You are on page 1of 1

Certificate of Analysis

Material 85517.260
Material description Buffer solution pH 9.18
Grade

Lot 21E284101
Expires end of 2023-Apr-26
Last Quality Control 2021-Jul-02

Additional information

Characteristics Specifications Measured values

pH (25°C) (tolerance ± 0.02) 9.16 - 9.20 9.18


pH laboratory uncertainty ± 0.007 (k=1) ± 0.007 (k=1)
pH homogeneity uncertainty ± 0.003 (k=1) ± 0.003 (k=1)
pH stability uncertainty ± 0.009 (k=1) ± 0.009 (k=1)
pH expanded, combined uncertainty ± 0.03 (k=2; 95 %) ± 0.03 (k=2; 95 %)

Signature

This document has been produced electronically and is


valid without a signature.

Anja Vanhalle, Head of Laboratory - Haasrode


VWR International bv; Geldenaaksebaan 464; BE-3001
Leuven; Belgium

Additional information

pH-Method: pH value is analyzed with a glass electrode after 4-point calibration following the validated standard procedure. The expanded
uncertainty relevant for the user contains contributions of bottle to bottle variation (inhomogeneity), stability over time and laboratory measurement
uncertainties as shown above and using a coverage factor k=2 for a 95 % coverage probability.

Preparation and use: This reference material is prepared gravimetrically from di-sodium tetraborate and high purity water. It is intended to be used
for calibrating pH measurement devices. The minimum sample size for use is 17.5 ml.

Accreditation: VWR International BVBA is accredited as reference material producer according to ISO 17034. The batch homogeneity has been
proven by analyzing minimum 6 samples distibuted over the entire production process.

The pH of this buffer solution is traceable to and verified against primary Standard Reference Materials (SRM) from National Institute of Standards
and Technology (NIST): SRM 186 I + II and SRM 191 I + II.

Store at +2°C to +25°C tightly closed in the original container under nitrogen. Keep this reference material always under nitrogen. Consume within
4 weeks after first opening.

For Professional use in Laboratory or Manufacturing. Not for use as an Active Pharmaceutical Ingredient or Food or Animal Feed. Suitability and intended use of the product remains the responsibility of the user. 57FBA5.190-E7F34190

VWR International LLC, Radnor Corporate Center, Building One, Suite 200, 100 Matsonford Road, Radnor, PA 19087, USA
85517.260 - 21E284101 - V. 1 - Page 1 / 1
VWR International bvba, Haasrode Research Park Zone 2020, Geldenaaksebaan 464, 3001 Leuven, Belgium

You might also like