You are on page 1of 1

5v

(NABL Accredited Loborotory os per I so/tEc 17O2s)


cc.3t26
A House Of Colibro tion
A-337, Sumel Business pork-4,
Kolupur-Norodo nooJ, Amdupuro-382345, Ahmedobod
M. 9974712437, 9173t691741 E_moil : orshe nterprisel
992@gmo il.com

Calibration Cefiiftcate
Discipline: Electro-Technical LJLR./Certificate No.: cc31262200001 5380F

1) Customer lnformation:
Customer Details: Vishakha Plastic Pipes Private Limited SRF No AETSRF l22l 1927
Block No.7924B, 28,2C-3, Sabaspur Receipt Date 08t10t2022
Road, Village - Motibhoyan, Ta-Kalol, Calibration Date o8t10t2022
Dist- Gandhinagar - 382721 Due Date 07 t10t2023
Gujarat lndia Date of lssue 09t10t2022
2) lnstrument lnformation:
Calibrated ltem Temperature Controller Visual lnspection ok
Make/Model Selec / PID-528 Type Digital / RT PT-100
Range 0 - 600"C Least Count 0.'l "c
Serial No./lO No.: 0709N08-13097 / === Calibration Performed At Site
User Location Carbon Black Content Tester Parameter Temperature
Machine Sr No DPPU2008-2009/113 Machine Make Deepak

3) Environmental Conditions:
Temperature 23!5 "C Calibration Ref Std No tTS-90
Relative Humidity 50 ! 10 % RH Calibration Procedure No. AE/CAUET/SOP-03
4) Refsrence Standards Used:
Reference Standard Serial No. Certificate No. Valid Up To Traceability
Multifunction Calibrator 994413890 AMC/00000410 14t02t2023 cc-2364
5) Calibration Results:

E)eanded
Set Reading On lvlaster Reading Obseved on ULJC Error (t) Uncertainty
(r"c)
00'c 00 L 0.0 'c
100 0 "c 100.1 'c 01'c
200.0 "c 200.3t"c 03"c 095

400 0 'c 400.5 "c 05"c


600.0'c 600 I 'c 0.8 l 'c

6) Notes:
1 For All Notes Kindly Refer Overleaf

/z=\
Calibrated By Authorised S atory

a Al.iiIEDA3AD fl
at fl c7 Arif
(Calibration Engineer) Oual Man
- End of Certifi

Forn No.A CAUGEN/F-36ReL(0t), Sep.l5, 2019 Page 1 of 1