WWW.BDITEST.COM
TECHNICAL SPECIFICATIONS
AUTOMATED STRAIN TRANSDUCER CALIBRATOR
Calibration Range ±2,000με (Typical)
Applied Displacement ±0.006 in (0.01524mm) @ ±2,000με
Strain Gage lengths 2.0 or 3.0 inch (50.8 or 76.2 mm)
Strain Gage Types 120Ω, 350Ω, 1,000Ω, and 5,000Ω - Wheatstone bridge sensors
Applied Excitation Voltage +2.5Vdc
Displacement Method Precision Stepper Motor Coupled to Linear Stage w/ 1.25 microns per step
Displacement Measurement Dual Precision AC LVDT’s, Output Averaged
Displacement Certification NIST Certificate #729970 & #821/274437-07
Linear Verification Technique Linear Correlation Coefficient of > 0.9998
Repeatability Verification Tech-
Standard Deviation <0.5% (of mean)
nique
Calibration Accuracy < ±1.0%
Calibration Certificate Automatic PDF generated report1
Operating Temperature +10°C to +35°C2
Dimensions & Weight 20” X 19” X 7”, 53lbs (50.8cm X 48.3cm X 17.8cm, 24kg)
1
Microsoft® Word required
2
Recommended calibration temperature between +15°C to