You are on page 1of 16

qty laboratory room

100 Lab-1 Spectrophotometer


Spectrophotometer (UV-VIS optional)
configuration : Double Beam
Optical System Is Ratio Beam System.
* Wave Length Range : varies (preferred that started from not less 200nm and not more 1100nm).
Spectral Band Width : ≤3nm.
Display : LCD or better.
Wave Length Accuracy : ±1nm or better.
Light Source : Halogen (Vis), deuterium (UV).
Absorbance: 0 -1,999A
Photometric Modes: ABS, T% & Cons.
Stability = 0.001 ABS / Hour.
PHOTODETECTOR : Solid state Detector Preferred
DATA PROCESSING : LIS Interface preferred
PRINTER : required built in printer
Supplied With Spare Lamps For Each Unit.
Supplied With Complete Accessories.
Standard:
Mains power 220- 240V ±6% , 50 Hz
Service manual
Operation manual
Service/ maintenance and technical documents inclusive of schematics, component diagrams, trouble
shooting and voltage/ wave form checks, diagnostics/ error codes and their interpretations, spare
parts ordering information etc.
Recommended spare parts list with prices & validity must be provided with the offer
Installation & Commissioning :
2 Years comprehensive warranty, from the date of installation and commissioning
User / Nurses training, by Specialist from the Suplier.
Shall comply to CE marked or FDA Certificate .
150 Lab-2 Hemoglobin Meter.
Measuring method : By Cyanmethemoglobin
Measuring range: 0 -30 g/dl
Three digits indication
Dilution ratio: 1:250
Wave lengh : 530 nm
Accuracy : 1%
Supplied with spare bulb and all accessories
150 Lab-3 PH Meter.
Measuring range: pH: 0.00 - 14.00pH
MV: -1,999 ~ +1,999mV
Temperature: 0 - 100.0°C
Resolution: pH: 0.0lpH, mV: ImV, temperature: 0.1 °C
Accuracy: pH: ± 0.0lpH, MV: ±0.1% FS
Temperature: ±0.5°C± 1 digit
Temperature of the measured solution: 0 - 60°C
Range of automatic temperature compensation: 0 - 100°C
150 Lab-4 Bilirubin Meter.
Measure Total Bilirubin Value.
Digital Display.
For Management Of Hyperbilirubinemia.
High Stability & Accuracy.
Supplied With Complete Accessories.
60 Lab-5 A- Laboratory Microscope, Binocular
Standard Laboratory Use Type.
With Binocular Observation Tube, Mechanical Stage, Condenser And Filter Holder.
10X, 20X, 40X & 100/Oil Achromatic Eyepieces.
6V, 20W Halogen Bulb Illumination System.
Supplied With Two Spare Bulbs For Each Unit.
Supplied With Complete Accessories.
Ground glass filter
Spare fuses
Main cable
One spare lamp
Bottle of Immersion oil
Dust cover
Instruction manual in English
Power supply 220-240 V 50Hz
80 Lab-6 Water Bath
Capacity About 14 liters
Temperature Accuracy = 0.5c.
Temperature = Up To75c.
Stainless Steel Made , Electronic control.
Supplied With Spare Heater For Each Unit.
Supplied With Complete Accessories.
20 Lab-7 Blood Bank Refrigerator
Temperature = + 3c 1c.
Doors: Thermal or Multipane Glass
Storage Capacity : 80 Bags (200ml) Minimum.
Microprocessor Temperature Control.
Complete Alarm System (Temp. low/high, power failure, unit with audiable alarms).
Chart Recorder Time is required
Recording thermometer, 7 days, electronic
Supplied With Complete Accessories.
Main UPS for system Operations 100-220V
Operation Manual
Service Manual
NOTE
All equipment needing consumables must allow the possibility to use generic and/or locally made
consumables and/or disposables. Compliance to this condition must be declared here by the bidders.
Service/ maintenance and technical documents inclusive of schematics, component diagrams, trouble
shooting , spare parts ordering information etc.
2 Years comprehensive warranty, from the date of installation and commissioning & Compressor 5
years Warranty
Installation & Commissioning must be done by manufacturer engineer
The system offered shall be designed to operate normally under the conditions of the purchaser's
country. The conditions include Power Supply, Climate, Temperature, Humidity, etc.
Users training, by Specialist from the Supplier.
Shall comply to CE marked or FDA Certificate
80 Lab-8 Bench Top Centrifuge
Table Top Centrifuge.
Maximum Revolution = 5,500 RPM.
Full Automatic Microprocessor Control.
Timer = From 0 - 60 Minute With Hold Mode .
Capacity = 15ml x 16 Tubes.
Supplied With Complete Accessories.
Supplied With One Year Recommended Spare Parts.
1 Year comprehensive warranty, from the date of installation and commissioning
User / Nurses training, by Specialist from the Suplier.
Shall comply to CE marked or FDA Certificate
40 Lab-9 Hematocrit Centrifuge
Micro Capillary Hematocrit Centrifuge.
Revolution = 12,000 - 14,000 RPM.
Digital Speed Display Indicator.
Capacity = 20 CapillaryTubes.
Supplied With Complete Accessories.
Supplied With One Year Recommended Spare Parts.
1 Year comprehensive warranty, from the date of installation and commissioning
User / Nurses training, by Specialist from the Suplier.
Shall comply to CE marked or FDA Certificate
Lab-
40 10 Differential Counter Digital
Light Weight, Count with Totalizer. For simplifying differential blood cell counts. This counter
eliminates the calculations necessary when making Leukocyte counts. The eight keys represent the
eight normal Schilling groups of Leucocytes, i.e.: Basso, Eos
Lab-
40 11 Water Distiller
Automatic Microprocessor Control. (Optional)
Capacity = 2- 5 Liter / Hour .
Water purification system included.
low water cut off
Water supply flow rate liters/hr 60lit approx.
Supplied With Complete Accessories & conection.
Lab-
40 12 Timer, Interval Timing 60 Minutes.
Three separate recording /display fields which can be used independent-three timers in one.
Stopwatch, timer and clock function available on all three fields. Magnetic clip aids convenient
handling, each uses a 24 h system displaying hours, minutes and sec.
Lab-
40 13 Laboratory Stool Swivel
Chrome-plated tubular framed stool with adjustable height
Wooden seat
Adjustable backrest, hand-controlled and foot circular rest
Shall be mobile on 5 stars stand, 5 castors
Lab-
40 14 a-Laboratory Medical Refrigerator
Maintains internal air temperature between +2°C and +8°C
Smooth, molded steel cabinet with rust-proof enamelled finish and bevelled edges
Easy non-electric push-button lighting.
Shall have no corners to trap dirt and bacteria.
The absorption-type cooling system shall be quiet and dependable, with no compressor or other
moving parts to wear out
Audio/visual alarm signal on temperature deviation, with remote alarm terminals providing mains
failure alarm signal
Shall be easy-to-clean and all-white interior features adjustable, removable shelves and door bins for
stocking flexibility
Internal capacity shall be minimum 250 lit.
Lockable door, 2 keys for security
Main power supply 100-240V
Supplied with :
5 removable shelves
Operation Manual
Service Manual
NOTE
All equipment needing consumables must allow the possibility to use generic and/or locally made
consumables and/or disposables. Compliance to this condition must be declared here by the bidders.
Service/ maintenance and technical documents inclusive of schematics, component diagrams, trouble
shooting , spare parts ordering information etc.
2 Years comprehensive warranty, from the date of installation and commissioning
Installation & Commissioning must be done by manufacturer engineer
Users training, by Specialist from the Suplier.
The system offered shall be designed to operate normally under the conditions of the purchaser's
country. The conditions include Power Supply, Climate, Temperature, Humidity, etc.
Lab-
100 15 Automated Micropipette, Fixed type (Different sizes)
1- 20 µl
2- 50 µl
3- 100 µl
4- 1000 µl
Lab-
100 16 Automated Micropipette, Variable type (Different Sizes)
Large, clear digital display, compatible with wide range of tips. High degree of accuracy and precision,
easy maintenance and calibration.
1-Single Variable:
a- 100 – 1000 µl
b- 20 – 200 µl
c- 10 – 50 µl
d- 2 - 20 µl
Lab-
40 18 Analytical Balance
Has rapid tarring over full 160 gm weighing range.
Digital read out to 0.01mg precision standard deviation 0.01mg complete with cord plus dust cover,
cleaning brush and instruction manual.
Accessories: light bulbs for one-year supply. With anti-vibration mat.
Lab-
40 19 Differential Counter manual
For simplifying differential blood cell counts, eliminates the calculations necessary when making
Leukocyte counts. The eight keys represent the eight normal Schilling groups of Leucocytes, i.e.:
Basso, Eosin, Myelo, Juven, stab seg,
Lymph, Mono. Above each key is the name and picture of particular cell. When a total of one
hungered cells is counted, a bell rings. The percentage of each type of cell is then read directly from
the window above each key. Counter clockwise rotate handle
Lab-
40 20 Stop Watch (Digital)
Lab-
20 21 Elisa Machine
Open system
Elisa Reader:
*
Two basic readers for 96-well micro plates,
Flexibility and reliability,
Fast reading: 5s for 96well plates,
Internal memory holds up to 64 assay,
Data manipulation through software,
Spectral range: 340-750 nm,
Read-out range: 0-3.5 Abs,
Measurement range: 0-2.5 at 405 nm,
Reading speed: 5 s entire plate,
Shaking: Linear shaking , 3 speed .
Photometric Method,
Optical System: measurement and reference Channels, multiple channels required,
Measurement modes: Single/dual wavelength, endpoint, Singe wavelength kinetic and two point.
Calculation modes: Ranged Grayscale Threshold, Floating cut-off, Point-to-point, Linear regression
and Cubic spline
Reading speed: 5 entire plate,
linearity: {( -2 )–(+2) }% Abs. CV<0.2%.
Resolution, OD: 0,01,
Compatible Microplates: roun/flat bottom preferred, programmable Assays is preferred.
with Display, interface.
Supply with: all required accessories to put Machine in function, Cover, Adapter, fuse, bulbs and
calibration Plate.
Elisa Washer:
Automated,
Compatible plates: 96-well plates,
Reliable and easy to use,
Co-axial wash heads for trouble-free washing and low residual volumes 8 and 12 way
interchangeable wash heads,
In-line rinse capability automatic pump shutdown.
Wash/waste bottle capacity: 2 L+ 2 L.
Additional wash bottle capacity: 2 L.
Priming consumption: 15-20 ml.
Wash heads: 8-and 12 ways.
Program cards: 1-4 washes.
Accuracy: ˃2%,
Max dispense volume (ml/well): ˂1
Power Supply / Input Power: 220 / 50 Hz
Bukup suitable UPS
Supplied with :
Operation Manual
Service Manual
Shall comply with CE marked or FDA Certificate.
Service/ maintenance and technical documents inclusive of schematics, component diagrams, trouble
shooting , spare parts ordering information etc.
2 Years comprehensive warranty, from the date of installation and commissioning
Installation & Commissioning must be done by manufacturer engineer
Users training, by Specialist from the Suplier.
The system offered shall be designed to operate normally under the conditions of the purchaser's
country. The conditions include Power Supply, Climate, Temperature, Humidity, etc.
Lab-
20 22 Automatic hematology Analyzer (cbc) complete blood counter
Principle: Electrical Impedance Method- WBC / RBC / PLT, Colorimetric- HGB
Open or Closed System, Diluent and reagents (3 or 5) , device must have an economic consumption
of reagents mechanism.
Reportable Parameters: 3 Part Differential, 23 Parameters - WBC, LYM#, MID#, GRA#, LYM
MID%,GRA%, RBC,HGB, HCT,MCV, MCH, MCHC, RDW-SD, RDW-CV, PL MPV, PDW, PCT,P-
LCR, WBC Histogram, RBC Histogram, PLT HistogramT
Operating Platform : Windows interface with mouse control (Not Compulsory).
Speed : 40-60 samples per hour
Calibration : Both Automatic and Manual Calibration mode. 3 sets of Calibration co -efficient for
Whole Blood, Anti-coagulant Peripheral Blood and Pre-diluted Peripheral Blood.
Quality Control : 3 QC Method; and X-R QC with appropriate QC Charts, Westgard
Results: Internal memory can store at liest 40,000 Test values with Histograms
Printer / Display: Builtin thermal printer with optional external printer, Color LCD, 9"
Interface: 2- 4 USB, 1 RS232 serial port, 1 Lan, 1 PS2 Mouse, 1 PS2
Power Supply / Input Power: 220 / 50 Hz
Backup suitable UPS
Supplied with :
Operation Manual
Service Manual
Service/ maintenance and technical documents inclusive of schematics, component diagrams, trouble
shooting , spare parts ordering information etc.
2 Years comprehensive warranty, from the date of installation and commissioning
Installation & Commissioning must be done by manufacturer engineer
Users training, by Specialist from the Suplier.
Shall comply with CE marked or FDA Certificate.
NOTE
Machine with close Or open system, the supplier must supply free quantity of reagents with machine
and must mention how many free kits and tests of reagents would be supplied. The cost of soluations
& consumbles per Test for one Year should also be offered.
Machine with close Or open system, should be put cost of soluations, consumbles per Test ,,should
be put cost of reagents kits for at least 5 years (price / kit ) and how many tests per kit. Device must
be supplied with manufacturer reference test sample for calibration and quality assurance.
Depended of above we well choise the system
The system offered shall be designed to operate normally under the conditions of the purchaser's
country. The conditions include Power Supply, Climate, Temperature, Humidity, etc.
Lab-
20 23 Semi-Automated biochemistry Analyzer
CONFIGURATION Benchtop
Connection to LAS track
Processing modes
System capacity
THROUGHPUT, max tests 50-70 /hr
SAMPLE TYPE
Serum/plasma
Urine
CSF
SAMPLE SIZE, 2- 30 µL
TEST MENU, min
Acid phosphatase ,Alanine transaminase (ALT) ,Albumin ,Aldolase,Alkaline phosphatase (ALP)
,Amylase ,Aspartate transaminase (AST) ,Bilirubin (direct/total) ,Blood urea nitrogen (BUN) ,C3 ,C4
,Calcium ,Chloride ,Cholesterol ,Cholinesterase ,Creatine kinase (CK) ,Creatinine ,CK-MB ,CO2 ,CSF
protein ,GGT ,Glucose ,IgA ,IgG ,IgM ,Inorganic phosphorus ,Iron ,Lactate dehydrogenase (LDH)
,Magnesium ,P-5-P ,Potassium
Total protein
Sodium
Thyroxine (T4)
Tyroxine uptake (TU)
Triglycerides
Urea
Uric acid
Urine protein
Others HBDH, lipase, neonatal bilirubin, RF, LD-1, ammonia, TIBC, LDL etc according to the device
range of tests offered.
PROGRAMMED 50 TESTS
USER-DEFINABLE TESTS
METHOD USED End point, rate, ISE OR Kinetic
Optical system (Photometric )
Light source (Tungsten-halogen )
REAGENTS, TYPE Liquid
Substitution
Refrigerated onboard
SYSTEM FEATURES
Closed-tube sampling
Direct sampling
Liquid-level sensing
Clot detection
Auto dilution
Abnormal values flag
Auto verification
Auto quality control
Auto calibration
DATA MANAGEMENT
Display LCD
Results stored
Computer interface
Current LIS vendor interfaces
Printer
Bar-code reader
LINE POWER, 220 VAC, 50Hz
Backup
WATER REQUIREMENTS, <15 L/hr
Onboard supply
Supplied with:
Set of 3 sterilisation containers adapted to the equipment proposed
Water distiller device
Operation Manual
Service Manual
Shall comply with CE marked or FDA Certificate.
NOTE
All equipment needing consumables must allow the possibility to use generic and/or locally made
consumables and/or disposables. Compliance to this condition must be declared here by the bidders.
Service/ maintenance and technical documents inclusive of schematics, component diagrams, trouble
shooting , spare parts ordering information etc.
2 Years comprehensive warranty, from the date of installation and commissioning
Installation & Commissioning must be done by manufacturer engineer
Users training, by Specialist from the Suplier.
The system offered shall be designed to operate normally under the conditions of the purchaser's
country. The conditions include Power Supply, Climate, Temperature, Humidity, etc.
Lab-
20 26 Blood bags Shaker
Microplate shaker provides reliable, regulated shaking for two or four microplates with a speed range
of 150 to 1000rpm.
Variable speed, horizontal microtitre plate shaker
Variable speed: 150 to 1000rpm or more
Orbit: 2mm
Integral timer, 1 minute to 24 hours, with automatic switch-off
Standard platform for two microplates, optional accessory platform for four microplates
Low voltage power supply enables safe coldroom operation
Operating ambient temperature range 4°C to 45°C, for use in coldroom or incubator
Low energy consumption.
The Specification attached with catalogue
Lab-
40 27 Flame Photometer
Digital, including integral dilutor, automatic digital readout, rapid measurement of Na, K and Li on a
minimum sample of approx1.6ml Complete with standard reagents, solutions, Integral diluter, air
compressor, hoses valve for gas cylinder
Lab-
100 28 Rack Test tubes.
Stainless Steel .
12 holes 13 mm 3 Tier
12 holes 20 mm 3 Tier
Lab-
20 29 Donor Chair
High Quality, Comfortable with storage drawers
WIth adjustable back rest, adjustable dual arm rest
nd not more 1100nm).

mponent diagrams, trouble


interpretations, spare
c and/or locally made
lared here by the bidders.
mponent diagrams, trouble

ioning & Compressor 5

ons of the purchaser's


nts. This counter
eight keys represent the

ee timers in one.
ip aids convenient
compressor or other

nals providing mains

shelves and door bins for

c and/or locally made


lared here by the bidders.
mponent diagrams, trouble

ons of the purchaser's

of accuracy and precision,

th cord plus dust cover,

essary when making


ps of Leucocytes, i.e.:
hen a total of one
s then read directly from

kinetic and two point.


point, Linear regression

s preferred.

pter, fuse, bulbs and

nd 12 way
mponent diagrams, trouble

ons of the purchaser's

n economic consumption

MID#, GRA#, LYM


L MPV, PDW, PCT,P-

tion co -efficient for

mponent diagrams, trouble

of reagents with machine


ed. The cost of soluations

umbles per Test ,,should


ests per kit. Device must
lity assurance.

ons of the purchaser's


phosphatase (ALP)
nitrogen (BUN) ,C3 ,C4
atinine ,CK-MB ,CO2 ,CSF
dehydrogenase (LDH)

according to the device


c and/or locally made
lared here by the bidders.
mponent diagrams, trouble

ons of the purchaser's

plates with a speed range

microplates

ent of Na, K and Li on a


s, Integral diluter, air