Calibration Certificate

Calibration
Customer Gas del Caribe, Grupo Tomza 30/9/2021
date
Contact person Douglas Montoya Next Service 1/4/2022
Issue date of
Calibration Certificate No. OA11351JIS 1/10/2021
Certificate
Instrument X-am 2500 Instrument part number Software version
Serial number ARLL-1381 8323900 V7.8
Configuration

CAT-Sensor EC-Sensor EC-Sensor EC-Sensor


Channel No. 1 Channel No. 2 Channel No. 3 Channel No. 4
Displayed gas ch4 O2 CO H2S
Part number 6812950 6810881 6813210 6811525
Serial number 10LL1381 11LL1381 12LL1381 13LL1381
Measuring range 100.00 %LEL 25.00 Vol% 2000.00 ppm 100.00 ppm
Last calibration 30/9/2021 30/9/2021 30/9/2021 30/9/2021
Next calibration 1/4/2022 1/4/2022 1/4/2022 1/4/2022
Alarm level A1 20.00 %LEL 19.00 Vol% 30.00 ppm 5.00 ppm
Alarm level A2 40.00 %LEL 23.00 Vol% 60.00 ppm 10.00 ppm
Hygiene Evaluation Mode - Inactivo Inactivo Inactivo
Mean Value Period - 15 min 15 min 15 min
STEL - 0.00 Vol% 60.00 ppm 5.00 ppm
TWA - 0.00 Vol% 30.00 ppm 5.00 ppm
Shift length - 8.00 h 8.00 h 8.00 h

Results of zero point / fresh air calibration


Calibration gas Fresh Air Fresh Air N2 N2
Calibration gas concentration - - 99.99 Vol% 99.99 Vol%
Calibration gas serial no. - - 955-356795 955-356795
Certificate / Lot. no. - - WO274645-21 WO274645-21 #N/A #N/A
Due date - - 30/10/25 30/10/25 #N/A #N/A
Set Value 0.00 %LEL 20.90 Vol% 0.00 ppm 0.00 ppm
Isvalue (before) -1.68 %LEL 20.27 Vol% 4.53 ppm 0.11 ppm
Isvalue (after) 0.00 %LEL 20.90 Vol% 0.00 ppm 0.00 ppm
Result OK OK OK OK

Results of span calibration


Calibration gas ch4 O2 CO H2S
Calibration gas concentration 56.82 %LEL 18.00 Vol% 100.00 ppm 25.00 ppm
Calibration gas serial no. 955-357633 955-357633 955-357633 955-357633
Certificate / Lot. no. WO290311-8 WO290311-8 WO290311-8 WO290311-8 #N/A #N/A
Due date 16/03/23 16/03/23 16/03/23 16/03/23 #N/A #N/A
Set Value 56.82 %LEL 18.00 Vol% 100.00 ppm 25.00 ppm
Isvalue (before) 64.46 %LEL 18.21 Vol% 99.69 ppm 27.54 ppm
Isvalue (after) 56.82 %LEL 18.00 Vol% 100.00 ppm 25.00 ppm
Result OK OK OK OK

The device has been checked and the measured values correspond to the specifications. The measuring equipment used for the
calibration is regularly adjusted and traceable to the national standards. If measuring method corresponds with the current technical
regulations and standards.

Dräger Safety AG & Co. KGaA


Service engineer
Signature
Jared Isaac Sauceda Alvarado
INTEK Honduras S.A. de C.V.

Dräger Safety AG & Co. KGaA Sitz der Gesellschaft. Lübeck Vorsitzender des Aufsichtsrates:
Revalstraße 1 Handelsregister: Stefan Lauer
D-23560 Lübeck Amtsgericht Lübeck WRB 4097 Komplementär
Telefon +49 451 8 82-0 Dräger Safety Verwaltungs AG
Telefax +49 451 8 82-2080 Vorstand:
Internet http://www.draeger.com Stefan Dräger (Vors./chairm.)

You might also like