You are on page 1of 2

Document number: D05604_04_applic_TPROT Plus, Date: 10/09, s.w. 7.

Reagent definition TPROT Plus


Reagent TPROT Plus
Stable on board (days) 21
Alarm limit (ml)
Vial volume 60
Barcode id 331
Syringe speed Normal

Alarm limit settings: 1 ml for 10 ml vial, 2 for 20 ml vial, 5 for 60 ml vial


Check the correct vial size.

Test definition TPROT Plus


Test type Photometric Test in use YES
Full name Total Protein Plus Low High Unit
Test limit, low (⌦) 3 250 g/l
Online name TPROT Plus Initial absorbance 0 1 A
Dilution limit * 100 g/l
Result unit g/l Secondary dil. 1 + 0 2
Number of decim. 0
Acceptance AUTOMATIC
Dilution 1 + 0

Sample type
Serum Plasma Urine Correction factor 1
CSF Other Correction bias 0 g/l

Konelab ™ / T Series TOTAL PROTEIN Plus - code: 981826 (10 x 60 ml), code: 981827 (12 x 20 ml)
Check the reagent insert for the specific information of the reagent(s), calibrator(s) and control(s).
CE-marked application for Konelab / T Series by Thermo
⌦ Determination limit (measuring range low).
Fisher Scientific Oy, applicable for use outside the US.
Value is lot dependent, refer to the calibrators value sheet.
T Series is distributed by Siemens Medical Solutions
Diagnostics in selected countries.

Page 1/2
Document number: D05604_04_applic_TPROT Plus, Date: 10/09, s.w. 7.2

Calibration parameters TPROT Plus


Calibration type LINEAR Factor Bias
Repeat time (d) 0 Abs. error (mA) 10 Bias correction in use NO
Points/Calibrator Duplicate Rel. error (%) 0 Bias corr. repeat time (dd:hh)
Acceptance MANUAL Response limit (mA) Bias corr. limit (mA)
Type of calibrators SEPARATE Min * Total
Curve direction ASCENDING Max * Incremental
Calibrator id Calibrator Conc. ( ) Dil. ratio 1 + Bias cal. id
Concentration Water 0 0
Dil. ratio 1 + sCal 69 0

Test flow TPROT Plus


Blank YES
Antigen excess NO Normal cuvette

Reagent Blank Sample Incubation End point

Reagent Resp. min Volume (µl) Time (sec.) Wavelength (nm)


TPROT Plus * 3 600 540
Volume (µl) Resp. max Disp. with Side wavel. (nm)
150 * WATER 700
Disp. with Volume (µl)
WATER 20
Volume (µl) Dilution with Res. net.abs.(A)
20 WATER 0
Wash reagent Wash reagent Meas. type
NONE NONE NORMAL

Konelab ™ / T Series TOTAL PROTEIN Plus - code: 981826 (10 x 60 ml), code: 981827 (12 x 20 ml)
Check the reagent insert for the specific information of the reagent(s), calibrator(s) and control(s).
CE-marked application for Konelab / T Series by Thermo
⌦ Determination limit (measuring range low).
Fisher Scientific Oy, applicable for use outside the US.
Value is lot dependent, refer to the calibrators value sheet.
T Series is distributed by Siemens Medical Solutions
Diagnostics in selected countries.