You are on page 1of 2

NDT Supply Com. Inc.

5805 Outlook, Mission, KS 66202, U.S.A.


T: 913-685-0675
www.ndtsupply.com

CERTIFICATE OF CONFORMANCE

Date: 2 September, 2016 Report No.: K1021


Customer: To Whom It May Concern .
Order Number: 033/NDT-JAK/VIII/16
Reference Number: 150824-1

Instrument Type: USM GO Digital Flaw Detector


Brand: GE Inspection Technologies
Series No.: 009080163
Date Calibration: September 2, 2016
Recalibration Date: September 1, 2017

I here certify, that on this date, NDT Supply Com, Inc., provided the supplies called for by the
above contract in accordance with all applicable requirements for shipment. I further certify
that the supplies are of the quality and quantity specified and are in all respects in
conformance with the requirements of ASME V Article 4 Mandatory, AWS D1.1 and ASTM
E317.

The unit was calibrated and certified in according to the manufacturer calibration and test
procedure. This procedure is traceable to the NIST requirement.

Tests were at an temperature of 78 degree F and a Relative Humidity of 28%

---Traceable Standards ---


Unit ID MFR PART NUMBER SERIAL NUMBER CALIBRATION DUE
1729 TEK TDS220 B047943 2017.05.20
1283 FLUKE 87 51810725 2017.07.26

QUALITY CONTROL DEPT.


NDT Supply Com. Inc.
5805 Outlook, Mission, KS 66202, U.S.A.
T: 913-685-0675
www.ndtsupply.com

Date: 2 September, 2016 Report No.: K1021


Customer: To Whom It May Concern .
Order Number: 033/NDT-JAK/VIII/16
Reference Number: 150824-1

Instrument Type: USM GO Digital Flaw Detector


Brand: GE Inspection Technologies
Series No.: 009080163
Date Calibration: September 2, 2016
Recalibration Date: September 1, 2017

Instrument
﹥55dB Horizontal Linearity
Sensitivity <0.1%

(dB) 0 2 4 6 8 10 12 14 16 18 20 22 24 26

V
E Amplitude % 100 80 65 50. 40/0 30 25.0 20.0 15.0 12.5 10.0 7.9 6.3 5.0

R
T
Actual % 99.5. 79.8 64.4 50.6 40.5 29.8 25.1 19.5 14.8 12.5 10.5 7.8 6.1 4.7
I
C
L Tolerance% -0.5 +0.2 -0.6 +0.6 +0.5 -0.2 +0.1 -0.5 -0.2 -0.0 +0.5 -0.1 -0.2 -0.3
E

Max. Tolerance 0.6 Min. Tolerance -0.5 Total Tolerance 1.1

Contact Probe Angle Beam Probe 22dB


28dB
resolution Resolution
Dead Zoom <4mm Dynamic Range 36dB

Max. Noise Level <10% RS232 transfer Normal


AVG curve OK 3 DAC Curve OK

AWS OK 4 DAC Curve OK

THE INSPECTION OF ABOVE INSTRUMENT IS ACCORDANCE TO THE MAKER SPECIFICATION

QUALITY CONTROL DEPT.

You might also like