K. MUHILAN - CALIBRATION TECHNICIAN R.KOWSELYA - CALIBRATION ENGINEER
HT / FT / 7.8 / 01 End of Certificate Page 1 of 1
CALIBRATION CERTIFICATE
Certificate No. HT/CC/201210-11/002 ULR No. ULR - CC215120000038202F
CUSTOMER DETAILS DEVICE UNDER CALIBRATION DETAILS
M/s. L&T Construction Research and Testing
Name Description Pressure Indicator
Centre,
L&T Construction, Make Geo Technik
PB.No.979, Model --
Address Mount Poonamallee Road,
Manapakkam, Range 0 to 3000 kPa
Chennai - 600 089. Least Count 0.1 kPa
Job No. 201210-11 Accuracy --
Job Receipt Date 10-Dec-2020 Identification S-BP-02 (TOP)
Issue Date 22-Dec-2020 Serial No. --
Received Condition of DUC Satisfactory Location Soil Lab
ENVIRONMENTAL CONDITION Calibration On 10-Dec-2020
Temperature 25 °C ± 5 °C Due On --(Customer Defined)
Humidity 30 % RH to 75 % RH Calibrated By V. Cibiraja
Discipline & Group Mechanical Calibration & Pressure and Vacuum
Calibration Procedure HT/WI/M/PV (Section 01) & DKD-R 6-1 (Sequence C ) by Comparison Method
STANDARD DETAILS
Description Process Calibrator
Identification No. M/PV/PC/160/01 & M/PV/PCI/01
Certificate No. HT/CC/200526-09/006
Valid Till 26-May-2021
CALIBRATION RESULT
Standard Reading in kPa Average of Expanded
Hysteresis
DUC Reading in kPa Standard Reading Error in kPa Uncertainty in ±
Increasing Decreasing in kPa
in kPa kPa
0.0 0.00000 0.00000 0.00000 0.00000 0.00000 --
25.0 24.9898 24.9860 24.9879 0.0121 0.0038 0.11
50.0 49.9726 49.9695 49.9711 0.0289 0.0031 0.11
100.0 99.9586 99.9560 99.9573 0.0427 0.0026 0.11
150.0 149.940 149.903 149.922 0.078 0.037 0.11
200.0 199.896 199.860 199.878 0.122 0.036 0.11
500.0 499.637 499.604 499.621 0.379 0.033 0.15
600.0 599.209 599.174 599.192 0.808 0.035 0.15
700.0 698.995 698.961 698.978 1.022 0.034 0.15
800.0 798.874 798.850 798.862 1.138 0.024 0.15
900.0 898.721 898.708 898.715 1.285 0.013 0.15
1000.0 998.593 998.580 998.587 1.413 0.013 0.15
2000.0 1998.18 1998.12 1998.15 1.85 0.06 0.36
2500.0 2497.96 2497.94 2497.95 2.05 0.02 0.35
Remarks :
• Pressure Conversion 1 kPa = 1000 Pa.
• The Calibration Certificate is valid for scientific & industrial purpose only.
• The Calibration is performed with a Temperature Change of ± 1 °C at Customer's Facility.
• The Error is provided as the Difference of DUC Reading and Average of Standard Reading.
• The Expanded uncertainty is expressed at 95.45 % confidence level with coverage factor k=2.
• Pressure Transmission Medium used is Water and the position of DUC during Calibration is Vertical.
• The Calibration Points are taken as per Customer's Request.
Checked By Approved By
K. MUHILAN - CALIBRATION TECHNICIAN R.KOWSELYA - CALIBRATION ENGINEER
HT / FT / 7.8 / 01 End of Certificate Page 1 of 1
CALIBRATION CERTIFICATE
Certificate No. HT/CC/201210-11/003 ULR No. ULR - CC215120000038203F
CUSTOMER DETAILS DEVICE UNDER CALIBRATION DETAILS
M/s. L&T Construction Research and Testing
Name Description Pressure Indicator
Centre,
L&T Construction, Make Geo Technik
PB.No.979, Model --
Address Mount Poonamallee Road,
Manapakkam, Range 0 to 3000 kPa
Chennai - 600 089. Least Count 0.1 kPa
Job No. 201210-11 Accuracy --
Job Receipt Date 10-Dec-2020 Identification S-BP-02 (Bottom)
Issue Date 22-Dec-2020 Serial No. --
Received Condition of DUC Satisfactory Location Soil Lab
ENVIRONMENTAL CONDITION Calibration On 10-Dec-2020
Temperature 25 °C ± 5 °C Due On --(Customer Defined)
Humidity 30 % RH to 75 % RH Calibrated By V. Cibiraja
Discipline & Group Mechanical Calibration & Pressure and Vacuum
Calibration Procedure HT/WI/M/PV (Section 01) & DKD-R 6-1 (Sequence C ) by Comparison Method
STANDARD DETAILS
Description Process Calibrator
Identification No. M/PV/PC/160/01 & M/PV/PCI/01
Certificate No. HT/CC/200526-09/006
Valid Till 26-May-2021
CALIBRATION RESULT
Standard Reading in kPa Average of Expanded
Hysteresis
DUC Reading in kPa Standard Reading Error in kPa Uncertainty in ±
Increasing Decreasing in kPa
in kPa kPa
0.0 0.00000 0.00000 0.00000 0.00000 0.00000 --
25.0 25.3145 25.3113 25.3129 -0.3129 0.0032 0.11
50.0 50.4030 50.4008 50.4019 -0.4019 0.0022 0.11
100.0 100.752 100.215 100.484 -0.484 0.537 0.33
150.0 150.792 150.769 150.781 -0.781 0.023 0.11
200.0 200.032 200.008 200.020 -0.020 0.024 0.11
500.0 499.093 499.050 499.072 0.928 0.043 0.15
600.0 599.227 599.208 599.218 0.782 0.019 0.15
700.0 698.997 698.978 698.988 1.012 0.019 0.15
800.0 798.026 798.012 798.019 1.981 0.014 0.15
900.0 898.067 898.052 898.060 1.940 0.015 0.15
1000.0 997.877 997.869 997.873 2.127 0.008 0.15
1500.0 1497.54 1497.46 1497.50 2.50 0.08 0.16
2000.0 1997.18 1997.10 1997.14 2.86 0.08 0.36
2500.0 2497.06 2497.01 2497.04 2.96 0.05 0.36
Remarks :
• Pressure Conversion 1 kPa = 1000 Pa.
• The Calibration Certificate is valid for scientific & industrial purpose only.
• The Calibration is performed with a Temperature Change of ± 1 °C at Customer's Facility.
• The Error is provided as the Difference of DUC Reading and Average of Standard Reading.
• The Expanded uncertainty is expressed at 95.45 % confidence level with coverage factor k=2.
• Pressure Transmission Medium used is Water and the position of DUC during Calibration is Vertical.
• The Calibration Points are taken as per Customer's Request.
Checked By Approved By
K. MUHILAN - CALIBRATION TECHNICIAN R.KOWSELYA - CALIBRATION ENGINEER
HT / FT / 7.8 / 01 End of Certificate Page 1 of 1
CALIBRATION CERTIFICATE
Certificate No. HT/CC/201210-11/004 ULR No. ULR - CC215120000038204F
CUSTOMER DETAILS DEVICE UNDER CALIBRATION DETAILS
M/s. L&T Construction Research and
Name Description LVDT Probe with Indicator
Testing Centre,
L&T Construction, Make GDS/RDP Model DTCH 200AG
PB.No.979, Range 0 to 10 mm Least Count 0.001 mm
Address Mount Poonamallee Road, Identification S-CON-08/S-LVDT-10
Manapakkam, Serial No. --
Chennai - 600 089. Location Soil Lab
Job No. 201210-11 Type Digital
Job Receipt Date 10-Dec-2020 Calibration On 10-Dec-2020
Issue Date 22-Dec-2020 Due On -- (Customer Defined)
Received Condition of DUC Satisfactory Calibrated By M. Deepan Ram
ENVIRONMENTAL CONDITION
Temperature 20 ± 2 °C Humidity 40 %RH to 60 %RH
Discipline & Group Mechanical Calibration & Dimension (Basic Measuring Instrument,Gauge Etc.)
Calibration Procedure HT/WI/M/D (Section 07) by Comparison Method
STANDARD DETAILS
Description Slip Gauge Sets (Tungsten Carbide)
Identification No. M / D / SG / 01
Certificate No. C-200717-5-1
Valid Till 17-July-2022
CALIBRATION RESULT
(All Values are in mm)
Standard Reading DUC Reading Error
0 0.000 0.000
0.5 0.497 -0.003
1 0.995 -0.005
1.5 1.490 -0.010
2 1.990 -0.010
3 2.988 -0.012
4 3.985 -0.015
5 4.986 -0.014
6 5.983 -0.017
7 6.982 -0.018
8 7.981 -0.019
9 8.980 -0.020
10 9.978 -0.022
Remarks :
• The Calibration is performed with a Temperature of 27.95 °C at Customer's Facility
• The Expanded uncertainty is ± 1.50 µm at 95.45 % confidence level with coverage factor k=2.
Checked By Approved By
K. MUHILAN - CALIBRATION TECHNICIAN R.R.HARI JAIRAM - TECHNICAL MANAGER
HT / FT / 7.8 / 01 End of Certificate Page 1 of 1
CALIBRATION CERTIFICATE
Certificate No. HT/CC/201210-11/005 ULR No. ULR - CC215120000038205F
CUSTOMER DETAILS DEVICE UNDER CALIBRATION DETAILS
M/s. L&T Construction Research and
Name Description LVDT Probe with Indicator
Testing Centre,
L&T Construction, Make RDP/GDS Model DTCH 200AG
PB.No.979, Range 0 to 10 mm Least Count 0.001 mm
Address Mount Poonamallee Road, Identification S-CON-05/S-LVDT-06
Manapakkam, Serial No. 166618
Chennai - 600 089. Location Soil Lab
Job No. 201210-11 Type Digital
Job Receipt Date 10-Dec-2020 Calibration On 10-Dec-2020
Issue Date 22-Dec-2020 Due On -- (Customer Defined)
Received Condition of DUC Satisfactory Calibrated By M. Deepan Ram
ENVIRONMENTAL CONDITION
Temperature 20 ± 2 °C Humidity 40 %RH to 60 %RH
Discipline & Group Mechanical Calibration & Dimension (Basic Measuring Instrument,Gauge Etc.)
Calibration Procedure HT/WI/M/D (Section 07) by Comparison Method
STANDARD DETAILS
Description Slip Gauge Sets (Tungsten Carbide)
Identification No. M / D / SG / 01
Certificate No. C-200717-5-1
Valid Till 17-July-2022
CALIBRATION RESULT
(All Values are in mm)
Standard Reading DUC Reading Error
0 0.000 0.000
0.5 0.495 -0.005
1 0.993 -0.007
1.5 1.491 -0.009
2 1.990 -0.010
3 2.985 -0.015
4 3.983 -0.017
5 4.980 -0.020
6 5.981 -0.019
7 6.979 -0.021
8 7.974 -0.026
9 8.975 -0.025
10 9.973 -0.027
Remarks :
• The Calibration is performed with a Temperature of 27.05 °C at Customer's Facility
• The Expanded uncertainty is ± 1.50 µm at 95.45 % confidence level with coverage factor k=2.
Checked By Approved By
K. MUHILAN - CALIBRATION TECHNICIAN R.R.HARI JAIRAM - TECHNICAL MANAGER
HT / FT / 7.8 / 01 End of Certificate Page 1 of 1
CALIBRATION CERTIFICATE
Certificate No. HT/CC/201210-11/006 ULR No. ULR - CC215120000038206F
CUSTOMER DETAILS DEVICE UNDER CALIBRATION DETAILS
M/s. L&T Construction Research and
Name Description LVDT Probe with Indicator
Testing Centre,
L&T Construction, Make GDS/RDP Model DTCH 200AG
PB.No.979, Range 0 to 10 mm Least Count 0.001 mm
Address Mount Poonamallee Road, Identification S-CON-03/S-LVDT-04
Manapakkam, Serial No. 17073
Chennai - 600 089. Location Soil Lab
Job No. 201210-11 Type Digital
Job Receipt Date 10-Dec-2020 Calibration On 10-Dec-2020
Issue Date 22-Dec-2020 Due On -- (Customer Defined)
Received Condition of DUC Satisfactory Calibrated By M. Deepan Ram
ENVIRONMENTAL CONDITION
Temperature 20 ± 2 °C Humidity 40 %RH to 60 %RH
Discipline & Group Mechanical Calibration & Dimension (Basic Measuring Instrument,Gauge Etc.)
Calibration Procedure HT/WI/M/D (Section 07) by Comparison Method
STANDARD DETAILS
Description Slip Gauge Sets (Tungsten Carbide)
Identification No. M / D / SG / 01
Certificate No. C-200717-5-1
Valid Till 17-July-2022
CALIBRATION RESULT
(All Values are in mm)
Standard Reading DUC Reading Error
0 0.000 0.000
0.5 0.497 -0.003
1 0.991 -0.009
1.5 1.492 -0.008
2 1.990 -0.010
3 2.990 -0.010
4 3.988 -0.012
5 4.988 -0.012
6 5.987 -0.013
7 6.987 -0.013
8 7.988 -0.012
9 8.988 -0.012
10 9.981 -0.019
Remarks :
• The Calibration is performed with a Temperature of 27.05 °C at Customer's Facility
• The Expanded uncertainty is ± 1.50 µm at 95.45 % confidence level with coverage factor k=2.
Checked By Approved By
K. MUHILAN - CALIBRATION TECHNICIAN R.R.HARI JAIRAM - TECHNICAL MANAGER
HT / FT / 7.8 / 01 End of Certificate Page 1 of 1
CALIBRATION CERTIFICATE
Certificate No. HT/CC/201210-11/007 ULR No. ULR - CC215120000038207F
CUSTOMER DETAILS DEVICE UNDER CALIBRATION DETAILS
M/s. L&T Construction Research and
Name Description LVDT Probe with Indicator
Testing Centre,
L&T Construction, Make GDS/RDP Model DTCH 200AG
PB.No.979, Range 0 to 10 mm Least Count 0.001 mm
Address Mount Poonamallee Road, Identification S-CON-04/S-LVDT-05
Manapakkam, Serial No. 16617
Chennai - 600 089. Location Soil Lab
Job No. 201210-11 Type Digital
Job Receipt Date 10-Dec-2020 Calibration On 10-Dec-2020
Issue Date 22-Dec-2020 Due On -- (Customer Defined)
Received Condition of DUC Satisfactory Calibrated By M. Deepan Ram
ENVIRONMENTAL CONDITION
Temperature 20 ± 2 °C Humidity 40 %RH to 60 %RH
Discipline & Group Mechanical Calibration & Dimension (Basic Measuring Instrument,Gauge Etc.)
Calibration Procedure HT/WI/M/D (Section 07) by Comparison Method
STANDARD DETAILS
Description Slip Gauge Sets (Tungsten Carbide)
Identification No. M / D / SG / 01
Certificate No. C-200717-5-1
Valid Till 17-July-2022
CALIBRATION RESULT
(All Values are in mm)
Standard Reading DUC Reading Error
0 0.000 0.000
0.5 0.496 -0.004
1 0.994 -0.006
1.5 1.490 -0.010
2 1.991 -0.009
3 2.985 -0.015
4 3.985 -0.015
5 4.975 -0.025
6 5.975 -0.025
7 6.976 -0.024
8 7.975 -0.025
9 8.972 -0.028
10 9.970 -0.030
Remarks :
• The Calibration is performed with a Temperature of 27.05 °C at Customer's Facility
• The Expanded uncertainty is ± 1.50 µm at 95.45 % confidence level with coverage factor k=2.
Checked By Approved By
K. MUHILAN - CALIBRATION TECHNICIAN R.R.HARI JAIRAM - TECHNICAL MANAGER
HT / FT / 7.8 / 01 End of Certificate Page 1 of 1
CALIBRATION CERTIFICATE
Certificate No. HT/CC/201210-11/008 ULR No. ULR - CC215120000038208F
CUSTOMER DETAILS DEVICE UNDER CALIBRATION DETAILS
M/s. L&T Construction Research and
Name Description LVDT Probe with Indicator
Testing Centre,
L&T Construction, Make GDS/RDP Model DTCH 200AG
PB.No.979, Range 0 to 10 mm Least Count 0.001 mm
Address Mount Poonamallee Road, Identification S-CON-07/S-LVDT-09
Manapakkam, Serial No. 191677
Chennai - 600 089. Location Soil Lab
Job No. 201210-11 Type Digital
Job Receipt Date 10-Dec-2020 Calibration On 10-Dec-2020
Issue Date 22-Dec-2020 Due On -- (Customer Defined)
Received Condition of DUC Satisfactory Calibrated By M. Deepan Ram
ENVIRONMENTAL CONDITION
Temperature 20 ± 2 °C Humidity 40 %RH to 60 %RH
Discipline & Group Mechanical Calibration & Dimension (Basic Measuring Instrument,Gauge Etc.)
Calibration Procedure HT/WI/M/D (Section 07) by Comparison Method
STANDARD DETAILS
Description Slip Gauge Sets (Tungsten Carbide)
Identification No. M / D / SG / 01
Certificate No. C-200717-5-1
Valid Till 17-July-2022
CALIBRATION RESULT
(All Values are in mm)
Standard Reading DUC Reading Error
0 0.000 0.000
0.5 0.495 -0.005
1 0.990 -0.010
1.5 1.491 -0.009
2 1.990 -0.010
3 2.991 -0.009
4 3.987 -0.013
5 4.985 -0.015
6 5.982 -0.018
7 6.983 -0.017
8 7.980 -0.020
9 8.981 -0.019
10 9.980 -0.020
Remarks :
• The Calibration is performed with a Temperature of 27.05 °C at Customer's Facility
• The Expanded uncertainty is ± 1.50 µm at 95.45 % confidence level with coverage factor k=2.
Checked By Approved By
K. MUHILAN - CALIBRATION TECHNICIAN R.R.HARI JAIRAM - TECHNICAL MANAGER
HT / FT / 7.8 / 01 End of Certificate Page 1 of 1
CALIBRATION CERTIFICATE
Certificate No. HT/CC/201210-11/009 ULR No. ULR - CC215120000038209F
CUSTOMER DETAILS DEVICE UNDER CALIBRATION DETAILS
M/s. L&T Construction Research and
Name Description LVDT Probe with Indicator
Testing Centre,
L&T Construction, Make GDS/RDP Model DTCH 200AG
PB.No.979, Range 0 to 10 mm Least Count 0.001 mm
Address Mount Poonamallee Road, Identification S-CON-06/S-LVDT-08
Manapakkam, Serial No. 191680
Chennai - 600 089. Location Soil Lab
Job No. 201210-11 Type Digital
Job Receipt Date 10-Dec-2020 Calibration On 10-Dec-2020
Issue Date 22-Dec-2020 Due On -- (Customer Defined)
Received Condition of DUC Satisfactory Calibrated By M. Deepan Ram
ENVIRONMENTAL CONDITION
Temperature 20 ± 2 °C Humidity 40 %RH to 60 %RH
Discipline & Group Mechanical Calibration & Dimension (Basic Measuring Instrument,Gauge Etc.)
Calibration Procedure HT/WI/M/D (Section 07) by Comparison Method
STANDARD DETAILS
Description Slip Gauge Sets (Tungsten Carbide)
Identification No. M / D / SG / 01
Certificate No. C-200717-5-1
Valid Till 17-July-2022
CALIBRATION RESULT
(All Values are in mm)
Standard Reading DUC Reading Error
0 0.000 0.000
0.5 0.496 -0.004
1 0.994 -0.006
1.5 1.492 -0.008
2 1.990 -0.010
3 2.991 -0.009
4 3.982 -0.018
5 4.985 -0.015
6 5.990 -0.010
7 6.994 -0.006
8 7.997 -0.003
9 8.998 -0.002
10 9.995 -0.005
Remarks :
• The Calibration is performed with a Temperature of 27.05 °C at Customer's Facility
• The Expanded uncertainty is ± 1.50 µm at 95.45 % confidence level with coverage factor k=2.
Checked By Approved By
K. MUHILAN - CALIBRATION TECHNICIAN R.R.HARI JAIRAM - TECHNICAL MANAGER
HT / FT / 7.8 / 01 End of Certificate Page 1 of 1
CALIBRATION CERTIFICATE
CALIBRATION RESULT
Expanded Uncertainty
Standard Reading in °C DUC Reading in °C Error in °C
in ± °C
298.94 300 1.06 0.69
548.55 550 1.45 1.50
623.36 625 1.64 1.86
798.14 800 1.86 1.86
998.06 1000 1.94 1.73
1197.94 1200 2.06 1.73
Remarks:
• The Calibration is performed at Customer's Facility.
• The Expanded Uncertainty is expressed at 95.45 % Confidence Level with Coverage Factor k=2.
Checked By Approved By
K. MUHILAN - CALIBRATION TECHNICIAN A. GOPINATH - ASSISTANT MANAGER
HT / FT / 7.8 / 01 End of Certificate Page 1 of 1
CALIBRATION CERTIFICATE
Checked By Approved By
K. MUHILAN - CALIBRATION TECHNICIAN R.R.HARI JAIRAM - TECHNICAL MANAGER
HT / FT / 7.8 / 01 End of Certificate Page 1 of 1
CALIBRATION CERTIFICATE
Certificate No. HT/CC/201210-11/012 ULR No. ULR - CC215120000038212F
CUSTOMER DETAILS DEVICE UNDER CALIBRATION DETAILS
M/s. L&T Construction Research and Testing
Name Description Stop Watch
Centre,
L&T Construction, Make Racer Model 630
PB.No.979,
Address Mount Poonamallee Road, Calibration Range Upto 2 h
Manapakkam,
Chennai - 600 089. Least Count 0.01/1 s
Job No. 201210-11 --
Accuracy
Job Receipt Date 10-Dec-2020 Identification B-SW-03
Issue Date 22-Dec-2020 Serial No. --
Received Condition of DUC SatisfactoryLocation --
ENVIRONMENTAL CONDITION Calibration On 11-Dec-2020
Temperature 25 °C ± 4 °C Due On --(Customer Defined)
Humidity 30 %RH to 75 %RH Calibrated By R. Shobana
Discipline & Group Electro Technical Calibration & Time and Frequency
Calibration Procedure HT / WI / E (Section 03) & NIST 960-12 by Comparison Method
STANDARD DETAILS
Description Stop Watch
Identification No. ET/SW/11
Certificate No. HT/CC/200725-17/001
Valid Till 26-July-2021
CALIBRATION RESULT
Expanded Uncertainty in ± h .
DUC Reading in min . s . cs Standard Reading in h. min . s . ms Error in h . min . s . ms
min . s . ms
00 . 06 . 15 000 . 00 . 06 . 272 (6.272 s) -000 . 00 . 00 . 122 (-0.122 s) 000 . 00 . 00 . 112 (0.112 s)
05 . 00 . 13 000 . 05 . 00 . 291 (300.291 s) -000 . 00 . 00 . 161 (-0.161 s) 000 . 00 . 00 . 263 (0.263 s)
10 . 00 . 19 000 . 10 . 00 . 328 (600.328 s) -000 . 00 . 00 . 138 (-0.138 s) 000 . 00 . 00 . 264 (0.264 s)
15 . 00 . 23 000 . 15 . 00 . 340 (900.340 s) -000 . 00 . 00 . 110 (-0.110 s) 000 . 00 . 00 . 264 (0.264 s)
20 . 00 . 23 000 . 20 . 00 . 389 (1200.389 s) -000 . 00 . 00 . 159 (-0.159 s) 000 . 00 . 01 . 038 (1.038 s)
25 . 00 . 13 000 . 25 . 00 . 419 (1500.419 s) -000 . 00 . 00 . 289 (-0.289 s) 000 . 00 . 01 . 038 (1.038 s)
Expanded Uncertainty in ± h .
DUC Reading in h . min . s Standard Reading in h. min . s . ms Error in h . min . s . ms
min . s . ms
00 . 30 . 00 000 . 30 . 00 . 470 (1800.470 s) -000 . 00 . 00 . 470 (-0.470 s) 000 . 00 . 01 . 188 (1.188 s)
00 . 35 . 00 000 . 35 . 00 . 495 (2100.495 s) -000 . 00 . 00 . 495 (-0.495 s) 000 . 00 . 01 . 188 (1.188 s)
00 . 40 . 00 000 . 40 . 00 . 527 (2400.527 s) -000 . 00 . 00 . 527 (-0.527 s) 000 . 00 . 01 . 188 (1.188 s)
00 . 45 . 00 000 . 45 . 00 . 561 (2700.561 s) -000 . 00 . 00 . 561 (-0.561 s) 000 . 00 . 01 . 188 (1.188 s)
Checked By Approved By
K. MUHILAN - CALIBRATION TECHNICIAN A. GOPINATH - ASSISTANT MANAGER
HT / FT / 7.8 / 01 Page 1 of 2
CALIBRATION CERTIFICATE
Certificate No. HT/CC/201210-11/012 ULR No. ULR - CC215120000038212F
CALIBRATION RESULT
Expanded Uncertainty in ± h .
DUC Reading in h . min . s Standard Reading in h. min . s . ms Error in h . min . s . ms
min . s . ms
00 . 50 . 00 000 . 50 . 00 . 587 (3000.587 s) -000 . 00 . 00 . 587 (-0.587 s) 000 . 00 . 01 . 188 (1.188 s)
00 . 55 . 00 000 . 55 . 00 . 616 (3300.616 s) -000 . 00 . 00 . 616 (-0.616 s) 000 . 00 . 01 . 188 (1.188 s)
01 . 00 . 00 001 . 00 . 00 . 655 (3600.655 s) -000 . 00 . 00 . 655 (-0.655 s) 000 . 00 . 01 . 188 (1.188 s)
01 . 05 . 00 001 . 05 . 00 . 685 (3900.685 s) -000 . 00 . 00 . 685 (-0.685 s) 000 . 00 . 01 . 188 (1.188 s)
01 . 10 . 00 001 . 10 . 00 . 724 (4200.724 s) -000 . 00 . 00 . 724 (-0.724 s) 000 . 00 . 01 . 188 (1.188 s)
01 . 20 . 00 001 . 20 . 00 . 746 (4800.746 s) -000 . 00 . 00 . 746 (-0.746 s) 000 . 00 . 01 . 188 (1.188 s)
01 . 25 . 00 001 . 25 . 00 . 784 (5100.784 s) -000 . 00 . 00 . 784 (-0.784 s) 000 . 00 . 01 . 536 (1.536 s)
01 . 30 . 00 001 . 30 . 00 . 897 (5400.897 s) -000 . 00 . 00 . 897 (-0.897 s) 000 . 00 . 01 . 536 (1.536 s)
01 . 35 . 00 001 . 35 . 00 . 969 (5700.969 s) -000 . 00 . 00 . 969 (-0.969 s) 000 . 00 . 01 . 536 (1.536 s)
01 . 40 . 00 001 . 40 . 01 . 126 (6001.126 s) -000 . 00 . 01 . 126 (-1.126 s) 000 . 00 . 01 . 536 (1.536 s)
01 . 45 . 00 001 . 45 . 01 . 170 (6301.170 s) -000 . 00 . 01 . 170 (-1.170 s) 000 . 00 . 01 . 536 (1.536 s)
01 . 50 . 00 001 . 50 . 01 . 224 (6601.224 s) -000 . 00 . 01 . 224 (-1.224 s) 000 . 00 . 01 . 536 (1.536 s)
01 . 55 . 00 001 . 55 . 01 . 233 (6901.233 s) -000 . 00 . 01 . 233 (-1.233 s) 000 . 00 . 01 . 536 (1.536 s)
02 . 00 . 00 002 . 00 . 01 . 325 (7201.325 s) -000 . 00 . 01 . 325 (-1.325 s) 000 . 00 . 01 . 536 (1.536 s)
Remarks :
• The Calibration is performed at Permanent Facility.
• The Expanded uncertainty is expressed at 95.45 % confidence level with coverage factor k=2.
Checked By Approved By
K. MUHILAN - CALIBRATION TECHNICIAN A. GOPINATH - ASSISTANT MANAGER
HT / FT / 7.8 / 01 End of Certificate Page 2 of 2
CALIBRATION CERTIFICATE
Certificate No. HT/CC/201210-11/013 ULR No. ULR - CC215120000038213F
CUSTOMER DETAILS DEVICE UNDER CALIBRATION DETAILS
M/s. L&T Construction Research and Testing
Name Description Stop Watch
Centre,
L&T Construction, Make Racer Model 630
PB.No.979,
Address Mount Poonamallee Road, Calibration Range Upto 6 h
Manapakkam,
Chennai - 600 089. Least Count 0.01/1 s
Job No. 201210-11 --
Accuracy
Job Receipt Date 10-Dec-2020 Identification C-SW-04
Issue Date 22-Dec-2020 Serial No. --
Received Condition of DUC Satisfactory Location --
ENVIRONMENTAL CONDITION Calibration On 11-Dec-2020 to 13-Dec-2020
Temperature 25 °C ± 4 °C Due On --(Customer Defined)
Humidity 30 %RH to 75 %RH Calibrated By R. Shobana
Discipline & Group Electro Technical Calibration & Time and Frequency
Calibration Procedure HT / WI / E (Section 03) & NIST 960-12 by Comparison Method
STANDARD DETAILS
Description Time Totaliser
Identification No. ET/TM/02
Certificate No. HT/4262/002 & 20/57 - 03
Valid Till 01-Jan-2021 & 11-Jun-2021
CALIBRATION RESULT
Expanded Uncertainty in ± h .
DUC Reading in min . s . cs Standard Reading in h. min . s . ms Error in h . min . s . ms
min . s . ms
05 . 00 . 14 000 . 05 . 00 . 241 (300.241 s) -000 . 00 . 00 . 101 (-0.101 s) 000 . 00 . 00 . 089 (0.089 s)
10 . 00 . 18 000 . 10 . 00 . 294 (600.294 s) -000 . 00 . 00 . 114 (-0.114 s) 000 . 00 . 00 . 088 (0.088 s)
15 . 00 . 20 000 . 15 . 00 . 360 (900.360 s) -000 . 00 . 00 . 160 (-0.160 s) 000 . 00 . 00 . 089 (0.089 s)
20 . 00 . 25 000 . 20 . 00 . 381 (1200.381 s) -000 . 00 . 00 . 131 (-0.131 s) 000 . 00 . 01 . 347 (1.347 s)
25 . 00 . 26 000 . 25 . 00 . 492 (1500.492 s) -000 . 00 . 00 . 232 (-0.232 s) 000 . 00 . 01 . 347 (1.347 s)
Expanded Uncertainty in ± h .
DUC Reading in h . min . s Standard Reading in h. min . s . ms Error in h . min . s . ms
min . s . ms
00 . 30 . 00 000 . 30 . 00 . 404 (1800.404 s) -000 . 00 . 00 . 404 (-0.404 s) 000 . 00 . 01 . 466 (1.466 s)
00 . 35 . 00 000 . 35 . 00 . 422 (2100.422 s) -000 . 00 . 00 . 422 (-0.422 s) 000 . 00 . 01 . 466 (1.466 s)
00 . 40 . 00 000 . 40 . 00 . 436 (2400.436 s) -000 . 00 . 00 . 436 (-0.436 s) 000 . 00 . 01 . 466 (1.466 s)
00 . 45 . 00 000 . 45 . 00 . 474 (2700.474 s) -000 . 00 . 00 . 474 (-0.474 s) 000 . 00 . 01 . 466 (1.466 s)
00 . 50 . 00 000 . 50 . 00 . 490 (3000.490 s) -000 . 00 . 00 . 490 (-0.490 s) 000 . 00 . 01 . 466 (1.466 s)
Checked By Approved By
K. MUHILAN - CALIBRATION TECHNICIAN A. GOPINATH - ASSISTANT MANAGER
HT / FT / 7.8 / 01 Page 1 of 2
CALIBRATION CERTIFICATE
Certificate No. HT/CC/201210-11/013 ULR No. ULR - CC215120000038213F
CALIBRATION RESULT
Expanded Uncertainty in ± h .
DUC Reading in h . min . s Standard Reading in h. min . s . ms Error in h . min . s . ms
min . s . ms
00 . 55 . 00 000 . 55 . 00 . 508 (3300.508 s) -000 . 00 . 00 . 508 (-0.508 s) 000 . 00 . 01 . 466 (1.466 s)
01 . 00 . 00 001 . 00 . 00 . 538 (3600.538 s) -000 . 00 . 00 . 538 (-0.538 s) 000 . 00 . 01 . 466 (1.466 s)
01 . 05 . 00 001 . 05 . 00 . 578 (3900.578 s) -000 . 00 . 00 . 578 (-0.578 s) 000 . 00 . 01 . 466 (1.466 s)
01 . 10 . 00 001 . 10 . 00 . 611 (4200.611 s) -000 . 00 . 00 . 611 (-0.611 s) 000 . 00 . 01 . 466 (1.466 s)
01 . 20 . 00 001 . 20 . 00 . 661 (4800.661 s) -000 . 00 . 00 . 661 (-0.661 s) 000 . 00 . 01 . 466 (1.466 s)
01 . 25 . 00 001 . 25 . 00 . 698 (5100.698 s) -000 . 00 . 00 . 698 (-0.698 s) 000 . 00 . 01 . 801 (1.801 s)
01 . 30 . 00 001 . 30 . 00 . 777 (5400.777 s) -000 . 00 . 00 . 777 (-0.777 s) 000 . 00 . 01 . 801 (1.801 s)
01 . 35 . 00 001 . 35 . 00 . 828 (5700.828 s) -000 . 00 . 00 . 828 (-0.828 s) 000 . 00 . 01 . 801 (1.801 s)
01 . 40 . 00 001 . 40 . 00 . 884 (6000.884 s) -000 . 00 . 00 . 884 (-0.884 s) 000 . 00 . 01 . 801 (1.801 s)
01 . 45 . 00 001 . 45 . 00 . 952 (6300.952 s) -000 . 00 . 00 . 952 (-0.952 s) 000 . 00 . 01 . 801 (1.801 s)
01 . 50 . 00 001 . 50 . 01 . 155 (6601.155 s) -000 . 00 . 01 . 155 (-1.155 s) 000 . 00 . 01 . 801 (1.801 s)
01 . 55 . 00 001 . 55 . 01 . 189 (6901.189 s) -000 . 00 . 01 . 189 (-1.189 s) 000 . 00 . 01 . 801 (1.801 s)
02 . 00 . 00 002 . 00 . 01 . 227 (7201.227 s) -000 . 00 . 01 . 227 (-1.227 s) 000 . 00 . 01 . 801 (1.801 s)
03 . 00 . 00 003 . 00 . 01 . 289 (10801.289 s) -000 . 00 . 01 . 289 (-1.289 s) 000 . 00 . 01 . 783 (1.783 s)
04 . 00 . 00 004 . 00 . 02 . 321 (14402.321 s) -000 . 00 . 02 . 321 (-2.321 s) 000 . 00 . 01 . 783 (1.783 s)
05 . 00 . 00 005 . 00 . 03 . 351 (18003.351 s) -000 . 00 . 03 . 351 (-3.351 s) 000 . 00 . 01 . 783 (1.783 s)
06 . 00 . 00 006 . 00 . 03 . 385 (21603.385 s) -000 . 00 . 03 . 385 (-3.385 s) 000 . 00 . 01 . 783 (1.783 s)
Remarks :
• The Calibration is performed at Permanent Facility.
• The Expanded uncertainty is expressed at 95.45 % confidence level with coverage factor k=2.
Checked By Approved By
K. MUHILAN - CALIBRATION TECHNICIAN A. GOPINATH - ASSISTANT MANAGER
HT / FT / 7.8 / 01 End of Certificate Page 2 of 2
CALIBRATION CERTIFICATE
Certificate No. HT/CC/201210-11/014 ULR No. ULR - CC215120000038214F
CUSTOMER DETAILS DEVICE UNDER CALIBRATION DETAILS
M/s. L&T Construction Research and Testing
Name Description Stop Watch
Centre,
L&T Construction, Make Racer Model 630
PB.No.979,
Address Mount Poonamallee Road, Calibration Range Upto 2 h
Manapakkam,
Chennai - 600 089. Least Count 0.01/1 s
Job No. 201210-11 --
Accuracy
Job Receipt Date 10-Dec-2020 Identification S-SW-09
Issue Date 22-Dec-2020 Serial No. --
Received Condition of DUC SatisfactoryLocation --
ENVIRONMENTAL CONDITION Calibration On 11-Dec-2020
Temperature 25 °C ± 4 °C Due On --(Customer Defined)
Humidity 30 %RH to 75 %RH Calibrated By R. Shobana
Discipline & Group Electro Technical Calibration & Time and Frequency
Calibration Procedure HT / WI / E (Section 03) & NIST 960-12 by Comparison Method
STANDARD DETAILS
Description Stop Watch
Identification No. ET/SW/11
Certificate No. HT/CC/200725-17/001
Valid Till 26-July-2021
CALIBRATION RESULT
Expanded Uncertainty in ± h .
DUC Reading in min . s . cs Standard Reading in h. min . s . ms Error in h . min . s . ms
min . s . ms
00 . 05 . 12 000 . 00 . 05 . 262 (5.262 s) -000 . 00 . 00 . 142 (-0.142 s) 000 . 00 . 00 . 115 (0.115 s)
05 . 00 . 18 000 . 05 . 00 . 324 (300.324 s) -000 . 00 . 00 . 144 (-0.144 s) 000 . 00 . 00 . 263 (0.263 s)
10 . 00 . 20 000 . 10 . 00 . 403 (600.403 s) -000 . 00 . 00 . 203 (-0.203 s) 000 . 00 . 00 . 264 (0.264 s)
15 . 00 . 31 000 . 15 . 00 . 433 (900.433 s) -000 . 00 . 00 . 123 (-0.123 s) 000 . 00 . 00 . 265 (0.265 s)
20 . 00 . 32 000 . 20 . 00 . 481 (1200.481 s) -000 . 00 . 00 . 161 (-0.161 s) 000 . 00 . 01 . 038 (1.038 s)
25 . 00 . 10 000 . 25 . 00 . 306 (1500.306 s) -000 . 00 . 00 . 206 (-0.206 s) 000 . 00 . 01 . 038 (1.038 s)
Expanded Uncertainty in ± h .
DUC Reading in h . min . s Standard Reading in h. min . s . ms Error in h . min . s . ms
min . s . ms
00 . 30 . 00 000 . 30 . 00 . 397 (1800.397 s) -000 . 00 . 00 . 397 (-0.397 s) 000 . 00 . 01 . 188 (1.188 s)
00 . 35 . 00 000 . 35 . 00 . 431 (2100.431 s) -000 . 00 . 00 . 431 (-0.431 s) 000 . 00 . 01 . 188 (1.188 s)
00 . 40 . 00 000 . 40 . 00 . 492 (2400.492 s) -000 . 00 . 00 . 492 (-0.492 s) 000 . 00 . 01 . 188 (1.188 s)
00 . 45 . 00 000 . 45 . 00 . 506 (2700.506 s) -000 . 00 . 00 . 506 (-0.506 s) 000 . 00 . 01 . 188 (1.188 s)
Checked By Approved By
K. MUHILAN - CALIBRATION TECHNICIAN A. GOPINATH - ASSISTANT MANAGER
HT / FT / 7.8 / 01 Page 1 of 2
CALIBRATION CERTIFICATE
Certificate No. HT/CC/201210-11/014 ULR No. ULR - CC215120000038214F
CALIBRATION RESULT
Expanded Uncertainty in ± h .
DUC Reading in h . min . s Standard Reading in h. min . s . ms Error in h . min . s . ms
min . s . ms
00 . 50 . 00 000 . 50 . 00 . 520 (3000.520 s) -000 . 00 . 00 . 520 (-0.520 s) 000 . 00 . 01 . 188 (1.188 s)
00 . 55 . 00 000 . 55 . 00 . 547 (3300.547 s) -000 . 00 . 00 . 547 (-0.547 s) 000 . 00 . 01 . 188 (1.188 s)
01 . 00 . 00 001 . 00 . 00 . 587 (3600.587 s) -000 . 00 . 00 . 587 (-0.587 s) 000 . 00 . 01 . 188 (1.188 s)
01 . 05 . 00 001 . 05 . 00 . 641 (3900.641 s) -000 . 00 . 00 . 641 (-0.641 s) 000 . 00 . 01 . 188 (1.188 s)
01 . 10 . 00 001 . 10 . 00 . 678 (4200.678 s) -000 . 00 . 00 . 678 (-0.678 s) 000 . 00 . 01 . 188 (1.188 s)
01 . 20 . 00 001 . 20 . 00 . 719 (4800.719 s) -000 . 00 . 00 . 719 (-0.719 s) 000 . 00 . 01 . 188 (1.188 s)
01 . 25 . 00 001 . 25 . 00 . 727 (5100.727 s) -000 . 00 . 00 . 727 (-0.727 s) 000 . 00 . 01 . 536 (1.536 s)
01 . 30 . 00 001 . 30 . 00 . 775 (5400.775 s) -000 . 00 . 00 . 775 (-0.775 s) 000 . 00 . 01 . 536 (1.536 s)
01 . 35 . 00 001 . 35 . 00 . 850 (5700.850 s) -000 . 00 . 00 . 850 (-0.850 s) 000 . 00 . 01 . 536 (1.536 s)
01 . 40 . 00 001 . 40 . 00 . 981 (6000.981 s) -000 . 00 . 00 . 981 (-0.981 s) 000 . 00 . 01 . 536 (1.536 s)
01 . 45 . 00 001 . 45 . 01 . 125 (6301.125 s) -000 . 00 . 01 . 125 (-1.125 s) 000 . 00 . 01 . 536 (1.536 s)
01 . 50 . 00 001 . 50 . 01 . 248 (6601.248 s) -000 . 00 . 01 . 248 (-1.248 s) 000 . 00 . 01 . 536 (1.536 s)
01 . 55 . 00 001 . 55 . 01 . 290 (6901.290 s) -000 . 00 . 01 . 290 (-1.290 s) 000 . 00 . 01 . 536 (1.536 s)
02 . 00 . 00 002 . 00 . 01 . 305 (7201.305 s) -000 . 00 . 01 . 305 (-1.305 s) 000 . 00 . 01 . 536 (1.536 s)
Remarks :
• The Calibration is performed at Permanent Facility.
• The Expanded uncertainty is expressed at 95.45 % confidence level with coverage factor k=2.
Checked By Approved By
K. MUHILAN - CALIBRATION TECHNICIAN A. GOPINATH - ASSISTANT MANAGER
HT / FT / 7.8 / 01 End of Certificate Page 2 of 2
CALIBRATION CERTIFICATE
Certificate No. HT/CC/201210-11/015 ULR No. ULR - CC215120000038215F
CUSTOMER DETAILS DEVICE UNDER CALIBRATION DETAILS
M/s. L&T Construction Research and
Name Description Weights
Testing Centre,
L&T Construction, Make Weigh India
PB.No.979, Range Refer Result
Address Mount Poonamallee Road, Accuracy M1 Class Serial No. Refer Result
Manapakkam, Identification Refer Result
Chennai - 600 089. Material Brass
Job No. 201210-11 Location --
Job Receipt Date 10-Dec-2020 Calibration On 14-Dec-2020
Issue Date 22-Dec-2020 Due On --(Customer Defined)
Received Condition of DUC Satisfactory Calibrated By M. Deepan Ram
ENVIRONMENTAL CONDITION
Temperature 25 °C ± 2 °C Humidity 40 % RH to 60 % RH Pressure 900 mbar to 1100 mbar
Discipline & Group Mechanical Calibration & Mass and Volume
Calibration Procedure HT/WI/M/MV (Section 01) & OIML R 111-1 by Direct Comparison Method using ABBA Cycle
STANDARD DETAILS
Description Standard Weights & Standard Weight Box
Identification M/MV/W/0.5,1,5,10/02, 2/03, 2/04, 20/17 & M/MV/WB/02
Certificate No. HT/CC/200518-01/003 & HT/CC/200518-01/004
Valid Till 20-May-2021 & 17-May-2021
Refer Next Page For Result
Identification No. Nominal Mass Conventional Mass in g Expanded Uncertainty in ± g
M-LW-16 20 g 19.99904 0.00009
M-LW-15 * 20 19.99619 0.00009
M-LW-14 50 49.99745 0.00021
M-LW-13 100 99.9943 0.00044
M-LW-12 200 199.9804 0.00085
M-LW-11 * 200 199.992 0.001
M-LW-10 500 499.992 0.0024
Remarks:
• Change in Temperature per hour is ± 1.5 °C.
• The Calibration is performed at Permanent Facility.
• Assumed Density of DUC Weights is 8400 ± 170 kg/m³ .
• The Calibration Certificate is valid for scientific & industrial purpose only.
• Assumed Density of Standard Weights used during calibration is 7950 ± 140 kg/m³.
• The Expanded Uncertainty is expressed at 95.45 % confidence level with coverage factor k=2.
Checked By Approved By
K. MUHILAN - CALIBRATION TECHNICIAN R.KOWSELYA - CALIBRATION ENGINEER
HT / FT / 7.8 / 01 Page 1 of 2
CALIBRATION CERTIFICATE
Certificate No. HT/CC/201210-11/015 ULR No. ULR - CC215120000038215F
CALIBRATION RESULT
Serial No. Identification No. Nominal Mass Conventional Mass in g Expanded Uncertainty in ± g
Checked By Approved By
K. MUHILAN - CALIBRATION TECHNICIAN R.KOWSELYA - CALIBRATION ENGINEER
HT / FT / 7.8 / 01 End of Certificate Page 2 of 2
CALIBRATION CERTIFICATE
Checked By Approved By
K. MUHILAN - CALIBRATION TECHNICIAN R.R.HARI JAIRAM - TECHNICAL MANAGER
HT / FT / 7.8 / 01 End of Certificate Page 1 of 1
CALIBRATION CERTIFICATE
Checked By Approved By
K. MUHILAN - CALIBRATION TECHNICIAN R.R.HARI JAIRAM - TECHNICAL MANAGER
HT / FT / 7.8 / 01 End of Certificate Page 1 of 1
CALIBRATION CERTIFICATE
Checked By Approved By
K. MUHILAN - CALIBRATION TECHNICIAN R.R.HARI JAIRAM - TECHNICAL MANAGER
HT / FT / 7.8 / 01 End of Certificate Page 1 of 1
CALIBRATION CERTIFICATE
Checked By Approved By
K. MUHILAN - CALIBRATION TECHNICIAN R.R.HARI JAIRAM - TECHNICAL MANAGER
HT / FT / 7.8 / 01 End of Certificate Page 1 of 1
CALIBRATION CERTIFICATE
Checked By Approved By
K. MUHILAN - CALIBRATION TECHNICIAN R.R.HARI JAIRAM - TECHNICAL MANAGER
HT / FT / 7.8 / 01 End of Certificate Page 1 of 1
CALIBRATION CERTIFICATE
Checked By Approved By
K. MUHILAN - CALIBRATION TECHNICIAN R.R.HARI JAIRAM - TECHNICAL MANAGER
HT / FT / 7.8 / 01 End of Certificate Page 1 of 1
CALIBRATION CERTIFICATE
Checked By Approved By
K. MUHILAN - CALIBRATION TECHNICIAN R.R.HARI JAIRAM - TECHNICAL MANAGER
HT / FT / 7.8 / 01 End of Certificate Page 1 of 1
CALIBRATION CERTIFICATE
Checked By Approved By
K. MUHILAN - CALIBRATION TECHNICIAN R.R.HARI JAIRAM - TECHNICAL MANAGER
HT / FT / 7.8 / 01 End of Certificate Page 1 of 1
CALIBRATION CERTIFICATE
Checked By Approved By
K. MUHILAN - CALIBRATION TECHNICIAN R.R.HARI JAIRAM - TECHNICAL MANAGER
HT / FT / 7.8 / 01 End of Certificate Page 1 of 1
CALIBRATION CERTIFICATE
Checked By Approved By
K. MUHILAN - CALIBRATION TECHNICIAN R.R.HARI JAIRAM - TECHNICAL MANAGER
HT / FT / 7.8 / 01 End of Certificate Page 1 of 1