SONEL S.A.

Testing and Calibration Laboratory


Wokulskiego 11
58-100 ŚWIDNICA
phone: +48 74 85 838 00, e-mail: laboratorium@sonel.pl

CALIBRATION CERTIFICATE
Date of issue: Certificate No: Page 1/7

OBJECT OF Power quality analyzer


CALIBRATION type: PQM-702 , serial No: , manufacturer: SONEL S.A.

APPLICANT

CALIBRATION According to IW01 ''Calibration of digital meters'', issue 2.1, date 07.10.2019 -
METHOD measurement form FP215/IW01/S08 date of issue 27.09.2018.

ENVIRONMENTAL Ambient temperature: (23 ÷ 23,6) °C


CONDITIONS Relative humidity: (28,3 ÷ 30) %.

DATE OF
CALIBRATION

TRACEABILITY This certificate provides traceability of measurement results to the International


System of Units (SI).

CALIBRATION Results for calibrated object have been presented on pages 2/7 to 7/7 including
RESULTS uncertainty of measurement.

UNCERTAINTY OF Uncertainty of measurement has been evaluated in compliance with EA-4/02


MEASUREMENTS M:2013. The expanded uncertainty assigned corresponds to a coverage probability
of 95 % and the coverage factor k = 2 .

This certificate without approval Laboratory shall not be reproduced except in full.
CALIBRATION CERTIFICATE issued by TESTING AND CALIBRATION LABORATORY SONEL S.A.

Date of issue Certificate No: Page 2/7

CALIBRATION Calibration results are the following:


RESULTS

1. DC voltage.
L1-N

Reference value of Measured value of Measurement


Measurement error
voltage voltage uncertainty

V V V V
100,0000 100,0088 0,0088 0,0061
-100,0000 -99,9857 0,0143 0,0061
500,000 499,998 -0,002 0,030
1000,000 1000,024 0,024 0,057
-1000,000 -1000,001 -0,001 0,057

L2-N

Reference value of Measured value of Measurement


Measurement error
voltage voltage uncertainty

V V V V
100,0000 99,9831 -0,0169 0,0061
-100,0000 -100,0130 -0,0130 0,0061
500,000 499,978 -0,022 0,030
1000,000 1000,011 0,011 0,057
-1000,000 -1000,041 -0,041 0,057

L3-N

Reference value of Measured value of Measurement


Measurement error
voltage voltage uncertainty

V V V V
100,0000 99,9725 -0,0275 0,0061
-100,0000 -100,0126 -0,0126 0,0061
500,000 499,982 -0,018 0,030
1000,000 1000,021 0,021 0,057
-1000,000 -1000,036 -0,036 0,057

PE-N

Reference value of Measured value of Measurement


Measurement error
voltage voltage uncertainty

V V V V
100,0000 100,0030 0,0030 0,0061
-100,0000 -100,0031 -0,0031 0,0061

Authorized by:
CALIBRATION CERTIFICATE issued by TESTING AND CALIBRATION LABORATORY SONEL S.A.

Date of issue Certificate No: Page 3/7

2. AC voltage.
L1-N

Reference value of Measured value of Measurement


Frequency Measurement error
voltage voltage uncertainty

Hz V V V V
20,0000 19,9994 -0,0006 0,0067
100,000 99,996 -0,004 0,054
50 230,00 229,99 -0,01 0,12
400,00 400,00 0,00 0,23
690,00 690,01 0,01 0,37

L2-N

Reference value of Measured value of Measurement


Frequency Measurement error
voltage voltage uncertainty

Hz V V V V
20,0000 19,9994 -0,0006 0,0067
100,000 99,996 -0,004 0,054
50 230,00 229,99 -0,01 0,12
400,00 400,00 0,00 0,23
690,00 690,02 0,02 0,37

L3-N

Reference value of Measured value of Measurement


Frequency Measurement error
voltage voltage uncertainty

Hz V V V V
20,0000 19,9991 -0,0009 0,0067
100,000 99,995 -0,005 0,054
50 230,00 229,99 -0,01 0,12
400,00 400,00 0,00 0,23
690,00 690,02 0,02 0,37

PE-N

Reference value of Measured value of Measurement


Frequency Measurement error
voltage voltage uncertainty

Hz V V V V
20,0000 20,0008 0,0008 0,0067
50 100,000 100,003 0,003 0,054
230,00 229,99 -0,01 0,12

Authorized by:
CALIBRATION CERTIFICATE issued by TESTING AND CALIBRATION LABORATORY SONEL S.A.

Date of issue Certificate No: Page 4/7

3. AC current (F-3 clamps)

L1

Reference value of Measured value of Measurement


Frequency Measurement error
current current uncertainty

Hz A A A A
5,000 5,063 0,063 0,015
100,00 101,05 1,05 0,30
50
1000,0 1010,0 10,0 3,0
1700,0 1716,7 16,7 5,1

L2

Reference value of Measured value of Measurement


Frequency Measurement error
current current uncertainty

Hz A A A A
5,000 5,076 0,076 0,015
100,00 101,32 1,32 0,30
50
1000,0 1013,6 13,6 3,0
1700,0 1721,5 21,5 5,1

L3

Reference value of Measured value of Measurement


Frequency Measurement error
current current uncertainty

Hz A A A A
5,000 5,042 0,042 0,015
100,00 100,77 0,77 0,30
50
1000,0 1007,9 7,9 3,0
1700,0 1712,9 12,9 5,1

Reference value of Measured value of Measurement


Frequency Measurement error
current current uncertainty

Hz A A A A
5,000 5,051 0,051 0,015
100,00 100,99 0,99 0,30
50
1000,0 1010,0 10,0 3,0
1700,0 1716,6 16,6 5,1

Authorized by:
CALIBRATION CERTIFICATE issued by TESTING AND CALIBRATION LABORATORY SONEL S.A.

Date of issue Certificate No: Page 5/7

4. AC current (C-4 clamps)

L1

Reference value of Measured value of Measurement


Frequency Measurement error
current current uncertainty

Hz A A A A
5,000 4,995 -0,005 0,015
50 100,00 99,99 -0,01 0,30
1000,0 999,6 -0,4 3,0

L2

Reference value of Measured value of Measurement


Frequency Measurement error
current current uncertainty

Hz A A A A
5,000 4,998 -0,002 0,015
50 100,00 99,93 -0,07 0,30
1000,0 1000,1 0,1 3,0

L3

Reference value of Measured value of Measurement


Frequency Measurement error
current current uncertainty

Hz A A A A
5,000 4,984 -0,016 0,015
50 100,00 99,88 -0,12 0,30
1000,0 1000,1 0,1 3,0

Reference value of Measured value of Measurement


Frequency Measurement error
current current uncertainty

Hz A A A A
5,000 4,985 -0,015 0,015
50 100,00 99,89 -0,11 0,30
1000,0 1000,2 0,2 3,0

Authorized by:
CALIBRATION CERTIFICATE issued by TESTING AND CALIBRATION LABORATORY SONEL S.A.

Date of issue Certificate No: Page 6/7

5. Frequency

Reference value of Measured value of Measurement


Nominal frequency Measurement error
frequency frequency uncertainty

Hz Hz Hz Hz Hz
40,0000 39,9994 -0,0006 0,0024
50 50,0000 49,9991 -0,0009 0,0024
60,0000 59,9989 -0,0011 0,0024
60,0000 59,9990 -0,0010 0,0024
60
70,0000 69,9987 -0,0013 0,0024

6. Voltage harmonics

Parameters of the test signal:


Fundamental (RMS) 100 V (60 Hz)
Harmonics (RMS) 5 V (n*60 Hz)
RMS value of distorted signal 100,12 V (60 Hz)

L1-N

Harmonic Reference value of Measured value of Measurement


Measurement error
No. harmonic voltage harmonic uncertainty

- V V V V
2 5,00 4,96 -0,04 0,19
20 5,00 5,00 0,00 0,19
50 5,00 4,89 -0,11 0,19

L2-N

Harmonic Reference value of Measured value of Measurement


Measurement error
No. harmonic voltage harmonic uncertainty

- V V V V
2 5,00 4,96 -0,04 0,19
20 5,00 5,00 0,00 0,19
50 5,00 4,97 -0,03 0,19

L3-N

Harmonic Reference value of Measured value of Measurement


Measurement error
No. harmonic voltage harmonic uncertainty

- V V V V
2 5,00 4,98 -0,02 0,19
20 5,00 5,00 0,00 0,19
50 5,00 4,97 -0,03 0,19

Authorized by:
CALIBRATION CERTIFICATE issued by TESTING AND CALIBRATION LABORATORY SONEL S.A.

Date of issue Certificate No: Page 7/7

7. Current harmonics

Parameters of the test signal:


Fundamental (RMS) 500 A (60 Hz)
Harmonics (RMS) 100 A (n*60 Hz)
RMS value of distorted signal 509,9 V (60 Hz)

L1

Harmonic Reference value of Measured value of Measurement


Measurement error
No. current harmonic current harmonic uncertainty

- A A A A
2 100,0 100,0 0,0 1,2
20 100,0 99,8 -0,2 1,2
50 100,0 100,1 0,1 1,2

L2

Harmonic Reference value of Measured value of Measurement


Measurement error
No. current harmonic current harmonic uncertainty

- A A A A
2 100,0 100,0 0,0 1,2
20 100,0 99,8 -0,2 1,2
50 100,0 100,1 0,1 1,2

L3

Harmonic Reference value of Measured value of Measurement


Measurement error
No. current harmonic current harmonic uncertainty

- A A A A
2 100,0 100,0 0,0 1,2
20 100,0 99,8 -0,2 1,2
50 100,0 100,1 0,1 1,2

Authorized by: