 % RH to 75 % RH Calibrated By V. Cibiraja
Discipline & Group Mechanical Calibration & Pressure and Vacuum
Calibration Procedure HT/WI/M/PV (Section 01) & DKD-R 6-1 (Sequence C ) by Comparison Method
STANDARD DETAILS
Description Process Calibrator
Identification No. M/PV/PC/160/01 & M/PV/PCI/01
Certificate No. HT/CC/200526-09/006
Valid Till 26-May-2021
CALIBRATION RESULT
Standard Reading in kPa Average of Expanded
Hysteresis
DUC Reading in kPa Standard Reading Error in kPa Uncertainty in ±
Increasing Decreasing in kPa
in kPa kPa
0.0 0.00000 0.00000 0.00000 0.00000 0.00000 --
25.0 25.3145 25.3113 25.3129 -0.3129 0.0032 0.11
50.0 50.4030 50.4008 50.4019 -0.4019 0.0022 0.11
100.0 100.752 100.215 100.484 -0.484 0.537 0.33
150.0 150.792 150.769 150.781 -0.781 0.023 0.11
200.0 200.032 200.008 200.020 -0.020 0.024 0.11
500.0 499.093 499.050 499.072 0.928 0.043 0.15
600.0 599.227 599.208 599.218 0.782 0.019 0.15
700.0 698.997 698.978 698.988 1.012 0.019 0.15
800.0 798.026 798.012 798.019 1.981 0.014 0.15
900.0 898.067 898.052 898.060 1.940 0.015 0.15
1000.0 997.877 997.869 997.873 2.127 0.008 0.15
1500.0 1497.54 1497.46 1497.50 2.50 0.08 0.16
2000.0 1997.18 1997.10 1997.14 2.86 0.08 0.36
2500.0 2497.06 2497.01 2497.04 2.96 0.05 0.36
Remarks :
• Pressure Conversion 1 kPa = 1000 Pa.
• The Calibration Certificate is valid for scientific & industrial purpose only.
• The Calibration is performed with a Temperature Change of ± 1 °C at Customer's Facility.
• The Error is provided as the Difference of DUC Reading and Average of Standard Reading.
• The Expanded uncertainty is expressed at 95.45 % confidence level with coverage factor k=2.
• Pressure Transmission Medium used is Water and the position of DUC during Calibration is Vertical.
• The Calibration Points are taken as per Customer's Request.

Checked By Approved By
K. MUHILAN - CALIBRATION TECHNICIAN R.KOWSELYA - CALIBRATION ENGINEER
HT / FT / 7.8 / 01 End of Certificate Page 1 of 1
CALIBRATION CERTIFICATE
Certificate No. HT/CC/201210-11/004 ULR No. ULR - CC215120000038204F
CUSTOMER DETAILS DEVICE UNDER CALIBRATION DETAILS
M/s. L&T Construction Research and
Name Description LVDT Probe with Indicator
Testing Centre,
L&T Construction, Make GDS/RDP Model DTCH 200AG
PB.No.979, Range 0 to 10 mm Least Count 0.001 mm
Address Mount Poonamallee Road, Identification S-CON-08/S-LVDT-10
Manapakkam, Serial No. --
Chennai - 600 089. Location Soil Lab
Job No. 201210-11 Type Digital
Job Receipt Date 10-Dec-2020 Calibration On 10-Dec-2020
Issue Date 22-Dec-2020 Due On -- (Customer Defined)
Received Condition of DUC Satisfactory Calibrated By M. Deepan Ram
ENVIRONMENTAL CONDITION
Temperature 20 ± 2 °C Humidity 40 %RH to 60 %RH
Discipline & Group Mechanical Calibration & Dimension (Basic Measuring Instrument,Gauge Etc.)
Calibration Procedure HT/WI/M/D (Section 07) by Comparison Method
STANDARD DETAILS
Description Slip Gauge Sets (Tungsten Carbide)
Identification No. M / D / SG / 01
Certificate No. C-200717-5-1
Valid Till 17-July-2022
CALIBRATION RESULT
(All Values are in mm)
Standard Reading DUC Reading Error
0 0.000 0.000
0.5 0.497 -0.003
1 0.995 -0.005
1.5 1.490 -0.010
2 1.990 -0.010
3 2.988 -0.012
4 3.985 -0.015
5 4.986 -0.014
6 5.983 -0.017
7 6.982 -0.018
8 7.981 -0.019
9 8.980 -0.020
10 9.978 -0.022
Remarks :
• The Calibration is performed with a Temperature of 27.95 °C at Customer's Facility
• The Expanded uncertainty is ± 1.50 µm at 95.45 % confidence level with coverage factor k=2.

Checked By Approved By
K. MUHILAN - CALIBRATION TECHNICIAN R.R.HARI JAIRAM - TECHNICAL MANAGER
HT / FT / 7.8 / 01 End of Certificate Page 1 of 1
CALIBRATION CERTIFICATE
Certificate No. HT/CC/201210-11/005 ULR No. ULR - CC215120000038205F
CUSTOMER DETAILS DEVICE UNDER CALIBRATION DETAILS
M/s. L&T Construction Research and
Name Description LVDT Probe with Indicator
Testing Centre,
L&T Construction, Make RDP/GDS Model DTCH 200AG
PB.No.979, Range 0 to 10 mm Least Count 0.001 mm
Address Mount Poonamallee Road, Identification S-CON-05/S-LVDT-06
Manapakkam, Serial No. 166618
Chennai - 600 089. Location Soil Lab
Job No. 201210-11 Type Digital
Job Receipt Date 10-Dec-2020 Calibration On 10-Dec-2020
Issue Date 22-Dec-2020 Due On -- (Customer Defined)
Received Condition of DUC Satisfactory Calibrated By M. Deepan Ram
ENVIRONMENTAL CONDITION
Temperature 20 ± 2 °C Humidity 40 %RH to 60 %RH
Discipline & Group Mechanical Calibration & Dimension (Basic Measuring Instrument,Gauge Etc.)
Calibration Procedure HT/WI/M/D (Section 07) by Comparison Method
STANDARD DETAILS
Description Slip Gauge Sets (Tungsten Carbide)
Identification No. M / D / SG / 01
Certificate No. C-200717-5-1
Valid Till 17-July-2022
CALIBRATION RESULT
(All Values are in mm)
Standard Reading DUC Reading Error
0 0.000 0.000
0.5 0.495 -0.005
1 0.993 -0.007
1.5 1.491 -0.009
2 1.990 -0.010
3 2.985 -0.015
4 3.983 -0.017
5 4.980 -0.020
6 5.981 -0.019
7 6.979 -0.021
8 7.974 -0.026
9 8.975 -0.025
10 9.973 -0.027
Remarks :
• The Calibration is performed with a Temperature of 27.05 °C at Customer's Facility
• The Expanded uncertainty is ± 1.50 µm at 95.45 % confidence level with coverage factor k=2.

Checked By Approved By
K. MUHILAN - CALIBRATION TECHNICIAN R.R.HARI JAIRAM - TECHNICAL MANAGER
HT / FT / 7.8 / 01 End of Certificate Page 1 of 1
CALIBRATION CERTIFICATE
Certificate No. HT/CC/201210-11/006 ULR No. ULR - CC215120000038206F
CUSTOMER DETAILS DEVICE UNDER CALIBRATION DETAILS
M/s. L&T Construction Research and
Name Description LVDT Probe with Indicator
Testing Centre,
L&T Construction, Make GDS/RDP Model DTCH 200AG
PB.No.979, Range 0 to 10 mm Least Count 0.001 mm
Address Mount Poonamallee Road, Identification S-CON-03/S-LVDT-04
Manapakkam, Serial No. 17073
Chennai - 600 089. Location Soil Lab
Job No. 201210-11 Type Digital
Job Receipt Date 10-Dec-2020 Calibration On 10-Dec-2020
Issue Date 22-Dec-2020 Due On -- (Customer Defined)
Received Condition of DUC Satisfactory Calibrated By M. Deepan Ram
ENVIRONMENTAL CONDITION
Temperature 20 ± 2 °C Humidity 40 %RH to 60 %RH
Discipline & Group Mechanical Calibration & Dimension (Basic Measuring Instrument,Gauge Etc.)
Calibration Procedure HT/WI/M/D (Section 07) by Comparison Method
STANDARD DETAILS
Description Slip Gauge Sets (Tungsten Carbide)
Identification No. M / D / SG / 01
Certificate No. C-200717-5-1
Valid Till 17-July-2022
CALIBRATION RESULT
(All Values are in mm)
Standard Reading DUC Reading Error
0 0.000 0.000
0.5 0.497 -0.003
1 0.991 -0.009
1.5 1.492 -0.008
2 1.990 -0.010
3 2.990 -0.010
4 3.988 -0.012
5 4.988 -0.012
6 5.987 -0.013
7 6.987 -0.013
8 7.988 -0.012
9 8.988 -0.012
10 9.981 -0.019
Remarks :
• The Calibration is performed with a Temperature of 27.05 °C at Customer's Facility
• The Expanded uncertainty is ± 1.50 µm at 95.45 % confidence level with coverage factor k=2.

Checked By Approved By
K. MUHILAN - CALIBRATION TECHNICIAN R.R.HARI JAIRAM - TECHNICAL MANAGER
HT / FT / 7.8 / 01 End of Certificate Page 1 of 1
CALIBRATION CERTIFICATE
Certificate No. HT/CC/201210-11/007 ULR No. ULR - CC215120000038207F
CUSTOMER DETAILS DEVICE UNDER CALIBRATION DETAILS
M/s. L&T Construction Research and
Name Description LVDT Probe with Indicator
Testing Centre,
L&T Construction, Make GDS/RDP Model DTCH 200AG
PB.No.979, Range 0 to 10 mm Least Count 0.001 mm
Address Mount Poonamallee Road, Identification S-CON-04/S-LVDT-05
Manapakkam, Serial No. 16617
Chennai - 600 089. Location Soil Lab
Job No. 201210-11 Type Digital
Job Receipt Date 10-Dec-2020 Calibration On 10-Dec-2020
Issue Date 22-Dec-2020 Due On -- (Customer Defined)
Received Condition of DUC Satisfactory Calibrated By M. Deepan Ram
ENVIRONMENTAL CONDITION
Temperature 20 ± 2 °C Humidity 40 %RH to 60 %RH
Discipline & Group Mechanical Calibration & Dimension (Basic Measuring Instrument,Gauge Etc.)
Calibration Procedure HT/WI/M/D (Section 07) by Comparison Method
STANDARD DETAILS
Description Slip Gauge Sets (Tungsten Carbide)
Identification No. M / D / SG / 01
Certificate No. C-200717-5-1
Valid Till 17-July-2022
CALIBRATION RESULT
(All Values are in mm)
Standard Reading DUC Reading Error
0 0.000 0.000
0.5 0.496 -0.004
1 0.994 -0.006
1.5 1.490 -0.010
2 1.991 -0.009
3 2.985 -0.015
4 3.985 -0.015
5 4.975 -0.025
6 5.975 -0.025
7 6.976 -0.024
8 7.975 -0.025
9 8.972 -0.028
10 9.970 -0.030
Remarks :
• The Calibration is performed with a Temperature of 27.05 °C at Customer's Facility
• The Expanded uncertainty is ± 1.50 µm at 95.45 % confidence level with coverage factor k=2.

Checked By Approved By
K. MUHILAN - CALIBRATION TECHNICIAN R.R.HARI JAIRAM - TECHNICAL MANAGER
HT / FT / 7.8 / 01 End of Certificate Page 1 of 1
CALIBRATION CERTIFICATE
Certificate No. HT/CC/201210-11/008 ULR No. ULR - CC215120000038208F
CUSTOMER DETAILS DEVICE UNDER CALIBRATION DETAILS
M/s. L&T Construction Research and
Name Description LVDT Probe with Indicator
Testing Centre,
L&T Construction, Make GDS/RDP Model DTCH 200AG
PB.No.979, Range 0 to 10 mm Least Count 0.001 mm
Address Mount Poonamallee Road, Identification S-CON-07/S-LVDT-09
Manapakkam, Serial No. 191677
Chennai - 600 089. Location Soil Lab
Job No. 201210-11 Type Digital
Job Receipt Date 10-Dec-2020 Calibration On 10-Dec-2020
Issue Date 22-Dec-2020 Due On -- (Customer Defined)
Received Condition of DUC Satisfactory Calibrated By M. Deepan Ram
ENVIRONMENTAL CONDITION
Temperature 20 ± 2 °C Humidity 40 %RH to 60 %RH
Discipline & Group Mechanical Calibration & Dimension (Basic Measuring Instrument,Gauge Etc.)
Calibration Procedure HT/WI/M/D (Section 07) by Comparison Method
STANDARD DETAILS
Description Slip Gauge Sets (Tungsten Carbide)
Identification No. M / D / SG / 01
Certificate No. C-200717-5-1
Valid Till 17-July-2022
CALIBRATION RESULT
(All Values are in mm)
Standard Reading DUC Reading Error
0 0.000 0.000
0.5 0.495 -0.005
1 0.990 -0.010
1.5 1.491 -0.009
2 1.990 -0.010
3 2.991 -0.009
4 3.987 -0.013
5 4.985 -0.015
6 5.982 -0.018
7 6.983 -0.017
8 7.980 -0.020
9 8.981 -0.019
10 9.980 -0.020
Remarks :
• The Calibration is performed with a Temperature of 27.05 °C at Customer's Facility
• The Expanded uncertainty is ± 1.50 µm at 95.45 % confidence level with coverage factor k=2.

Checked By Approved By
K. MUHILAN - CALIBRATION TECHNICIAN R.R.HARI JAIRAM - TECHNICAL MANAGER
HT / FT / 7.8 / 01 End of Certificate Page 1 of 1
CALIBRATION CERTIFICATE
Certificate No. HT/CC/201210-11/009 ULR No. ULR - CC215120000038209F
CUSTOMER DETAILS DEVICE UNDER CALIBRATION DETAILS
M/s. L&T Construction Research and
Name Description LVDT Probe with Indicator
Testing Centre,
L&T Construction, Make GDS/RDP Model DTCH 200AG
PB.No.979, Range 0 to 10 mm Least Count 0.001 mm
Address Mount Poonamallee Road, Identification S-CON-06/S-LVDT-08
Manapakkam, Serial No. 191680
Chennai - 600 089. Location Soil Lab
Job No. 201210-11 Type Digital
Job Receipt Date 10-Dec-2020 Calibration On 10-Dec-2020
Issue Date 22-Dec-2020 Due On -- (Customer Defined)
Received Condition of DUC Satisfactory Calibrated By M. Deepan Ram
ENVIRONMENTAL CONDITION
Temperature 20 ± 2 °C Humidity 40 %RH to 60 %RH
Discipline & Group Mechanical Calibration & Dimension (Basic Measuring Instrument,Gauge Etc.)
Calibration Procedure HT/WI/M/D (Section 07) by Comparison Method
STANDARD DETAILS
Description Slip Gauge Sets (Tungsten Carbide)
Identification No. M / D / SG / 01
Certificate No. C-200717-5-1
Valid Till 17-July-2022
CALIBRATION RESULT
(All Values are in mm)
Standard Reading DUC Reading Error
0 0.000 0.000
0.5 0.496 -0.004
1 0.994 -0.006
1.5 1.492 -0.008
2 1.990 -0.010
3 2.991 -0.009
4 3.982 -0.018
5 4.985 -0.015
6 5.990 -0.010
7 6.994 -0.006
8 7.997 -0.003
9 8.998 -0.002
10 9.995 -0.005
Remarks :
• The Calibration is performed with a Temperature of 27.05 °C at Customer's Facility
• The Expanded uncertainty is ± 1.50 µm at 95.45 % confidence level with coverage factor k=2.

Checked By Approved By
K. MUHILAN - CALIBRATION TECHNICIAN R.R.HARI JAIRAM - TECHNICAL MANAGER
HT / FT / 7.8 / 01 End of Certificate Page 1 of 1
CALIBRATION CERTIFICATE

Certificate No. HT/CC/201210-11/010 ULR No. ULR - CC215120000038210F


CUSTOMER DETAILS DEVICE UNDER CALIBRATION DETAILS
M/s. L&T Construction Research and Testing Temperature Controller with Sensor of Muffle
Name Description
Centre, Furnace
L&T Construction, Make Nubertherm Model LT/15/12-B410
PB.No.979,
Address Mount Poonamallee Road, Range 30 to 3000 °C
Manapakkam,
Chennai - 600 089. Least Count 1 °C
Job No. 201210-11 Accuracy --
Job Receipt Date 10-Dec-2020 Identification C-MF-03
Issue Date 22-Dec-2020 Serial No. --
Received Condition of DUC Satisfactory Location Chemistry Lab
ENVIRONMENTAL CONDITION Calibration On 11-Dec-2020
Temperature 28 °C ± 7 °C Due On --(Customer Defined)
Humidity 30 %RH to 75 %RH Calibrated By K. Peter Remigius
Discipline & Group Thermal Calibration & Temperature
Calibration Procedure HT / WI / T (Section 05) & AMS 2750 E by Comparison Method
STANDARD DETAILS
Description Multifunction Calibrator with Thermocouple
Identification No. T/TS-1/MFC-02
Certificate No. HT/CC/201112-12/006
Valid Till 12-Nov-2021

CALIBRATION RESULT

Expanded Uncertainty
Standard Reading in °C DUC Reading in °C Error in °C
in ± °C
298.94 300 1.06 0.69
548.55 550 1.45 1.50
623.36 625 1.64 1.86
798.14 800 1.86 1.86
998.06 1000 1.94 1.73
1197.94 1200 2.06 1.73
Remarks:
• The Calibration is performed at Customer's Facility.
• The Expanded Uncertainty is expressed at 95.45 % Confidence Level with Coverage Factor k=2.

Checked By Approved By
K. MUHILAN - CALIBRATION TECHNICIAN A. GOPINATH - ASSISTANT MANAGER
HT / FT / 7.8 / 01 End of Certificate Page 1 of 1
CALIBRATION CERTIFICATE

Certificate No. HT/CC/201210-11/011 ULR No. ULR - CC215120000038211F


CUSTOMER DETAILS DEVICE UNDER CALIBRATION DETAILS
M/s. L&T Construction Research and
Name Description Thickness Gauge
Testing Centre,
L&T Construction, Make --
PB.No.979, Range 0 to 12.7 mm Least Count 0.001 mm
Address Mount Poonamallee Road, Identification C-LCA-02
Manapakkam, Serial No. 1612191
Chennai - 600 089. Location --
Job No. 201210-11 Type Digital
Job Receipt Date 10-Dec-2020 Calibration On 11-Dec-2020
Issue Date 22-Dec-2020 Due On -- (Customer Defined)
Received Condition of DUC Satisfactory Calibrated By S.Muthamizhselvi
ENVIRONMENTAL CONDITION
Temperature 20 ± 2 °C Humidity 40 %RH to 60 %RH
Discipline & Group Mechanical Calibration & Dimension (Basic Measuring Instrument,Gauge Etc.)
Calibration Procedure HT/WI/M/D (Section 10) by Comparison Method
STANDARD DETAILS
Description Slip Gauge Sets (Tungsten Carbide)
Identification No. M / D / SG / 02
Certificate No. C-200522-4-1
Valid Till 22-May-2022
CALIBRATION RESULT
(All Values are in mm)
Standard Reading DUC Reading Error
0 0.000 0.000
0.5 0.499 -0.001
1 1.001 0.001
1.5 1.499 -0.001
2 2.000 0.000
4 4.000 0.000
6 5.999 -0.001
8 7.999 -0.001
10 9.998 -0.002
12 11.998 -0.002
Parallelism of Measuring Face : 0.002 mm
Remarks :
• The Calibration is performed at Permanent Facility.
• The Expanded uncertainty is ± 0.70 µm at 95.45 % confidence level with coverage factor k=2.

Checked By Approved By
K. MUHILAN - CALIBRATION TECHNICIAN R.R.HARI JAIRAM - TECHNICAL MANAGER
HT / FT / 7.8 / 01 End of Certificate Page 1 of 1
CALIBRATION CERTIFICATE
Certificate No. HT/CC/201210-11/012 ULR No. ULR - CC215120000038212F
CUSTOMER DETAILS DEVICE UNDER CALIBRATION DETAILS
M/s. L&T Construction Research and Testing
Name Description Stop Watch
Centre,
L&T Construction, Make Racer Model 630
PB.No.979,
Address Mount Poonamallee Road, Calibration Range Upto 2 h
Manapakkam,
Chennai - 600 089. Least Count 0.01/1 s
Job No. 201210-11 --
Accuracy
Job Receipt Date 10-Dec-2020 Identification B-SW-03
Issue Date 22-Dec-2020 Serial No. --
Received Condition of DUC SatisfactoryLocation --
ENVIRONMENTAL CONDITION Calibration On 11-Dec-2020
Temperature 25 °C ± 4 °C Due On --(Customer Defined)
Humidity 30 %RH to 75 %RH Calibrated By R. Shobana
Discipline & Group Electro Technical Calibration & Time and Frequency
Calibration Procedure HT / WI / E (Section 03) & NIST 960-12 by Comparison Method
STANDARD DETAILS
Description Stop Watch
Identification No. ET/SW/11
Certificate No. HT/CC/200725-17/001
Valid Till 26-July-2021
CALIBRATION RESULT
Expanded Uncertainty in ± h .
DUC Reading in min . s . cs Standard Reading in h. min . s . ms Error in h . min . s . ms
min . s . ms
00 . 06 . 15 000 . 00 . 06 . 272 (6.272 s) -000 . 00 . 00 . 122 (-0.122 s) 000 . 00 . 00 . 112 (0.112 s)
05 . 00 . 13 000 . 05 . 00 . 291 (300.291 s) -000 . 00 . 00 . 161 (-0.161 s) 000 . 00 . 00 . 263 (0.263 s)
10 . 00 . 19 000 . 10 . 00 . 328 (600.328 s) -000 . 00 . 00 . 138 (-0.138 s) 000 . 00 . 00 . 264 (0.264 s)
15 . 00 . 23 000 . 15 . 00 . 340 (900.340 s) -000 . 00 . 00 . 110 (-0.110 s) 000 . 00 . 00 . 264 (0.264 s)
20 . 00 . 23 000 . 20 . 00 . 389 (1200.389 s) -000 . 00 . 00 . 159 (-0.159 s) 000 . 00 . 01 . 038 (1.038 s)
25 . 00 . 13 000 . 25 . 00 . 419 (1500.419 s) -000 . 00 . 00 . 289 (-0.289 s) 000 . 00 . 01 . 038 (1.038 s)
Expanded Uncertainty in ± h .
DUC Reading in h . min . s Standard Reading in h. min . s . ms Error in h . min . s . ms
min . s . ms
00 . 30 . 00 000 . 30 . 00 . 470 (1800.470 s) -000 . 00 . 00 . 470 (-0.470 s) 000 . 00 . 01 . 188 (1.188 s)
00 . 35 . 00 000 . 35 . 00 . 495 (2100.495 s) -000 . 00 . 00 . 495 (-0.495 s) 000 . 00 . 01 . 188 (1.188 s)
00 . 40 . 00 000 . 40 . 00 . 527 (2400.527 s) -000 . 00 . 00 . 527 (-0.527 s) 000 . 00 . 01 . 188 (1.188 s)
00 . 45 . 00 000 . 45 . 00 . 561 (2700.561 s) -000 . 00 . 00 . 561 (-0.561 s) 000 . 00 . 01 . 188 (1.188 s)

Checked By Approved By
K. MUHILAN - CALIBRATION TECHNICIAN A. GOPINATH - ASSISTANT MANAGER
HT / FT / 7.8 / 01 Page 1 of 2
CALIBRATION CERTIFICATE
Certificate No. HT/CC/201210-11/012 ULR No. ULR - CC215120000038212F
CALIBRATION RESULT
Expanded Uncertainty in ± h .
DUC Reading in h . min . s Standard Reading in h. min . s . ms Error in h . min . s . ms
min . s . ms
00 . 50 . 00 000 . 50 . 00 . 587 (3000.587 s) -000 . 00 . 00 . 587 (-0.587 s) 000 . 00 . 01 . 188 (1.188 s)
00 . 55 . 00 000 . 55 . 00 . 616 (3300.616 s) -000 . 00 . 00 . 616 (-0.616 s) 000 . 00 . 01 . 188 (1.188 s)
01 . 00 . 00 001 . 00 . 00 . 655 (3600.655 s) -000 . 00 . 00 . 655 (-0.655 s) 000 . 00 . 01 . 188 (1.188 s)
01 . 05 . 00 001 . 05 . 00 . 685 (3900.685 s) -000 . 00 . 00 . 685 (-0.685 s) 000 . 00 . 01 . 188 (1.188 s)
01 . 10 . 00 001 . 10 . 00 . 724 (4200.724 s) -000 . 00 . 00 . 724 (-0.724 s) 000 . 00 . 01 . 188 (1.188 s)
01 . 20 . 00 001 . 20 . 00 . 746 (4800.746 s) -000 . 00 . 00 . 746 (-0.746 s) 000 . 00 . 01 . 188 (1.188 s)
01 . 25 . 00 001 . 25 . 00 . 784 (5100.784 s) -000 . 00 . 00 . 784 (-0.784 s) 000 . 00 . 01 . 536 (1.536 s)
01 . 30 . 00 001 . 30 . 00 . 897 (5400.897 s) -000 . 00 . 00 . 897 (-0.897 s) 000 . 00 . 01 . 536 (1.536 s)
01 . 35 . 00 001 . 35 . 00 . 969 (5700.969 s) -000 . 00 . 00 . 969 (-0.969 s) 000 . 00 . 01 . 536 (1.536 s)
01 . 40 . 00 001 . 40 . 01 . 126 (6001.126 s) -000 . 00 . 01 . 126 (-1.126 s) 000 . 00 . 01 . 536 (1.536 s)
01 . 45 . 00 001 . 45 . 01 . 170 (6301.170 s) -000 . 00 . 01 . 170 (-1.170 s) 000 . 00 . 01 . 536 (1.536 s)
01 . 50 . 00 001 . 50 . 01 . 224 (6601.224 s) -000 . 00 . 01 . 224 (-1.224 s) 000 . 00 . 01 . 536 (1.536 s)
01 . 55 . 00 001 . 55 . 01 . 233 (6901.233 s) -000 . 00 . 01 . 233 (-1.233 s) 000 . 00 . 01 . 536 (1.536 s)
02 . 00 . 00 002 . 00 . 01 . 325 (7201.325 s) -000 . 00 . 01 . 325 (-1.325 s) 000 . 00 . 01 . 536 (1.536 s)

Remarks :
• The Calibration is performed at Permanent Facility.
• The Expanded uncertainty is expressed at 95.45 % confidence level with coverage factor k=2.

Checked By Approved By
K. MUHILAN - CALIBRATION TECHNICIAN A. GOPINATH - ASSISTANT MANAGER
HT / FT / 7.8 / 01 End of Certificate Page 2 of 2
CALIBRATION CERTIFICATE
Certificate No. HT/CC/201210-11/013 ULR No. ULR - CC215120000038213F
CUSTOMER DETAILS DEVICE UNDER CALIBRATION DETAILS
M/s. L&T Construction Research and Testing
Name Description Stop Watch
Centre,
L&T Construction, Make Racer Model 630
PB.No.979,
Address Mount Poonamallee Road, Calibration Range Upto 6 h
Manapakkam,
Chennai - 600 089. Least Count 0.01/1 s
Job No. 201210-11 --
Accuracy
Job Receipt Date 10-Dec-2020 Identification C-SW-04
Issue Date 22-Dec-2020 Serial No. --
Received Condition of DUC Satisfactory Location --
ENVIRONMENTAL CONDITION Calibration On 11-Dec-2020 to 13-Dec-2020
Temperature 25 °C ± 4 °C Due On --(Customer Defined)
Humidity 30 %RH to 75 %RH Calibrated By R. Shobana
Discipline & Group Electro Technical Calibration & Time and Frequency
Calibration Procedure HT / WI / E (Section 03) & NIST 960-12 by Comparison Method
STANDARD DETAILS
Description Time Totaliser
Identification No. ET/TM/02
Certificate No. HT/4262/002 & 20/57 - 03
Valid Till 01-Jan-2021 & 11-Jun-2021
CALIBRATION RESULT
Expanded Uncertainty in ± h .
DUC Reading in min . s . cs Standard Reading in h. min . s . ms Error in h . min . s . ms
min . s . ms
05 . 00 . 14 000 . 05 . 00 . 241 (300.241 s) -000 . 00 . 00 . 101 (-0.101 s) 000 . 00 . 00 . 089 (0.089 s)
10 . 00 . 18 000 . 10 . 00 . 294 (600.294 s) -000 . 00 . 00 . 114 (-0.114 s) 000 . 00 . 00 . 088 (0.088 s)
15 . 00 . 20 000 . 15 . 00 . 360 (900.360 s) -000 . 00 . 00 . 160 (-0.160 s) 000 . 00 . 00 . 089 (0.089 s)
20 . 00 . 25 000 . 20 . 00 . 381 (1200.381 s) -000 . 00 . 00 . 131 (-0.131 s) 000 . 00 . 01 . 347 (1.347 s)
25 . 00 . 26 000 . 25 . 00 . 492 (1500.492 s) -000 . 00 . 00 . 232 (-0.232 s) 000 . 00 . 01 . 347 (1.347 s)
Expanded Uncertainty in ± h .
DUC Reading in h . min . s Standard Reading in h. min . s . ms Error in h . min . s . ms
min . s . ms
00 . 30 . 00 000 . 30 . 00 . 404 (1800.404 s) -000 . 00 . 00 . 404 (-0.404 s) 000 . 00 . 01 . 466 (1.466 s)
00 . 35 . 00 000 . 35 . 00 . 422 (2100.422 s) -000 . 00 . 00 . 422 (-0.422 s) 000 . 00 . 01 . 466 (1.466 s)
00 . 40 . 00 000 . 40 . 00 . 436 (2400.436 s) -000 . 00 . 00 . 436 (-0.436 s) 000 . 00 . 01 . 466 (1.466 s)
00 . 45 . 00 000 . 45 . 00 . 474 (2700.474 s) -000 . 00 . 00 . 474 (-0.474 s) 000 . 00 . 01 . 466 (1.466 s)
00 . 50 . 00 000 . 50 . 00 . 490 (3000.490 s) -000 . 00 . 00 . 490 (-0.490 s) 000 . 00 . 01 . 466 (1.466 s)

Checked By Approved By
K. MUHILAN - CALIBRATION TECHNICIAN A. GOPINATH - ASSISTANT MANAGER
HT / FT / 7.8 / 01 Page 1 of 2
CALIBRATION CERTIFICATE
Certificate No. HT/CC/201210-11/013 ULR No. ULR - CC215120000038213F
CALIBRATION RESULT
Expanded Uncertainty in ± h .
DUC Reading in h . min . s Standard Reading in h. min . s . ms Error in h . min . s . ms
min . s . ms
00 . 55 . 00 000 . 55 . 00 . 508 (3300.508 s) -000 . 00 . 00 . 508 (-0.508 s) 000 . 00 . 01 . 466 (1.466 s)
01 . 00 . 00 001 . 00 . 00 . 538 (3600.538 s) -000 . 00 . 00 . 538 (-0.538 s) 000 . 00 . 01 . 466 (1.466 s)
01 . 05 . 00 001 . 05 . 00 . 578 (3900.578 s) -000 . 00 . 00 . 578 (-0.578 s) 000 . 00 . 01 . 466 (1.466 s)
01 . 10 . 00 001 . 10 . 00 . 611 (4200.611 s) -000 . 00 . 00 . 611 (-0.611 s) 000 . 00 . 01 . 466 (1.466 s)
01 . 20 . 00 001 . 20 . 00 . 661 (4800.661 s) -000 . 00 . 00 . 661 (-0.661 s) 000 . 00 . 01 . 466 (1.466 s)
01 . 25 . 00 001 . 25 . 00 . 698 (5100.698 s) -000 . 00 . 00 . 698 (-0.698 s) 000 . 00 . 01 . 801 (1.801 s)
01 . 30 . 00 001 . 30 . 00 . 777 (5400.777 s) -000 . 00 . 00 . 777 (-0.777 s) 000 . 00 . 01 . 801 (1.801 s)
01 . 35 . 00 001 . 35 . 00 . 828 (5700.828 s) -000 . 00 . 00 . 828 (-0.828 s) 000 . 00 . 01 . 801 (1.801 s)
01 . 40 . 00 001 . 40 . 00 . 884 (6000.884 s) -000 . 00 . 00 . 884 (-0.884 s) 000 . 00 . 01 . 801 (1.801 s)
01 . 45 . 00 001 . 45 . 00 . 952 (6300.952 s) -000 . 00 . 00 . 952 (-0.952 s) 000 . 00 . 01 . 801 (1.801 s)
01 . 50 . 00 001 . 50 . 01 . 155 (6601.155 s) -000 . 00 . 01 . 155 (-1.155 s) 000 . 00 . 01 . 801 (1.801 s)
01 . 55 . 00 001 . 55 . 01 . 189 (6901.189 s) -000 . 00 . 01 . 189 (-1.189 s) 000 . 00 . 01 . 801 (1.801 s)
02 . 00 . 00 002 . 00 . 01 . 227 (7201.227 s) -000 . 00 . 01 . 227 (-1.227 s) 000 . 00 . 01 . 801 (1.801 s)
03 . 00 . 00 003 . 00 . 01 . 289 (10801.289 s) -000 . 00 . 01 . 289 (-1.289 s) 000 . 00 . 01 . 783 (1.783 s)
04 . 00 . 00 004 . 00 . 02 . 321 (14402.321 s) -000 . 00 . 02 . 321 (-2.321 s) 000 . 00 . 01 . 783 (1.783 s)
05 . 00 . 00 005 . 00 . 03 . 351 (18003.351 s) -000 . 00 . 03 . 351 (-3.351 s) 000 . 00 . 01 . 783 (1.783 s)
06 . 00 . 00 006 . 00 . 03 . 385 (21603.385 s) -000 . 00 . 03 . 385 (-3.385 s) 000 . 00 . 01 . 783 (1.783 s)

Remarks :
• The Calibration is performed at Permanent Facility.
• The Expanded uncertainty is expressed at 95.45 % confidence level with coverage factor k=2.

Checked By Approved By
K. MUHILAN - CALIBRATION TECHNICIAN A. GOPINATH - ASSISTANT MANAGER
HT / FT / 7.8 / 01 End of Certificate Page 2 of 2
CALIBRATION CERTIFICATE
Certificate No. HT/CC/201210-11/014 ULR No. ULR - CC215120000038214F
CUSTOMER DETAILS DEVICE UNDER CALIBRATION DETAILS
M/s. L&T Construction Research and Testing
Name Description Stop Watch
Centre,
L&T Construction, Make Racer Model 630
PB.No.979,
Address Mount Poonamallee Road, Calibration Range Upto 2 h
Manapakkam,
Chennai - 600 089. Least Count 0.01/1 s
Job No. 201210-11 --
Accuracy
Job Receipt Date 10-Dec-2020 Identification S-SW-09
Issue Date 22-Dec-2020 Serial No. --
Received Condition of DUC SatisfactoryLocation --
ENVIRONMENTAL CONDITION Calibration On 11-Dec-2020
Temperature 25 °C ± 4 °C Due On --(Customer Defined)
Humidity 30 %RH to 75 %RH Calibrated By R. Shobana
Discipline & Group Electro Technical Calibration & Time and Frequency
Calibration Procedure HT / WI / E (Section 03) & NIST 960-12 by Comparison Method
STANDARD DETAILS
Description Stop Watch
Identification No. ET/SW/11
Certificate No. HT/CC/200725-17/001
Valid Till 26-July-2021
CALIBRATION RESULT
Expanded Uncertainty in ± h .
DUC Reading in min . s . cs Standard Reading in h. min . s . ms Error in h . min . s . ms
min . s . ms
00 . 05 . 12 000 . 00 . 05 . 262 (5.262 s) -000 . 00 . 00 . 142 (-0.142 s) 000 . 00 . 00 . 115 (0.115 s)
05 . 00 . 18 000 . 05 . 00 . 324 (300.324 s) -000 . 00 . 00 . 144 (-0.144 s) 000 . 00 . 00 . 263 (0.263 s)
10 . 00 . 20 000 . 10 . 00 . 403 (600.403 s) -000 . 00 . 00 . 203 (-0.203 s) 000 . 00 . 00 . 264 (0.264 s)
15 . 00 . 31 000 . 15 . 00 . 433 (900.433 s) -000 . 00 . 00 . 123 (-0.123 s) 000 . 00 . 00 . 265 (0.265 s)
20 . 00 . 32 000 . 20 . 00 . 481 (1200.481 s) -000 . 00 . 00 . 161 (-0.161 s) 000 . 00 . 01 . 038 (1.038 s)
25 . 00 . 10 000 . 25 . 00 . 306 (1500.306 s) -000 . 00 . 00 . 206 (-0.206 s) 000 . 00 . 01 . 038 (1.038 s)
Expanded Uncertainty in ± h .
DUC Reading in h . min . s Standard Reading in h. min . s . ms Error in h . min . s . ms
min . s . ms
00 . 30 . 00 000 . 30 . 00 . 397 (1800.397 s) -000 . 00 . 00 . 397 (-0.397 s) 000 . 00 . 01 . 188 (1.188 s)
00 . 35 . 00 000 . 35 . 00 . 431 (2100.431 s) -000 . 00 . 00 . 431 (-0.431 s) 000 . 00 . 01 . 188 (1.188 s)
00 . 40 . 00 000 . 40 . 00 . 492 (2400.492 s) -000 . 00 . 00 . 492 (-0.492 s) 000 . 00 . 01 . 188 (1.188 s)
00 . 45 . 00 000 . 45 . 00 . 506 (2700.506 s) -000 . 00 . 00 . 506 (-0.506 s) 000 . 00 . 01 . 188 (1.188 s)

Checked By Approved By
K. MUHILAN - CALIBRATION TECHNICIAN A. GOPINATH - ASSISTANT MANAGER
HT / FT / 7.8 / 01 Page 1 of 2
CALIBRATION CERTIFICATE
Certificate No. HT/CC/201210-11/014 ULR No. ULR - CC215120000038214F
CALIBRATION RESULT
Expanded Uncertainty in ± h .
DUC Reading in h . min . s Standard Reading in h. min . s . ms Error in h . min . s . ms
min . s . ms
00 . 50 . 00 000 . 50 . 00 . 520 (3000.520 s) -000 . 00 . 00 . 520 (-0.520 s) 000 . 00 . 01 . 188 (1.188 s)
00 . 55 . 00 000 . 55 . 00 . 547 (3300.547 s) -000 . 00 . 00 . 547 (-0.547 s) 000 . 00 . 01 . 188 (1.188 s)
01 . 00 . 00 001 . 00 . 00 . 587 (3600.587 s) -000 . 00 . 00 . 587 (-0.587 s) 000 . 00 . 01 . 188 (1.188 s)
01 . 05 . 00 001 . 05 . 00 . 641 (3900.641 s) -000 . 00 . 00 . 641 (-0.641 s) 000 . 00 . 01 . 188 (1.188 s)
01 . 10 . 00 001 . 10 . 00 . 678 (4200.678 s) -000 . 00 . 00 . 678 (-0.678 s) 000 . 00 . 01 . 188 (1.188 s)
01 . 20 . 00 001 . 20 . 00 . 719 (4800.719 s) -000 . 00 . 00 . 719 (-0.719 s) 000 . 00 . 01 . 188 (1.188 s)
01 . 25 . 00 001 . 25 . 00 . 727 (5100.727 s) -000 . 00 . 00 . 727 (-0.727 s) 000 . 00 . 01 . 536 (1.536 s)
01 . 30 . 00 001 . 30 . 00 . 775 (5400.775 s) -000 . 00 . 00 . 775 (-0.775 s) 000 . 00 . 01 . 536 (1.536 s)
01 . 35 . 00 001 . 35 . 00 . 850 (5700.850 s) -000 . 00 . 00 . 850 (-0.850 s) 000 . 00 . 01 . 536 (1.536 s)
01 . 40 . 00 001 . 40 . 00 . 981 (6000.981 s) -000 . 00 . 00 . 981 (-0.981 s) 000 . 00 . 01 . 536 (1.536 s)
01 . 45 . 00 001 . 45 . 01 . 125 (6301.125 s) -000 . 00 . 01 . 125 (-1.125 s) 000 . 00 . 01 . 536 (1.536 s)
01 . 50 . 00 001 . 50 . 01 . 248 (6601.248 s) -000 . 00 . 01 . 248 (-1.248 s) 000 . 00 . 01 . 536 (1.536 s)
01 . 55 . 00 001 . 55 . 01 . 290 (6901.290 s) -000 . 00 . 01 . 290 (-1.290 s) 000 . 00 . 01 . 536 (1.536 s)
02 . 00 . 00 002 . 00 . 01 . 305 (7201.305 s) -000 . 00 . 01 . 305 (-1.305 s) 000 . 00 . 01 . 536 (1.536 s)

Remarks :
• The Calibration is performed at Permanent Facility.
• The Expanded uncertainty is expressed at 95.45 % confidence level with coverage factor k=2.

Checked By Approved By
K. MUHILAN - CALIBRATION TECHNICIAN A. GOPINATH - ASSISTANT MANAGER
HT / FT / 7.8 / 01 End of Certificate Page 2 of 2
CALIBRATION CERTIFICATE
Certificate No. HT/CC/201210-11/015 ULR No. ULR - CC215120000038215F
CUSTOMER DETAILS DEVICE UNDER CALIBRATION DETAILS
M/s. L&T Construction Research and
Name Description Weights
Testing Centre,
L&T Construction, Make Weigh India
PB.No.979, Range Refer Result
Address Mount Poonamallee Road, Accuracy M1 Class Serial No. Refer Result
Manapakkam, Identification Refer Result
Chennai - 600 089. Material Brass
Job No. 201210-11 Location --
Job Receipt Date 10-Dec-2020 Calibration On 14-Dec-2020
Issue Date 22-Dec-2020 Due On --(Customer Defined)
Received Condition of DUC Satisfactory Calibrated By M. Deepan Ram
ENVIRONMENTAL CONDITION
Temperature 25 °C ± 2 °C Humidity 40 % RH to 60 % RH Pressure 900 mbar to 1100 mbar
Discipline & Group Mechanical Calibration & Mass and Volume
Calibration Procedure HT/WI/M/MV (Section 01) & OIML R 111-1 by Direct Comparison Method using ABBA Cycle
STANDARD DETAILS
Description Standard Weights & Standard Weight Box
Identification M/MV/W/0.5,1,5,10/02, 2/03, 2/04, 20/17 & M/MV/WB/02
Certificate No. HT/CC/200518-01/003 & HT/CC/200518-01/004
Valid Till 20-May-2021 & 17-May-2021
Refer Next Page For Result
Identification No. Nominal Mass Conventional Mass in g Expanded Uncertainty in ± g
M-LW-16 20 g 19.99904 0.00009
M-LW-15 * 20 19.99619 0.00009
M-LW-14 50 49.99745 0.00021
M-LW-13 100 99.9943 0.00044
M-LW-12 200 199.9804 0.00085
M-LW-11 * 200 199.992 0.001
M-LW-10 500 499.992 0.0024
Remarks:
• Change in Temperature per hour is ± 1.5 °C.
• The Calibration is performed at Permanent Facility.
• Assumed Density of DUC Weights is 8400 ± 170 kg/m³ .
• The Calibration Certificate is valid for scientific & industrial purpose only.
• Assumed Density of Standard Weights used during calibration is 7950 ± 140 kg/m³.
• The Expanded Uncertainty is expressed at 95.45 % confidence level with coverage factor k=2.
Checked By Approved By
K. MUHILAN - CALIBRATION TECHNICIAN R.KOWSELYA - CALIBRATION ENGINEER
HT / FT / 7.8 / 01 Page 1 of 2
CALIBRATION CERTIFICATE
Certificate No. HT/CC/201210-11/015 ULR No. ULR - CC215120000038215F

CALIBRATION RESULT

Serial No. Identification No. Nominal Mass Conventional Mass in g Expanded Uncertainty in ± g

1 M-LW-16 20 g 19.99904 0.00009


2 M-LW-15 * 20 19.99619 0.00009
-- M-LW-14 50 49.99745 0.00021
-- M-LW-13 100 99.9943 0.00044
2 M-LW-12 200 199.9804 0.00085
1 M-LW-11 * 200 199.992 0.0010
-- M-LW-10 500 499.992 0.0024
-- M-LW-09 1 999.999 0.079
1 M-LW-08 2 1999.99 0.16
2 M-LW-07 *2 1999.99 0.16
2 M-LW-06 5 5000.00 0.39
1 M-LW-05 *5 5000.00 0.39
4 M-LW-04 10 10000.08 0.79
3 M-LW-03 * 10 10000.04 0.79
2 M-LW-02 10 10000.06 0.79
1 M-LW-01 10 10000.06 0.79
Remarks:
• Change in Temperature per hour is ± 1.5 °C.
• The Calibration is performed at Permanent Facility.
• Assumed Density of DUC Weights is 8400 ± 170 kg/m³ .
• Thermal Stabilization time given before calibration : 30 mins.
• The Calibration Certificate is valid for scientific & industrial purpose only.
• Assumed Density of Standard Weights used during calibration is 7950 ± 140 kg/m³.
• The Expanded Uncertainty is expressed at 95.45 % confidence level with coverage factor k=2.

Checked By Approved By
K. MUHILAN - CALIBRATION TECHNICIAN R.KOWSELYA - CALIBRATION ENGINEER
HT / FT / 7.8 / 01 End of Certificate Page 2 of 2
CALIBRATION CERTIFICATE

Certificate No. HT/CC/201210-11/016 ULR No. ULR - CC215120000038216F


CUSTOMER DETAILS DEVICE UNDER CALIBRATION DETAILS
M/s. L&T Construction Research and Testing
Name Description Test Sieves
Centre,
Make Aimil Ltd
L&T Construction,
PB.No.979, Range 300 µm
Address Mount Poonamallee Road, Type Woven-Wire Cloth
Manapakkam, Material Brass
Chennai - 600 089.
Identification C-S16-45
Job No. 201210-11 Serial No. 165245
Job Receipt Date 10-Dec-2020 Calibration On 16-Dec-2020
Issue Date 22-Dec-2020 Due On -- (Customer Defined)
Received Condition of DUC Satisfactory Calibrated By S. Muthamizhselvi
ENVIRONMENTAL CONDITION
Temperature 20 ± 2 °C Humidity 40 %RH to 60 %RH
Discipline & Group Mechanical Calibration & Dimension (Basic Measuring Instrument, Gauge etc.)
Calibration Procedure HT/WI/M/D (Section 27) & IS 460 Part I, III by Comparison Method
STANDARD DETAILS
Description Profile Projector
Identification No. M / D / PP / 01
Certificate No. HT/1089/001
Valid Till 04-Jun-2021
CALIBRATION RESULT
Observed Value in mm
Description Nominal value in µm
Weft Direction Warp Direction
Average of apertures
300 0.307 0.312
over the entire sieve (y)
Maximum of aperture
300 0.313 0.318
over the entire sieve (x)
Wire Diameter 200 0.177 0.170
The Total No. of Apertures lies between (+x) & (+z) 0
Remarks :
• The Expanded uncertainty is ± 4.2 µm at 95.45 % confidence level with coverage factor k=2.
• Tolerance for Average of Aperture Size (y) is + 0.012 mm
• Maximum Tolerance for any one Aperture (x) is + 0.065 mm
• Tolerance for total number of aperture shall have sizes between (+x) & (+z) is 18
• The Calibration is performed at Permanent Facility

Checked By Approved By
K. MUHILAN - CALIBRATION TECHNICIAN R.R.HARI JAIRAM - TECHNICAL MANAGER
HT / FT / 7.8 / 01 End of Certificate Page 1 of 1
CALIBRATION CERTIFICATE

Certificate No. HT/CC/201210-11/017 ULR No. ULR - CC215120000038217F


CUSTOMER DETAILS DEVICE UNDER CALIBRATION DETAILS
M/s. L&T Construction Research and
Name Description Test Sieves
Testing Centre,
L&T Construction, Make Aimil Ltd
PB.No.979, Range 1.18 mm
Address Mount Poonamallee Road, Type Woven-Wire Cloth
Manapakkam, Material Brass
Chennai - 600 089. Identification C-SIE-43
Job No. 201210-11 Serial No. 165184
Job Receipt Date 10-Dec-2020 Calibration On 16-Dec-2020
Issue Date 22-Dec-2020 Due On -- (Customer Defined)
Received Condition of DUC Satisfactory Calibrated By S. Muthamizhselvi
ENVIRONMENTAL CONDITION
Temperature 20 ± 2 °C Humidity 40 %RH to 60 %RH
Discipline & Group Mechanical Calibration & Dimension (Basic Measuring Instrument, Gauge etc.)
Calibration Procedure HT/WI/M/D (Section 27) & IS 460 Part I, III by Comparison Method
STANDARD DETAILS
Description Profile Projector
Identification No. M / D / PP / 01
Certificate No. HT/1089/001
Valid Till 04-Jun-2021
CALIBRATION RESULT
Observed Value in mm
Description Nominal value in mm
Weft Direction Warp Direction
Average of apertures
1.18 1.050 1.060
over the entire sieve (y)
Maximum of aperture
1.18 1.095 1.069
over the entire sieve (x)
Wire Diameter 0.63 0.443 0.441
The Total No. of Apertures lies between (+x) & (+z) 0
Remarks :
• The Expanded uncertainty is ± 4.2 µm at 95.45 % confidence level with coverage factor k=2.
• Tolerance for Average of Aperture Size (y) is + 0.04 mm
• Maximum Tolerance for any one Aperture (x) is + 0.16 mm
• Tolerance for total number of aperture shall have sizes between (+x) & (+z) is 5
• The Calibration is performed at Permanent Facility

Checked By Approved By
K. MUHILAN - CALIBRATION TECHNICIAN R.R.HARI JAIRAM - TECHNICAL MANAGER
HT / FT / 7.8 / 01 End of Certificate Page 1 of 1
CALIBRATION CERTIFICATE

Certificate No. HT/CC/201210-11/018 ULR No. ULR - CC215120000038218F


CUSTOMER DETAILS DEVICE UNDER CALIBRATION DETAILS
M/s. L&T Construction Research and
Name Description Test Sieves
Testing Centre,
L&T Construction, Make Aimil Ltd
PB.No.979, Range 600 µm
Address Mount Poonamallee Road, Type Woven-Wire Cloth
Manapakkam, Material Brass
Chennai - 600 089. Identification C-SIE-43
Job No. 201210-11 Serial No. 165184
Job Receipt Date 10-Dec-2020 Calibration On 16-Dec-2020
Issue Date 22-Dec-2020 Due On -- (Customer Defined)
Received Condition of DUC Satisfactory Calibrated By S. Muthamizhselvi
ENVIRONMENTAL CONDITION
Temperature 20 ± 2 °C Humidity 40 %RH to 60 %RH
Discipline & Group Mechanical Calibration & Dimension (Basic Measuring Instrument, Gauge etc.)
Calibration Procedure HT/WI/M/D (Section 27) & IS 460 Part I, III by Comparison Method
STANDARD DETAILS
Description Profile Projector
Identification No. M / D / PP / 01
Certificate No. HT/1089/001
Valid Till 04-Jun-2021
CALIBRATION RESULT
Observed Value in mm
Description Nominal value in µm
Weft Direction Warp Direction
Average of apertures
600 0.615 0.616
over the entire sieve (y)
Maximum of aperture
600 0.618 0.619
over the entire sieve (x)
Wire Diameter 400 0.287 0.296
The Total No. of Apertures lies between (+x) & (+z) 0
Remarks :
• The Expanded uncertainty is ± 4.2 µm at 95.45 % confidence level with coverage factor k=2.
• Tolerance for Average of Aperture Size (y) is + 0.021 mm
• Maximum Tolerance for any one Aperture (x) is + 0.101 mm
• Tolerance for total number of aperture shall have sizes between (+x) & (+z) is 9
• The Calibration is performed at Permanent Facility

Checked By Approved By
K. MUHILAN - CALIBRATION TECHNICIAN R.R.HARI JAIRAM - TECHNICAL MANAGER
HT / FT / 7.8 / 01 End of Certificate Page 1 of 1
CALIBRATION CERTIFICATE

Certificate No. HT/CC/201210-11/019 ULR No. ULR - CC215120000038219F


CUSTOMER DETAILS DEVICE UNDER CALIBRATION DETAILS
M/s. L&T Construction Research and
Name Description Test Sieves
Testing Centre,
L&T Construction, Make Aimil Ltd
PB.No.979, Range 150 µm
Address Mount Poonamallee Road, Type Woven-Wire Cloth
Manapakkam, Material Brass
Chennai - 600 089. Identification C-SIE-46
Job No. 201210-11 Serial No. 165420
Job Receipt Date 10-Dec-2020 Calibration On 16-Dec-2020
Issue Date 22-Dec-2020 Due On -- (Customer Defined)
Received Condition of DUC Satisfactory Calibrated By S. Muthamizhselvi
ENVIRONMENTAL CONDITION
Temperature 20 ± 2 °C Humidity 40 %RH to 60 %RH
Discipline & Group Mechanical Calibration & Dimension (Basic Measuring Instrument, Gauge etc.)
Calibration Procedure HT/WI/M/D (Section 27) & IS 460 Part I, III by Comparison Method
STANDARD DETAILS
Description Profile Projector
Identification No. M / D / PP / 01
Certificate No. HT/1089/001
Valid Till 04-Jun-2021
CALIBRATION RESULT
Observed Value in mm
Description Nominal value in µm
Weft Direction Warp Direction
Average of apertures
150 0.151 0.152
over the entire sieve (y)
Maximum of aperture
150 0.156 0.155
over the entire sieve (x)
Wire Diameter 100 0.096 0.102
The Total No. of Apertures lies between (+x) & (+z) 0
Remarks :
• The Expanded uncertainty is ± 4.2 µm at 95.45 % confidence level with coverage factor k=2.
• Tolerance for Average of Aperture Size (y) is + 0.0066 mm
• Maximum Tolerance for any one Aperture (x) is + 0.043 mm
• Tolerance for total number of aperture shall have sizes between (+x) & (+z) is 18
• The Calibration is performed at Permanent Facility

Checked By Approved By
K. MUHILAN - CALIBRATION TECHNICIAN R.R.HARI JAIRAM - TECHNICAL MANAGER
HT / FT / 7.8 / 01 End of Certificate Page 1 of 1
CALIBRATION CERTIFICATE

Certificate No. HT/CC/201210-11/020 ULR No. ULR - CC215120000038220F


CUSTOMER DETAILS DEVICE UNDER CALIBRATION DETAILS
M/s. L&T Construction Research and
Name Description Test Sieves
Testing Centre,
L&T Construction, Make Aimil Ltd
PB.No.979, Range 1.7 mm
Address Mount Poonamallee Road, Type Woven-Wire Cloth
Manapakkam, Material Brass
Chennai - 600 089. Identification C-SIE-42
Job No. 201210-11 Serial No. 165052
Job Receipt Date 10-Dec-2020 Calibration On 16-Dec-2020
Issue Date 22-Dec-2020 Due On -- (Customer Defined)
Received Condition of DUC Satisfactory Calibrated By S. Muthamizhselvi
ENVIRONMENTAL CONDITION
Temperature 20 ± 2 °C Humidity 40 %RH to 60 %RH
Discipline & Group Mechanical Calibration & Dimension (Basic Measuring Instrument, Gauge etc.)
Calibration Procedure HT/WI/M/D (Section 27) & IS 460 Part I, III by Comparison Method
STANDARD DETAILS
Description Profile Projector
Identification No. M / D / PP / 01
Certificate No. HT/1089/001
Valid Till 04-Jun-2021
CALIBRATION RESULT
Observed Value in mm
Description Nominal value in mm
Weft Direction Warp Direction
Average of apertures
1.7 1.670 1.680
over the entire sieve (y)
Maximum of aperture
1.7 1.675 1.687
over the entire sieve (x)
Wire Diameter 0.8 0.435 0.440
The Total No. of Apertures lies between (+x) & (+z) 0
Remarks :
• The Expanded uncertainty is ± 4.2 µm at 95.45 % confidence level with coverage factor k=2.
• Tolerance for Average of Aperture Size (y) is + 0.06 mm
• Maximum Tolerance for any one Aperture (x) is + 0.2 mm
• Tolerance for total number of aperture shall have sizes between (+x) & (+z) is 3
• The Calibration is performed at Permanent Facility

Checked By Approved By
K. MUHILAN - CALIBRATION TECHNICIAN R.R.HARI JAIRAM - TECHNICAL MANAGER
HT / FT / 7.8 / 01 End of Certificate Page 1 of 1
CALIBRATION CERTIFICATE

Certificate No. HT/CC/201210-11/021 ULR No. ULR - CC215120000038221F


CUSTOMER DETAILS DEVICE UNDER CALIBRATION DETAILS
M/s. L&T Construction Research and
Name Description Test Sieves
Testing Centre,
L&T Construction, Make Jayant
PB.No.979, Range 600 µm
Address Mount Poonamallee Road, Type Woven-Wire Cloth
Manapakkam, Material Brass
Chennai - 600 089. Identification B-SIE-01
Job No. 201210-11 Serial No. --
Job Receipt Date 10-Dec-2020 Calibration On 16-Dec-2020
Issue Date 22-Dec-2020 Due On -- (Customer Defined)
Received Condition of DUC Satisfactory Calibrated By S. Muthamizhselvi
ENVIRONMENTAL CONDITION
Temperature 20 ± 2 °C Humidity 40 %RH to 60 %RH
Discipline & Group Mechanical Calibration & Dimension (Basic Measuring Instrument, Gauge etc.)
Calibration Procedure HT/WI/M/D (Section 27) & IS 460 Part I, III by Comparison Method
STANDARD DETAILS
Description Profile Projector
Identification No. M / D / PP / 01
Certificate No. HT/1089/001
Valid Till 04-Jun-2021
CALIBRATION RESULT
Observed Value in mm
Description Nominal value in µm
Weft Direction Warp Direction
Average of apertures
600 0.704 0.678
over the entire sieve (y)
Maximum of aperture
600 0.731 0.731
over the entire sieve (x)
Wire Diameter 400 0.358 0.362
The Total No. of Apertures lies between (+x) & (+z) 79
Remarks :
• The Expanded uncertainty is ± 4.2 µm at 95.45 % confidence level with coverage factor k=2.
• Tolerance for Average of Aperture Size (y) is + 0.021 mm
• Maximum Tolerance for any one Aperture (x) is + 0.101 mm
• Tolerance for total number of aperture shall have sizes between (+x) & (+z) is 9
• The Calibration is performed at Permanent Facility

Checked By Approved By
K. MUHILAN - CALIBRATION TECHNICIAN R.R.HARI JAIRAM - TECHNICAL MANAGER
HT / FT / 7.8 / 01 End of Certificate Page 1 of 1
CALIBRATION CERTIFICATE

Certificate No. HT/CC/201210-11/022 ULR No. ULR - CC215120000038222F


CUSTOMER DETAILS DEVICE UNDER CALIBRATION DETAILS
M/s. L&T Construction Research and
Name Description Test Sieves
Testing Centre,
L&T Construction, Make Jayant
PB.No.979, Range 1.4 mm
Address Mount Poonamallee Road, Type Woven-Wire Cloth
Manapakkam, Material Brass
Chennai - 600 089. Identification B-SIE-05
Job No. 201210-11 Serial No. --
Job Receipt Date 10-Dec-2020 Calibration On 16-Dec-2020
Issue Date 22-Dec-2020 Due On -- (Customer Defined)
Received Condition of DUC Satisfactory Calibrated By S. Muthamizhselvi
ENVIRONMENTAL CONDITION
Temperature 20 ± 2 °C Humidity 40 %RH to 60 %RH
Discipline & Group Mechanical Calibration & Dimension (Basic Measuring Instrument, Gauge etc.)
Calibration Procedure HT/WI/M/D (Section 27) & IS 460 Part I, III by Comparison Method
STANDARD DETAILS
Description Profile Projector
Identification No. M / D / PP / 01
Certificate No. HT/1089/001
Valid Till 04-Jun-2021
CALIBRATION RESULT
Observed Value in mm
Description Nominal value in mm
Weft Direction Warp Direction
Average of apertures
1.4 1.506 1.491
over the entire sieve (y)
Maximum of aperture
1.4 1.510 1.503
over the entire sieve (x)
Wire Diameter 0.71 0.651 0.688
The Total No. of Apertures lies between (+x) & (+z) 0
Remarks :
• The Expanded uncertainty is ± 4.2 µm at 95.45 % confidence level with coverage factor k=2.
• Tolerance for Average of Aperture Size (y) is + 0.05 mm
• Maximum Tolerance for any one Aperture (x) is + 0.18 mm
• Tolerance for total number of aperture shall have sizes between (+x) & (+z) is 4
• The Calibration is performed at Permanent Facility

Checked By Approved By
K. MUHILAN - CALIBRATION TECHNICIAN R.R.HARI JAIRAM - TECHNICAL MANAGER
HT / FT / 7.8 / 01 End of Certificate Page 1 of 1
CALIBRATION CERTIFICATE

Certificate No. HT/CC/201210-11/023 ULR No. ULR - CC215120000038223F


CUSTOMER DETAILS DEVICE UNDER CALIBRATION DETAILS
M/s. L&T Construction Research and
Name Description Test Sieves
Testing Centre,
L&T Construction, Make Jayant
PB.No.979, Range 1.4 mm
Address Mount Poonamallee Road, Type Woven-Wire Cloth
Manapakkam, Material Brass
Chennai - 600 089. Identification B-SIE-04
Job No. 201210-11 Serial No. --
Job Receipt Date 10-Dec-2020 Calibration On 16-Dec-2020
Issue Date 22-Dec-2020 Due On -- (Customer Defined)
Received Condition of DUC Satisfactory Calibrated By S. Muthamizhselvi
ENVIRONMENTAL CONDITION
Temperature 20 ± 2 °C Humidity 40 %RH to 60 %RH
Discipline & Group Mechanical Calibration & Dimension (Basic Measuring Instrument, Gauge etc.)
Calibration Procedure HT/WI/M/D (Section 27) & IS 460 Part I, III by Comparison Method
STANDARD DETAILS
Description Profile Projector
Identification No. M / D / PP / 01
Certificate No. HT/1089/001
Valid Till 04-Jun-2021
CALIBRATION RESULT
Observed Value in mm
Description Nominal value in mm
Weft Direction Warp Direction
Average of apertures
1.4 1.401 1.404
over the entire sieve (y)
Maximum of aperture
1.4 1.421 1.444
over the entire sieve (x)
Wire Diameter 0.71 0.658 0.686
The Total No. of Apertures lies between (+x) & (+z) 0
Remarks :
• The Expanded uncertainty is ± 4.2 µm at 95.45 % confidence level with coverage factor k=2.
• Tolerance for Average of Aperture Size (y) is + 0.05 mm
• Maximum Tolerance for any one Aperture (x) is + 0.18 mm
• Tolerance for total number of aperture shall have sizes between (+x) & (+z) is 4
• The Calibration is performed at Permanent Facility

Checked By Approved By
K. MUHILAN - CALIBRATION TECHNICIAN R.R.HARI JAIRAM - TECHNICAL MANAGER
HT / FT / 7.8 / 01 End of Certificate Page 1 of 1
CALIBRATION CERTIFICATE

Certificate No. HT/CC/201210-11/024 ULR No. ULR - CC215120000038224F


CUSTOMER DETAILS DEVICE UNDER CALIBRATION DETAILS
M/s. L&T Construction Research and
Name Description Test Sieves
Testing Centre,
L&T Construction, Make Aimil Ltd
PB.No.979, Range 2.36 mm
Address Mount Poonamallee Road, Type Woven-Wire Cloth
Manapakkam, Material Brass
Chennai - 600 089. Identification C-SIE-41
Job No. 201210-11 Serial No. --
Job Receipt Date 10-Dec-2020 Calibration On 16-Dec-2020
Issue Date 22-Dec-2020 Due On -- (Customer Defined)
Received Condition of DUC Satisfactory Calibrated By S. Muthamizhselvi
ENVIRONMENTAL CONDITION
Temperature 20 ± 2 °C Humidity 40 %RH to 60 %RH
Discipline & Group Mechanical Calibration & Dimension (Basic Measuring Instrument, Gauge etc.)
Calibration Procedure HT/WI/M/D (Section 27) & IS 460 Part I, III by Comparison Method
STANDARD DETAILS
Description Profile Projector
Identification No. M / D / PP / 01
Certificate No. HT/1089/001
Valid Till 04-Jun-2021
CALIBRATION RESULT
Observed Value in mm
Description Nominal value in mm
Weft Direction Warp Direction
Average of apertures
2.36 2.549 2.558
over the entire sieve (y)
Maximum of aperture
2.36 2.585 2.568
over the entire sieve (x)
Wire Diameter 1 0.640 0.713
The Total No. of Apertures lies between (+x) & (+z) 39
Remarks :
• The Expanded uncertainty is ± 4.2 µm at 95.45 % confidence level with coverage factor k=2.
• Tolerance for Average of Aperture Size (y) is + 0.08 mm
• Maximum Tolerance for any one Aperture (x) is + 0.25 mm
• Tolerance for total number of aperture shall have sizes between (+x) & (+z) is 2
• The Calibration is performed at Permanent Facility

Checked By Approved By
K. MUHILAN - CALIBRATION TECHNICIAN R.R.HARI JAIRAM - TECHNICAL MANAGER
HT / FT / 7.8 / 01 End of Certificate Page 1 of 1
CALIBRATION CERTIFICATE

Certificate No. HT/CC/201210-11/025 ULR No. ULR - CC215120000038225F


CUSTOMER DETAILS DEVICE UNDER CALIBRATION DETAILS
M/s. L&T Construction Research and
Name Description Test Sieves
Testing Centre,
L&T Construction, Make Aimil Ltd
PB.No.979, Range 4.75 mm
Address Mount Poonamallee Road, Type Perforated Plate
Manapakkam, Material Brass
Chennai - 600 089. Identification C-SIE-40
Job No. 201210-11 Serial No. --
Job Receipt Date 10-Dec-2020 Calibration On 16-Dec-2020
Issue Date 22-Dec-2020 Due On -- (Customer Defined)
Received Condition of DUC Satisfactory Calibrated By S. Muthamizhselvi
ENVIRONMENTAL CONDITION
Temperature 20 ± 2 °C Humidity 40 %RH to 60 %RH
Discipline & Group Mechanical Calibration & Dimension (Basic Measuring Instrument, Gauge etc.)
Calibration Procedure HT/WI/M/D (Section 27) & IS 460 Part II, III by Comparison Method
STANDARD DETAILS
Description Profile Projector
Identification No. M / D / PP / 01
Certificate No. HT/1089/001
Valid Till 04-Jun-2021
CALIBRATION RESULT
Observed Value in mm
Description Nominal value in mm
Weft Direction Warp Direction
Average of apertures
4.75 4.714 4.708
over the entire sieve (y)
Maximum of aperture
4.75 4.774 4.777
over the entire sieve (x)
Pitch Distance 6.6 6.948 7.086
The Total No. of Apertures lies between (+x) & (+z) 0
Remarks :
• The Expanded uncertainty is ± 4.2 µm at 95.45 % confidence level with coverage factor k=2.
• Tolerance for Average of Aperture Size (y) is ± 0.14 mm
• The Calibration is performed at Permanent Facility

Checked By Approved By
K. MUHILAN - CALIBRATION TECHNICIAN R.R.HARI JAIRAM - TECHNICAL MANAGER
HT / FT / 7.8 / 01 End of Certificate Page 1 of 1