Electrosurgery
Principles, Applications and Risks
Fundamentals of Electricity
Principles of Electrosurgery
Electrosurgical Technologies
• Grounded, Isolated,Tissue Response,CQM,Bipolar Cut,
EndoCut, Randomize Spray,Programming,Error Codes
Electrocautery
Operates on DC currents.
Indirect Heating.
No direct heating of tissue, hence more losses
Electrosurgery
XcelLance Medical Technologies Pvt. Ltd.
Advance Energy Sources
Ultrasonic energy is a mechanical high frequency
Ultrasonic Surgery
which achieves cutting, coagulation and
hemostasis at low temperatures with
minimal lateral thermal spread
XcelLance Medical Technologies Pvt. Ltd.
Advance Energy Sources
Ultrasonic energy is a mechanical high frequency
Lasers
Cautery
Electrosurgery
Surgical Diathermy
Electrosurgical Generator
Electrosurgical Unit
ESU
-
Time
-
Time
100KHz to 500KHz
0.1MHz to 0.5MHz
HF Range
550KHz to 1550KHz
0.55MHz to 1.55MHz
Radio Wave
1600KHz to 4200KHz
1.6MHz to 4.2MHz
RF Range
XcelLance Medical Technologies Pvt. Ltd.
Current Concentration/Density
High Current
Concentration
Low Current
Concentration
Bipolar Monopolar
Bipolar Monopolar
Microbipolar - Desiccation
Continuous waveform
Low voltage
Self limiting (100 ohms)
Return
electrode at
another site
Current flows
through the body
between the
electrodes
Manipulation of Electrode
Waveform
Size of Electrode
Speed of Incision
Therapeutic Current
Interrogation
Current
Safety Tips
Capacitive Coupling
XcelLance Medical Technologies Pvt. Ltd.
Why do O.R. team members sometimes get
burned while holding hemostats?
the voltage
of the
waveform
and
Leakage current path
the construction
characteristics of
the generator
It is recommended
that:
• Cords not be wrapped
around metal instruments
• Cords not be bundled
together
Gall bladder
Buzz hemostats
Desiccate - stop bleeding
resistance resistance
when bloody when dry
Resistance in Ohms at Radio Frequency
when Cutting or Desiccating Increasing tissue impedance
Programme Key :
Preset programme
cover most of the
surgical procedures
Recall Key :
Previous case
settings can be
recalled
XcelLance Medical Technologies Pvt. Ltd.
Basic ESU Product
Bicoag Alarm
EndoCut function availability
APC upgradable
Programmability
Shalya Lx
Shalya Sigma (200W)
Wireless Footswitch
User Program 30 30
UNLOCK CODE
All domestic units will be dispatched with 1 month usage period from the day
of start of unit.
Once the 1 month is completed , the machine will work for 5 days with
notification of 5 days remain to lock the unit, 4days, , 3 days…….
In this days you need to give the installation & take the final password from
us.
If the unit is on demo you will require further use, you need to discuss with
concern and take temporary password.
The unit will be on lock code system still the final installation is not received
by us.
RF Scalpel
XcelLance Medical Technologies Pvt. Ltd.
Argon Plasma Coagulator
Technology
Colonoscopy
Applications
APCof Broncoscopy
Application
Shalya
ArGON
Laprosurgery
OncoSurgery
• Allows longer resection with safety for the larger prostate, avoids open
prostatectomy
• Less blood loss due to superficial hemostasis
TCRE
Applications of Saline
Plasma Bipolar Resection
Hysteroscopy
With SHALYA TURoSeal,
conventional TURP set can
Arthroscopic be used for Bipolar TURP
surgery surgery with additional
attachments.
Pressure and
Energy
Electrosurgical
Technology That To Create Seal
Combines
control
Senses that tissue response is complete and stops
the cycle
XcelLance Medical Technologies Pvt. Ltd.
Features of Vessel Sealing
• Tissue seal can withstand up to 3 times
of normal systolic pressure
• Seals artery, vein or tissue bundles up
to 7mm
• Minimal thermal spread to adjacent
tissue
Lap-Assisted Vaginal
Gastric Bypass Hemorrhoidectomy Hysterectomy
(LAVH)
Whipple
Thermal Spreading
2.5
2 mm
2
0.5
0
LigaSure Harmonic Thunderbeat RF Scalpel
Scalpel
Features
•300W ESU
•Digital model
•1 Monopolar
•1 Bipolar/ 1 Saline Plasma Bipolar
•1 Vessel sealing
•30 user defined program
Features
•400W ESU
•Touch screen model
•1 Monopolar (2 monopolar in E+)
•1 Bipolar/ 1 Saline Plasma Bipolar
•1 Vessel sealing
•100 user defined program
Penta
Features
•400W ESU
•2 monopolar
•1 Bipolar/1 Saline Plasma Bipolar
•2 Vessel sealing
•Inbuilt argon port
•100 user defined program
•Autocut mode
•Internet enabled
Parameter Surf
Cut Mode Pure Sine wave- 1.7MHz
Pure- Continues sine wave,100 W at 100 ohm ,CF 1.6,
Blend – Brust 10Khz, 100W at 100 ohm, CF 2.2
Reusable
2.
Instruments
MODEL : Force Triad
MODEL : Shalya Penta
Shalya Endojet
CO2 Insufflator
CO2 Insufflator
with Smoke Evacuation
1 Product Image
7 Display 5” LCD touch screen display Digital & Bar graph display Digital Display
Veress Veress
Veress Mode
Pediatric Semi-continuous flow
8 Modes Trocar Mode
Endoscopic Intermittent flow
Bariatric
Touch screen LCD Interface &
9 Control Panel Keypad Keypad
Keypad
Smoke & Mist
10 Available NA NA
Evacuator
Gas Cylinder Supply
11 Available Available Available
Indicator
XcelLance Medical Technologies Pvt. Ltd.
KARL STORZ STRYKER-30 L HIGH
Sr.
Model SHALYA ENDOJET-30L THERMOFLATOR- FLOW
No
30L INSUFFLATOR
Percentage indication of remnant
12 Available NA NA
gas
Over Pressure alarm Over Pressure
Tube obstruction Alarm alarm
Over pressure alarm
13 Special Alarms Virtual/External Pressure Tube Occlusion
(SECUVENT feature)
Alarm alarm
Leakage alarm
14 Surgery History 100 surgery records NA NA
Necessitates Stryker
Inbuilt 2 stage Gas heating control OPTITHERM heater Gas Heating tube
15 Gas Heating System
system system assembly connection
to the unit
16 Surgical Timing Indicator Available NA NA
Audio-visual indication of critical
17 Available NA Available
error warnings
Shalya Endojet - E
CO2 Insufflator
Features:
High Flow Insufflation upto 30LPM
Accurate Flow control: For delicate & critical surgeries, Flow can be
set in 1 LPM/step
Clear & un-obstructed view: Warms CO2 gas upto body temperature
to avoid fogging over the camera due to temperature variation.
Touch Screen LCD Interface
Over-pressure detection
Indicator for consumed gas
Automatic pressure regulation with internal exhaust port
Virtual / External Back Pressure Alarm
XcelLance Medical Technologies Pvt. Ltd.
LED Endoscopy Light Source
Shalya EndoLUX
80k
60k
40k
20k
Life in
hours
XcelLance Medical Technologies Pvt. Ltd.
BETTER SUBSTITUTE TO XENON SOURCE
Luminous output, CCT
and CRI of the light are
comparable with a
typical 300W Xenon
Lamp, while consuming
less power.
Light Output
1800 Lumens Light Output
1800 Lumens
Features:
LED light source
Better substitute to xenon source
No undesired heat
Images
Pressure Range Suction: 50 – 500 Suction : 450 mmHg Suction: 0- 600 Suction : 450
mmHg (±10 %) mmHg mmHg
Simultaneous Suction
Present Present Present Present
–Irrigation process
Hysteroscopy
Hysteroscopy
Surgery mode ,laparoscopy Laparoscopy Laparoscopy
,laparoscopy
,Arthroscopy, Urology
5 kg 4 kg 8 kg 8 kg
Weight
Wolf
Fluid control LAP 2216
Features:
Camera Head buttons for complete surgeon control
Surgeon as well as Surgery Specific Custom Menu
Sensitivity 2000lx at F8
Functions:
Image mirror, Flip & Rotate
Electrical Zoom (max. 2.5 )
Electronic Shutter
Shalya iLUX
Electronic focusing
High CRI
Self-Diagnosis
10 LED
bulb x 1W
1 LED bulb x 10W
Power consumed
by conventional
halogen OT Light
Power
Consumed by
Shalya iLUX
Advantage : Even if the light field diameter is changed, the central illuminance
Ec remains same.
Thus the intensity & quality of light never degrades even on increasing the
focal spot.
No motorized movement ensures low maintenance
Ra> 97
It is termed as describing
the colour characteristics of
light whether it is warm
white (yellowish) or
Cool white (bluish) &
expressing it in kelvin.
Obstacle
Detection
Sensor
Advantage: No loose
hangings & part falling, avoid
unintentional opening for safety
purpose.
Prices:
Single Arm Pendant :
600mm : Rs. 518,000/- & onwards
Shalya Integra
OR Integration System
Shalya Flex