You are on page 1of 22

Catalog

1. Instrumentintroduction.................................................................................................. 2
2. Instrument parameter.................................................................................................... 2
3. Instrument installation instructions......................................................................... 2
4. Method and procedure of handle operation........................................................... 2
5. Matters needing attention
.................................................................................................................................................................. 1
6. Matters needing attention........................................................................................ 4
7. Maintenance and repair........................................................................................... 8
8. Common faults and troubleshooting methods....................................................... 9
9. Manufacturer's responsibility....................................................................................... 9
10. Warranty..............................................................................................................................9
11. Replacement of fittings.......................................................................................... 10
12. Duration of use....................................................................................................... 13
13. Warranty card............................................................................................................ 14

Preface
First of all, let's express our heartfelt thanks to our customers who purchased
our instruments.This manual is designed for customers to better use this instrument.
We hope that the company's products and services will honor you.
Notice
This product must be used strictly in accordance with the method in this manual.
The company shall not be liable for any consequences arising from the misuse of any
method other than the instruction.
The Company does not warrant in any form, including (but is not limited to)
implied warranties of their actual sale and suitability for a particular purpose. The
Company shall not be liable for any incidental or indirect damage caused by any error
contained in this information or by the actual performance or use of this manual as a
result of its provision.
This manual contains proprietary information protected by patent law. Copyright
reserved and no photography, photocopy, reproduction or translation of any part of
this manual into any other language is permitted without the written consent of the
Company.
The contents contained in the manual can be changed without notice to the user.
Manufacturer's responsibility
The company is only responsible for the safety and reliability of the instrument
under the following circumstances. That is, the assembly operation, re-adjustment,
improvement and maintenance are carried out by the company's accredited personnel,
the relevant electrical equipment meets the national standards, and the instrument is
used in accordance with operational guidance.
Guarantee
Manufacturing Process and Raw Materials: Our company guarantees that there
are no production process and raw material faults within one year from the date of
shipment under the condition of frequent use and maintenance of instruments (except
spare parts). Any test of the company is not perfect.
In this manual
This user manual contains all the general information of this instrument. This is
the starting point for new users to start using the instrument, because it gives a
general introduction to the system, describes how it works, introduces the installation
of the instrument and various methods of operation. You can find the required
information through the directory.
This manual is for those who are familiar with the cosmetic knowledge of this
instrument and have experience in using cosmetic operation equipment.
Please use safely
The following symbols are used in the instrument and instructions. In order to
use the instrument correctly, prevent damage to operators and patients, and protect
the instrument from damage, so as to avoid property losses. Please note the types and
meanings of the following symbols.
warning It indicates that the operation ● Mark
will cause death and serious indicates
injury. mandatory
date.
danger It indicates that the operation Marks
will cause death and serious indicate
injury. prohibited
items.
notice It is possible to cause Marks
personnel damage and indicate
instrument failure due to attention
incorrect operation. items.
The
notice Please consult the protection
instructions. type of the
equipment is
B.
Note Remind attention, emphasize
important information
1. Instrument introduction

1
Treatment Theory
Exfoliate and provide water at the same time, the machine can exfoliate keratin and
hair follicle foreign, can effectively eliminate the accumulation and aging of
keratinocytes, effectively remove follicular funnel departmently ( hair follicle hole )
and healthy skin and hair follicles hole without any impact. During cleansing, it
produces a lot of moisture and provides lasting skin moisturization.

Handle name Theory


Hydra facial handle here are three hole in the tip, two hole for
sucking the dirt and waste of the skin, one
hole for outflow the water and nutrients.
A&D: Dirt suction hole
B: Water and nutrients outflow
C: Cutin cleaning edge

Skin scrubber High Frequency Ultrasonic Vibration


The 28,000 times per second microseism
Created by two power-change slices can
effectively help discharge toxins and
absorb
nutrition to solve skin problems and beautify
skin.

Low Frequency Ultrasonic Vibration


The aging problems on the surface of skin
and the slow metabolism, together with the
influence of external environment lead to
melanin precipitation and the formation of
dark spots and freckles. By low frequency
ultrasonic vibration, it can massage skin,
promote blood circulation, promote
detoxification and make the skin restore
elasticity.

Anion leading-in Nutrition


The regular cleaning of corneous layer needs
to match with anion function for better
absorption of leading-in skin lotic and cream.
In anion function, skin surface will form a
local electric field to make the ionized
nutrition deeply penetrate the barrier layer and
let skin fully absorb nutrients.

2
Ultrasonic handle
1. By the high Frequency vibration
(1MHZ), generate ultrasonic with strong
penetrability. It is a special instrument
that could radiate an alternate wave and
spread around. It is much stronger that the
normal sonic wave, with high frequency,
good directive, strong penetrability and
great expand ability.
2. Because of the high frequency vibration,
it will do soft and micro massage to skin
cells, accelerate the blood circulation,
facilitate the metabolism, active the cells and
improved the adsorption of nutrition.
3.High frequency micro massage could
eliminate the aged cells,expel the toxin
and fine the wrinkles. Accompany usage
with cosmetic product or medicine,it could
treat and improve the skin problems.
4, The Frequency of ultrasonic could also
generate resonance vibration of the cells,
consume the fat and improve water absorbing
capacity of cells, rejuvenate the skin and
recover the elasticity.

Bio RF handle High frequency wave permeates epidemis


and take effect on the collagen-rich dermal
tissue directly which makes water molecules
in the skin move. The move produces
controllable heat and results in contraction of
collagen and generation of new collagen and
dermal remodeling .as a result, the skin looks
tighter, wrinkle reduce gradually .Radio
Frequency: energy is transmitted to the place
between three electrodes; the current
permeates target tissue and generate heat.

3
H2 O2 Polymer atomizing pen Nutrition or cosmetic product can be filled
into the plastic bottle. Witth high pressure,
the nutrition is sprayed on the skin area which
greatly emhance nutrition to be absorbed
maximally, skin cleaning and skin
rejuvenation

Cold hammer handle Ice skin, reducing pores and calming skin.

Product Functions
Handle Function Before and after
name
Skin whitening,
Dead skin removal,
Blackhead removal,
Essence of skin
whitening lead-in.

Lighten color patch and


hyper pigmentation,
Shrink pores,
Improve skin problems
like dark, yellow,
blackhead,
Deep cleaning and
refresh skin,
Make skin more clastic,
lustrous

4
moisturizing

Moisturizing
Collagen enrichment,
Wrinkle remove

Dilute fine wrinkle


remove,
Skin lift,
Skin tightening.

Swelling
remove, Light
dark circles
remove,
Eye bag
remove. Shrink
pore.

5
2. Instrument parameter

Voltage AC110V/220V 50-60Hz


Power 200W
Screen 9.2 inches
Bio RF handle 1Mhz
H2 O2 Polymer atomizing pen Impact pressure 100Kpa
Cold hammer handle -10℃ ~ 5℃
Skin scrubber 1Mhz
Ultrasonic handle 40K, 28.3Mhz
Hydra facial handle Vacuum pressure 100Kpa
Material ABS plastic
N.W. 5kgs
Package size 45*42*25cm

6
3. Instrument installation instructions
3.1 Open the box and check
3.1.1 carefully remove the instrument and accessories from the box and keep the
packaging materials for later transportation or preservation. Check the attachment.
3.1.2 to check whether there is any mechanical damage.
note: if there is any problem, contact our sales department or agent
immediately.
3.2 The instrument should be placed in a clean and ventilated room with a temperature
of 5440C and a mixing degree not exceeding 80% to avoid direct sunlight. And
ensure that ventilation has at least 5CM gap. Remember not to place on the
instrument.
3.3 Insert the accessories into the corresponding interface according to the text and
picture information above the interface on both sides of the instrument. Then put the
accessories neatly on the accessories rack.
3.4 instrument back and side device description.

7
note:
● Please pour out the liquid in A, B.C, D bottles and waste liquid bottles before
carrying the instrument, otherwise the liquid will enter the instrument during the
process of transportation and cause damage to the instrument.
note:
● During use, the buzzer emits a buzzing sound. When the icon of the waste liquid
bottle on the interface flickers, if it flickers, please remove the waste liquid from
the waste liquid bottle and reinstall the waste liquid bottle.
4. Method and procedure of handle operation
4.1 Use of water mill cleaning handle

4.1.1 Cleansing
4.1.2 Choose a suitable water grinder according to the skin and rotate the screen
forward to the bottle A, B, or 0.
4.1.3 Adjust negative pressure suction and water flow to make the face clean and
clean (10 minutes).
note:

When using the D bottle , turn the knob to D, then press and twist,
wait 3 minutes for the bottle to produce enough for reuse.
4.1.1After the whole face is cleaned, turn the knob to the position to clean the
water mill pipe. First, adjust the water output to the maximum (instrument front
knob) suction to the maximum. Open the instrument and cover the water mill
head and the pipe with a random gift of the cover.

8
note:
When the whole bottle of diluted raw liquor is exhausted, the A, B, C, D bottles
must be loaded with warm water and cleaned again in accordance with the
method of washing the water mill pipe to prevent pipe clogging.
4.2 Use of H2 O2 Polymer atomizing pen

4.1.1 Clean the parts that need to be operated.


4.1.2 Pour the diluted essence into the container on the spray
pen. 4.1.3Press the button on the screen.
4.1.4The pen is aligned to the operation part, and the operation is started by
pressing the opening key on the handle.
4.3 Use of cold hammer handle

4.3.1 Clean the parts that need to be


operated. 4.3.2Press start button.
4.3.3Operate the front part of the handle.
4.4 Use of Bipolar RF handle

9
4.4.1 Clean the parts that need to be operated.
4.4.2Apply proper amount of nutritious gel on the
part.
4.4.3 Adjusting working hours, working intensity and mode of RF.
4.4.4 Click on the RF icon on the screen or press the start button on the handle to
activate the RF function.
4.4.5 According to the facial lines, the neck needs customers to lift their necks
from the bottom up. The operation time of each part is 10 -15 minutes.
4.5 Use of ultrasonic handle

4.5.1 Clean the parts that need to be


operated. 4.5.2Apply the right amount of the
essence.
4.5.3 Adjust working time, sound intensity and mode parameters.
4.5.4 Click the sound wave icon on the screen and press the start button on the
handle to start the sound wave function.
4.5.5 Circle slowly along the lines of the position for 5-8 minutes.
4.6 Use of shovel skin

10
4.6.1 Clean the parts that need to be operated.
4.6.2 Adjust the intensity of work, select the required working mode, press the
start button, and start working.
4.6.3 Cleanliness is the way to remove surface dirt.
1) Cleansing skin with facial cleanser
2) Disinfect the spade head and hold the stick in the hands of the user.
3) First, apply enough pure water to the skin.
4) The head of the shovel is 45 degrees sloping close to the skin, moving
from bottom to top, with moderate speed.
5) The operation time is 20 minutes.
4.6.4Stroking is a method of lymphatic
massage.
1) Cleansing skin with facial cleanser
2) Disinfect the spade head and hold the stick in the hands of the
user. 3)First apply enough rose water to the skin.
4)Flat skin sliding according to the texture of the muscle.
5)The operation time is 15 minutes.

11
5. Matters needing attention
5.1 Points for attention in
use warning:
● 1)Please do not use well connected socket to avoid fire.
2)Switch off the power switch when replacing the fuse, so as not to cause a fire.
3)The power supply and effective grounding socket of 220V 22V.50HZ 1HZ will
be used.
4) Do not use damaged or processed power lines to avoid fire and danger of electric
shock
5) Bend over and twist the power cord to avoid fire and electric shock.
6) Please contact the company in time if the instrument is loose, falling or
damaged. 7)Do not use this instrument in unstable environments such as tilt,
vibration and
impact.
8) No water or sundries can be found inside the instrument. When there are ten
thousand foreign objects entering, please contact our company in time.
9) When the instrument moves, it must first unplug the power
supply. 10)When the instrument moves, avoid vibration as far as
possible.
11) The instrument should be placed in a stable place. A small area of 5CM or more
should be used to ensure ventilation.
12) The instrument can not be covered with anything, so as to prevent it from
blocking.
13) Avoid using this instrument in the following environments, where the sun
shines; high temperature environment: near water environment; vibration; inclined
place; strong magnetic field environment; electromagnetic wave, place where the
impact voltage occurs; chemical preservation place; corrosive gas place
14) This instrument should not be placed near radio, TV, duplicator, fax machine,
etc.
15) It should not be used with other instruments (such as microwave), so as not to
interfere with each other and cause misoperation.
5.2 Precautions before
use warning:
1) Read the instructions carefully before starting the machine. The operator must
be familiar with the instructions and operation methods. Only the device must be
managed by a special person.
2) setting the operating parameters under the guidance of
professionals. Notice:
3)confirm whether the receiver is in normal operation before using, confirm
whether the earth wire is connected correctly.
4)make sure that all wires are properly connected and safe.
5)When the instrument is used together, it must be used correctly. Otherwise, there
may be danger.
5.3 Precautions in
use warning:
12
1) When an abnormal condition occurs during the operation of the instrument or

13
patient, the operator immediately removes the operating handle from the patient,
presses the stop switch, and stops the use of the instrument.
Notice:
2) During the operation of the instrument, it is necessary to constantly observe the
situation of the beauty guests and instruments, and keep an eye on whether any
abnormal conditions have occurred.
3) A skilled operator shall use this instrument correctly, and no person other than
the operator shall use this instrument.
5.4 Points for attention after
use Notice:
1) after the operation is finished, the control handle is put back to the control
handle rack.
2) the probe should be cleaned after using the
handle. 5.5Failure, safekeeping, precautions during
inspection
warning:
1) When the instrument is abnormal, please do not use it, in case of smoke, coke
smell, can not operate the situation, if continue to use, there is a risk of fire shock.
Please press the stop button quickly, turn off the power supply, unplug the power
plug, and immediately contact the company or distributor.
2) personnel other than technicians, never repair and refit the instrument. The
instrument must not be opened inside.
Notice:
3) Instruments and components must be inspected regularly.
4) Do not use cleaning tools that damage the protective coatings of the instruments.
5) If the instruments are not used for more than 2 weeks, the instruments must be
carefully inspected for re-use.
5.6Warning and prompting instructions
warning:
When the high frequency equipment and the instrument are connected to the
same patient at the same time, the electrode position of the instrument may cause
skin burns or damage to the instrument. When the instrument is used near a
shortwave or microwave equipment (1M), the output of the instrument may be
unstable due to microwave interference.
Using electrodes near the chest will increase the risk of fibrillation.
6. Taboo population
1). sensitive skin. Severe hemorrhoids.
2) rhythm pacemaker, input cardioverter defibrillator or other electromagnetic
implantable medical device.
3) severe diabetes, hypertension and
hyperlipidemia. 4)people with skin infections.
7. Maintenance and repair
7.1 maintain
7.1.1 Routine examination
Managers must check instruments and components.
14
A) is there a reliable connection between the power outlet and the ground
B) Whether the power supply wire is deformed or broken: through the naked
eye inspection, whether the wire is deteriorated or deformed, if any, may cause fire
because of leakage of electricity. Please replace the new power cord immediately.
C) Whether the control handle is abnormal or broken: through the naked eye
inspection,
whether the output line is deteriorated or deformed, if any, please replace
immediately.
D) Please check whether the instrument is correct before use, if not, please
stop using.
7.1.2 cleaning
Dust must be wiped with a dry, soft cloth while cleaning the instrument. The
stolen goods must be wiped with a soft, wet cloth. Before cleaning the
instrument, the power switch must be turned off and the power cord must be
disconnected.
7.1.3 Replacement of fuse
A) specification of power fuse 8A (65 * 20M)
B) change the fuse. Please click the steps.
1) close the power switch;
2) unplug the power plug;
3) pry the fuse housing with a screwdriver;
4) remove the broken fuse and replace it with a new one;
5) put the fuse housing back to the original position 6) and plug in the power
plug.
warning:
In order to prevent electric shock, you must turn off the power supply and
unplug the power supply when changing the fuse.
7.1.4 For a long time (more than one year), the use of cleaning equipment
should be carried out by professional maintenance personnel.
7.1.5 When the instrument is not used, it should be placed in a ventilated and
dry place, with a dust-proof cover covering it.
7.2 Repair
7.2.1 After purchasing the instrument, it is necessary to fill in the warranty card
and send it to the manufacturer or agent. It is guaranteed to be effective. It is
free of charge for one year since the date of purchasing the instrument and
Issuing the instrument.
7.2.2 After the warranty period, please contact the manufacturer directly to
provide service guarantee.
7.2.3 Maintenance of this machine must contact our customer service
department or dealers, do not take out repairs, self-sealing, the company does
not warrant repair.
8. Common faults and troubleshooting methods
8.1 Instrument can not work properly, do not easily disassemble and assemble, should
be timely with the company's customer service department or dealers, after finding

15
out the cause of failure.
8.2 Before you ask for maintenance service, please check the instrument according to
the table to see if you can handle the problem yourself.
Elimination and solution
Failure phenomenon Causes methods
No response after boot. The power plug is not Check if plugs and sockets
inserted into the socket, are inserted.
The power switch is not
opened.

the power fuse is broken


A handle can not be used Start button without Click the start button of
normally. clicking on handle. the handle.
The probe is not installed Check whether probe
correctly. installation is correct.
The working head has Replace working heads or
been damaged. aprons.
The interface behind the Reinstall
instrument is not properly
installed.
Leaky liquid Damaged or missing apron Replacement of silicone
of sealing position. apron for sealing.
Improper installation
Return to the manufacturer
for repair.
9. Manufacturer's responsibility
8.3.1 The product should conform to the product standard, and if it does not meet
our company's responsibility, it will be responsible for replacement and return.
8.3.2 Our company shall bear the corresponding liability for personal and property
damage caused by defects in the manufacture of products, but we shall not be liable
for any of the following circumstances
A) Buyer, operator, user not operating according to the instructions, not
according to safety precautions, warning instructions used
B) Violation of product transportation, installation, use, maintenance, storage
and other provisions cause damage to products, personal and property.
10. Our company carries out lifelong maintenance and technical services for the
products. The quality problems occurred during the stipulated quality assurance
period will be repaired free of charge for the users.
11. Replacement of fittings
Storage: Instruments should be stored in the original packaging box, placed in a
well-ventilated clean room, packing box should be padded high, environmental
temperature - 10C ~ 40C, relative temperature is not more than 93%, do not allow
harmful gases, flammable, explosive materials and corrosive gases.
Transportation: Instruments under packaged conditions suitable for highway,

16
railway, aviation and waterway transport shall be protected from violent vibration and
shock during loading, unloading and transportation, and shall not be damped, and
shall not be mixed with inflammable and corrosive substances, and shall be
specifically required to comply with the provisions of the order contract.
12. Duration of use
12.1 The warranty period of this instrument is 1 years. (Note: except for man-made
damage).
12.2 The warranty period for the parts is 3 months (except for man-made damage).
13.3If the instrument exceeds the service life, it must be approved by the engineer,
and the engineer should inspect the performance of the instrument regularly in the
subsequent use of the well.
12.4Discarded instruments and accessories exceeding the time limit should be
handled according to local regulations.
13. Warranty card

Product name : Purchase date: deadline:

Seller: Add.: Tel.:

Customer: Add.: Tel.:

Repair date Reason of the fault Repair content

Maintain list