You are on page 1of 2

Document number: 841088_13_applic_UA AOX, Date: 10/09, s.w. 7.

Reagent definition UA AOX


Reagent UA AOX A UA AOX B
Stable on board (days) 30 30
Alarm limit (ml)
Vial volume 60 20
Barcode id 071 072
Syringe speed Normal Normal

Alarm limit settings: 1 ml for 10 ml vial, 2 ml for 20 ml vial, 5 ml for 60 ml vial.


Check the correct vial size.

Test definition UA AOX


Test type Photometric Test in use YES
Full name Uric Acid, AOX 2 reag. Low High Unit
system
Test limit, low (⌦) 10 7200 µmol/l
Online name UA AOX Initial absorbance 0 0.8 A
Dilution limit * 1200 µmol/l
Result unit µmol/l Secondary dil. 1 + 0 5
Number of decim. 0
Acceptance AUTOMATIC
Dilution 1 + 0

Sample type
Serum Plasma Urine Correction factor 1
CSF Other Correction bias 0 µmol/l

Konelab ™ / T Series URIC ACID (AOX) - code: 981391 (7 x 60 ml), code: 981788 (8 x 20 ml)
Check the reagent insert for the specific information of the reagent(s), calibrator(s) and control(s).
CE-marked application for Konelab / T Series by Thermo
⌦ Determination limit (measuring range low)
Fisher Scientific Oy, applicable for use outside the US.
Value is lot dependent, refer to the calibrators value sheet.
T Series is distributed by Siemens Medical Solutions
Diagnostics in selected countries.

Page 1/2
Document number: 841088_13_applic_UA AOX, Date: 10/09, s.w. 7.2

Calibration parameters UA AOX


Calibration type LINEAR Factor Bias
Repeat time (d) 0 Abs. error (mA) 4 Bias correction in use NO
Points/Calibrator Duplicate Rel. error (%) 5 Bias corr. repeat time (dd:hh)
Acceptance MANUAL Response limit (mA) Bias corr. limit (mA)
Type of calibrators SEPARATE Min * Total
Curve direction ASCENDING Max * Incremental
Calibrator id Calibrator Conc. ( ) Dil. ratio 1 + Bias cal. id
Concentration Water 0 0
Dil. ratio 1 + sCal 376 0

Test flow UA AOX


Blank YES
Antigen excess NO Normal cuvette

Reagent Sample Incubation Blank Reagent Incubation End point

Reagent Volume (µl) Time (sec.) Resp. min Reagent Time (sec.) Wavelength (nm)
UA AOX A 3 300 * UA AOX B 300 540
Volume (µl) Disp. with Resp. max Volume (µl) Side wavel. (nm)
120 WATER * 30 700
Disp. with Volume (µl) Disp. with Res. net.abs.(A)
WATER 20 WATER 0
Volume (µl) Dilution with Volume (µl)
20 WATER 10
Wash reagent Wash reagent Wash reagent Meas. type
UA AOX A NONE UA AOX B NORMAL

Before disp. Before disp.

Konelab ™ / T Series URIC ACID (AOX) - code: 981391 (7 x 60 ml), code: 981788 (8 x 20 ml)
Check the reagent insert for the specific information of the reagent(s), calibrator(s) and control(s).
CE-marked application for Konelab / T Series by Thermo
⌦ Determination limit (measuring range low)
Fisher Scientific Oy, applicable for use outside the US.
Value is lot dependent, refer to the calibrators value sheet.
T Series is distributed by Siemens Medical Solutions
Diagnostics in selected countries.

Page 2/2