Certificate of Calibration

Calibration Date : 17/02/2024 Certificate Number : JCMS/A8385-08/2024

Customer Req. Cal Due : 16/02/2025 Certificate Issue Date : 19/02/2024

Customer Details

Name : Sharjah Steel Pipe Manufacturing Co L.L.C

Address : E 88 - Alsajaa Industrial - Sharjah

Equipment Description

Equipment : Manual UT Range : 0.2 mm to 100 mm

Manufacturer : GE Resolution : 0.1 mm

Model No : USM-35 Asset ID : MUS-01

Serial No. : 4651a Calibrated by : Karunakaran B

Environmental Conditions

Temperature (°C ) Start 23.1 End 23.3

Humidity (% RH) Start 44 End 43

Procedure Used For Calibration

JCMS/WP/37

Traceability
Measurements are traceable to the International System of Units (SI) through calibration laboratories accredited to ISO 17025
by an Accrediatation Body which is a signatory of the ILAC-MRA or National Metrology Institutes that are signatories of the
CIPM-MRA.

Standard Used For Calibration

V1 IIW1 Standard Block , Serial No: 11014 ;

Gauge Block Set , Serial No: 1106518; Certificate No: 361185

(Calibrated by DCL , traceable to Mikes - National Standard Laboratory- Finland)

Digital Multimeter , Serial No: 16150083; Certificate No: JCMS/IC1087-01/2023

(Calibrated by EIAC Accredited Lab , traceable to EMI, Abu Dhabi)

Uncertainty Statement :- The uncertainty reported is based on a standard uncertainty multiplied by a coverage factor k=2,
providing a level of confidence of approximately 95%.

Approved by : Arshad MT
JCMS is not liable for any damage caused by or resulting from improper or negligent use.The results of calibration refers only to the particular items submitted for
calibration and does not carry any intimation regarding the continual stability of the equipment.The results reported in this certificate are valid at the time & under the
stated conditions of measurement for the particular items & values identified above. The calibration due date is recorded in the certificate as per customer requirement.
This Certificate may not be reproduced other than in full, except with the prior written permission of Jubilant Calibration & Measurement Services LLC.

JCMS/CC/FRM/22 Rev:0.1 Issue Date:23/04/2018 Page 1 of 2


Certificate of Calibration
Calibration Date : 17/02/2024 Certificate Number : JCMS/A8385-08/2024

Customer Req. Cal Due : 16/02/2025 Certificate Issue Date : 19/02/2024

Results obtained

Horzontal Linearity

Applied Value Indicated Value Deviation


(Master) (mm) (UUC) (mm) (mm)

25.00 25.00 0.00

50.00 50.01 0.01

75.00 75.03 0.03

100.00 100.05 0.05

Maximum deviation : 0.05 mm

Uncertainty of Measurement: ± 0.01 mm

Vertical Linearity

Set % Full Screen Height Indicated Value


dB Control Change Specification
(FSH) (UUC) (mm)

80% -6 39%
35% to 45%
80% -12 21%
15% to 25%
40% +6 81%
65% to 95%
20% +12 81%
65% to 95%

Pulse Repetition Frequency (PRF) Test

Indicated Value
Set (Hz) Error (Hz)
(UUC) (mm)

100 100.00 Hz 0.00

500 500.01 Hz 0.01

1000 1.000 kHz 0.00

1500 1.500 kHz 0.00

2000 2.000 kHz 0.00

*** End of Certificate ***