You are on page 1of 19

CC - 2646

Certificate No AACPL / 04435F Field Elec. Tech. Group Page No. 1 of 1


SRF No & Date 210428.1 DT - 28 / 04 / 2021 ULR No. CC2646210000 04435F
Calibration Method : Direct Method Accuracy
Company Name M/s Atcc Test Labs Calibration Date 29.04.2021
Address 45/6, Village Prahaladpur Bangar, Cali. Due On (As Sugg. By Cust.) 28.04.2022
Industrial Area Delhi-110042 Certificate Issue Dt. 10.05.2021
UUC Condition Good

Calibration Instrument Detail


Sl Instrument Name Range Least Count Make/Sl.No Party ID / Location
1 Dig. Frequency Meter 45 - 99 Hz 0.001 Hz Everfine/PF 310 UID-194
Power Analyzer

Standard Equipments Used ( Traceable to National Standard )


Sl Instrument Name Make/Sl.No Calibrated At Cal Certificate No Cal. Valid Upto
Fluke
1 Multi Product Calibrator AIPL AIPL/2020/0756 27/12/2021
3811802

Reference Standard Calibration Purpose Humidity Temperature Calibration Performed At


IS : 13875 - Part 1 & 2 Frequency Measurement (50 ± 20)%RH (25 ± 10) ˚C Site

Calibration Procedure : As per Calibration Procedure Manual AACPL / CP / 03, Section No : CPET - 06

Calibration Result
Sl Selected Range Standard Value Measured Value Deviation Uncertainty at approx 95% confidence
( In Hz ) ( In Hz ) on UUC (in Hz) ( In Hz ) Level and coverage factor k = 2 is ±
1 45 45.00 45.001 0.001 0.02 Hz
2 70 70.00 70.001 0.001 0.02 Hz
3 95 95.00 95.001 0.001 0.02 Hz

Note
1 The calibration results reported in this calibration certificate are valid at the time of & under stated condition
2 This certificate cannot be reproduced except in full without our prior permission in writing
3 This certificate refers only to the particular item(s) submitted for calibration
4 UUC - Unit under calibration
DOCUMENT UNCERTAINTY ISSUE NO. - 08
NO. AACPL / CP / 03 PROCEDURE MANUAL DATE : 11.01.2020
REVISION NO. - 00 UNCERTAINTY CALCULATION OF SECTION NO. UPET - 06
DATE : -- -- -- -- FREQUENCY PAGE 1 OF 3

Frequency Meter
Calibration Range 45 Hz Least Count: 0.001 Hz

Unit of Measurement = Hz Range Up to 99 Hz

Multiproduct Calibrator Fluke 5502 is used for calibration


UUC Average 45.001
Std.Uncert.of Multiproduct calibrator 0.00001 Hz Std Average 45.00000
Temp.coeffice.of Multiproduct calibrator Deviation 0.001
Observed Accu.of Multiproduct calibrator 0.0035 Hz Uncertainty 0.00411
in % 0.00913744

Five Readings are taken and the deviation from the nominal value is as follows-
n

Mean Devitation x = (å xj)¸n


j=1

Standard Value UUC Value UUC Average (xj-x) (xj-x)2


Hz Hz Hz Hz Hz
45.00 45.000 -0.0006000 0.0000004
45.00 45.001 0.0004000 0.0000002
45.00 45.000 45.001 -0.0006000 0.0000004
45.00 45.001 0.0004000 0.0000002
45.00 45.001 0.0004000 0.0000002
Average 45.00 45.001

Standard Devitation , S(X) = å (Xj - X)2


n-1
= 0.0005 Hz

[S(X)]2
Standard Devitation of the Mean, S ( X) = n
= 0.00024 Hz

Standard Uncertainty in Type 'A' Evalutation, UA = 0.00024 Hz

PREPARED BY APPROVED BY COPY STATUS

Name : BHOOPESH KUMAR Name : ANAND CHAURASIA


Designation : TECHNICAL MANAGER Designation : CEO
DOCUMENT UNCERTAINTY ISSUE NO. - 08
NO. AACPL / CP / 03 PROCEDURE MANUAL DATE : 11.01.2020
REVISION NO. - 00 UNCERTAINTY CALCULATION OF SECTION NO. UPET - 06
DATE : -- -- -- -- FREQUENCY PAGE 2 OF 3
Type 'B' Evaluation
1. Standard Uncertainty Due to the uncertainty of Standard Equipment used i.e. Multiproduct calibrator
The value of uncertainty is taken from its calibration certificate
Assuming Normal distribution Coverage Factor of Calibrating Lab k =2
0.00001
U1 = = 0.00001 Hz
2

2. Standard Uncertainty Due to the Resolution of UUC


Considering half of the least count & assuming reactangular distribution

0.0005
U2 = = 0.0003 Hz
3
3. Standard Uncertainty Due to the Accuracy of Multiproduct calibrator
Assuming reactangular distribution
0.0035
U3 = = 0.0020 Hz
3

4. Standard Uncertainty Due to the Temp.coeficient of Multiproduct calibrator


Assuming reactangular distribution
0
U4 = = 0.0000 Hz
3

Combined Uncertainty:
Uc = (UA)2+(U1)2+(U2)2+(U3)2+(U4)2

Uc = 0.0021 Hz

Degree of Freedom, (Veff)


uc(y)4
Veff =
ån(ui(y)4)
j=1
Vi
Veff = 19852

Expanded Uncertainty of Overall Uncertainty or Uncertainty of Measurement :


From thestudent's distribution table, for the Confidence Level approximately 95%, the Coverage
Factor, k = 2
U = k x Uc 2 = 2 x 0.002 Hz
= 0.004 Hz

Therefore uncertainty in above measurement is = ± 0.004 Hz

Unit Average UUC Reading Average Measured Reading (Std.) Error Uncertainty ±
Hz 45.000 45.00060 -0.001 0.004
PREPARED BY APPROVED BY COPY STATUS
Name : BHOOPESH KUMAR Name : ANAND CHAURASIA
Designation : TECHNICAL MANAGER Designation : CEO
DOCUMENT UNCERTAINTY ISSUE NO. - 08
NO. AACPL / CP / 03 PROCEDURE MANUAL DATE : 11.01.2020
REVISION NO. - 00 UNCERTAINTY CALCULATION OF SECTION NO. UPET - 06
DATE : -- -- -- -- FREQUENCY PAGE 3 OF 3

X1 is reading of Reference master Minimum


X2 is reading of Reference master Maximum
Formula For Calculating True Value Y1 is reading of AACPL master Minimum
X1+((Y-Y1)/(Y2-Y1)*(X2-X1)) Y2 is reading of AACPL master Maximum
Y is the Voltage at which calibration is performed
X1 X2 Y1 Y2 Y
10.000 50.000 10.00000 50.00000 45.00000
Standard Corrected Reading 45.000002
Standard Corrected Reading 45.00
UUC Average Reading 45.001

Source of Estimate Limit Probability Standard Sensitivity Uncertainty Degre of


Uncertainty ± Distribution Uncertainty Coefficient Contribution Freedom

Ua 0.000548 0.0003 NORMAL√5 0.0002 1 0.000244949 4


U1 0.0000 0.0000 NORMAL 2 0.0000 1 0.000005 ¥
U2 0.0005 0.0003 RECT√3 0.0003 1 0.0003 ¥
U3 0.0035 0.0018 RECT√3 0.0020 1 0.0020 ¥
U4 0 0.0000 RECT√3 0.0000 1 0.0000 ¥
U5
Uc Combined Uncertainty 0.0021
Ue Expanded Uncertainty 0.0041
Veff Effective Degree of Freedom 19852

PREPARED BY APPROVED BY COPY STATUS

Name : BHOOPESH KUMAR Name : ANAND CHAURASIA


Designation : TECHNICAL MANAGER Designation : CEO
Range MeasurementPPM + mHz in Hz
Hz 119.99 45.0000 0.0025 1 0.001
Value x x 0.0025
119.99
Total = 0.00350 accuracy

Range MeasurementPPM + mHz in Hz


Hz 1199.9 45.0000 0.0025 1 0.001
Value x x 0.0025
1199.9
Total = 0.00350 accuracy

Range MeasurementPPM + mHz in Hz


Hz 1199.9 45.0000 0.0025 1 0.001
Value x 0.0025
1199.9
Total = 0.00350 accuracy

Range Measurement PPM + mHz in Hz


Hz 1199.9 45.0000 0.0025 1 0.001
Value x x 0.0025
1199.9
Total = 0.00350 accuracy
28.12.2020 to 27.12.2021
Multiproduct
Reading Uncertainty
10 0.000002
50.00 0.00001 NABL Uncertaint
100.00 0.00001 Range Uncertainty
1000.00 0.0001 20 0.02
1000.00 0.0001 100 0.02
1000.00 0.0001 1000 0.02
1000.00 0.0001 Min Unc 0.02
1000.00 0.0001

Multiproduct calibrator
Compare X Value Y Value
10 10.00001 10.000
50.00 50.00000 50.000
100.00 100.00002 100.000
1000.00 1000.00060 1000.000
1000.00 1000.00060 1000.000
1000.00 1000.00060 1000.000
1000.00 1000.00060 1000.000
1000.00 1000.00060 1000.000
DOCUMENT UNCERTAINTY ISSUE NO. - 08
NO. AACPL / CP / 03 PROCEDURE MANUAL DATE : 11.01.2020
REVISION NO. - 00 UNCERTAINTY CALCULATION OF SECTION NO. UPET - 06
DATE : -- -- -- -- FREQUENCY PAGE 1 OF 3

Frequency Meter
Calibration Range 70 Hz Least Count: 0.001 Hz

Unit of Measurement = Hz Range Up to 99 Hz

Multiproduct Calibrator Fluke 5502 is used for calibration


UUC Average 70.001
Std.Uncert.of Multiproduct calibrator 0.00001 Hz Std Average 70.00000
Temp.coeffice.of Multiproduct calibrator Deviation 0.001
Observed Accu.of Multiproduct calibrator 0.0035 Hz Uncertainty 0.00408
in % 0.005832224

Five Readings are taken and the deviation from the nominal value is as follows-
n

Mean Devitation x = (å xj)¸n


j=1

Standard Value UUC Value UUC Average (xj-x) (xj-x)2


Hz Hz Hz Hz Hz
70.00 70.001 0.0000000 0.0000000
70.00 70.001 0.0000000 0.0000000
70.00 70.001 70.001 0.0000000 0.0000000
70.00 70.001 0.0000000 0.0000000
70.00 70.001 0.0000000 0.0000000
Average 70.00 70.001

Standard Devitation , S(X) = å (Xj - X)2


n-1
= 0.0000 Hz

[S(X)]2
Standard Devitation of the Mean, S ( X) = n
= 0.00000 Hz

Standard Uncertainty in Type 'A' Evalutation, UA = 0.00000 Hz

PREPARED BY APPROVED BY COPY STATUS

Name : BHOOPESH KUMAR Name : ANAND CHAURASIA


Designation : TECHNICAL MANAGER Designation : CEO
DOCUMENT UNCERTAINTY ISSUE NO. - 08
NO. AACPL / CP / 03 PROCEDURE MANUAL DATE : 11.01.2020
REVISION NO. - 00 UNCERTAINTY CALCULATION OF SECTION NO. UPET - 06
DATE : -- -- -- -- FREQUENCY PAGE 2 OF 3
Type 'B' Evaluation
1. Standard Uncertainty Due to the uncertainty of Standard Equipment used i.e. Multiproduct calibrator
The value of uncertainty is taken from its calibration certificate
Assuming Normal distribution Coverage Factor of Calibrating Lab k =2
0.00001
U1 = = 0.00001 Hz
2

2. Standard Uncertainty Due to the Resolution of UUC


Considering half of the least count & assuming reactangular distribution

0.0005
U2 = = 0.0003 Hz
3
3. Standard Uncertainty Due to the Accuracy of Multiproduct calibrator
Assuming reactangular distribution
0.0035
U3 = = 0.0020 Hz
3

4. Standard Uncertainty Due to the Temp.coeficient of Multiproduct calibrator


Assuming reactangular distribution
0
U4 = = 0.0000 Hz
3

Combined Uncertainty:
Uc = (UA)2+(U1)2+(U2)2+(U3)2+(U4)2

Uc = 0.0020 Hz

Degree of Freedom, (Veff)


uc(y)4
Veff =
ån(ui(y)4)
j=1
Vi
Veff = #DIV/0!

Expanded Uncertainty of Overall Uncertainty or Uncertainty of Measurement :


From thestudent's distribution table, for the Confidence Level approximately 95%, the Coverage
Factor, k = 2
U = k x Uc 2 = 2 x 0.002 Hz
= 0.004 Hz

Therefore uncertainty in above measurement is = ± 0.004 Hz

Unit Average UUC Reading Average Measured Reading (Std.) Error Uncertainty ±
Hz 70.000 70.00100 -0.001 0.004
PREPARED BY APPROVED BY COPY STATUS
Name : BHOOPESH KUMAR Name : ANAND CHAURASIA
Designation : TECHNICAL MANAGER Designation : CEO
DOCUMENT UNCERTAINTY ISSUE NO. - 08
NO. AACPL / CP / 03 PROCEDURE MANUAL DATE : 11.01.2020
REVISION NO. - 00 UNCERTAINTY CALCULATION OF SECTION NO. UPET - 06
DATE : -- -- -- -- FREQUENCY PAGE 3 OF 3

X1 is reading of Reference master Minimum


X2 is reading of Reference master Maximum
Formula For Calculating True Value Y1 is reading of AACPL master Minimum
X1+((Y-Y1)/(Y2-Y1)*(X2-X1)) Y2 is reading of AACPL master Maximum
Y is the Voltage at which calibration is performed
X1 X2 Y1 Y2 Y
50.000 100.000 50.00000 100.00000 70.00000
Standard Corrected Reading 70.000008
Standard Corrected Reading 70.00
UUC Average Reading 70.001

Source of Estimate Limit Probability Standard Sensitivity Uncertainty Degre of


Uncertainty ± Distribution Uncertainty Coefficient Contribution Freedom

Ua 0.000000 0.0000 NORMAL√5 0.0000 1 0 4


U1 0.0000 0.0000 NORMAL 2 0.0000 1 0.000005 ¥
U2 0.0005 0.0003 RECT√3 0.0003 1 0.0003 ¥
U3 0.0035 0.0018 RECT√3 0.0020 1 0.0020 ¥
U4 0 0.0000 RECT√3 0.0000 1 0.0000 ¥
U5
Uc Combined Uncertainty 0.0020
Ue Expanded Uncertainty 0.0041
Veff Effective Degree of Freedom #DIV/0!

PREPARED BY APPROVED BY COPY STATUS

Name : BHOOPESH KUMAR Name : ANAND CHAURASIA


Designation : TECHNICAL MANAGER Designation : CEO
Range MeasurementPPM + mHz in Hz
Hz 119.99 70.0000 0.0025 1 0.001
Value x x 0.0025
119.99
Total = 0.00350 accuracy

Range MeasurementPPM + mHz in Hz


Hz 1199.9 70.0000 0.0025 1 0.001
Value x x 0.0025
1199.9
Total = 0.00350 accuracy

Range MeasurementPPM + mHz in Hz


Hz 1199.9 70.0000 0.0025 1 0.001
Value x 0.0025
1199.9
Total = 0.00350 accuracy

Range Measurement PPM + mHz in Hz


Hz 1199.9 70.0000 0.0025 1 0.001
Value x x 0.0025
1199.9
Total = 0.00350 accuracy
28.12.2020 to 27.12.2021
Multiproduct
Reading Uncertainty
10 0.000002
50 0.00001 NABL Uncertaint
100 0.00001 Range Uncertainty
1000 0.0001 20 0.02
1000 0.0001 100 0.02
1000 0.0001 1000 0.02
1000 0.0001 Min Unc 0.02
1000 0.0001

Multiproduct calibrator
Compare X Value Y Value
10 10.00001 10.000
50 50.00000 50.000
100 100.00002 100.000
1000 1000.00060 1000.000
1000 1000.00060 1000.000
1000 1000.00060 1000.000
1000 1000.00060 1000.000
1000 1000.00060 1000.000
DOCUMENT UNCERTAINTY ISSUE NO. - 08
NO. AACPL / CP / 03 PROCEDURE MANUAL DATE : 11.01.2020
REVISION NO. - 00 UNCERTAINTY CALCULATION OF SECTION NO. UPET - 06
DATE : -- -- -- -- FREQUENCY PAGE 1 OF 3

Frequency Meter
Calibration Range 95 Hz Least Count: 0.001 Hz

Unit of Measurement = Hz Range Up to 99 Hz

Multiproduct Calibrator Fluke 5502 is used for calibration


UUC Average 95.001
Std.Uncert.of Multiproduct calibrator 0.00001 Hz Std Average 95.00000
Temp.coeffice.of Multiproduct calibrator Deviation 0.001
Observed Accu.of Multiproduct calibrator 0.0035 Hz Uncertainty 0.00408
in % 0.004297444

Five Readings are taken and the deviation from the nominal value is as follows-
n

Mean Devitation x = (å xj)¸n


j=1

Standard Value UUC Value UUC Average (xj-x) (xj-x)2


Hz Hz Hz Hz Hz
95.00 95.001 0.0000000 0.0000000
95.00 95.001 0.0000000 0.0000000
95.00 95.001 95.001 0.0000000 0.0000000
95.00 95.001 0.0000000 0.0000000
95.00 95.001 0.0000000 0.0000000
Average 95.00 95.001

Standard Devitation , S(X) = å (Xj - X)2


n-1
= 0.0000 Hz

[S(X)]2
Standard Devitation of the Mean, S ( X) = n
= 0.00000 Hz

Standard Uncertainty in Type 'A' Evalutation, UA = 0.00000 Hz

PREPARED BY APPROVED BY COPY STATUS

Name : BHOOPESH KUMAR Name : ANAND CHAURASIA


Designation : TECHNICAL MANAGER Designation : CEO
DOCUMENT UNCERTAINTY ISSUE NO. - 08
NO. AACPL / CP / 03 PROCEDURE MANUAL DATE : 11.01.2020
REVISION NO. - 00 UNCERTAINTY CALCULATION OF SECTION NO. UPET - 06
DATE : -- -- -- -- FREQUENCY PAGE 2 OF 3
Type 'B' Evaluation
1. Standard Uncertainty Due to the uncertainty of Standard Equipment used i.e. Multiproduct calibrator
The value of uncertainty is taken from its calibration certificate
Assuming Normal distribution Coverage Factor of Calibrating Lab k =2
0.00001
U1 = = 0.00001 Hz
2

2. Standard Uncertainty Due to the Resolution of UUC


Considering half of the least count & assuming reactangular distribution

0.0005
U2 = = 0.0003 Hz
3
3. Standard Uncertainty Due to the Accuracy of Multiproduct calibrator
Assuming reactangular distribution
0.0035
U3 = = 0.0020 Hz
3

4. Standard Uncertainty Due to the Temp.coeficient of Multiproduct calibrator


Assuming reactangular distribution
0
U4 = = 0.0000 Hz
3

Combined Uncertainty:
Uc = (UA)2+(U1)2+(U2)2+(U3)2+(U4)2

Uc = 0.0020 Hz

Degree of Freedom, (Veff)


uc(y)4
Veff =
ån(ui(y)4)
j=1
Vi
Veff = #DIV/0!

Expanded Uncertainty of Overall Uncertainty or Uncertainty of Measurement :


From thestudent's distribution table, for the Confidence Level approximately 95%, the Coverage
Factor, k = 2
U = k x Uc 2 = 2 x 0.002 Hz
= 0.004 Hz

Therefore uncertainty in above measurement is = ± 0.004 Hz

Unit Average UUC Reading Average Measured Reading (Std.) Error Uncertainty ±
Hz 95.000 95.00100 -0.001 0.004
PREPARED BY APPROVED BY COPY STATUS
Name : BHOOPESH KUMAR Name : ANAND CHAURASIA
Designation : TECHNICAL MANAGER Designation : CEO
DOCUMENT UNCERTAINTY ISSUE NO. - 08
NO. AACPL / CP / 03 PROCEDURE MANUAL DATE : 11.01.2020
REVISION NO. - 00 UNCERTAINTY CALCULATION OF SECTION NO. UPET - 06
DATE : -- -- -- -- FREQUENCY PAGE 3 OF 3

X1 is reading of Reference master Minimum


X2 is reading of Reference master Maximum
Formula For Calculating True Value Y1 is reading of AACPL master Minimum
X1+((Y-Y1)/(Y2-Y1)*(X2-X1)) Y2 is reading of AACPL master Maximum
Y is the Voltage at which calibration is performed
X1 X2 Y1 Y2 Y
50.000 100.000 50.00000 100.00000 95.00000
Standard Corrected Reading 95.000018
Standard Corrected Reading 95.00
UUC Average Reading 95.001

Source of Estimate Limit Probability Standard Sensitivity Uncertainty Degre of


Uncertainty ± Distribution Uncertainty Coefficient Contribution Freedom

Ua 0.000000 0.0000 NORMAL√5 0.0000 1 0 4


U1 0.0000 0.0000 NORMAL 2 0.0000 1 0.000005 ¥
U2 0.0005 0.0003 RECT√3 0.0003 1 0.0003 ¥
U3 0.0035 0.0018 RECT√3 0.0020 1 0.0020 ¥
U4 0 0.0000 RECT√3 0.0000 1 0.0000 ¥
U5
Uc Combined Uncertainty 0.0020
Ue Expanded Uncertainty 0.0041
Veff Effective Degree of Freedom #DIV/0!

PREPARED BY APPROVED BY COPY STATUS

Name : BHOOPESH KUMAR Name : ANAND CHAURASIA


Designation : TECHNICAL MANAGER Designation : CEO
Range MeasurementPPM + mHz in Hz
Hz 119.99 95.0000 0.0025 1 0.001
Value x x 0.0025
119.99
Total = 0.00350 accuracy

Range MeasurementPPM + mHz in Hz


Hz 1199.9 95.0000 0.0025 1 0.001
Value x x 0.0025
1199.9
Total = 0.00350 accuracy

Range MeasurementPPM + mHz in Hz


Hz 1199.9 95.0000 0.0025 1 0.001
Value x 0.0025
1199.9
Total = 0.00350 accuracy

Range Measurement PPM + mHz in Hz


Hz 1199.9 95.0000 0.0025 1 0.001
Value x x 0.0025
1199.9
Total = 0.00350 accuracy
28.12.2020 to 27.12.2021
Multiproduct
Reading Uncertainty
10 0.000002
50 0.00001 NABL Uncertaint
100 0.00001 Range Uncertainty
1000 0.0001 20 0.02
1000 0.0001 100 0.02
1000 0.0001 1000 0.02
1000 0.0001 Min Unc 0.02
1000 0.0001

Multiproduct calibrator
Compare X Value Y Value
10 10.00001 10.000
50 50.00000 50.000
100 100.00002 100.000
1000 1000.00060 1000.000
1000 1000.00060 1000.000
1000 1000.00060 1000.000
1000 1000.00060 1000.000
1000 1000.00060 1000.000

You might also like